A sensor for spirometric feedback in ventilation maneuvers during cardiopulmonary resuscitation training.

Detalhes bibliográficos
Autor(a) principal: Leocádio, Rodolfo Rocha Vieira
Data de Publicação: 2019
Outros Autores: Rêgo Segundo, Alan Kardek, Louzada, Cibelle Ferreira
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Institucional da UFOP
Texto Completo: http://www.repositorio.ufop.br/handle/123456789/12711
https://doi.org/10.3390/s19235095
Resumo: This work proposes adapting an existing sensor and embedding it on mannequins used in cardiopulmonary resuscitation (CPR) training to accurately measure the amount of air supplied to the lungs during ventilation. Mathematical modeling, calibration, and validation of the sensor along with metrology, statistical inference, and spirometry techniques were used as a base for aquiring scientific knowledge of the system. The system directly measures the variable of interest (air volume) and refers to spirometric techniques in the elaboration of its model. This improves the realism of the dummies during the CPR training, because it estimates, in real-time, not only the volume of air entering in the lungs but also the Forced Vital Capacity (FVC), Forced Expiratory Volume (FEVt) and Medium Forced Expiratory Flow (FEF20–75%). The validation of the sensor achieved results that address the requirements for this application, that is, the error below 3.4% of full scale. During the spirometric tests, the system presented the measurement results of (305 ± 22, 450 ± 23, 603 ± 24, 751 ± 26, 922 ± 27, 1021 ± 30, 1182 ± 33, 1326 ± 36, 1476 ± 37, 1618 ± 45 and 1786 ± 56) × 10−6 m3 for reference values of (300, 450, 600, 750, 900, 1050, 1200, 1350, 1500, 1650 and 1800) × 10−6 m3 , respectively. Therefore, considering the spirometry and pressure boundary conditions of the manikin lungs, the system achieves the objective of simulating valid spirometric data for debriefings, that is, there is an agreement between the measurement results when compared to the signal generated by a commercial spirometer (Koko brand). The main advantages that this work presents in relation to the sensors commonly used for this purpose are: (i) the reduced cost, which makes it possible, for the first time, to use a respiratory volume sensor in medical simulators or training dummies; (ii) the direct measurement of air entering the lung using a noninvasive method, which makes it possible to use spirometry parameters to characterize simulated human respiration during the CPR training; and (iii) the measurement of spirometric parameters (FVC, FEVt, and FEF20–75%), in real-time, during the CPR training, to achieve optimal ventilation performance. Therefore, the system developed in this work addresses the minimum requirements for the practice of ventilation in the CPR maneuvers and has great potential in several future applications.
id UFOP_510e8307afe4946c6500f7d921d5a17e
oai_identifier_str oai:localhost:123456789/12711
network_acronym_str UFOP
network_name_str Repositório Institucional da UFOP
repository_id_str 3233
