Estudo da otimização em mamografia para mamas espessas utilizando diferentes filtrações

Detalhes bibliográficos
Autor(a) principal: Azevedo, Davi Silveira
Data de Publicação: 2022
Outros Autores: Campos, Laélia, Estácio, Marcela Costa Alcântara, Ferreira, Cássio Costa, Valença, João Vinícius Batista, Contassot, Raíssa Xavier
Tipo de documento: Artigo
Idioma: por
Título da fonte: Research, Society and Development
Texto Completo: https://rsdjournal.org/index.php/rsd/article/view/36357
Resumo: Mammography is the gold standard exam for breast cancer screening, being recommended by the Brazilian Ministry of Health every 2 years for women aged 50 to 69 years. As there is a great demand in diagnostic imaging services and due to their clinical importance, quality control and the optimization process are indispensable tools to guarantee the quality of the mammography exam. The mammography unit evaluated in this work is limited, in automatic mode, to a maximum voltage of 30 kVp, causing damage to images of thick breasts, in which higher voltages are required to produce an image with diagnostic quality. Thus, the present study aimed to evaluate the contrast-to-noise ratio (CNR), mean glandular dose (MGD) and image quality, in an optimization process directed to the use of more energetic beams in thick breasts (equivalent to 7 cm of PMMA). The results showed that with the use of manual mode to increase the beam energy (32 and 35 kVp), the RCRrel was higher, consequently increasing the DGM. A downward trend in the Figure of Merit (FOM) was observed as the current-time product (mAs) increased sufficiently to meet the minimum national requirement. Regarding the image quality compared to the use of automatic mode, it was possible to observe a greater amount of microcalcification groups and fibers.
id UNIFEI_a98f88e2e95ab0be5daee87b122698f9
oai_identifier_str oai:ojs.pkp.sfu.ca:article/36357
network_acronym_str UNIFEI
network_name_str Research, Society and Development
repository_id_str
spelling Estudo da otimização em mamografia para mamas espessas utilizando diferentes filtrações Study of optimization in mammography for thick breasts using different filtrationsEstudio de optimización en mamografía para mamas gruesas utilizando diferentes filtracionesControl de calidadOptimizaciónDosisMamografíaCáncer de mama.Controle de qualidadeOtimizaçãoDoseMamografiaCâncer de mama.Quality controlOptimizationDoseMamographyBreast cancer.Mammography is the gold standard exam for breast cancer screening, being recommended by the Brazilian Ministry of Health every 2 years for women aged 50 to 69 years. As there is a great demand in diagnostic imaging services and due to their clinical importance, quality control and the optimization process are indispensable tools to guarantee the quality of the mammography exam. The mammography unit evaluated in this work is limited, in automatic mode, to a maximum voltage of 30 kVp, causing damage to images of thick breasts, in which higher voltages are required to produce an image with diagnostic quality. Thus, the present study aimed to evaluate the contrast-to-noise ratio (CNR), mean glandular dose (MGD) and image quality, in an optimization process directed to the use of more energetic beams in thick breasts (equivalent to 7 cm of PMMA). The results showed that with the use of manual mode to increase the beam energy (32 and 35 kVp), the RCRrel was higher, consequently increasing the DGM. A downward trend in the Figure of Merit (FOM) was observed as the current-time product (mAs) increased sufficiently to meet the minimum national requirement. Regarding the image quality compared to the use of automatic mode, it was possible to observe a greater amount of microcalcification groups and fibers.A mamografia é o exame padrão ouro para rastreamento do câncer de mama, sendo recomendada pelo Ministério da Saúde do Brasil a cada 2 anos para mulheres de 50 a 69 anos. Por ter uma grande demanda nos serviços de diagnóstico por imagem e pela sua importância clínica, o controle de qualidade e o processo de otimização são ferramentas indispensáveis para a garantia da qualidade do exame de mamografia. O mamógrafo avaliado neste trabalho é limitado, no modo automático, a uma tensão máxima de 30 