Development and validation of a data logger for thermal characterization in laying hen facilities

Detalhes bibliográficos
Autor(a) principal: Freitas,Letícia C. da S. R.
Data de Publicação: 2019
Outros Autores: Tinôco,Ilda de F. F., Gates,Richard S., Barbari,Matteo, Cândido,Márcia G. L., Toledo,João V.
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Revista Brasileira de Engenharia Agrícola e Ambiental (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1415-43662019001000787
Resumo: ABSTRACT The environmental monitoring in animal facilities that includes collected data storage in a robust, practical and feasible way is a constant challenge. The aim of this study was to develop a reliable data logger for monitoring the air temperature and air relative humidity of aviaries and to assess the adequacy of the design using commercially available reference standard instruments. The experimental data logger was installed together with a commercial data logger, a mercury thermometer and a calibrated Vaisala HMP110 air relative humidity probe in a meteorological shelter. Linear regression analysis was performed with the collected air temperature and air relative humidity to develop calibration equations. The Nash-Sutcliffe Index and the relative error were calculated to validate the experimental data logger. The air temperature and the air relative humidity calibration equations presented Nash-Sutcliffe of 0.993 and -0.281 for the commercial data logger, and 0.913 and 0.932 for the experimental data. The mean relative error of the air temperature readings was 3 and 1% and for air relative humidity 5 and 20%, for the experimental and commercial logger, respectively. The experimental data logger reliably stored all collected data without error to the micro-SD card. The experimental data logger can be considered low-cost and sufficiently accurate for monitoring air temperature and air relative humidity in aviaries, presenting field performance very close to the commercial data logger for air temperature measurement, and better performance than the commercial data logger for the measurement of air relative humidity.
id UFCG-1_b7b0dd2f2fbca22cc09e45c2cec08436
oai_identifier_str oai:scielo:S1415-43662019001000787
network_acronym_str UFCG-1
network_name_str Revista Brasileira de Engenharia Agrícola e Ambiental (Online)
repository_id_str
spelling Development and validation of a data logger for thermal characterization in laying hen facilitieselectronic deviceenvironmental monitoringpoultry facilitiesABSTRACT The environmental monitoring in animal facilities that includes collected data storage in a robust, practical and feasible way is a constant challenge. The aim of this study was to develop a reliable data logger for monitoring the air temperature and air relative humidity of aviaries and to assess the adequacy of the design using commercially available reference standard instruments. The experimental data logger was installed together with a commercial data logger, a mercury thermometer and a calibrated Vaisala HMP110 air relative humidity probe in a meteorological shelter. Linear regression analysis was performed with the collected air temperature and air relative humidity to develop calibration equations. The Nash-Sutcliffe Index and the relative error were calculated to validate the experimental data logger. The air temperature and the air relative humidity calibration equations presented Nash-Sutcliffe of 0.993 and -0.281 for the commercial data logger, and 0.913 and 0.932 for the experimental data. The mean relative error of the air temperature readings was 3 and 1% and for air relative humidity 5 and 20%, for the experimental and commercial logger, respectively. The experimental data logger reliably stored all collected data without error to the micro-SD card. The experimental data logger can be considered low-cost and sufficiently accurate for monitoring air temperature and air relative humidity in aviaries, presenting field performance very close to the commercial data logger for air temperature measurement, and better performance than the commercial data logger for the measurement of air relative humidity.Departamento de Engenharia Agrícola - UFCG2019-10-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S1415-43662019001000787Revista Brasileira de Engenharia Agrícola e Ambiental v.23 n.10 2019reponame:Revista Brasileira de Engenharia Agrícola e Ambiental (Online)instname:Universidade Federal de Campina Grande (UFCG)instacron:UFCG10.1590/1807-1929/agriambi.v23n10p787-793info:eu-repo/semantics/openAccessFreitas,Letícia C. da S. R.Tinôco,Ilda de F. F.Gates,Richard S.Barbari,MatteoCândido,Márcia G. L.Toledo,João V.eng2019-09-03T00:00:00Zoai:scielo:S1415-43662019001000787Revistahttp://www.scielo.br/rbeaaPUBhttps://old.scielo.br/oai/scielo-oai.php||agriambi@agriambi.com.br1807-19291415-4366opendoar:2019-09-03T00:00Revista Brasileira de Engenharia Agrícola e Ambiental (Online) - Universidade Federal de Campina Grande (UFCG)false
dc.title.none.fl_str_mv Development and validation of a data logger for thermal characterization in laying hen facilities
title Development and validation of a data logger for thermal characterization in laying hen facilities
spellingShingle Development and validation of a data logger for thermal characterization in laying hen facilities
Freitas,Letícia C. da S. R.
