Biomechanical evaluation of internal and external hexagon platform switched implant-abutment connections: an in vitro laboratory and three-dimensional finite element analysis

Detalhes bibliográficos
Autor(a) principal: Freitas Junior, Amilcar Chagas
Data de Publicação: 2012
Outros Autores: Rocha, Eduardo P., Bonfante, Estevam A., Almeida, Erika Oliveira de, Anchieta, Rodolfo Bruniera, Martini, Ana P., Assunção, Wirley G., Silva, Nelson R. F. A., Coelho, Paulo G.
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Institucional da UFRN
Texto Completo: https://repositorio.ufrn.br/jspui/handle/123456789/23800
https://doi.org/10.1016/j.dental.2012.05.004
Resumo: Objectives The aim of this study was to assess the effect of abutment's diameter shifting on reliability and stress distribution within the implant-abutment connection for internal and external hexagon implants. The postulated hypothesis was that platform-switched implants would result in increased stress concentration within the implant-abutment connection, leading to the systems’ lower reliability. Methods Eighty-four implants were divided in four groups (n = 21): REG-EH and SWT-EH (regular and switched-platform implants with external connection, respectively); REG-IH and SWT-IH (regular and switched-platform implants with internal connection, respectively). The corresponding abutments were screwed to the implants and standardized maxillary central incisor metal crowns were cemented and subjected to step-stress accelerated life testing. Use-level probability Weibull curves and reliability were calculated. Four finite element models reproducing the characteristics of specimens used in laboratory testing were created. The models were full constrained on the bottom and lateral surface of the cylinder of acrylic resin and one 30° off-axis load (300 N) was applied on the lingual side of the crown (close to the incisal edge) in order to evaluate the stress distribution (svM) within the implant-abutment complex. Results The Beta values for groups SWT-EH (1.31), REG-EH (1.55), SWT-IH (1.83) and REG-IH (1.82) indicated that fatigue accelerated the failure of all groups. The higher levels of σvM within the implant-abutment connection observed for platform-switched implants (groups SWT-EH and SWT-IH) were in agreement with the lower reliability observed for the external hex implants, but not for the internal hex implants. The reliability 90% confidence intervals (50,000 cycles at 300 N) were 0.53(0.33–0.70), 0.93(0.80–0.97), 0.99(0.93–0.99) and 0.99(0.99–1.00), for the SWT-EH, REG-EH, SWT-IH, and REH-IH, respectively. Significance The postulated hypothesis was partially accepted. The higher levels of stress observed within implant-abutment connection when reducing abutment diameter (cross-sectional area) resulted in lower reliability for external hex implants, but not for internal hex implants.
id UFRN_3ae592b04c54ef113d45d703f4478f9d
oai_identifier_str oai:https://repositorio.ufrn.br:123456789/23800
network_acronym_str UFRN