spelling Leocádio, Rodolfo Rocha VieiraRêgo Segundo, Alan KardekLouzada, Cibelle Ferreira2020-09-10T16:29:00Z2020-09-10T16:29:00Z2019LEOCÁDIO, R. R. V.; RÊGO SEGUNDO, A. K.; LOUZADA, C. F. A sensor for spirometric feedback in ventilation maneuvers during cardiopulmonary resuscitation training. Sensor, v. 19, n. 23, 2019. Disponível em: <https://www.mdpi.com/1424-8220/19/23/5095>. Acesso em: 10 mar. 2020.1424-8220http://www.repositorio.ufop.br/handle/123456789/12711https://doi.org/10.3390/s19235095This work proposes adapting an existing sensor and embedding it on mannequins used in cardiopulmonary resuscitation (CPR) training to accurately measure the amount of air supplied to the lungs during ventilation. Mathematical modeling, calibration, and validation of the sensor along with metrology, statistical inference, and spirometry techniques were used as a base for aquiring scientific knowledge of the system. The system directly measures the variable of interest (air volume) and refers to spirometric techniques in the elaboration of its model. This improves the realism of the dummies during the CPR training, because it estimates, in real-time, not only the volume of air entering in the lungs but also the Forced Vital Capacity (FVC), Forced Expiratory Volume (FEVt) and Medium Forced Expiratory Flow (FEF20–75%). The validation of the sensor achieved results that address the requirements for this application, that is, the error below 3.4% of full scale. During the spirometric tests, the system presented the measurement results of (305 ± 22, 450 ± 23, 603 ± 24, 751 ± 26, 922 ± 27, 1021 ± 30, 1182 ± 33, 1326 ± 36, 1476 ± 37, 1618 ± 45 and 1786 ± 56) × 10−6 m3 for reference values of (300, 450, 600, 750, 900, 1050, 1200, 1350, 1500, 1650 and 1800) × 10−6 m3 , respectively. Therefore, considering the spirometry and pressure boundary conditions of the manikin lungs, the system achieves the objective of simulating valid spirometric data for debriefings, that is, there is an agreement between the measurement results when compared to the signal generated by a commercial spirometer (Koko brand). The main advantages that this work presents in relation to the sensors commonly used for this purpose are: (i) the reduced cost, which makes it possible, for the first time, to use a respiratory volume sensor in medical simulators or training dummies; (ii) the direct measurement of air entering the lung using a noninvasive method, which makes it possible to use spirometry parameters to characterize simulated human respiration during the CPR training; and (iii) the measurement of spirometric parameters (FVC, FEVt, and FEF20–75%), in real-time, during the CPR training, to achieve optimal ventilation performance. Therefore, the system developed in this work addresses the minimum requirements for the practice of ventilation in the CPR maneuvers and has great potential in several future applications.This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/). Fonte: o próprio artigo.info:eu-repo/semantics/openAccessMedical mannequinsSpirometric techniquesA sensor for spirometric feedback in ventilation maneuvers during cardiopulmonary resuscitation training.info:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articleengreponame:Repositório Institucional da UFOPinstname:Universidade Federal de Ouro Preto (UFOP)instacron:UFOPLICENSElicense.txtlicense.txttext/plain; charset=utf-8924http://www.repositorio.ufop.br/bitstream/123456789/12711/2/license.txt62604f8d955274beb56c80ce1ee5dcaeMD52ORIGINALARTIGO_SensorSpirometricFeedback.pdfARTIGO_SensorSpirometricFeedback.pdfapplication/pdf2616168http://www.repositorio.ufop.br/bitstream/123456789/12711/1/ARTIGO_SensorSpirometricFeedback.pdf18deacb1809932a88d3ea0ed200598a6MD51123456789/127112020-09-10 