Radial dilation of nephrostomy balloons: a comparative analysis

Detalhes bibliográficos
Autor(a) principal: Hendlin,Kari
Data de Publicação: 2008
Outros Autores: Monga,Manoj
Tipo de documento: Artigo
Idioma: eng
Título da fonte: International Braz J Urol (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1677-55382008000500002
Resumo: PURPOSE: The dynamics of percutaneous balloon expansion may differ with increasing extrinsic compressive forces and increasing inflation pressures. This study compares the ability of percutaneous balloons to expand under different radial constrictive forces. MATERIALS AND METHODS: Three 30F nephrostomy balloons were tested: Bard X-Force™, Boston Scientific Microvasive Amplatz Tractmaster™, and Cook Ultraxx™. With a super stiff guidewire in place, the balloon tip was secured by elevated vice grips on either side of the balloon. A string was wrapped around the balloon center once, and incremental increases in load were added (2g, 42g, 82g, and 122g) to represent increasing extrinsic compression. The balloon was inflated with a contrast agent and circumference changes were measured at increments of 4 ATM, 10 ATM, and burst pressure. Balloons were tested in triplicate for each load. RESULTS: All balloons were unable to reach 90% of their expected diameter with larger constrictive loads (122g) at low (4 ATM) and nominal (10 ATM) inflation pressures. Only the Bard and Cook balloons reached at least 90% of the expected diameter with a coefficient of variance (CV) less than 10% at burst pressure under the larger constrictive load (122g), 94.3% ± 6.7%, CV 7.1% and 96.3% ± 2.9%, CV 3.0% respectively. All balloons performed well under low constriction forces and reached at least 80% of the expected diameter by 10 ATM under all constrictive loads. CONCLUSIONS: The Bard X-Force and Cook Ultraxx percutaneous nephrostomy balloons achieved the most reliable radial dilation against large constrictive forces simulating fascial or retroperitoneal scar tissue.
id SBU-1_ff951caccf2b0602a3bb767a2b08e0cf
oai_identifier_str oai:scielo:S1677-55382008000500002
network_acronym_str SBU-1
network_name_str International Braz J Urol (Online)
repository_id_str
spelling Radial dilation of nephrostomy balloons: a comparative analysisballoon dilationpercutaneous nephrolithotomyrenal calculiPURPOSE: The dynamics of percutaneous balloon expansion may differ with increasing extrinsic compressive forces and increasing inflation pressures. This study compares the ability of percutaneous balloons to expand under different radial constrictive forces. MATERIALS AND METHODS: Three 30F nephrostomy balloons were tested: Bard X-Force™, Boston Scientific Microvasive Amplatz Tractmaster™, and Cook Ultraxx™. With a super stiff guidewire in place, the balloon tip was secured by elevated vice grips on either side of the balloon. A string was wrapped around the balloon center once, and incremental increases in load were added (2g, 42g, 82g, and 122g) to represent increasing extrinsic compression. The balloon was inflated with a contrast agent and circumference changes were measured at increments of 4 ATM, 10 ATM, and burst pressure. Balloons were tested in triplicate for each load. RESULTS: All balloons were unable to reach 90% of their expected diameter with larger constrictive loads (122g) at low (4 ATM) and nominal (10 ATM) inflation pressures. Only the Bard and Cook balloons reached at least 90% of the expected diameter with a coefficient of variance (CV) less than 10% at burst pressure under the larger constrictive load (122g), 94.3% ± 6.7%, CV 7.1% and 96.3% ± 2.9%, CV 3.0% respectively. All balloons performed well under low constriction forces and reached at least 80% of the expected diameter by 10 ATM under all constrictive loads. CONCLUSIONS: The Bard X-Force and Cook Ultraxx percutaneous nephrostomy balloons achieved the most reliable radial dilation against large constrictive forces simulating fascial or retroperitoneal scar tissue.Sociedade Brasileira de Urologia2008-10-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S1677-55382008000500002International braz j urol v.34 n.5 2008reponame:International Braz J Urol (Online)instname:Sociedade Brasileira de Urologia (SBU)instacron:SBU10.1590/S1677-55382008000500002info:eu-repo/semantics/openAccessHendlin,KariMonga,Manojeng2008-12-11T00:00:00Zoai:scielo:S1677-55382008000500002Revistahttp://www.brazjurol.com.br/ONGhttps://old.scielo.br/oai/scielo-oai.php||brazjurol@brazjurol.com.br1677-61191677-5538opendoar:2008-12-11T00:00International Braz J Urol (Online) - Sociedade Brasileira de Urologia (SBU)false
