Air kerma area product in cone beam computed tomography

Detalhes bibliográficos
Autor(a) principal: Oliveira, Marcus Vinicius Linhares de
Data de Publicação: 2014
Outros Autores: Batista, Wilson Otto Gomes, Soares, Maria Rosângela, Campos, Paulo Sergio Flores
Tipo de documento: Artigo
Idioma: por
Título da fonte: Repositório Institucional da UFBA
Texto Completo: http://repositorio.ufba.br/ri/handle/ri/23181
Resumo: Abstract Objective: to evaluate the influence of FOV in air kerma-area product (KAP) and the constancy of exposure parameters on cone beam computed tomography equipments. Methodology: Two cone beam CT (GENDEX CBX 500 and i-CAT Classic) were used and seven exposures with the following FOVs were performed: (A) 14cm x 8,5cm, (B) 14cm x 6cm, (C) 8,5cm x 8.5cm e (D) 8,5cm x 6cm, for CBX 500; and (E) 14cm x 6.cm, (F) 14cm x 8cm e (G) 14cm x 13cm, for the i-CAT. The technical exposure factors (kV, mA, mAs and voxel), remained constant. The dosimetric evaluation was performed with air KAP meter manufactured by IBA dosimetry, model kerma X plus TinO, positioned at the output of the X-ray beam. To evaluate the constancy of the exposure parameters a semiconductor (Radcal, Rapidose) fixed in front of the tomography image receptor was used. Result: The KAP values obtained ranged between 360.1 mGy.cm² and 1031.2 mGy.cm². The FOV height had a substantial influence on the radiation dose. Repeatability and accuracy of the tube voltage varied less than 10%. Conclusion: The radiation dose is directly related to the height and inversely related with the FOV diameter; even within the recommended limits, the percentage variation of repeatability and accuracy of kV, for the tomography equipments tested, points to the regular equipment calibration, in order to reduce radiation dose to the patient to a minimum. Keywords: Cone beam computed tomography. Radiation. Dosimetry.
id UFBA-2_0a1f75f7bacf49255772e93f531184d5
oai_identifier_str oai:repositorio.ufba.br:ri/23181
network_acronym_str UFBA-2
network_name_str Repositório Institucional da UFBA
repository_id_str 1932
spelling Oliveira, Marcus Vinicius Linhares deBatista, Wilson Otto GomesSoares, Maria RosângelaCampos, Paulo Sergio Flores2017-06-20T13:30:04Z2017-06-20T13:30:04Z2014-09Rev. Ciênc. Méd. Biol., Salvador, v. 13, n. 3 – especial, p. 309-312, set./dez. 2014.2236-5222http://repositorio.ufba.br/ri/handle/ri/23181v.13, n.3Abstract Objective: to evaluate the influence of FOV in air kerma-area product (KAP) and the constancy of exposure parameters on cone beam computed tomography equipments. Methodology: Two cone beam CT (GENDEX CBX 500 and i-CAT Classic) were used and seven exposures with the following FOVs were performed: (A) 14cm x 8,5cm, (B) 14cm x 6cm, (C) 8,5cm x 8.5cm e (D) 8,5cm x 6cm, for CBX 500; and (E) 14cm x 6.cm, (F) 14cm x 8cm e (G) 14cm x 13cm, for the i-CAT. The technical exposure factors (kV, mA, mAs and voxel), remained constant. The dosimetric evaluation was performed with air KAP meter manufactured by IBA dosimetry, model kerma X plus TinO, positioned at the output of the X-ray beam. To evaluate the constancy of the exposure parameters a semiconductor (Radcal, Rapidose) fixed in front of the tomography image receptor was used. Result: The KAP values obtained ranged between 360.1 mGy.cm² and 1031.2 mGy.cm². The FOV height had a substantial influence on the radiation dose. Repeatability and accuracy of the tube voltage varied less than 10%. Conclusion: The radiation dose is directly related to the height and inversely related with the FOV diameter; even within the recommended limits, the percentage variation of repeatability and accuracy of kV, for the tomography equipments tested, points to the regular equipment calibration, in order to reduce radiation dose to the patient to a minimum. Keywords: Cone beam computed tomography. Radiation. Dosimetry.Objetivo: avaliar a influência do FOV no produto kerma-área (PKA) no ar e a constância dos parâmetros de exposição em equipamentos de tomografia computadorizada feixe cônico. Metodologia: Foram utilizados dois tomógrafos de feixe cônico (GENDEX CBX 500 e i-CAT