kVp, acarretando um prejuízo para imagens de mamas espessas, em que é necessário o uso de tensões maiores para produzir uma imagem com qualidade diagnóstica. Desta forma, o presente trabalho teve como objetivo avaliar a relação contraste-ruído (RCR), a dose glandular média (DGM) e a qualidade da imagem, em um processo de otimização direcionado para o uso de feixes mais energéticos em mamas espessas (equivalentes a 7 cm de PMMA). Os resultados mostraram que com o uso do modo manual para aumentar a energia do feixe (32 e 35 kVp), a RCRrel foi maior, aumentando consequentemente a DGM. Foi observada uma tendência de redução da Figura de Mérito (FOM) à medida que o produto corrente-tempo (mAs) aumentou de forma suficiente para atender à exigência nacional mínima. Em relação à qualidade da imagem em comparação ao uso do modo automático, foi possível observar uma maior quantidade de grupos de microcalcificações e fibras.La mamografía es el examen estándar de oro para la detección del cáncer de mama, recomendado por el Ministerio de Salud de Brasil cada 2 años para mujeres de 50 a 69 años. Debido a que existe una gran demanda en los servicios de diagnóstico por imágenes y por su importancia clínica, el control de calidad y el proceso de optimización son herramientas indispensables para garantizar la calidad del examen de mamografía. La unidad de mamografía evaluada en este trabajo está limitada, en modo automático, a un voltaje máximo de 30 kVp, provocando daños en las imágenes de mamas gruesas, en las que se requieren voltajes más altos para producir una imagen con calidad diagnóstica. Así, el presente estudio tuvo como objetivo evaluar la relación contraste-ruido (CNR), dosis glandular media (MGD) y calidad de imagen, en un proceso de optimización dirigido al uso de haces más energéticos en mamas gruesas (equivalentes a 7 cm de PMMA). Los resultados mostraron que con el uso del modo manual para aumentar la energía del haz (32 y 35 kVp), la RCRrel fue mayor, aumentando consecuentemente la DGM. Se observó una tendencia a la baja en la Figura de Mérito (FOM) a medida que el producto de tiempo actual (mAs) aumentó lo suficiente como para cumplir con el requisito mínimo nacional. En cuanto a la calidad de imagen en comparación con el uso del modo automático, se pudo observar una mayor cantidad de grupos de microcalcificación y fibras.Research, Society and Development2022-10-31info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionapplication/pdfhttps://rsdjournal.org/index.php/rsd/article/view/3635710.33448/rsd-v11i14.36357Research, Society and Development; Vol. 11 No. 14; e405111436357Research, Society and Development; Vol. 11 Núm. 14; e405111436357Research, Society and Development; v. 11 n. 14; e4051114363572525-3409reponame:Research, Society and Developmentinstname:Universidade Federal de Itajubá (UNIFEI)instacron:UNIFEIporhttps://rsdjournal.org/index.php/rsd/article/view/36357/30477Copyright (c) 2022 Davi Silveira Azevedo; Laélia Campos; Marcela Costa Alcântara Estácio; Cássio Costa Ferreira; João Vinícius Batista Valença; Raíssa Xavier Contassothttps://creativecommons.org/licenses/by/4.0info:eu-repo/semantics/openAccessAzevedo, Davi Silveira Campos, Laélia Estácio, Marcela Costa Alcântara Ferreira, Cássio Costa Valença, João Vinícius Batista Contassot, Raíssa Xavier2022-11-08T13:36:27Zoai:ojs.pkp.sfu.ca:article/36357Revistahttps://rsdjournal.org/index.php/rsd/indexPUBhttps://rsdjournal.org/index.php/rsd/oairsd.articles@gmail.com2525-34092525-3409opendoar:2024-01-17T09:50:57.122010Research, Society and Development - Universidade Federal de Itajubá (UNIFEI)false
dc.title.none.fl_str_mv Estudo da otimização em mamografia para mamas espessas utilizando diferentes filtrações
Study of optimization in mammography for thick breasts using different filtrations
Estudio de optimización en mamografía para mamas gruesas utilizando diferentes filtraciones
title Estudo da otimização em mamografia para mamas espessas utilizando diferentes filtrações
spellingShingle Estudo da otimização em mamografia para mamas espessas utilizando diferentes filtrações
Azevedo, Davi Silveira
Control de calidad
Optimización
Dosis
Mamografía
Cáncer de mama.
Controle de qualidade
Otimização
Dose
Mamografia
Câncer de mama.
