Data analyses of fatigue tests by extensometry in hip prosthesis of the Co-28Cr-6Mo alloy

Detalhes bibliográficos
Autor(a) principal: Souza, Caique Movio Pereira de
Data de Publicação: 2022
Outros Autores: Santos , Raul Gaspari, Souza, Renato Chaves, Militão , Vanderlei Araujo, Silva, Isaias Gouveia, Seriacopi, Vanessa, Silva Junior, Wilson Carlos da
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Research, Society and Development
Texto Completo: https://rsdjournal.org/index.php/rsd/article/view/27854
Resumo: ANVISA made changes to its regulations, setting new values for the fatigue tests, increasing the number of cycles that prostheses must endure without presenting failures. The standard used is ABNT NBR 7206-6:2013, in which all prostheses being commercialized must be tested in the new parameters. In this article, studies were conducted based on the requirement to revalidate the Co-26Cr-6Mo metallic alloy (ASTM F75). Microstructural characterizations (Optical and Scanning Electron Microscopies), and fatigue tests were carried out, with the aim of obtaining the mechanical behavior and features of the materials used in hip prostheses and comparing them with the standard. The grain size and inclusion contents were found to be controlled. Likewise, the prosthesis has withstand 10,000,000 cycles, and has not shown cracks, nor plastic deformations, enabling the biomedical use of this alloy according to the new regulations. Tensile test and liquid penetrant inspection were also carried out to take the parameters of the related physical properties. In the analysis with strain gauges, it was possible to detect the purely elastic deformation that occurred with the application of the load during fatigue tests, resulting in a slight variation of stress in the data acquisition system.
id UNIFEI_31b6f35a2adb944ce1e010805f0fe109
oai_identifier_str oai:ojs.pkp.sfu.ca:article/27854
network_acronym_str UNIFEI
network_name_str Research, Society and Development
repository_id_str
spelling Data analyses of fatigue tests by extensometry in hip prosthesis of the Co-28Cr-6Mo alloy Análisis de datos de ensayos de fatiga por extensometría en prótesis de cadera de la aleación Co-28Cr-6MoAnálise de dados de ensaio de fadiga por Extensometria em prótese de quadril da liga Co-28Cr-6Mo Prótesis de CaderaASTM F75Prueba de FatigaExtensometríaABNT NBR 7206-6:2013.Próteses de QuadrilASTM F75Ensaio de FadigaExtensômetriaABNT NBR 7206-6:2013.Hip ProsthesisASTM F75Fatigue TestsExtensometryABNT NBR 7206-6: 2013 standard.ANVISA made changes to its regulations, setting new values for the fatigue tests, increasing the number of cycles that prostheses must endure without presenting failures. The standard used is ABNT NBR 7206-6:2013, in which all prostheses being commercialized must be tested in the new parameters. In this article, studies were conducted based on the requirement to revalidate the Co-26Cr-6Mo metallic alloy (ASTM F75). Microstructural characterizations (Optical and Scanning Electron Microscopies), and fatigue tests were carried out, with the aim of obtaining the mechanical behavior and features of the materials used in hip prostheses and comparing them with the standard. The grain size and inclusion contents were found to be controlled. Likewise, the prosthesis has withstand 10,000,000 cycles, and has not shown cracks, nor plastic deformations, enabling the biomedical use of this alloy according to the new regulations. Tensile test and liquid penetrant inspection were also carried out to take the parameters of the related physical properties. In the analysis with strain gauges, it was possible to detect the purely elastic deformation that occurred with the application of the load during fatigue tests, resulting in a slight variation of stress in the data acquisition system.ANVISA realizó modificaciones en su normativa, fijando nuevos valores para las pruebas de fatiga, aumentando el número de ciclos que las prótesis deben soportar sin presentar fallas. La norma utilizada es la ABNT NBR 7206-6: 2013, en la que todas las prótesis que se comercializan deben ser probadas en los nuevos parámetros. En este artículo se realizaron estudios en base al requisito de revalidar la aleación metálica Co-26Cr-6Mo (ASTM F75). Se realizaron caracterizaciones