Confiabilidade do equipamento S-FLEX® para verificação da pressão de sucção não nutritiva em recém-nascidos

Detalhes bibliográficos
Autor(a) principal: Nascimento, Marlove Duarte
Data de Publicação: 2018
Tipo de documento: Dissertação
Idioma: por
Título da fonte: Manancial - Repositório Digital da UFSM
dARK ID: ark:/26339/0013000000zpx
Texto Completo: http://repositorio.ufsm.br/handle/1/20238
Resumo: The evaluation of non-nutritive sucking in newborns is part of the rotine in many Neonatal Intensive Care Units, and it is performed with the gloved finger technique, by an examiner, at most of times. The purpose to this study was to establish the reliability of S-FLEX® equipment to verify the non-nutritive sucking pressure in term newborns. Thirty healthy term newborns, from both sexes, were evaluated between the first and the third day of life. Two evaluations were executed in each newborn, by two different examiners (examiner A and examiner B), with an interval of 3 to 4 hours between evaluations. In each evaluation, maximal and mean non-nutritive sucking pressure were determined in three consecutive moments. Repeatability and reproducibility of the found measures were examined to verify reliability. To calculate repeatability, three consecutive measures, in each one of the moments, were compared through Analysis of Variance (ANOVA) for repeated measures. The maximal sucking pressure, in the first evaluation (examiner A) was 10.44 (±2.25), 10.12 (±2.25) and 9.73 (±2.61) mmHg, (p=0.120), respectively in first, second and third sequential measurements, and, in the second evaluation (examiner B), the values were 10.10 (±2.35), 9.98 (±2.02) and 10.37 (±2.29), (p=0.404). The mean for sucking pressure in the first evaluation (examiner A) was 3.64 (±1.04), 3.59 (±1.04) and 3.56 (±1.03), (p=0.852), and, in the second evaluation (examiner B), it was 3.59 (±1.04), 3.51 (±0.92) and 3.63 (±1.04), (p=0.721). Reproducibility was verified according to analysis of the obtained measures by different examiners, through interclass correlation coefficient (ICC). To maximal sucking pressure, correlation coefficient was 0.708 (p=0.001) and, to mean pressure, it was 0.828 (p<0.001), which demonstrates good and excellent correlation, respectively. S-FLEX® proved itself a reliable equipment to determine maximal and mean non-nutritive sucking pressure in healthy term newborns.
id UFSM_0f28907ed8e870b2901871e00e7f5c93
oai_identifier_str oai:repositorio.ufsm.br:1/20238
network_acronym_str UFSM
network_name_str Manancial - Repositório Digital da UFSM
repository_id_str
spelling Confiabilidade do equipamento S-FLEX® para verificação da pressão de sucção não nutritiva em recém-nascidosReliability of S-FLEX® equipment for verification of non-nutritive sucking pressure in newbornsComportamento de sucçãoRecém-nascidoConfiabilidadeSuction behaviorNewbornReliabilityCNPQ::CIENCIAS DA SAUDE::FONOAUDIOLOGIAThe evaluation of non-nutritive sucking in newborns is part of the rotine in many Neonatal Intensive Care Units, and it is performed with the gloved finger technique, by an examiner, at most of times. The purpose to this study was to establish the reliability of S-FLEX® equipment to verify the non-nutritive sucking pressure in term newborns. Thirty healthy term newborns, from both sexes, were evaluated between the first and the third day of life. Two evaluations were executed in each newborn, by two different examiners (examiner A and examiner B), with an interval of 3 to 4 hours between evaluations. In each evaluation, maximal and mean non-nutritive sucking pressure were determined in three consecutive moments. Repeatability and reproducibility of the found measures were examined to verify reliability. To calculate repeatability, three consecutive measures, in each one of the moments, were compared through Analysis of Variance (ANOVA) for repeated measures. The maximal sucking pressure, in the first evaluation (examiner A) was 10.44 (±2.25), 10.12 (±2.25) and 9.73 (±2.61) mmHg, (p=0.120), respectively in first, second and third sequential measurements, and, in the second evaluation (examiner B), the values were 10.10 (±2.35), 9.98 (±2.02) and 10.37 (±2.29), (p=0.404). The mean for sucking pressure in the first evaluation (examiner A) was 3.64 (±1.04), 3.59 (±1.04) and 3.56 (±1.03), (p=0.852), and, in the second evaluation (examiner B), it was 3.59 (±1.04), 3.51 (±0.92) and 3.63 (±1.04), (p=0.721). Reproducibility was verified according to analysis of the obtained measures by different examiners, through interclass correlation coefficient (ICC). To maximal sucking