12:29:00.018oai:localhost:123456789/12711RGVjbGFyYcOnw6NvIGRlIGRpc3RyaWJ1acOnw6NvIG7Do28tZXhjbHVzaXZhCgpPIHJlZmVyaWRvIGF1dG9yOgoKYSlEZWNsYXJhIHF1ZSBvIGRvY3VtZW50byBlbnRyZWd1ZSDDqSBzZXUgdHJhYmFsaG8gb3JpZ2luYWwgZSBxdWUgZGV0w6ltIG8gZGlyZWl0byBkZSBjb25jZWRlciBvcyBkaXJlaXRvcyBjb250aWRvcyBuZXN0YSBsaWNlbsOnYS4gRGVjbGFyYSB0YW1iw6ltIHF1ZSBhIGVudHJlZ2EgZG8gZG9jdW1lbnRvIG7Do28gaW5mcmluZ2UsIHRhbnRvIHF1YW50byBsaGUgw6kgcG9zc8OtdmVsIHNhYmVyLCBvcyBkaXJlaXRvcyBkZSBxdWFscXVlciBwZXNzb2Egb3UgZW50aWRhZGUuCgpiKVNlIG8gZG9jdW1lbnRvIGVudHJlZ3VlIGNvbnTDqW0gbWF0ZXJpYWwgZG8gcXVhbCBuw6NvIGRldMOpbSBvcyBkaXJlaXRvcyBkZSBhdXRvciwgZGVjbGFyYSBxdWUgb2J0ZXZlIGF1dG9yaXphw6fDo28gZG8gZGV0ZW50b3IgZG9zIGRpcmVpdG9zIGRlIGF1dG9yIHBhcmEgY29uY2VkZXIgw6AgVW5pdmVyc2lkYWRlIEZlZGVyYWwgZGUgT3VybyBQcmV0by9VRk9QIG9zIGRpcmVpdG9zIHJlcXVlcmlkb3MgcG9yIGVzdGEgbGljZW7Dp2EgZSBxdWUgZXNzZSBtYXRlcmlhbCwgY3Vqb3MgZGlyZWl0b3Mgc8OjbyBkZSB0ZXJjZWlyb3MsIGVzdMOhIGNsYXJhbWVudGUgaWRlbnRpZmljYWRvIGUgcmVjb25oZWNpZG8gbm8gdGV4dG8gb3UgY29udGXDumRvcyBkbyBkb2N1bWVudG8gZW50cmVndWUuCgpjKVNlIG8gZG9jdW1lbnRvIGVudHJlZ3VlIMOpIGJhc2VhZG8gZW0gdHJhYmFsaG8gZmluYW5jaWFkbyBvdSBhcG9pYWRvIHBvciBvdXRyYSBpbnN0aXR1acOnw6NvIHF1ZSBuw6NvIGEgVUZPUCwgZGVjbGFyYSBxdWUgY3VtcHJpdSBxdWFpc3F1ZXIgb2JyaWdhw6fDtWVzIGV4aWdpZGFzIHBlbG8gY29udHJhdG8gb3UgYWNvcmRvLgoKRepositório InstitucionalPUBhttp://www.repositorio.ufop.br/oai/requestrepositorio@ufop.edu.bropendoar:32332020-09-10T16:29Repositório Institucional da UFOP - Universidade Federal de Ouro Preto (UFOP)false
dc.title.pt_BR.fl_str_mv A sensor for spirometric feedback in ventilation maneuvers during cardiopulmonary resuscitation training.
title A sensor for spirometric feedback in ventilation maneuvers during cardiopulmonary resuscitation training.
spellingShingle A sensor for spirometric feedback in ventilation maneuvers during cardiopulmonary resuscitation training.
Leocádio, Rodolfo Rocha Vieira
Medical mannequins
Spirometric techniques
title_short A sensor for spirometric feedback in ventilation maneuvers during cardiopulmonary resuscitation training.
title_full A sensor for spirometric feedback in ventilation maneuvers during cardiopulmonary resuscitation training.
title_fullStr A sensor for spirometric feedback in ventilation maneuvers during cardiopulmonary resuscitation training.
title_full_unstemmed A sensor for spirometric feedback in ventilation maneuvers during cardiopulmonary resuscitation training.
title_sort A sensor for spirometric feedback in ventilation maneuvers during cardiopulmonary resuscitation training.
author Leocádio, Rodolfo Rocha Vieira
author_facet Leocádio, Rodolfo Rocha Vieira
Rêgo Segundo, Alan Kardek
Louzada, Cibelle Ferreira
author_role author
author2 Rêgo Segundo, Alan Kardek
Louzada, Cibelle Ferreira
author2_role author
author
dc.contributor.author.fl_str_mv Leocádio, Rodolfo Rocha Vieira
Rêgo Segundo, Alan Kardek
Louzada, Cibelle Ferreira
dc.subject.por.fl_str_mv Medical mannequins
Spirometric techniques
topic Medical mannequins
Spirometric techniques
description This work proposes adapting an existing sensor and embedding it on mannequins used in cardiopulmonary resuscitation (CPR) training to accurately measure the amount of air supplied to the lungs during ventilation. Mathematical modeling, calibration, and validation of the sensor along with metrology, statistical inference, and spirometry techniques were used as a base for aquiring scientific knowledge of the system. The system directly measures the variable of interest (air volume) and refers to spirometric techniques in the elaboration of its model. This improves the realism of the dummies during the CPR training, because it estimates, in real-time, not only the volume of air entering in the lungs but also the Forced Vital Capacity (FVC), Forced Expiratory Volume (FEVt) and Medium Forced Expiratory Flow (FEF20–75%). The validation of the sensor