network_name_str Repositório Institucional da UFRN
repository_id_str
spelling Freitas Junior, Amilcar ChagasRocha, Eduardo P.Bonfante, Estevam A.Almeida, Erika Oliveira deAnchieta, Rodolfo BrunieraMartini, Ana P.Assunção, Wirley G.Silva, Nelson R. F. A.Coelho, Paulo G.2017-09-11T12:46:47Z2017-09-11T12:46:47Z2012FREITAS JUNIOR, Amilcar C. et al. Biomechanical Evaluation of Internal and External Hexagon Platform Switched Implant-Abutment Connections: An In Vitro Laboratory and Three-Dimensional Finite Element Analysis. Dental Materials, v. 7, n. 10, p. 1-8, 2012.https://repositorio.ufrn.br/jspui/handle/123456789/23800https://doi.org/10.1016/j.dental.2012.05.004engELSEVIERDental implantsPlatform switchingBiomechanicsReliabilityFinite element analysisBiomechanical evaluation of internal and external hexagon platform switched implant-abutment connections: an in vitro laboratory and three-dimensional finite element analysisinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articleObjectives The aim of this study was to assess the effect of abutment's diameter shifting on reliability and stress distribution within the implant-abutment connection for internal and external hexagon implants. The postulated hypothesis was that platform-switched implants would result in increased stress concentration within the implant-abutment connection, leading to the systems’ lower reliability. Methods Eighty-four implants were divided in four groups (n = 21): REG-EH and SWT-EH (regular and switched-platform implants with external connection, respectively); REG-IH and SWT-IH (regular and switched-platform implants with internal connection, respectively). The corresponding abutments were screwed to the implants and standardized maxillary central incisor metal crowns were cemented and subjected to step-stress accelerated life testing. Use-level probability Weibull curves and reliability were calculated. Four finite element models reproducing the characteristics of specimens used in laboratory testing were created. The models were full constrained on the bottom and lateral surface of the cylinder of acrylic resin and one 30° off-axis load (300 N) was applied on the lingual side of the crown (close to the incisal edge) in order to evaluate the stress distribution (svM) within the implant-abutment complex. Results The Beta values for groups SWT-EH (1.31), REG-EH (1.55), SWT-IH (1.83) and REG-IH (1.82) indicated that fatigue accelerated the failure of all groups. The higher levels of σvM within the implant-abutment connection observed for platform-switched implants (groups SWT-EH and SWT-IH) were in agreement with the lower reliability observed for the external hex implants, but not for the internal hex implants. The reliability 90% confidence intervals (50,000 cycles at 300 N) were 0.53(0.33–0.70), 0.93(0.80–0.97), 0.99(0.93–0.99) and 0.99(0.99–1.00), for the SWT-EH, REG-EH, SWT-IH, and REH-IH, respectively. Significance The postulated hypothesis was partially accepted. The higher levels