pressure, correlation coefficient was 0.708 (p=0.001) and, to mean pressure, it was 0.828 (p<0.001), which demonstrates good and excellent correlation, respectively. S-FLEX® proved itself a reliable equipment to determine maximal and mean non-nutritive sucking pressure in healthy term newborns.A avaliação da sucção não nutritiva em recém-nascidos é parte da rotina de muitas Unidades de Terapia Intensiva Neonatal e é realizada, na maioria das vezes, com a técnica do dedo enluvado do examinador. O objetivo do presente estudo foi determinar a confiabilidade do equipamento S-FLEX® para verificar pressão de sucção não nutritiva em recém-nascidos a termo. Foram avaliados 30 recém-nascidos a termo saudáveis, de ambos os sexos, que se encontravam entre o primeiro e o terceiro dia de vida. Foram realizadas duas avaliações com os recém-nascidos, por dois avaliadores (avaliador A e avaliador B), com intervalo de 3 a 4 horas entre a 1ª e a 2ª avaliação. Em cada avaliação, foram determinadas a pressão máxima e a pressão média de sucção não nutritiva em três momentos consecutivos. Para verificar a confiabilidade, foram avaliadas a repetibilidade e a reprodutibilidade das medidas produzidas. Para o cálculo de repetibilidade, foram comparadas as médias das três medidas sucessivas, em cada um dos momentos, através da Análise de Variância (ANOVA) para medidas repetidas. A pressão máxima de sucção, na 1ª avaliação (avaliador A), foi de 10,44 (±2,25), 10,12 (±2,25) e 9,73 (±2,61) mmHg, (p=0,120), respectivamente para a 1ª, 2ª e 3ª medidas realizadas em sequência e, na 2ª avaliação (avaliador B), foi de 10,10 (±2,35), 9,98 (±2,02) e 10,37 (±2,29), (p=0,404). A pressão média de sucção na 1ª avaliação (avaliador A) foi de 3,64 (±1,04), 3,59 (±1,04) e 3,56 (±1,03), (p=0,852) e, na 2ª avaliação (avaliador B) foi 3,59 (±1,04), 3,51 (±0,92) e 3,63 (±1,04), (p=0,721). A reprodutibilidade foi verificada com a análise da concordância das medidas obtidas pelos distintos avaliadores através do coeficiente de correlação intra-classe (ICC). Para a pressão máxima de sucção, o coeficiente de correlação foi de 0,708 (p=0,001) e, para a pressão média, foi de 0,828 (p<0,001), evidenciando boa e excelente correlação, respectivamente. O S-FLEX® mostrou ser um equipamento confiável para determinar a pressão máxima e média de sucção não nutritiva em recém-nascidos a termo saudáveis.Universidade Federal de Santa MariaBrasilFonoaudiologiaUFSMPrograma de Pós-Graduação em Distúrbios da Comunicação HumanaCentro de Ciências da SaúdeWeinmann, Angela Regina Macielhttp://lattes.cnpq.br/9151119377173425Bolzan, Geovana de Paulahttp://lattes.cnpq.br/6911505466941984Rodrigues, Analu Lopeshttp://lattes.cnpq.br/2403912790384543Berwig, Luana Cristinahttp://lattes.cnpq.br/0187578798812108Nascimento, Marlove Duarte2021-01-04T09:15:27Z2021-01-04T09:15:27Z2018-07-30info:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/masterThesisapplication/pdfhttp://repositorio.ufsm.br/handle/1/20238ark:/26339/0013000000zpxporAttribution-NonCommercial-NoDerivatives 4.0 Internationalhttp://creativecommons.org/licenses/by-nc-nd/4.0/info:eu-repo/semantics/openAccessreponame:Manancial - Repositório Digital da UFSMinstname:Universidade Federal de Santa Maria (UFSM)instacron:UFSM2021-01-05T06:00:51Zoai:repositorio.ufsm.br:1/20238Biblioteca Digital de Teses e Dissertaçõeshttps://repositorio.ufsm.br/ONGhttps://repositorio.ufsm.br/oai/requestatendimento.sib@ufsm.br||tedebc@gmail.comopendoar:2021-01-05T06:00:51Manancial - Repositório Digital da UFSM - Universidade Federal de Santa Maria (UFSM)false
dc.title.none.fl_str_mv Confiabilidade do equipamento S-FLEX® para verificação da pressão de sucção não nutritiva em recém-nascidos
Reliability of S-FLEX® equipment for verification of non-nutritive sucking pressure in newborns
title Confiabilidade do equipamento S-FLEX® para verificação da pressão de sucção não nutritiva em recém-nascidos
spellingShingle Confiabilidade do equipamento S-FLEX® para verificação da pressão de sucção não nutritiva em recém-nascidos
Nascimento, Marlove Duarte
Comportamento de sucção
Recém-nascido
Confiabilidade
Suction behavior
Newborn
Reliability
CNPQ::CIENCIAS DA SAUDE::FONOAUDIOLOGIA
title_short Confiabilidade do equipamento S-FLEX® para verificação da pressão de sucção não nutritiva em recém-nascidos
title_full Confiabilidade do equipamento S-FLEX® para verificação da pressão de sucção não nutritiva em recém-nascidos
title_fullStr Confiabilidade do equipamento S-FLEX® para verificação da pressão de sucção não nutritiva em recém-nascidos