achieved results that address the requirements for this application, that is, the error below 3.4% of full scale. During the spirometric tests, the system presented the measurement results of (305 ± 22, 450 ± 23, 603 ± 24, 751 ± 26, 922 ± 27, 1021 ± 30, 1182 ± 33, 1326 ± 36, 1476 ± 37, 1618 ± 45 and 1786 ± 56) × 10−6 m3 for reference values of (300, 450, 600, 750, 900, 1050, 1200, 1350, 1500, 1650 and 1800) × 10−6 m3 , respectively. Therefore, considering the spirometry and pressure boundary conditions of the manikin lungs, the system achieves the objective of simulating valid spirometric data for debriefings, that is, there is an agreement between the measurement results when compared to the signal generated by a commercial spirometer (Koko brand). The main advantages that this work presents in relation to the sensors commonly used for this purpose are: (i) the reduced cost, which makes it possible, for the first time, to use a respiratory volume sensor in medical simulators or training dummies; (ii) the direct measurement of air entering the lung using a noninvasive method, which makes it possible to use spirometry parameters to characterize simulated human respiration during the CPR training; and (iii) the measurement of spirometric parameters (FVC, FEVt, and FEF20–75%), in real-time, during the CPR training, to achieve optimal ventilation performance. Therefore, the system developed in this work addresses the minimum requirements for the practice of ventilation in the CPR maneuvers and has great potential in several future applications.
publishDate 2019
dc.date.issued.fl_str_mv 2019
dc.date.accessioned.fl_str_mv 2020-09-10T16:29:00Z
dc.date.available.fl_str_mv 2020-09-10T16:29:00Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.citation.fl_str_mv LEOCÁDIO, R. R. V.; RÊGO SEGUNDO, A. K.; LOUZADA, C. F. A sensor for spirometric feedback in ventilation maneuvers during cardiopulmonary resuscitation training. Sensor, v. 19, n. 23, 2019. Disponível em: <https://www.mdpi.com/1424-8220/19/23/5095>. Acesso em: 10 mar. 2020.
dc.identifier.uri.fl_str_mv http://www.repositorio.ufop.br/handle/123456789/12711
dc.identifier.issn.none.fl_str_mv 1424-8220
dc.identifier.doi.pt_BR.fl_str_mv https://doi.org/10.3390/s19235095
identifier_str_mv LEOCÁDIO, R. R. V.; RÊGO SEGUNDO, A. K.; LOUZADA, C. F. A sensor for spirometric feedback in ventilation maneuvers during cardiopulmonary resuscitation training. Sensor, v. 19, n. 23, 2019. Disponível em: <https://www.mdpi.com/1424-8220/19/23/5095>. Acesso em: 10 mar. 2020.
1424-8220
url http://www.repositorio.ufop.br/handle/123456789/12711
https://doi.org/10.3390/s19235095
dc.language.iso.fl_str_mv eng
language eng
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.source.none.fl_str_mv reponame:Repositório Institucional da UFOP
instname:Universidade Federal de Ouro Preto (UFOP)
instacron:UFOP
instname_str Universidade Federal de Ouro Preto (UFOP)
instacron_str UFOP
institution UFOP
reponame_str Repositório Institucional da UFOP
collection Repositório Institucional da UFOP
bitstream.url.fl_str_mv http://www.repositorio.ufop.br/bitstream/123456789/12711/2/license.txt
http://www.repositorio.ufop.br/bitstream/123456789/12711/1/ARTIGO_SensorSpirometricFeedback.pdf
bitstream.checksum.fl_str_mv 62604f8d955274beb56c80ce1ee5dcae
18deacb1809932a88d3ea0ed200598a6
bitstream.checksumAlgorithm.fl_str_mv MD5
MD5
repository.name.fl_str_mv Repositório Institucional da UFOP - Universidade Federal de Ouro Preto (UFOP)
repository.mail.fl_str_mv repositorio@ufop.edu.br
_version_ 1801685746790170624