Optimal parameters for lateral oblique radiographs of rat mandibles

Detalhes bibliográficos
Autor(a) principal: Mahl, Célia Regina Winck
Data de Publicação: 2008
Outros Autores: Fontanella, Vania Regina Camargo
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Institucional da UFRGS
Texto Completo: http://hdl.handle.net/10183/115823
Resumo: Objective: To establish optimal exposure parameters for lateral oblique radiographs of rat mandibles using an intraoral X-ray machine. Methods: A positioning apparatus, previously tested for its reproducibility, was used to obtain radiographs of four heads of formaldehyde-preserved Wistar rats. Radiographs were exposed at 50 kV and 8 mA using four size 2 films (Insight, Ektaspeed Plus, Ultraspeed and D-Speed; Eastman Kodak Co., Rochester, NY), two focal distances (30 cm and 60 cm) and seven exposure times (0.6 s, 0.7 s, 0.8 s, 1.0 s, 1.5 s, 2.0 s and 2.5 s). The radiographs were evaluated by two examiners and scored for image quality: 1, very poor; 2, poor; 3, fair; 4, good; and 5, excellent. The evaluations were repeated at 30 days. Results: Intraobserver reproducibility was substantial (kappa 5 0.798 and 0.667). Also, the two examiners showed substantial agreement (kappa 5 0.726). There was a significant difference (a 5 1%) between the scores for the films under study. At a 30 cm distance, Espeed films had mean scores of 4 and 5 at 0.8 s and 1 s exposure times, and D-speed films, at 1.5 s, 2 s and 2.5 s exposure times. At a 60 cm distance, the best results were found for Espeed films at 2.0 s and 2.5 s exposure times. Conclusion: The films under study did not show any significant difference in the quality of rat mandible images as long as exposure times were adequate to their sensitivity and focal distance.
id UFRGS-2_e40d4972d17815766eeec764ac70dc2f
oai_identifier_str oai:www.lume.ufrgs.br:10183/115823
network_acronym_str UFRGS-2
network_name_str Repositório Institucional da UFRGS
repository_id_str
spelling Mahl, Célia Regina WinckFontanella, Vania Regina Camargo2015-05-05T01:58:36Z20080250-832Xhttp://hdl.handle.net/10183/115823000952398Objective: To establish optimal exposure parameters for lateral oblique radiographs of rat mandibles using an intraoral X-ray machine. Methods: A positioning apparatus, previously tested for its reproducibility, was used to obtain radiographs of four heads of formaldehyde-preserved Wistar rats. Radiographs were exposed at 50 kV and 8 mA using four size 2 films (Insight, Ektaspeed Plus, Ultraspeed and D-Speed; Eastman Kodak Co., Rochester, NY), two focal distances (30 cm and 60 cm) and seven exposure times (0.6 s, 0.7 s, 0.8 s, 1.0 s, 1.5 s, 2.0 s and 2.5 s). The radiographs were evaluated by two examiners and scored for image quality: 1, very poor; 2, poor; 3, fair; 4, good; and 5, excellent. The evaluations were repeated at 30 days. Results: Intraobserver reproducibility was substantial (kappa 5 0.798 and 0.667). Also, the two examiners showed substantial agreement (kappa 5 0.726). There was a significant difference (a 5 1%) between the scores for the films under study. At a 30 cm distance, Espeed films had mean scores of 4 and 5 at 0.8 s and 1 s exposure times, and D-speed films, at 1.5 s, 2 s and 2.5 s exposure times. At a 60 cm distance, the best results were found for Espeed films at 2.0 s and 2.5 s exposure times. Conclusion: The films under study did not show any significant difference in the quality of rat mandible images as long as exposure times were adequate to their sensitivity and focal distance.application/pdfengDentomaxillofacial radiology. Oxford, UK. Vol. 37, no. 4 (May 2008), p. 224-227RadiografiaMandíbulaRatos WistarOdontologiaRadiographyRats, WistarMandibleX-ray filmOptimal parameters for lateral oblique radiographs of rat mandiblesEstrangeiroinfo:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/openAccessreponame:Repositório Institucional da UFRGSinstname:Universidade Federal do Rio Grande do Sul (UFRGS)instacron:UFRGSORIGINAL000952398.pdf000952398.pdfTexto completo (inglês)application/pdf99147http://www.lume.ufrgs.br/bitstream/10183/115823/1/000952398.pdf56340820fe4454670e979bfe65830900MD51TEXT000952398.pdf.txt000952398.pdf.txtExtracted Texttext/plain16955http://www.lume.ufrgs.br/bitstream/10183/115823/2/000952398.pdf.txt780173de073724bc4eb0a7e497e2ee58MD52THUMBNAIL000952398.pdf.jpg000952398.pdf.jpgGenerated Thumbnailimage/jpeg1789http://www.lume.ufrgs.br/bitstream/10183/115823/3/000952398.pdf.jpg7cf10c776802c3e305fa788b8c61c310MD5310183/1158232018-10-22 07:39:02.343oai:www.lume.ufrgs.br:10183/115823Repositório de PublicaçõesPUBhttps://lume.ufrgs.br/oai/requestopendoar:2018-10-22T10:39:02Repositório Institucional da UFRGS - Universidade Federal do Rio Grande do Sul (UFRGS)false