microestructurales (Microscopía Óptica y Electrónica de Barrido), y ensayos de fatiga, con el objetivo de obtener el comportamiento mecánico y las características de los materiales utilizados en las prótesis de cadera y compararlos con el estándar. Se encontró que el tamaño de grano y los contenidos de inclusión estaban controlados. Asimismo, la prótesis ha resistido 10.000.000 de ciclos, y no ha presentado grietas, ni deformaciones plásticas, lo que permite el uso biomédico de esta aleación de acuerdo con la nueva normativa. También se realizaron pruebas de tracción e inspección por líquidos penetrantes para tomar los parámetros de las propiedades físicas relacionadas. En el análisis con galgas extensiométricas se pudo detectar la deformación puramente elástica que se presentaba con la aplicación de la carga durante los ensayos de fatiga, resultando en una ligera variación de esfuerzos en el sistema de adquisición de datos.A ANVISA realizou alterações em sua regulamentação, estabelecendo novos valores para os ensaios de fadiga, aumentando o número de ciclos que as próteses devem suportar sem apresentar falhas. A norma utilizada é a ABNT NBR 7206-6:2013, na qual todas as próteses que estão sendo comercializadas devem ser testadas nos novos parâmetros. Neste artigo, foram realizados estudos com base na exigência de revalidação da liga metálica Co-26Cr-6Mo (ASTM F75). Foram realizadas caracterizações microestruturais (Microscopias Óptica e Eletrônica de Varredura) e ensaios de fadiga, com o objetivo de obter o comportamento mecânico e as características dos materiais utilizados nas próteses de quadril e compare-los com a norma. O tamanho de grão e os teores de inclusão foram controlados. Da mesa forma, as próteses resistiram a 10.000.000 de ciclos, e não apresentando trincas, nem deformações plásticas, possibilitando o uso biomédico desta liga de acordo com as novas regulamentações. Os ensaios de tração e o ensaio de líquido penetrante também foram realizados para obtenção dos parâmetros das propriedades físicas relacionadas. Na análise com extensômetros foi possível detector a deformação puramente elástica que ocorreu com a aplicação da carga durante os ensaios de fadiga, resultando em uma pequena variação de tensão no sistema de aquisição de dados.Research, Society and Development2022-03-25info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionapplication/pdfhttps://rsdjournal.org/index.php/rsd/article/view/2785410.33448/rsd-v11i4.27854Research, Society and Development; Vol. 11 No. 4; e52011427854Research, Society and Development; Vol. 11 Núm. 4; e52011427854Research, Society and Development; v. 11 n. 4; e520114278542525-3409reponame:Research, Society and Developmentinstname:Universidade Federal de Itajubá (UNIFEI)instacron:UNIFEIenghttps://rsdjournal.org/index.php/rsd/article/view/27854/24158Copyright (c) 2022 Caique Movio Pereira de Souza; Raul Gaspari Santos ; Renato Chaves Souza; Vanderlei Araujo Militão ; Isaias Gouveia Silva; Vanessa Seriacopi; Wilson Carlos da Silva Juniorhttps://creativecommons.org/licenses/by/4.0info:eu-repo/semantics/openAccessSouza, Caique Movio Pereira deSantos , Raul Gaspari Souza, Renato Chaves Militão , Vanderlei Araujo Silva, Isaias Gouveia Seriacopi, Vanessa Silva Junior, Wilson Carlos da 2022-03-27T17:17:09Zoai:ojs.pkp.sfu.ca:article/27854Revistahttps://rsdjournal.org/index.php/rsd/indexPUBhttps://rsdjournal.org/index.php/rsd/oairsd.articles@gmail.com2525-34092525-3409opendoar:2024-01-17T09:45:27.296117Research, Society and Development - Universidade Federal de Itajubá (UNIFEI)false
dc.title.none.fl_str_mv Data analyses of fatigue tests by extensometry in hip prosthesis of the Co-28Cr-6Mo alloy
Análisis de datos de ensayos de fatiga por extensometría en prótesis de cadera de la aleación Co-28Cr-6Mo
Análise de dados de ensaio de fadiga por Extensometria em prótese de quadril da liga Co-28Cr-6Mo
title Data analyses of fatigue tests by extensometry in hip prosthesis of the Co-28Cr-6Mo alloy
spellingShingle Data analyses of fatigue tests by extensometry in hip prosthesis of the Co-28Cr-6Mo alloy
Souza, Caique Movio Pereira de
Prótesis de Cadera
ASTM F75
Prueba de Fatiga
Extensometría
ABNT NBR 7206-6:2013.
Próteses de Quadril
ASTM F75
Ensaio de Fadiga
Extensômetria
ABNT NBR 7206-6:2013.