dc.title.none.fl_str_mv Radial dilation of nephrostomy balloons: a comparative analysis
title Radial dilation of nephrostomy balloons: a comparative analysis
spellingShingle Radial dilation of nephrostomy balloons: a comparative analysis
Hendlin,Kari
balloon dilation
percutaneous nephrolithotomy
renal calculi
title_short Radial dilation of nephrostomy balloons: a comparative analysis
title_full Radial dilation of nephrostomy balloons: a comparative analysis
title_fullStr Radial dilation of nephrostomy balloons: a comparative analysis
title_full_unstemmed Radial dilation of nephrostomy balloons: a comparative analysis
title_sort Radial dilation of nephrostomy balloons: a comparative analysis
author Hendlin,Kari
author_facet Hendlin,Kari
Monga,Manoj
author_role author
author2 Monga,Manoj
author2_role author
dc.contributor.author.fl_str_mv Hendlin,Kari
Monga,Manoj
dc.subject.por.fl_str_mv balloon dilation
percutaneous nephrolithotomy
renal calculi
topic balloon dilation
percutaneous nephrolithotomy
renal calculi
description PURPOSE: The dynamics of percutaneous balloon expansion may differ with increasing extrinsic compressive forces and increasing inflation pressures. This study compares the ability of percutaneous balloons to expand under different radial constrictive forces. MATERIALS AND METHODS: Three 30F nephrostomy balloons were tested: Bard X-Force™, Boston Scientific Microvasive Amplatz Tractmaster™, and Cook Ultraxx™. With a super stiff guidewire in place, the balloon tip was secured by elevated vice grips on either side of the balloon. A string was wrapped around the balloon center once, and incremental increases in load were added (2g, 42g, 82g, and 122g) to represent increasing extrinsic compression. The balloon was inflated with a contrast agent and circumference changes were measured at increments of 4 ATM, 10 ATM, and burst pressure. Balloons were tested in triplicate for each load. RESULTS: All balloons were unable to reach 90% of their expected diameter with larger constrictive loads (122g) at low (4 ATM) and nominal (10 ATM) inflation pressures. Only the Bard and Cook balloons reached at least 90% of the expected diameter with a coefficient of variance (CV) less than 10% at burst pressure under the larger constrictive load (122g), 94.3% ± 6.7%, CV 7.1% and 96.3% ± 2.9%, CV 3.0% respectively. All balloons performed well under low constriction forces and reached at least 80% of the expected diameter by 10 ATM under all constrictive loads. CONCLUSIONS: The Bard X-Force and Cook Ultraxx percutaneous nephrostomy balloons achieved the most reliable radial dilation against large constrictive forces simulating fascial or retroperitoneal scar tissue.
publishDate 2008
dc.date.none.fl_str_mv 2008-10-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1677-55382008000500002
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1677-55382008000500002
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/S1677-55382008000500002
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Sociedade Brasileira de Urologia
publisher.none.fl_str_mv Sociedade Brasileira de Urologia
dc.source.none.fl_str_mv International braz j urol v.34 n.5 2008
reponame:International Braz J Urol (Online)
instname:Sociedade Brasileira de Urologia (SBU)
instacron:SBU
instname_str Sociedade Brasileira de Urologia (SBU)
instacron_str SBU
institution SBU
reponame_str International Braz J Urol (Online)
collection International Braz J Urol (Online)
repository.name.fl_str_mv International Braz J Urol (Online) - Sociedade Brasileira de Urologia (SBU)
repository.mail.fl_str_mv ||brazjurol@brazjurol.com.br
_version_ 1750318070892593152