Ex vivo and in vivo study of Kowa HA-2 applanation tonometer in the measurement of intraocular pressure in dogs

Detalhes bibliográficos
Autor(a) principal: Andrade, Silvia Franco
Data de Publicação: 2016
Outros Autores: Andrade, Alexandre Lima de, Kanashiro, Glaucia Prada, Silva, Danielle Alves, Barbero, Rafael Cabral, Kuhn, Jacqueline Marcussi Pereira, Silva, Andressa Caroline da
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Semina. Ciências Agrárias (Online)
Texto Completo: https://ojs.uel.br/revistas/uel/index.php/semagrarias/article/view/20434
Resumo: The objective of this study was to evaluate the use of the Kowa HA-2 applanation tonometer in measuring intraocular pressure (IOP) in dogs. Twenty eyes were used in an ex vivo study in which the calibration curve for manometry vs. tonometry was determined by artificially raising the IOP in 5 mmHg increments up to 60 mmHg (10-60 mmHg). Both eyes of 10 anesthetized dogs were studied in vivo to compare manometry vs. tonometry. In the ambulatory study, 168 healthy eyes, 74 eyes with glaucoma and 60 eyes with uveitis were evaluated by tonometry, which was performed with topical anesthesia and 1% fluorescein eye drops for the formation of fluorescein semicircles. The ex vivo study showed an excellent correlation coefficient (r2= 0.993) between the aneroid manometer and the Kowa HA-2 tonometer. In the in vivo study, there was no significant difference (P>0.05) between the IOP values by manometry and tonometry, showing the excellent accuracy of the Kowa HA-2 tonometer. In the ambulatory study using the Kowa HA-2 tonometer, the IOP values (mean±SD, in mmHg) were 15.1±1.8 (12.0 – 20.0) for the healthy eyes, 25.2±4.0 (20.0 – 38.0) for glaucomatous eyes and 10.1±2.3 (5.0 – 13.7) for eyes with uveitis. There was a strong correlation between the IOP values obtained by direct ocular manometry and those from the Kowa HA-2 tonometer. In the ambulatory study, the IOP values measured by the tonometer were compatible for healthy eyes and for eyes with glaucoma or uveitis. We conclude that Kowa HA-2 applanation tonometer is accurate and practical for IOP measurement in dogs.
id UEL-11_2464188414ee4d3ce1de2ca99b0b27ba
oai_identifier_str oai:ojs.pkp.sfu.ca:article/20434
network_acronym_str UEL-11
network_name_str Semina. Ciências Agrárias (Online)
repository_id_str
spelling Ex vivo and in vivo study of Kowa HA-2 applanation tonometer in the measurement of intraocular pressure in dogsEstudo ex vivo e in vivo do uso do tonômetro de aplanação Kowa HA-2 na medição da pressão intraocular em cãesApplanation tonometerDogsGlaucomaKowa ha-2 tonometerOcular manometryUveitis.CãesGlaucomaKowa ha-2Manometria ocularTonômetro de aplanaçãoUveíte.The objective of this study was to evaluate the use of the Kowa HA-2 applanation tonometer in measuring intraocular pressure (IOP) in dogs. Twenty eyes were used in an ex vivo study in which the calibration curve for manometry vs. tonometry was determined by artificially raising the IOP in 5 mmHg increments up to 60 mmHg (10-60 mmHg). Both eyes of 10 anesthetized dogs were studied in vivo to compare manometry vs. tonometry. In the ambulatory study, 168 healthy eyes, 74 eyes with glaucoma and 60 eyes with uveitis were evaluated by tonometry, which was performed with topical anesthesia and 1% fluorescein eye drops for the formation of fluorescein semicircles. The ex vivo study showed an excellent correlation coefficient (r2= 0.993) between the aneroid manometer and the Kowa HA-2 tonometer. In the in vivo study, there was no significant difference (P>0.05) between the IOP values by manometry and tonometry, showing the excellent accuracy of the Kowa HA-2 tonometer. In the ambulatory study using the Kowa HA-2 tonometer, the IOP values (mean±SD, in mmHg) were 15.1±1.8 (12.0 – 20.0) for the healthy