Measurement of Peri-Implant Bone Width with and without Metal Artifact Reduction Algorithm Using Two Cone-Beam Computed Tomography Software Programs

Detalhes bibliográficos
Autor(a) principal: Hedayatipanah, Morad
Data de Publicação: 2021
Outros Autores: Salemi, Fatemeh, Kamyari, Naser, Yalpanian, Ashkan
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Pesquisa Brasileira em Odontopediatria e Clínica Integrada (Online)
Texto Completo: https://revista.uepb.edu.br/PBOCI/article/view/741
Resumo: Objective: To assess the effect of metal artifact reduction (MAR) algorithm of two cone-beam computed tomography (CBCT) software programs on the accuracy of peri-implant bone width measurements. Material and Methods: This in vitro study evaluated 35 bovine rib bone blocks measuring 8 x 8 x 11 mm. Titanium implants were inserted in bone blocks and placed in a wax model of mandible. CBCT scans were obtained with ProMax 3D and Cranex 3D CBCT systems with and without the MAR algorithm. The width of buccal and lingual bone plates surrounding the implant was measured on CBCT scans by two observers. The Cronbach’s alpha was calculated to assess inter-observer agreement. The area under the curve (AUC), sensitivity and specificity values were calculated and compared. Results: The two observers had an excellent agreement. The accuracy of Cranex 3D was higher than that of ProMax 3D (p<0.05). Both CBCT systems showed higher accuracy when the MAR algorithm was not used (p<0.05). Both CBCT systems showed higher accuracy for measuring the width of the lingual plate than buccal plate (p<0.05). Conclusion: The application of MAR algorithm did not notably increase the measurement accuracy in any CBCT system. Cranex 3D showed generally higher measurement accuracy than ProMax 3D.
id UEPB-4_c57c3f20382ff382ac34a6ba53aefc4d
oai_identifier_str oai:ojs.pkp.sfu.ca:article/741
network_acronym_str UEPB-4
network_name_str Pesquisa Brasileira em Odontopediatria e Clínica Integrada (Online)
repository_id_str
spelling Measurement of Peri-Implant Bone Width with and without Metal Artifact Reduction Algorithm Using Two Cone-Beam Computed Tomography Software ProgramsCone-Beam Computed TomographyDental ImplantsArtifactsObjective: To assess the effect of metal artifact reduction (MAR) algorithm of two cone-beam computed tomography (CBCT) software programs on the accuracy of peri-implant bone width measurements. Material and Methods: This in vitro study evaluated 35 bovine rib bone blocks measuring 8 x 8 x 11 mm. Titanium implants were inserted in bone blocks and placed in a wax model of mandible. CBCT scans were obtained with ProMax 3D and Cranex 3D CBCT systems with and without the MAR algorithm. The width of buccal and lingual bone plates surrounding the implant was measured on CBCT scans by two observers. The Cronbach’s alpha was calculated to assess inter-observer agreement. The area under the curve (AUC), sensitivity and specificity values were calculated and compared. Results: The two observers had an excellent agreement. The accuracy of Cranex 3D was higher than that of ProMax 3D (p<0.05). Both CBCT systems showed higher accuracy when the MAR algorithm was not used (p<0.05). Both CBCT systems showed higher accuracy for measuring the width of the lingual plate than buccal plate (p<0.05). Conclusion: The application of MAR algorithm did not notably increase the measurement accuracy in any CBCT system. Cranex 3D showed generally higher measurement accuracy than ProMax 3D.EDUEPB - EDITORA DA UNIVERSIDADE ESTADUAL DA PARAÍBA2021-11-16info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionapplication/pdfhttps://revista.uepb.edu.br/PBOCI/article/view/741Pesquisa Brasileira em Odontopediatria e Clínica Integrada; Vol. 21 (2021); e0050Pesquisa Brasileira em Odontopediatria e Clínica Integrada; v. 21 (2021); e00501983-46321519-0501reponame:Pesquisa Brasileira em Odontopediatria e Clínica Integrada (Online)instname:Universidade Estadual da Paraíba (UEPB)instacron:UEPBenghttps://revista.uepb.edu.br/PBOCI/article/view/741/428Copyright (c) 2021 Pesquisa Brasileira em Odontopediatria e Clínica Integradahttps://creativecommons.org/licenses/by-nc/4.0info:eu-repo/semantics/openAccessHedayatipanah, Morad Salemi, Fatemeh Kamyari, NaserYalpanian, Ashkan 2021-11-16T15:18:24Zoai:ojs.pkp.sfu.ca:article/741Revistahttps://revista.uepb.edu.br/PBOCI/PUBhttps://revista.uepb.edu.br/PBOCI/oaipboci.editoria@gmail.com || alessandrouepb@gmail.com1983-46321519-0501opendoar:2021-11-16T15:18:24Pesquisa Brasileira em Odontopediatria e Clínica Integrada (Online) - Universidade Estadual da Paraíba (UEPB)false
dc.title.none.fl_str_mv Measurement of Peri-Implant Bone Width with and without Metal Artifact Reduction Algorithm Using Two Cone-Beam Computed Tomography Software Programs