of stress observed within implant-abutment connection when reducing abutment diameter (cross-sectional area) resulted in lower reliability for external hex implants, but not for internal hex implants.info:eu-repo/semantics/openAccessreponame:Repositório Institucional da UFRNinstname:Universidade Federal do Rio Grande do Norte (UFRN)instacron:UFRNLICENSElicense.txtlicense.txttext/plain; charset=utf-81748https://repositorio.ufrn.br/bitstream/123456789/23800/2/license.txt8a4605be74aa9ea9d79846c1fba20a33MD52TEXTBiomechanical Evaluation of Internal_2012.pdf.txtBiomechanical Evaluation of Internal_2012.pdf.txtExtracted texttext/plain48091https://repositorio.ufrn.br/bitstream/123456789/23800/5/Biomechanical%20Evaluation%20of%20Internal_2012.pdf.txtf68e7907f83e237f91753c732c42cc38MD55THUMBNAILBiomechanical Evaluation of Internal_2012.pdf.jpgBiomechanical Evaluation of Internal_2012.pdf.jpgIM Thumbnailimage/jpeg8538https://repositorio.ufrn.br/bitstream/123456789/23800/6/Biomechanical%20Evaluation%20of%20Internal_2012.pdf.jpg4628afc433355f8aee92a3060ae5d30dMD56123456789/238002023-02-02 17:03:10.056oai:https://repositorio.ufrn.br:123456789/23800Tk9URTogUExBQ0UgWU9VUiBPV04gTElDRU5TRSBIRVJFClRoaXMgc2FtcGxlIGxpY2Vuc2UgaXMgcHJvdmlkZWQgZm9yIGluZm9ybWF0aW9uYWwgcHVycG9zZXMgb25seS4KCk5PTi1FWENMVVNJVkUgRElTVFJJQlVUSU9OIExJQ0VOU0UKCkJ5IHNpZ25pbmcgYW5kIHN1Ym1pdHRpbmcgdGhpcyBsaWNlbnNlLCB5b3UgKHRoZSBhdXRob3Iocykgb3IgY29weXJpZ2h0Cm93bmVyKSBncmFudHMgdG8gRFNwYWNlIFVuaXZlcnNpdHkgKERTVSkgdGhlIG5vbi1leGNsdXNpdmUgcmlnaHQgdG8gcmVwcm9kdWNlLAp0cmFuc2xhdGUgKGFzIGRlZmluZWQgYmVsb3cpLCBhbmQvb3IgZGlzdHJpYnV0ZSB5b3VyIHN1Ym1pc3Npb24gKGluY2x1ZGluZwp0aGUgYWJzdHJhY3QpIHdvcmxkd2lkZSBpbiBwcmludCBhbmQgZWxlY3Ryb25pYyBmb3JtYXQgYW5kIGluIGFueSBtZWRpdW0sCmluY2x1ZGluZyBidXQgbm90IGxpbWl0ZWQgdG8gYXVkaW8gb3IgdmlkZW8uCgpZb3UgYWdyZWUgdGhhdCBEU1UgbWF5LCB3aXRob3V0IGNoYW5naW5nIHRoZSBjb250ZW50LCB0cmFuc2xhdGUgdGhlCnN1Ym1pc3Npb24gdG8gYW55IG1lZGl1bSBvciBmb3JtYXQgZm9yIHRoZSBwdXJwb3NlIG9mIHByZXNlcnZhdGlvbi4KCllvdSBhbHNvIGFncmVlIHRoYXQgRFNVIG1heSBrZWVwIG1vcmUgdGhhbiBvbmUgY29weSBvZiB0aGlzIHN1Ym1pc3Npb24gZm9yCnB1cnBvc2VzIG9mIHNlY3VyaXR5LCBiYWNrLXVwIGFuZCBwcmVzZXJ2YXRpb24uCgpZb3UgcmVwcmVzZW50IHRoYXQgdGhlIHN1Ym1pc3Npb24gaXMgeW91ciBvcmlnaW5hbCB3b3JrLCBhbmQgdGhhdCB5b3UgaGF2ZQp0aGUgcmlnaHQgdG8gZ3JhbnQgdGhlIHJpZ2h0cyBjb250YWluZWQgaW4gdGhpcyBsaWNlbnNlLiBZb3UgYWxzbyByZXByZXNlbnQKdGhhdCB5b3VyIHN1Ym1pc3Npb24gZG9lcyBub3QsIHRvIHRoZSBiZXN0IG9mIHlvdXIga25vd2xlZGdlLCBpbmZyaW5nZSB1cG9uCmFueW9uZSdzIGNvcHlyaWdodC4KCklmIHRoZSBzdWJtaXNzaW9uIGNvbnRhaW5zIG1hdGVyaWFsIGZvciB3aGljaCB5b3UgZG8gbm90IGhvbGQgY29weXJpZ2h0LAp5b3UgcmVwcmVzZW50IHRoYXQgeW91IGhhdmUgb2J0YWluZWQgdGhlIHVucmVzdHJpY3RlZCBwZXJtaXNzaW9uIG9mIHRoZQpjb3B5cmlnaHQgb3duZXIgdG8gZ3JhbnQgRFNVIHRoZSByaWdodHMgcmVxdWlyZWQgYnkgdGhpcyBsaWNlbnNlLCBhbmQgdGhhdApzdWNoIHRoaXJkLXBhcnR5IG93bmVkIG1hdGVyaWFsIGlzIGNsZWFybHkgaWRlbnRpZmllZCBhbmQgYWNrbm93bGVkZ2VkCndpdGhpbiB0aGUgdGV4dCBvciBjb250ZW50IG9mIHRoZSBzdWJtaXNzaW9uLgoKSUYgVEhFIFNVQk1JU1NJT04gSVMgQkFTRUQgVVBPTiBXT1JLIFRIQVQgSEFTIEJFRU4gU1BPTlNPUkVEIE9SIFNVUFBPUlRFRApCWSBBTiBBR0VOQ1kgT1IgT1JHQU5JWkFUSU9OIE9USEVSIFRIQU4gRFNVLCBZT1UgUkVQUkVTRU5UIFRIQVQgWU9VIEhBVkUKRlVMRklMTEVEIEFOWSBSSUdIVCBPRiBSRVZJRVcgT1IgT1RIRVIgT0JMSUdBVElPTlMgUkVRVUlSRUQgQlkgU1VDSApDT05UUkFDVCBPUiBBR1JFRU1FTlQuCgpEU1Ugd2lsbCBjbGVhcmx5IGlkZW50aWZ5IHlvdXIgbmFtZShzKSBhcyB0aGUgYXV0aG9yKHMpIG9yIG93bmVyKHMpIG9mIHRoZQpzdWJtaXNzaW9uLCBhbmQgd2lsbCBub3QgbWFrZSBhbnkgYWx0ZXJhdGlvbiwgb3RoZXIgdGhhbiBhcyBhbGxvd2VkIGJ5IHRoaXMKbGljZW5zZSwgdG8geW91ciBzdWJtaXNzaW9uLgo=Repositório de PublicaçõesPUBhttp://repositorio.ufrn.br/oai/opendoar:2023-02-02T20:03:10Repositório Institucional da UFRN - Universidade Federal do Rio Grande do Norte (UFRN)false
dc.title.pt_BR.fl_str_mv Biomechanical evaluation of internal and external hexagon platform switched implant-abutment connections: an in vitro laboratory and three-dimensional finite element analysis
title Biomechanical evaluation of internal and external hexagon platform switched implant-abutment connections: an in vitro laboratory and three-dimensional finite element analysis
spellingShingle Biomechanical evaluation of internal and external hexagon platform switched implant-abutment connections: an in vitro laboratory and three-dimensional finite element analysis
Freitas Junior, Amilcar Chagas
Dental implants
Platform switching
Biomechanics
Reliability
Finite element analysis
title_short Biomechanical evaluation of internal and external hexagon platform switched implant-abutment connections: an in vitro laboratory and three-dimensional finite element analysis
title_full Biomechanical evaluation of internal and external hexagon platform switched implant-abutment connections: an in vitro laboratory and three-dimensional finite element analysis
title_fullStr Biomechanical evaluation of internal and external hexagon platform switched implant-abutment connections: an in vitro laboratory and three-dimensional finite element analysis
title_full_unstemmed Biomechanical evaluation of internal and external hexagon platform switched implant-abutment connections: an in vitro laboratory and three-dimensional finite element analysis
title_sort Biomechanical evaluation of internal and external hexagon platform switched implant-abutment connections: an in vitro laboratory and three-dimensional finite element analysis
author Freitas Junior, Amilcar Chagas
author_facet Freitas Junior, Amilcar Chagas
Rocha, Eduardo P.
Bonfante, Estevam A.
Almeida, Erika Oliveira de
Anchieta, Rodolfo Bruniera
Martini, Ana P.
Assunção, Wirley G.
Silva, Nelson R. F. A.
Coelho, Paulo G.
author_role author
author2 Rocha, Eduardo P.
Bonfante, Estevam A.
Almeida, Erika Oliveira de
Anchieta, Rodolfo Bruniera
Martini, Ana P.
Assunção, Wirley G.
Silva, Nelson R. F. A.
Coelho, Paulo G.
author2_role author
author
author
author
author
author
author
author
dc.contributor.author.fl_str_mv Freitas Junior, Amilcar Chagas
Rocha, Eduardo P.
Bonfante, Estevam A.
Almeida, Erika Oliveira de
Anchieta, Rodolfo Bruniera
Martini, Ana P.
Assunção, Wirley G.
Silva, Nelson R. F. A.
Coelho, Paulo G.
