Evaluation of the mixtures ammonium phosphate/magnesium nitrate and palladium nitrate/magnesium nitrate as modifiers for simultaneous determination of Cd, Cr, Ni and Pb in mineral water by GFAAS

Detalhes bibliográficos
Autor(a) principal: Amorim Filho,Volnei R
Data de Publicação: 2004
Outros Autores: Fernandes,Kelly G, Moraes,Mercedes de, Gomes Neto,José Anchieta
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Journal of the Brazilian Chemical Society (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-50532004000100006
Resumo: A method is described for the simultaneous determination of Cd, Cr, Ni and Pb in mineral water samples by graphite furnace atomic absorption spectrometry with a transversely heated graphite atomizer (THGA) and a longitudinal Zeeman-effect background correction system. The electrothermal behavior of analytes during pyrolysis and atomization steps was studied without modifier, in presence of 5 mug Pd and 3 mug Mg(NO3)2 and in presence of 50 mug NH4H2PO4 and 3 mug Mg(NO3)2. A volume of 20 muL of a 0.028 mol L-1 HNO3 solution containing 50 mug L-1 Ni and Pb, 10 mug L-1 Cr and 5 mug L-1 Cd was dispensed into the graphite tube at 20 °C. The mixture palladium/magnesium was selected as the optimum modifier. The pyrolysis and atomization temperatures were fixed at 1000 °C and 2300 °C, respectively. The characteristic masses were calculated as 2.2 pg Cd, 10 pg Cr, 42 pg Ni and 66 pg Pb and the lifetime of the graphite tube was around 600 firings. Limits of detection based on integrated absorbance were 0.02 mug L-1 Cd, 0.94 mug L-1 Cr, 0.45 mug L-1 Ni and 0.75 mug L-1 Pb, which exceeded the requirements of Brazilian Food Regulation that establish the maximum permissible level for Cd, Cr, Ni and Pb at 3 mug L-1, 50 mug L-1, 20 mug L-1 and 10 mug L-1, respectively. The recoveries of Cd, Cr, Ni and Pb added to mineral water samples varied within the 93-108%, 96-104%, 87-101% and 98-108% ranges, respectively. Results of analysis of standard reference materials (National Institute of Standards and Technology: 1640-Trace Elements in Natural Water; 1643d-Trace Elements in Water) were in agreement with certified values at the 95% confidence level.
id SBQ-2_7820ca3a6c6760c819fa47da1d255fe9
oai_identifier_str oai:scielo:S0103-50532004000100006
network_acronym_str SBQ-2
network_name_str Journal of the Brazilian Chemical Society (Online)
repository_id_str
spelling Evaluation of the mixtures ammonium phosphate/magnesium nitrate and palladium nitrate/magnesium nitrate as modifiers for simultaneous determination of Cd, Cr, Ni and Pb in mineral water by GFAASGFAASsimultaneous determinationmineral waterinorganic trace elementsA method is described for the simultaneous determination of Cd, Cr, Ni and Pb in mineral water samples by graphite furnace atomic absorption spectrometry with a transversely heated graphite atomizer (THGA) and a longitudinal Zeeman-effect background correction system. The electrothermal behavior of analytes during pyrolysis and atomization steps was studied without modifier, in presence of 5 mug Pd and 3 mug Mg(NO3)2 and in presence of 50 mug NH4H2PO4 and 3 mug Mg(NO3)2. A volume of 20 muL of a 0.028 mol L-1 HNO3 solution containing 50 mug L-1 Ni and Pb, 10 mug L-1 Cr and 5 mug L-1 Cd was dispensed into the graphite tube at 20 °C. The mixture palladium/magnesium was selected as the optimum modifier. The pyrolysis and atomization temperatures were fixed at 1000 °C and 2300 °C, respectively. The characteristic masses were calculated as 2.2 pg Cd, 10 pg Cr, 42 pg Ni and 66 pg Pb and the lifetime of the graphite tube was around 600 firings. Limits of detection based on integrated absorbance were 0.02 mug L-1 Cd, 0.94 mug L-1 Cr, 0.45 mug L-1 Ni and 0.75 mug L-1 Pb, which exceeded the requirements of Brazilian Food Regulation that establish the maximum permissible level for Cd, Cr, Ni and Pb at 3 mug L-1, 50 mug L-1, 20 mug L-1 and 10 mug L-1, respectively. The recoveries of Cd, Cr, Ni and Pb added to mineral water samples varied within the 93-108%, 96-104%, 87-101% and 98-108% ranges, respectively. Results of analysis of standard reference materials (National Institute of Standards and Technology: 1640-Trace Elements in Natural Water; 1643d-Trace Elements in Water) were in agreement with certified values at the 95% confidence level.Sociedade Brasileira de Química2004-02-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-50532004000100006Journal of the Brazilian Chemical Society v.15 n.1 2004reponame:Journal of the Brazilian Chemical Society (Online)instname:Sociedade Brasileira de Química (SBQ)instacron:SBQ10.1590/S0103-50532004000100006info:eu-repo/semantics/openAccessAmorim Filho,Volnei RFernandes,Kelly GMoraes,Mercedes deGomes Neto,José Anchietaeng2004-05-25T00:00:00Zoai:scielo:S0103-50532004000100006Revistahttp://jbcs.sbq.org.brONGhttps://old.scielo.br/oai/scielo-oai.php||office@jbcs.sbq.org.br1678-47900103-5053opendoar:2004-05-25T00:00Journal of the Brazilian Chemical Society (Online) - Sociedade Brasileira de Química (SBQ)false