Quality control
Optimization
Dose
Mamography
Breast cancer.
title_short Estudo da otimização em mamografia para mamas espessas utilizando diferentes filtrações
title_full Estudo da otimização em mamografia para mamas espessas utilizando diferentes filtrações
title_fullStr Estudo da otimização em mamografia para mamas espessas utilizando diferentes filtrações
title_full_unstemmed Estudo da otimização em mamografia para mamas espessas utilizando diferentes filtrações
title_sort Estudo da otimização em mamografia para mamas espessas utilizando diferentes filtrações
author Azevedo, Davi Silveira
author_facet Azevedo, Davi Silveira
Campos, Laélia
Estácio, Marcela Costa Alcântara
Ferreira, Cássio Costa
Valença, João Vinícius Batista
Contassot, Raíssa Xavier
author_role author
author2 Campos, Laélia
Estácio, Marcela Costa Alcântara
Ferreira, Cássio Costa
Valença, João Vinícius Batista
Contassot, Raíssa Xavier
author2_role author
author
author
author
author
dc.contributor.author.fl_str_mv Azevedo, Davi Silveira
Campos, Laélia
Estácio, Marcela Costa Alcântara
Ferreira, Cássio Costa
Valença, João Vinícius Batista
Contassot, Raíssa Xavier
dc.subject.por.fl_str_mv Control de calidad
Optimización
Dosis
Mamografía
Cáncer de mama.
Controle de qualidade
Otimização
Dose
Mamografia
Câncer de mama.
Quality control
Optimization
Dose
Mamography
Breast cancer.
topic Control de calidad
Optimización
Dosis
Mamografía
Cáncer de mama.
Controle de qualidade
Otimização
Dose
Mamografia
Câncer de mama.
Quality control
Optimization
Dose
Mamography
Breast cancer.
description Mammography is the gold standard exam for breast cancer screening, being recommended by the Brazilian Ministry of Health every 2 years for women aged 50 to 69 years. As there is a great demand in diagnostic imaging services and due to their clinical importance, quality control and the optimization process are indispensable tools to guarantee the quality of the mammography exam. The mammography unit evaluated in this work is limited, in automatic mode, to a maximum voltage of 30 kVp, causing damage to images of thick breasts, in which higher voltages are required to produce an image with diagnostic quality. Thus, the present study aimed to evaluate the contrast-to-noise ratio (CNR), mean glandular dose (MGD) and image quality, in an optimization process directed to the use of more energetic beams in thick breasts (equivalent to 7 cm of PMMA). The results showed that with the use of manual mode to increase the beam energy (32 and 35 kVp), the RCRrel was higher, consequently increasing the DGM. A downward trend in the Figure of Merit (FOM) was observed as the current-time product (mAs) increased sufficiently to meet the minimum national requirement. Regarding the image quality compared to the use of automatic mode, it was possible to observe a greater amount of microcalcification groups and fibers.
publishDate 2022
dc.date.none.fl_str_mv 2022-10-31
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv https://rsdjournal.org/index.php/rsd/article/view/36357
10.33448/rsd-v11i14.36357
url https://rsdjournal.org/index.php/rsd/article/view/36357
identifier_str_mv 10.33448/rsd-v11i14.36357
dc.language.iso.fl_str_mv por
language por
dc.relation.none.fl_str_mv https://rsdjournal.org/index.php/rsd/article/view/36357/30477
dc.rights.driver.fl_str_mv https://creativecommons.org/licenses/by/4.0
info:eu-repo/semantics/openAccess
rights_invalid_str_mv https://creativecommons.org/licenses/by/4.0
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Research, Society and Development
publisher.none.fl_str_mv Research, Society and Development
dc.source.none.fl_str_mv Research, Society and Development; Vol. 11 No. 14; e405111436357
Research, Society and Development; Vol. 11 Núm. 14; e405111436357
Research, Society and Development; v. 11 n. 14; e405111436357
2525-3409
reponame:Research, Society and Development
instname:Universidade Federal de Itajubá (UNIFEI)
instacron:UNIFEI
instname_str Universidade Federal de Itajubá (UNIFEI)
instacron_str UNIFEI
institution UNIFEI
reponame_str Research, Society and Development
collection Research, Society and Development
repository.name.fl_str_mv Research, Society and Development - Universidade Federal de Itajubá (UNIFEI)
repository.mail.fl_str_mv rsd.articles@gmail.com
_version_ 1797052773495209984