electronic device
environmental monitoring
poultry facilities
title_short Development and validation of a data logger for thermal characterization in laying hen facilities
title_full Development and validation of a data logger for thermal characterization in laying hen facilities
title_fullStr Development and validation of a data logger for thermal characterization in laying hen facilities
title_full_unstemmed Development and validation of a data logger for thermal characterization in laying hen facilities
title_sort Development and validation of a data logger for thermal characterization in laying hen facilities
author Freitas,Letícia C. da S. R.
author_facet Freitas,Letícia C. da S. R.
Tinôco,Ilda de F. F.
Gates,Richard S.
Barbari,Matteo
Cândido,Márcia G. L.
Toledo,João V.
author_role author
author2 Tinôco,Ilda de F. F.
Gates,Richard S.
Barbari,Matteo
Cândido,Márcia G. L.
Toledo,João V.
author2_role author
author
author
author
author
dc.contributor.author.fl_str_mv Freitas,Letícia C. da S. R.
Tinôco,Ilda de F. F.
Gates,Richard S.
Barbari,Matteo
Cândido,Márcia G. L.
Toledo,João V.
dc.subject.por.fl_str_mv electronic device
environmental monitoring
poultry facilities
topic electronic device
environmental monitoring
poultry facilities
description ABSTRACT The environmental monitoring in animal facilities that includes collected data storage in a robust, practical and feasible way is a constant challenge. The aim of this study was to develop a reliable data logger for monitoring the air temperature and air relative humidity of aviaries and to assess the adequacy of the design using commercially available reference standard instruments. The experimental data logger was installed together with a commercial data logger, a mercury thermometer and a calibrated Vaisala HMP110 air relative humidity probe in a meteorological shelter. Linear regression analysis was performed with the collected air temperature and air relative humidity to develop calibration equations. The Nash-Sutcliffe Index and the relative error were calculated to validate the experimental data logger. The air temperature and the air relative humidity calibration equations presented Nash-Sutcliffe of 0.993 and -0.281 for the commercial data logger, and 0.913 and 0.932 for the experimental data. The mean relative error of the air temperature readings was 3 and 1% and for air relative humidity 5 and 20%, for the experimental and commercial logger, respectively. The experimental data logger reliably stored all collected data without error to the micro-SD card. The experimental data logger can be considered low-cost and sufficiently accurate for monitoring air temperature and air relative humidity in aviaries, presenting field performance very close to the commercial data logger for air temperature measurement, and better performance than the commercial data logger for the measurement of air relative humidity.
publishDate 2019
dc.date.none.fl_str_mv 2019-10-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1415-43662019001000787
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1415-43662019001000787
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/1807-1929/agriambi.v23n10p787-793
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Departamento de Engenharia Agrícola - UFCG
publisher.none.fl_str_mv Departamento de Engenharia Agrícola - UFCG
dc.source.none.fl_str_mv Revista Brasileira de Engenharia Agrícola e Ambiental v.23 n.10 2019
reponame:Revista Brasileira de Engenharia Agrícola e Ambiental (Online)
instname:Universidade Federal de Campina Grande (UFCG)
instacron:UFCG
instname_str Universidade Federal de Campina Grande (UFCG)
instacron_str UFCG
institution UFCG
reponame_str Revista Brasileira de Engenharia Agrícola e Ambiental (Online)
collection Revista Brasileira de Engenharia Agrícola e Ambiental (Online)
repository.name.fl_str_mv Revista Brasileira de Engenharia Agrícola e Ambiental (Online) - Universidade Federal de Campina Grande (UFCG)
repository.mail.fl_str_mv ||agriambi@agriambi.com.br
_version_ 1750297686955786240