Hip Prosthesis
ASTM F75
Fatigue Tests
Extensometry
ABNT NBR 7206-6: 2013 standard.
title_short Data analyses of fatigue tests by extensometry in hip prosthesis of the Co-28Cr-6Mo alloy
title_full Data analyses of fatigue tests by extensometry in hip prosthesis of the Co-28Cr-6Mo alloy
title_fullStr Data analyses of fatigue tests by extensometry in hip prosthesis of the Co-28Cr-6Mo alloy
title_full_unstemmed Data analyses of fatigue tests by extensometry in hip prosthesis of the Co-28Cr-6Mo alloy
title_sort Data analyses of fatigue tests by extensometry in hip prosthesis of the Co-28Cr-6Mo alloy
author Souza, Caique Movio Pereira de
author_facet Souza, Caique Movio Pereira de
Santos , Raul Gaspari
Souza, Renato Chaves
Militão , Vanderlei Araujo
Silva, Isaias Gouveia
Seriacopi, Vanessa
Silva Junior, Wilson Carlos da
author_role author
author2 Santos , Raul Gaspari
Souza, Renato Chaves
Militão , Vanderlei Araujo
Silva, Isaias Gouveia
Seriacopi, Vanessa
Silva Junior, Wilson Carlos da
author2_role author
author
author
author
author
author
dc.contributor.author.fl_str_mv Souza, Caique Movio Pereira de
Santos , Raul Gaspari
Souza, Renato Chaves
Militão , Vanderlei Araujo
Silva, Isaias Gouveia
Seriacopi, Vanessa
Silva Junior, Wilson Carlos da
dc.subject.por.fl_str_mv Prótesis de Cadera
ASTM F75
Prueba de Fatiga
Extensometría
ABNT NBR 7206-6:2013.
Próteses de Quadril
ASTM F75
Ensaio de Fadiga
Extensômetria
ABNT NBR 7206-6:2013.
Hip Prosthesis
ASTM F75
Fatigue Tests
Extensometry
ABNT NBR 7206-6: 2013 standard.
topic Prótesis de Cadera
ASTM F75
Prueba de Fatiga
Extensometría
ABNT NBR 7206-6:2013.
Próteses de Quadril
ASTM F75
Ensaio de Fadiga
Extensômetria
ABNT NBR 7206-6:2013.
Hip Prosthesis
ASTM F75
Fatigue Tests
Extensometry
ABNT NBR 7206-6: 2013 standard.
description ANVISA made changes to its regulations, setting new values for the fatigue tests, increasing the number of cycles that prostheses must endure without presenting failures. The standard used is ABNT NBR 7206-6:2013, in which all prostheses being commercialized must be tested in the new parameters. In this article, studies were conducted based on the requirement to revalidate the Co-26Cr-6Mo metallic alloy (ASTM F75). Microstructural characterizations (Optical and Scanning Electron Microscopies), and fatigue tests were carried out, with the aim of obtaining the mechanical behavior and features of the materials used in hip prostheses and comparing them with the standard. The grain size and inclusion contents were found to be controlled. Likewise, the prosthesis has withstand 10,000,000 cycles, and has not shown cracks, nor plastic deformations, enabling the biomedical use of this alloy according to the new regulations. Tensile test and liquid penetrant inspection were also carried out to take the parameters of the related physical properties. In the analysis with strain gauges, it was possible to detect the purely elastic deformation that occurred with the application of the load during fatigue tests, resulting in a slight variation of stress in the data acquisition system.
publishDate 2022
dc.date.none.fl_str_mv 2022-03-25
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv https://rsdjournal.org/index.php/rsd/article/view/27854
10.33448/rsd-v11i4.27854
url https://rsdjournal.org/index.php/rsd/article/view/27854
identifier_str_mv 10.33448/rsd-v11i4.27854
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv https://rsdjournal.org/index.php/rsd/article/view/27854/24158
dc.rights.driver.fl_str_mv https://creativecommons.org/licenses/by/4.0
info:eu-repo/semantics/openAccess
rights_invalid_str_mv https://creativecommons.org/licenses/by/4.0
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Research, Society and Development
publisher.none.fl_str_mv Research, Society and Development
dc.source.none.fl_str_mv Research, Society and Development; Vol. 11 No. 4; e52011427854
Research, Society and Development; Vol. 11 Núm. 4; e52011427854
Research, Society and Development; v. 11 n. 4; e52011427854
2525-3409
reponame:Research, Society and Development
instname:Universidade Federal de Itajubá (UNIFEI)
instacron:UNIFEI
instname_str Universidade Federal de Itajubá (UNIFEI)
instacron_str UNIFEI
institution UNIFEI
reponame_str Research, Society and Development
collection Research, Society and Development
repository.name.fl_str_mv Research, Society and Development - Universidade Federal de Itajubá (UNIFEI)
repository.mail.fl_str_mv rsd.articles@gmail.com
_version_ 1797052793972850688