title_full_unstemmed Confiabilidade do equipamento S-FLEX® para verificação da pressão de sucção não nutritiva em recém-nascidos
title_sort Confiabilidade do equipamento S-FLEX® para verificação da pressão de sucção não nutritiva em recém-nascidos
author Nascimento, Marlove Duarte
author_facet Nascimento, Marlove Duarte
author_role author
dc.contributor.none.fl_str_mv Weinmann, Angela Regina Maciel
http://lattes.cnpq.br/9151119377173425
Bolzan, Geovana de Paula
http://lattes.cnpq.br/6911505466941984
Rodrigues, Analu Lopes
http://lattes.cnpq.br/2403912790384543
Berwig, Luana Cristina
http://lattes.cnpq.br/0187578798812108
dc.contributor.author.fl_str_mv Nascimento, Marlove Duarte
dc.subject.por.fl_str_mv Comportamento de sucção
Recém-nascido
Confiabilidade
Suction behavior
Newborn
Reliability
CNPQ::CIENCIAS DA SAUDE::FONOAUDIOLOGIA
topic Comportamento de sucção
Recém-nascido
Confiabilidade
Suction behavior
Newborn
Reliability
CNPQ::CIENCIAS DA SAUDE::FONOAUDIOLOGIA
description The evaluation of non-nutritive sucking in newborns is part of the rotine in many Neonatal Intensive Care Units, and it is performed with the gloved finger technique, by an examiner, at most of times. The purpose to this study was to establish the reliability of S-FLEX® equipment to verify the non-nutritive sucking pressure in term newborns. Thirty healthy term newborns, from both sexes, were evaluated between the first and the third day of life. Two evaluations were executed in each newborn, by two different examiners (examiner A and examiner B), with an interval of 3 to 4 hours between evaluations. In each evaluation, maximal and mean non-nutritive sucking pressure were determined in three consecutive moments. Repeatability and reproducibility of the found measures were examined to verify reliability. To calculate repeatability, three consecutive measures, in each one of the moments, were compared through Analysis of Variance (ANOVA) for repeated measures. The maximal sucking pressure, in the first evaluation (examiner A) was 10.44 (±2.25), 10.12 (±2.25) and 9.73 (±2.61) mmHg, (p=0.120), respectively in first, second and third sequential measurements, and, in the second evaluation (examiner B), the values were 10.10 (±2.35), 9.98 (±2.02) and 10.37 (±2.29), (p=0.404). The mean for sucking pressure in the first evaluation (examiner A) was 3.64 (±1.04), 3.59 (±1.04) and 3.56 (±1.03), (p=0.852), and, in the second evaluation (examiner B), it was 3.59 (±1.04), 3.51 (±0.92) and 3.63 (±1.04), (p=0.721). Reproducibility was verified according to analysis of the obtained measures by different examiners, through interclass correlation coefficient (ICC). To maximal sucking pressure, correlation coefficient was 0.708 (p=0.001) and, to mean pressure, it was 0.828 (p<0.001), which demonstrates good and excellent correlation, respectively. S-FLEX® proved itself a reliable equipment to determine maximal and mean non-nutritive sucking pressure in healthy term newborns.
publishDate 2018
dc.date.none.fl_str_mv 2018-07-30
2021-01-04T09:15:27Z
2021-01-04T09:15:27Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/masterThesis
format masterThesis
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://repositorio.ufsm.br/handle/1/20238
dc.identifier.dark.fl_str_mv ark:/26339/0013000000zpx
url http://repositorio.ufsm.br/handle/1/20238
identifier_str_mv ark:/26339/0013000000zpx
dc.language.iso.fl_str_mv por
language por
dc.rights.driver.fl_str_mv Attribution-NonCommercial-NoDerivatives 4.0 International
http://creativecommons.org/licenses/by-nc-nd/4.0/
info:eu-repo/semantics/openAccess
rights_invalid_str_mv Attribution-NonCommercial-NoDerivatives 4.0 International
http://creativecommons.org/licenses/by-nc-nd/4.0/
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Universidade Federal de Santa Maria
Brasil
Fonoaudiologia
UFSM
Programa de Pós-Graduação em Distúrbios da Comunicação Humana
Centro de Ciências da Saúde
publisher.none.fl_str_mv Universidade Federal de Santa Maria
Brasil
Fonoaudiologia
UFSM
Programa de Pós-Graduação em Distúrbios da Comunicação Humana
Centro de Ciências da Saúde
dc.source.none.fl_str_mv reponame:Manancial - Repositório Digital da UFSM
instname:Universidade Federal de Santa Maria (UFSM)
instacron:UFSM
instname_str Universidade Federal de Santa Maria (UFSM)
instacron_str UFSM
institution UFSM
reponame_str Manancial - Repositório Digital da UFSM
collection Manancial - Repositório Digital da UFSM
repository.name.fl_str_mv Manancial - Repositório Digital da UFSM - Universidade Federal de Santa Maria (UFSM)
repository.mail.fl_str_mv atendimento.sib@ufsm.br||tedebc@gmail.com
_version_ 1815172260904828928