dc.title.pt_BR.fl_str_mv Optimal parameters for lateral oblique radiographs of rat mandibles
title Optimal parameters for lateral oblique radiographs of rat mandibles
spellingShingle Optimal parameters for lateral oblique radiographs of rat mandibles
Mahl, Célia Regina Winck
Radiografia
Mandíbula
Ratos Wistar
Odontologia
Radiography
Rats, Wistar
Mandible
X-ray film
title_short Optimal parameters for lateral oblique radiographs of rat mandibles
title_full Optimal parameters for lateral oblique radiographs of rat mandibles
title_fullStr Optimal parameters for lateral oblique radiographs of rat mandibles
title_full_unstemmed Optimal parameters for lateral oblique radiographs of rat mandibles
title_sort Optimal parameters for lateral oblique radiographs of rat mandibles
author Mahl, Célia Regina Winck
author_facet Mahl, Célia Regina Winck
Fontanella, Vania Regina Camargo
author_role author
author2 Fontanella, Vania Regina Camargo
author2_role author
dc.contributor.author.fl_str_mv Mahl, Célia Regina Winck
Fontanella, Vania Regina Camargo
dc.subject.por.fl_str_mv Radiografia
Mandíbula
Ratos Wistar
Odontologia
topic Radiografia
Mandíbula
Ratos Wistar
Odontologia
Radiography
Rats, Wistar
Mandible
X-ray film
dc.subject.eng.fl_str_mv Radiography
Rats, Wistar
Mandible
X-ray film
description Objective: To establish optimal exposure parameters for lateral oblique radiographs of rat mandibles using an intraoral X-ray machine. Methods: A positioning apparatus, previously tested for its reproducibility, was used to obtain radiographs of four heads of formaldehyde-preserved Wistar rats. Radiographs were exposed at 50 kV and 8 mA using four size 2 films (Insight, Ektaspeed Plus, Ultraspeed and D-Speed; Eastman Kodak Co., Rochester, NY), two focal distances (30 cm and 60 cm) and seven exposure times (0.6 s, 0.7 s, 0.8 s, 1.0 s, 1.5 s, 2.0 s and 2.5 s). The radiographs were evaluated by two examiners and scored for image quality: 1, very poor; 2, poor; 3, fair; 4, good; and 5, excellent. The evaluations were repeated at 30 days. Results: Intraobserver reproducibility was substantial (kappa 5 0.798 and 0.667). Also, the two examiners showed substantial agreement (kappa 5 0.726). There was a significant difference (a 5 1%) between the scores for the films under study. At a 30 cm distance, Espeed films had mean scores of 4 and 5 at 0.8 s and 1 s exposure times, and D-speed films, at 1.5 s, 2 s and 2.5 s exposure times. At a 60 cm distance, the best results were found for Espeed films at 2.0 s and 2.5 s exposure times. Conclusion: The films under study did not show any significant difference in the quality of rat mandible images as long as exposure times were adequate to their sensitivity and focal distance.
publishDate 2008
dc.date.issued.fl_str_mv 2008
dc.date.accessioned.fl_str_mv 2015-05-05T01:58:36Z
dc.type.driver.fl_str_mv Estrangeiro
info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://hdl.handle.net/10183/115823
dc.identifier.issn.pt_BR.fl_str_mv 0250-832X
dc.identifier.nrb.pt_BR.fl_str_mv 000952398
identifier_str_mv 0250-832X
000952398
url http://hdl.handle.net/10183/115823
dc.language.iso.fl_str_mv eng
language eng
dc.relation.ispartof.pt_BR.fl_str_mv Dentomaxillofacial radiology. Oxford, UK. Vol. 37, no. 4 (May 2008), p. 224-227
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.source.none.fl_str_mv reponame:Repositório Institucional da UFRGS
instname:Universidade Federal do Rio Grande do Sul (UFRGS)
instacron:UFRGS
instname_str Universidade Federal do Rio Grande do Sul (UFRGS)
instacron_str UFRGS
institution UFRGS
reponame_str Repositório Institucional da UFRGS
collection Repositório Institucional da UFRGS
bitstream.url.fl_str_mv http://www.lume.ufrgs.br/bitstream/10183/115823/1/000952398.pdf
http://www.lume.ufrgs.br/bitstream/10183/115823/2/000952398.pdf.txt
http://www.lume.ufrgs.br/bitstream/10183/115823/3/000952398.pdf.jpg
bitstream.checksum.fl_str_mv 56340820fe4454670e979bfe65830900
780173de073724bc4eb0a7e497e2ee58
7cf10c776802c3e305fa788b8c61c310
bitstream.checksumAlgorithm.fl_str_mv MD5
MD5
MD5
repository.name.fl_str_mv Repositório Institucional da UFRGS - Universidade Federal do Rio Grande do Sul (UFRGS)
repository.mail.fl_str_mv
_version_ 1815447580684845056