dc.subject.por.fl_str_mv Dental implants
Platform switching
Biomechanics
Reliability
Finite element analysis
topic Dental implants
Platform switching
Biomechanics
Reliability
Finite element analysis
description Objectives The aim of this study was to assess the effect of abutment's diameter shifting on reliability and stress distribution within the implant-abutment connection for internal and external hexagon implants. The postulated hypothesis was that platform-switched implants would result in increased stress concentration within the implant-abutment connection, leading to the systems’ lower reliability. Methods Eighty-four implants were divided in four groups (n = 21): REG-EH and SWT-EH (regular and switched-platform implants with external connection, respectively); REG-IH and SWT-IH (regular and switched-platform implants with internal connection, respectively). The corresponding abutments were screwed to the implants and standardized maxillary central incisor metal crowns were cemented and subjected to step-stress accelerated life testing. Use-level probability Weibull curves and reliability were calculated. Four finite element models reproducing the characteristics of specimens used in laboratory testing were created. The models were full constrained on the bottom and lateral surface of the cylinder of acrylic resin and one 30° off-axis load (300 N) was applied on the lingual side of the crown (close to the incisal edge) in order to evaluate the stress distribution (svM) within the implant-abutment complex. Results The Beta values for groups SWT-EH (1.31), REG-EH (1.55), SWT-IH (1.83) and REG-IH (1.82) indicated that fatigue accelerated the failure of all groups. The higher levels of σvM within the implant-abutment connection observed for platform-switched implants (groups SWT-EH and SWT-IH) were in agreement with the lower reliability observed for the external hex implants, but not for the internal hex implants. The reliability 90% confidence intervals (50,000 cycles at 300 N) were 0.53(0.33–0.70), 0.93(0.80–0.97), 0.99(0.93–0.99) and 0.99(0.99–1.00), for the SWT-EH, REG-EH, SWT-IH, and REH-IH, respectively. Significance The postulated hypothesis was partially accepted. The higher levels of stress observed within implant-abutment connection when reducing abutment diameter (cross-sectional area) resulted in lower reliability for external hex implants, but not for internal hex implants.
publishDate 2012
dc.date.issued.fl_str_mv 2012
dc.date.accessioned.fl_str_mv 2017-09-11T12:46:47Z
dc.date.available.fl_str_mv 2017-09-11T12:46:47Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.citation.fl_str_mv FREITAS JUNIOR, Amilcar C. et al. Biomechanical Evaluation of Internal and External Hexagon Platform Switched Implant-Abutment Connections: An In Vitro Laboratory and Three-Dimensional Finite Element Analysis. Dental Materials, v. 7, n. 10, p. 1-8, 2012.
dc.identifier.uri.fl_str_mv https://repositorio.ufrn.br/jspui/handle/123456789/23800
dc.identifier.doi.none.fl_str_mv https://doi.org/10.1016/j.dental.2012.05.004
identifier_str_mv FREITAS JUNIOR, Amilcar C. et al. Biomechanical Evaluation of Internal and External Hexagon Platform Switched Implant-Abutment Connections: An In Vitro Laboratory and Three-Dimensional Finite Element Analysis. Dental Materials, v. 7, n. 10, p. 1-8, 2012.
url https://repositorio.ufrn.br/jspui/handle/123456789/23800
https://doi.org/10.1016/j.dental.2012.05.004
dc.language.iso.fl_str_mv eng
language eng
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.publisher.none.fl_str_mv ELSEVIER
publisher.none.fl_str_mv ELSEVIER
dc.source.none.fl_str_mv reponame:Repositório Institucional da UFRN
instname:Universidade Federal do Rio Grande do Norte (UFRN)
instacron:UFRN
instname_str Universidade Federal do Rio Grande do Norte (UFRN)
instacron_str UFRN
institution UFRN
reponame_str Repositório Institucional da UFRN
collection Repositório Institucional da UFRN
bitstream.url.fl_str_mv https://repositorio.ufrn.br/bitstream/123456789/23800/2/license.txt
https://repositorio.ufrn.br/bitstream/123456789/23800/5/Biomechanical%20Evaluation%20of%20Internal_2012.pdf.txt
https://repositorio.ufrn.br/bitstream/123456789/23800/6/Biomechanical%20Evaluation%20of%20Internal_2012.pdf.jpg
bitstream.checksum.fl_str_mv 8a4605be74aa9ea9d79846c1fba20a33
f68e7907f83e237f91753c732c42cc38
4628afc433355f8aee92a3060ae5d30d
bitstream.checksumAlgorithm.fl_str_mv MD5
MD5
MD5
repository.name.fl_str_mv Repositório Institucional da UFRN - Universidade Federal do Rio Grande do Norte (UFRN)
repository.mail.fl_str_mv
_version_ 1797777028806606848