Classic) e foram realizadas sete exposições com os seguintes FOVs: (A) 14cm x 8,5cm, (B) 14cm x 6cm, (C) 8,5cm x 8.5cm e (D) 8,5cm x 6cm, para o CBX 500; e (E) 14cm x 6.cm, (F) 14cm x 8cm e (G) 14cm x 13cm, para o i-CAT. Os fatores técnicos de exposição (kV, mA, mAs e voxel), permaneceram constantes. A avaliação dosimétrica foi realizada com um medidor do produto kerma-área no ar fabricado pela IBA dosimetry, modelo kerma X plus TinO, posicionado na saída do feixe de raios X. Para a avaliação da constância dos parâmetros de exposição foi utilizado um multimedidor do tipo semicondutor, Radcal, modelo Rapidose fixado na entrada do receptor de imagem do tomógrafo. Resultados: Os valores de PKA obtidos variaram entre 360,1 mGy.cm² e 1.031,2 mGy.cm². A altura do FOV teve influência substantiva na dose de radiação. A repetibilidade e exatidão da tensão do tubo variaram abaixo dos 10%. Conclusões: a dose de radiação tem relação direta com a altura e inversa com o diâmetro do FOV; embora dentro das margens recomendadas, a variação dos percentuais de repetibilidade e exatidão do kV, para os tomógrafos testados, aponta para a calibragem periódica dos equipamentos, com vistas a reduzir ao mínimo necessário a dose de radiação para o paciente.Submitted by ROBERTO PAULO CORREIA DE ARAÚJO (ppgorgsistem@ufba.br) on 2017-06-20T13:30:04Z No. of bitstreams: 1 8_v.13_3.pdf: 667079 bytes, checksum: 4bb9cc9e39591f313692aa5eb8347146 (MD5)Made available in DSpace on 2017-06-20T13:30:04Z (GMT). No. of bitstreams: 1 8_v.13_3.pdf: 667079 bytes, checksum: 4bb9cc9e39591f313692aa5eb8347146 (MD5) Previous issue date: 2014-09SalvadorInstituto de Ciências da Saúde/ Universidade Federal da BahiaBrasilhttp://www.portalseer.ufba.br/index.php/cmbio/article/view/12936/9339reponame:Repositório Institucional da UFBAinstname:Universidade Federal da Bahia (UFBA)instacron:UFBATomografia Computadorizada de Feixe Cônico.Radiação.Dosimetria.Air kerma area product in cone beam computed tomographyRevista de Ciências Médicas e Biológicasinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articleinfo:eu-repo/semantics/openAccessporORIGINAL8_v.13_3.pdf8_v.13_3.pdfapplication/pdf667079https://repositorio.ufba.br/bitstream/ri/23181/1/8_v.13_3.pdf4bb9cc9e39591f313692aa5eb8347146MD51LICENSElicense.txtlicense.txttext/plain1345https://repositorio.ufba.br/bitstream/ri/23181/2/license.txtff6eaa8b858ea317fded99f125f5fcd0MD52TEXT8_v.13_3.pdf.txt8_v.13_3.pdf.txtExtracted texttext/plain15238https://repositorio.ufba.br/bitstream/ri/23181/3/8_v.13_3.pdf.txtf57208fc30f418ec9b877ac8ff2aca7dMD53ri/231812022-03-10 14:40:48.813oai:repositorio.ufba.br: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Repositório InstitucionalPUBhttp://192.188.11.11:8080/oai/requestopendoar:19322022-03-10T17:40:48Repositório Institucional da UFBA - Universidade Federal da Bahia (UFBA)false
dc.title.pt_BR.fl_str_mv Air kerma area product in cone beam computed tomography
dc.title.alternative.pt_BR.fl_str_mv Revista de Ciências Médicas e Biológicas
title Air kerma area product in cone beam computed tomography
spellingShingle Air kerma area product in cone beam computed tomography
Oliveira, Marcus Vinicius Linhares de
Tomografia Computadorizada de Feixe Cônico.
Radiação.
Dosimetria.
title_short Air kerma area product in cone beam computed tomography
title_full Air kerma area product in cone beam computed tomography
title_fullStr Air kerma area product in cone beam computed tomography
title_full_unstemmed Air kerma area product in cone beam computed tomography
title_sort Air kerma area product in cone beam computed tomography
author Oliveira, Marcus Vinicius Linhares de
author_facet Oliveira, Marcus Vinicius Linhares de
Batista, Wilson Otto Gomes
Soares, Maria Rosângela
Campos, Paulo Sergio Flores
author_role author
author2 Batista, Wilson Otto Gomes
Soares, Maria Rosângela
Campos, Paulo Sergio Flores
author2_role author
author
author
dc.contributor.author.fl_str_mv Oliveira, Marcus Vinicius Linhares de
Batista, Wilson Otto Gomes
Soares, Maria Rosângela
Campos, Paulo Sergio Flores
dc.subject.por.fl_str_mv Tomografia Computadorizada de Feixe Cônico.