eyes, 25.2±4.0 (20.0 – 38.0) for glaucomatous eyes and 10.1±2.3 (5.0 – 13.7) for eyes with uveitis. There was a strong correlation between the IOP values obtained by direct ocular manometry and those from the Kowa HA-2 tonometer. In the ambulatory study, the IOP values measured by the tonometer were compatible for healthy eyes and for eyes with glaucoma or uveitis. We conclude that Kowa HA-2 applanation tonometer is accurate and practical for IOP measurement in dogs.O objetivo deste estudo foi avaliar o uso do tonômetro de aplanação Kowa HA-2 na medição da pressão intraocular (PIO) em cães. Vinte olhos foram utilizados no estudo ex vivo onde a curva de calibração para a manometria versus tonometria foi determinada aumentando artificialmente a PIO em 5 mmHg até 60 mmHg (10 – 60 mmHg). Ambos os olhos de 10 cães anestesiados foram estudados no estudo in vivo para comparação da manometria versus tonometria. No estudo ambulatorial, 168 olhos saudáveis, 74 olhos com glaucoma e 60 olhos com uveíte foram avaliados pela tonometria, que foi realizada com anestesia tópica e colírio de fluoresceína 1% para a formação de semicírculos de fluoresceína. O estudo ex vivo mostrou um excelente coeficiente de correlação (r2= 0,993) entre o manômetro aneroide e o tonômetro Kowa HA-2. No estudo in vivo, não houve diferença significativa (P>0,05) entre os valores de PIO da manometria e os da tonometria demonstrando uma excelente acurácia. No estudo ambulatorial, utilizando o tonômetro Kowa HA-2, os valores de PIO (média±desvio padrão, em mmHg) foram para os olhos sadios 15,1±1,8 (12,0 – 20,0), para os olhos com glaucoma 25,2±4,0 (20,0 – 38,0), e para os olhos com uveíte,10,1±2,3 (5,0 – 13,7). Houve uma forte correlação dos valores de PIO obtidos pela manometria ocular direta com os obtidos com o tonômetro Kowa HA-2. No estudo ambulatorial, os valores mensurados pelo tonômetro foram compatíveis para olhos saudáveis e com glaucoma e uveíte. Nós concluímos que o tonômetro de aplanação Kowa HA-2 é acurado e prático para a mensuração da PIO em cães.UEL2016-04-26info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionPesquisa Científicaapplication/pdfhttps://ojs.uel.br/revistas/uel/index.php/semagrarias/article/view/2043410.5433/1679-0359.2016v37n2p797Semina: Ciências Agrárias; Vol. 37 No. 2 (2016); 797-806Semina: Ciências Agrárias; v. 37 n. 2 (2016); 797-8061679-03591676-546Xreponame:Semina. Ciências Agrárias (Online)instname:Universidade Estadual de Londrina (UEL)instacron:UELenghttps://ojs.uel.br/revistas/uel/index.php/semagrarias/article/view/20434/18598http://creativecommons.org/licenses/by-nc/4.0info:eu-repo/semantics/openAccessAndrade, Silvia FrancoAndrade, Alexandre Lima deKanashiro, Glaucia PradaSilva, Danielle AlvesBarbero, Rafael CabralKuhn, Jacqueline Marcussi PereiraSilva, Andressa Caroline da2022-12-01T16:21:18Zoai:ojs.pkp.sfu.ca:article/20434Revistahttp://www.uel.br/revistas/uel/index.php/semagrariasPUBhttps://ojs.uel.br/revistas/uel/index.php/semagrarias/oaisemina.agrarias@uel.br1679-03591676-546Xopendoar:2022-12-01T16:21:18Semina. Ciências Agrárias (Online) - Universidade Estadual de Londrina (UEL)false
dc.title.none.fl_str_mv Ex vivo and in vivo study of Kowa HA-2 applanation tonometer in the measurement of intraocular pressure in dogs
Estudo ex vivo e in vivo do uso do tonômetro de aplanação Kowa HA-2 na medição da pressão intraocular em cães
title Ex vivo and in vivo study of Kowa HA-2 applanation tonometer in the measurement of intraocular pressure in dogs
spellingShingle Ex vivo and in vivo study of Kowa HA-2 applanation tonometer in the measurement of intraocular pressure in dogs
Andrade, Silvia Franco
Applanation tonometer
Dogs
Glaucoma
Kowa ha-2 tonometer
Ocular manometry
Uveitis.
Cães
Glaucoma
Kowa ha-2
Manometria ocular
Tonômetro de aplanação
Uveíte.