title Measurement of Peri-Implant Bone Width with and without Metal Artifact Reduction Algorithm Using Two Cone-Beam Computed Tomography Software Programs
spellingShingle Measurement of Peri-Implant Bone Width with and without Metal Artifact Reduction Algorithm Using Two Cone-Beam Computed Tomography Software Programs
Hedayatipanah, Morad
Cone-Beam Computed Tomography
Dental Implants
Artifacts
title_short Measurement of Peri-Implant Bone Width with and without Metal Artifact Reduction Algorithm Using Two Cone-Beam Computed Tomography Software Programs
title_full Measurement of Peri-Implant Bone Width with and without Metal Artifact Reduction Algorithm Using Two Cone-Beam Computed Tomography Software Programs
title_fullStr Measurement of Peri-Implant Bone Width with and without Metal Artifact Reduction Algorithm Using Two Cone-Beam Computed Tomography Software Programs
title_full_unstemmed Measurement of Peri-Implant Bone Width with and without Metal Artifact Reduction Algorithm Using Two Cone-Beam Computed Tomography Software Programs
title_sort Measurement of Peri-Implant Bone Width with and without Metal Artifact Reduction Algorithm Using Two Cone-Beam Computed Tomography Software Programs
author Hedayatipanah, Morad
author_facet Hedayatipanah, Morad
Salemi, Fatemeh
Kamyari, Naser
Yalpanian, Ashkan
author_role author
author2 Salemi, Fatemeh
Kamyari, Naser
Yalpanian, Ashkan
author2_role author
author
author
dc.contributor.author.fl_str_mv Hedayatipanah, Morad
Salemi, Fatemeh
Kamyari, Naser
Yalpanian, Ashkan
dc.subject.por.fl_str_mv Cone-Beam Computed Tomography
Dental Implants
Artifacts
topic Cone-Beam Computed Tomography
Dental Implants
Artifacts
description Objective: To assess the effect of metal artifact reduction (MAR) algorithm of two cone-beam computed tomography (CBCT) software programs on the accuracy of peri-implant bone width measurements. Material and Methods: This in vitro study evaluated 35 bovine rib bone blocks measuring 8 x 8 x 11 mm. Titanium implants were inserted in bone blocks and placed in a wax model of mandible. CBCT scans were obtained with ProMax 3D and Cranex 3D CBCT systems with and without the MAR algorithm. The width of buccal and lingual bone plates surrounding the implant was measured on CBCT scans by two observers. The Cronbach’s alpha was calculated to assess inter-observer agreement. The area under the curve (AUC), sensitivity and specificity values were calculated and compared. Results: The two observers had an excellent agreement. The accuracy of Cranex 3D was higher than that of ProMax 3D (p<0.05). Both CBCT systems showed higher accuracy when the MAR algorithm was not used (p<0.05). Both CBCT systems showed higher accuracy for measuring the width of the lingual plate than buccal plate (p<0.05). Conclusion: The application of MAR algorithm did not notably increase the measurement accuracy in any CBCT system. Cranex 3D showed generally higher measurement accuracy than ProMax 3D.
publishDate 2021
dc.date.none.fl_str_mv 2021-11-16
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv https://revista.uepb.edu.br/PBOCI/article/view/741
url https://revista.uepb.edu.br/PBOCI/article/view/741
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv https://revista.uepb.edu.br/PBOCI/article/view/741/428
dc.rights.driver.fl_str_mv Copyright (c) 2021 Pesquisa Brasileira em Odontopediatria e Clínica Integrada
https://creativecommons.org/licenses/by-nc/4.0
info:eu-repo/semantics/openAccess
rights_invalid_str_mv Copyright (c) 2021 Pesquisa Brasileira em Odontopediatria e Clínica Integrada
https://creativecommons.org/licenses/by-nc/4.0
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv EDUEPB - EDITORA DA UNIVERSIDADE ESTADUAL DA PARAÍBA
publisher.none.fl_str_mv EDUEPB - EDITORA DA UNIVERSIDADE ESTADUAL DA PARAÍBA
dc.source.none.fl_str_mv Pesquisa Brasileira em Odontopediatria e Clínica Integrada; Vol. 21 (2021); e0050
Pesquisa Brasileira em Odontopediatria e Clínica Integrada; v. 21 (2021); e0050
1983-4632
1519-0501
reponame:Pesquisa Brasileira em Odontopediatria e Clínica Integrada (Online)
instname:Universidade Estadual da Paraíba (UEPB)
instacron:UEPB
instname_str Universidade Estadual da Paraíba (UEPB)
instacron_str UEPB
institution UEPB
reponame_str Pesquisa Brasileira em Odontopediatria e Clínica Integrada (Online)
collection Pesquisa Brasileira em Odontopediatria e Clínica Integrada (Online)
repository.name.fl_str_mv Pesquisa Brasileira em Odontopediatria e Clínica Integrada (Online) - Universidade Estadual da Paraíba (UEPB)
repository.mail.fl_str_mv pboci.editoria@gmail.com || alessandrouepb@gmail.com
_version_ 1803387862799876096