Radiação.
Dosimetria.
topic Tomografia Computadorizada de Feixe Cônico.
Radiação.
Dosimetria.
description Abstract Objective: to evaluate the influence of FOV in air kerma-area product (KAP) and the constancy of exposure parameters on cone beam computed tomography equipments. Methodology: Two cone beam CT (GENDEX CBX 500 and i-CAT Classic) were used and seven exposures with the following FOVs were performed: (A) 14cm x 8,5cm, (B) 14cm x 6cm, (C) 8,5cm x 8.5cm e (D) 8,5cm x 6cm, for CBX 500; and (E) 14cm x 6.cm, (F) 14cm x 8cm e (G) 14cm x 13cm, for the i-CAT. The technical exposure factors (kV, mA, mAs and voxel), remained constant. The dosimetric evaluation was performed with air KAP meter manufactured by IBA dosimetry, model kerma X plus TinO, positioned at the output of the X-ray beam. To evaluate the constancy of the exposure parameters a semiconductor (Radcal, Rapidose) fixed in front of the tomography image receptor was used. Result: The KAP values obtained ranged between 360.1 mGy.cm² and 1031.2 mGy.cm². The FOV height had a substantial influence on the radiation dose. Repeatability and accuracy of the tube voltage varied less than 10%. Conclusion: The radiation dose is directly related to the height and inversely related with the FOV diameter; even within the recommended limits, the percentage variation of repeatability and accuracy of kV, for the tomography equipments tested, points to the regular equipment calibration, in order to reduce radiation dose to the patient to a minimum. Keywords: Cone beam computed tomography. Radiation. Dosimetry.
publishDate 2014
dc.date.issued.fl_str_mv 2014-09
dc.date.accessioned.fl_str_mv 2017-06-20T13:30:04Z
dc.date.available.fl_str_mv 2017-06-20T13:30:04Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.citation.fl_str_mv Rev. Ciênc. Méd. Biol., Salvador, v. 13, n. 3 – especial, p. 309-312, set./dez. 2014.
dc.identifier.uri.fl_str_mv http://repositorio.ufba.br/ri/handle/ri/23181
dc.identifier.issn.none.fl_str_mv 2236-5222
dc.identifier.number.pt_BR.fl_str_mv v.13, n.3
identifier_str_mv Rev. Ciênc. Méd. Biol., Salvador, v. 13, n. 3 – especial, p. 309-312, set./dez. 2014.
2236-5222
v.13, n.3
url http://repositorio.ufba.br/ri/handle/ri/23181
dc.language.iso.fl_str_mv por
language por
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.publisher.none.fl_str_mv Instituto de Ciências da Saúde/ Universidade Federal da Bahia
dc.publisher.country.fl_str_mv Brasil
publisher.none.fl_str_mv Instituto de Ciências da Saúde/ Universidade Federal da Bahia
dc.source.pt_BR.fl_str_mv http://www.portalseer.ufba.br/index.php/cmbio/article/view/12936/9339
dc.source.none.fl_str_mv reponame:Repositório Institucional da UFBA
instname:Universidade Federal da Bahia (UFBA)
instacron:UFBA
instname_str Universidade Federal da Bahia (UFBA)
instacron_str UFBA
institution UFBA
reponame_str Repositório Institucional da UFBA
collection Repositório Institucional da UFBA
bitstream.url.fl_str_mv https://repositorio.ufba.br/bitstream/ri/23181/1/8_v.13_3.pdf
https://repositorio.ufba.br/bitstream/ri/23181/2/license.txt
https://repositorio.ufba.br/bitstream/ri/23181/3/8_v.13_3.pdf.txt
bitstream.checksum.fl_str_mv 4bb9cc9e39591f313692aa5eb8347146
ff6eaa8b858ea317fded99f125f5fcd0
f57208fc30f418ec9b877ac8ff2aca7d
bitstream.checksumAlgorithm.fl_str_mv MD5
MD5
MD5
repository.name.fl_str_mv Repositório Institucional da UFBA - Universidade Federal da Bahia (UFBA)
repository.mail.fl_str_mv
_version_ 1801502583582359552