title_short Ex vivo and in vivo study of Kowa HA-2 applanation tonometer in the measurement of intraocular pressure in dogs
title_full Ex vivo and in vivo study of Kowa HA-2 applanation tonometer in the measurement of intraocular pressure in dogs
title_fullStr Ex vivo and in vivo study of Kowa HA-2 applanation tonometer in the measurement of intraocular pressure in dogs
title_full_unstemmed Ex vivo and in vivo study of Kowa HA-2 applanation tonometer in the measurement of intraocular pressure in dogs
title_sort Ex vivo and in vivo study of Kowa HA-2 applanation tonometer in the measurement of intraocular pressure in dogs
author Andrade, Silvia Franco
author_facet Andrade, Silvia Franco
Andrade, Alexandre Lima de
Kanashiro, Glaucia Prada
Silva, Danielle Alves
Barbero, Rafael Cabral
Kuhn, Jacqueline Marcussi Pereira
Silva, Andressa Caroline da
author_role author
author2 Andrade, Alexandre Lima de
Kanashiro, Glaucia Prada
Silva, Danielle Alves
Barbero, Rafael Cabral
Kuhn, Jacqueline Marcussi Pereira
Silva, Andressa Caroline da
author2_role author
author
author
author
author
author
dc.contributor.author.fl_str_mv Andrade, Silvia Franco
Andrade, Alexandre Lima de
Kanashiro, Glaucia Prada
Silva, Danielle Alves
Barbero, Rafael Cabral
Kuhn, Jacqueline Marcussi Pereira
Silva, Andressa Caroline da
dc.subject.por.fl_str_mv Applanation tonometer
Dogs
Glaucoma
Kowa ha-2 tonometer
Ocular manometry
Uveitis.
Cães
Glaucoma
Kowa ha-2
Manometria ocular
Tonômetro de aplanação
Uveíte.
topic Applanation tonometer
Dogs
Glaucoma
Kowa ha-2 tonometer
Ocular manometry
Uveitis.
Cães
Glaucoma
Kowa ha-2
Manometria ocular
Tonômetro de aplanação
Uveíte.
description The objective of this study was to evaluate the use of the Kowa HA-2 applanation tonometer in measuring intraocular pressure (IOP) in dogs. Twenty eyes were used in an ex vivo study in which the calibration curve for manometry vs. tonometry was determined by artificially raising the IOP in 5 mmHg increments up to 60 mmHg (10-60 mmHg). Both eyes of 10 anesthetized dogs were studied in vivo to compare manometry vs. tonometry. In the ambulatory study, 168 healthy eyes, 74 eyes with glaucoma and 60 eyes with uveitis were evaluated by tonometry, which was performed with topical anesthesia and 1% fluorescein eye drops for the formation of fluorescein semicircles. The ex vivo study showed an excellent correlation coefficient (r2= 0.993) between the aneroid manometer and the Kowa HA-2 tonometer. In the in vivo study, there was no significant difference (P>0.05) between the IOP values by manometry and tonometry, showing the excellent accuracy of the Kowa HA-2 tonometer. In the ambulatory study using the Kowa HA-2 tonometer, the IOP values (mean±SD, in mmHg) were 15.1±1.8 (12.0 – 20.0) for the healthy eyes, 25.2±4.0 (20.0 – 38.0) for glaucomatous eyes and 10.1±2.3 (5.0 – 13.7) for eyes with uveitis. There was a strong correlation between the IOP values obtained by direct ocular manometry and those from the Kowa HA-2 tonometer. In the ambulatory study, the IOP values measured by the tonometer were compatible for healthy eyes and for eyes with glaucoma or uveitis. We conclude that Kowa HA-2 applanation tonometer is accurate and practical for IOP measurement in dogs.
publishDate 2016
dc.date.none.fl_str_mv 2016-04-26
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
Pesquisa Científica
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv https://ojs.uel.br/revistas/uel/index.php/semagrarias/article/view/20434
10.5433/1679-0359.2016v37n2p797
url https://ojs.uel.br/revistas/uel/index.php/semagrarias/article/view/20434
identifier_str_mv 10.5433/1679-0359.2016v37n2p797
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv https://ojs.uel.br/revistas/uel/index.php/semagrarias/article/view/20434/18598
dc.rights.driver.fl_str_mv http://creativecommons.org/licenses/by-nc/4.0
info:eu-repo/semantics/openAccess
rights_invalid_str_mv http://creativecommons.org/licenses/by-nc/4.0
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv UEL
publisher.none.fl_str_mv UEL
dc.source.none.fl_str_mv Semina: Ciências Agrárias; Vol. 37 No. 2 (2016); 797-806
Semina: Ciências Agrárias; v. 37 n. 2 (2016); 797-806
1679-0359
1676-546X
reponame:Semina. Ciências Agrárias (Online)
instname:Universidade Estadual de Londrina (UEL)
instacron:UEL
instname_str Universidade Estadual de Londrina (UEL)
instacron_str UEL
institution UEL
reponame_str Semina. Ciências Agrárias (Online)
collection Semina. Ciências Agrárias (Online)
repository.name.fl_str_mv Semina. Ciências Agrárias (Online) - Universidade Estadual de Londrina (UEL)
repository.mail.fl_str_mv semina.agrarias@uel.br
_version_ 1799306055957610496