dc.title.none.fl_str_mv Evaluation of the mixtures ammonium phosphate/magnesium nitrate and palladium nitrate/magnesium nitrate as modifiers for simultaneous determination of Cd, Cr, Ni and Pb in mineral water by GFAAS
title Evaluation of the mixtures ammonium phosphate/magnesium nitrate and palladium nitrate/magnesium nitrate as modifiers for simultaneous determination of Cd, Cr, Ni and Pb in mineral water by GFAAS
spellingShingle Evaluation of the mixtures ammonium phosphate/magnesium nitrate and palladium nitrate/magnesium nitrate as modifiers for simultaneous determination of Cd, Cr, Ni and Pb in mineral water by GFAAS
Amorim Filho,Volnei R
GFAAS
simultaneous determination
mineral water
inorganic trace elements
title_short Evaluation of the mixtures ammonium phosphate/magnesium nitrate and palladium nitrate/magnesium nitrate as modifiers for simultaneous determination of Cd, Cr, Ni and Pb in mineral water by GFAAS
title_full Evaluation of the mixtures ammonium phosphate/magnesium nitrate and palladium nitrate/magnesium nitrate as modifiers for simultaneous determination of Cd, Cr, Ni and Pb in mineral water by GFAAS
title_fullStr Evaluation of the mixtures ammonium phosphate/magnesium nitrate and palladium nitrate/magnesium nitrate as modifiers for simultaneous determination of Cd, Cr, Ni and Pb in mineral water by GFAAS
title_full_unstemmed Evaluation of the mixtures ammonium phosphate/magnesium nitrate and palladium nitrate/magnesium nitrate as modifiers for simultaneous determination of Cd, Cr, Ni and Pb in mineral water by GFAAS
title_sort Evaluation of the mixtures ammonium phosphate/magnesium nitrate and palladium nitrate/magnesium nitrate as modifiers for simultaneous determination of Cd, Cr, Ni and Pb in mineral water by GFAAS
author Amorim Filho,Volnei R
author_facet Amorim Filho,Volnei R
Fernandes,Kelly G
Moraes,Mercedes de
Gomes Neto,José Anchieta
author_role author
author2 Fernandes,Kelly G
Moraes,Mercedes de
Gomes Neto,José Anchieta
author2_role author
author
author
dc.contributor.author.fl_str_mv Amorim Filho,Volnei R
Fernandes,Kelly G
Moraes,Mercedes de
Gomes Neto,José Anchieta
dc.subject.por.fl_str_mv GFAAS
simultaneous determination
mineral water
inorganic trace elements
topic GFAAS
simultaneous determination
mineral water
inorganic trace elements
description A method is described for the simultaneous determination of Cd, Cr, Ni and Pb in mineral water samples by graphite furnace atomic absorption spectrometry with a transversely heated graphite atomizer (THGA) and a longitudinal Zeeman-effect background correction system. The electrothermal behavior of analytes during pyrolysis and atomization steps was studied without modifier, in presence of 5 mug Pd and 3 mug Mg(NO3)2 and in presence of 50 mug NH4H2PO4 and 3 mug Mg(NO3)2. A volume of 20 muL of a 0.028 mol L-1 HNO3 solution containing 50 mug L-1 Ni and Pb, 10 mug L-1 Cr and 5 mug L-1 Cd was dispensed into the graphite tube at 20 °C. The mixture palladium/magnesium was selected as the optimum modifier. The pyrolysis and atomization temperatures were fixed at 1000 °C and 2300 °C, respectively. The characteristic masses were calculated as 2.2 pg Cd, 10 pg Cr, 42 pg Ni and 66 pg Pb and the lifetime of the graphite tube was around 600 firings. Limits of detection based on integrated absorbance were 0.02 mug L-1 Cd, 0.94 mug L-1 Cr, 0.45 mug L-1 Ni and 0.75 mug L-1 Pb, which exceeded the requirements of Brazilian Food Regulation that establish the maximum permissible level for Cd, Cr, Ni and Pb at 3 mug L-1, 50 mug L-1, 20 mug L-1 and 10 mug L-1, respectively. The recoveries of Cd, Cr, Ni and Pb added to mineral water samples varied within the 93-108%, 96-104%, 87-101% and 98-108% ranges, respectively. Results of analysis of standard reference materials (National Institute of Standards and Technology: 1640-Trace Elements in Natural Water; 1643d-Trace Elements in Water) were in agreement with certified values at the 95% confidence level.
publishDate 2004
dc.date.none.fl_str_mv 2004-02-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-50532004000100006
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-50532004000100006
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/S0103-50532004000100006
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Sociedade Brasileira de Química
publisher.none.fl_str_mv Sociedade Brasileira de Química
dc.source.none.fl_str_mv Journal of the Brazilian Chemical Society v.15 n.1 2004
reponame:Journal of the Brazilian Chemical Society (Online)
instname:Sociedade Brasileira de Química (SBQ)
instacron:SBQ
instname_str Sociedade Brasileira de Química (SBQ)
instacron_str SBQ
institution SBQ
reponame_str Journal of the Brazilian Chemical Society (Online)
collection Journal of the Brazilian Chemical Society (Online)
repository.name.fl_str_mv Journal of the Brazilian Chemical Society (Online) - Sociedade Brasileira de Química (SBQ)
repository.mail.fl_str_mv ||office@jbcs.sbq.org.br
_version_ 1750318165404942336