Effects of cuff size and position on the agreement between arterial blood pressure measured by Doppler ultrasound and through a dorsal pedal artery catheter in anesthetized cats

Detalhes bibliográficos
Autor(a) principal: Cerejo, Sofia A. [UNESP]
Data de Publicação: 2020
Outros Autores: Teixeira-Neto, Francisco J. [UNESP], Garofalo, Natache A. [UNESP], Pimenta, Eutálio LM., Zanuzzo, Felipe S. [UNESP], Klein, Adriana V. [UNESP]
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Institucional da UNESP
Texto Completo: http://dx.doi.org/10.1016/j.vaa.2019.11.001
http://hdl.handle.net/11449/198468
Resumo: Objective: To compare the effects of cuff size/position on the agreement between arterial blood pressure measured by Doppler ultrasound (ABPDoppler) and dorsal pedal artery catheter measurements of systolic (SAPinvasive) and mean arterial pressure (MAPinvasive) in anesthetized cats. Study design: Prospective study. Animals: A total of eight cats (3.0–3.8 kg) for neutering. Methods: During isoflurane anesthesia, before surgery, changes in end-tidal isoflurane concentrations and/or administration of dopamine were performed to achieve SAPinvasive within 60–150 mmHg. Cuff sizes 1, 2 and 3 (bladder width: 20, 25 and 35 mm, respectively) were placed on distal third of the antebrachium, above the tarsus and below the tarsus for ABPDoppler measurements. Agreement between ABPDoppler and SAPinvasive or between ABPDoppler and MAPinvasive was compared with reference standards for noninvasive blood pressure devices used in humans and small animals. Results: Mean bias and precision (±standard deviation) between ABPDoppler and SAPinvasive met veterinary standards (≤10 ± 15 mmHg), but not human standards (≤5 ± 8 mmHg), with cuffs 1 and 2 placed on the thoracic limb (7.4 ± 13.9 and –5.8 ± 9.5 mmHg, respectively), and with cuff 2 placed proximal to the tarsus (7.2 ± 12.4 mmHg). Cuff width-to-limb circumference ratios resulting in acceptable agreement between ABPDoppler and SAPinvasive were 0.31 ± 0.04 (cuff 1) and 0.42 ± 0.05 (cuff 2) on the thoracic limb, and 0.43 ± 0.05 (cuff 2) above the tarsus. ABPDoppler showed no acceptable agreement with MAPinvasive by any reference standard. Conclusions and clinical relevance: The agreement between ABPDoppler and SAPinvasive can be optimized by placing the occlusive cuff on the distal third of the antebrachium and above the tarsus. In these locations, cuff width should approach 40% of limb circumference to provide clinically acceptable estimations of SAPinvasive. Doppler ultrasound cannot be used to estimate MAPinvasive in cats.
id UNSP_14c7bdabf273a393edf393822cebe724
oai_identifier_str oai:repositorio.unesp.br:11449/198468
network_acronym_str UNSP
network_name_str Repositório Institucional da UNESP
repository_id_str 2946
spelling Effects of cuff size and position on the agreement between arterial blood pressure measured by Doppler ultrasound and through a dorsal pedal artery catheter in anesthetized catscatsDopplernoninvasive blood pressureObjective: To compare the effects of cuff size/position on the agreement between arterial blood pressure measured by Doppler ultrasound (ABPDoppler) and dorsal pedal artery catheter measurements of systolic (SAPinvasive) and mean arterial pressure (MAPinvasive) in anesthetized cats. Study design: Prospective study. Animals: A total of eight cats (3.0–3.8 kg) for neutering. Methods: During isoflurane anesthesia, before surgery, changes in end-tidal isoflurane concentrations and/or administration of dopamine were performed to achieve SAPinvasive within 60–150 mmHg. Cuff sizes 1, 2 and 3 (bladder width: 20, 25 and 35 mm, respectively) were placed on distal third of the antebrachium, above the tarsus and below the tarsus for ABPDoppler measurements. Agreement between ABPDoppler and SAPinvasive or between ABPDoppler and MAPinvasive was compared with reference standards for noninvasive blood pressure devices used in humans and small animals. Results: Mean bias and precision (±standard deviation) between ABPDoppler and SAPinvasive met veterinary standards (≤10 ± 15 mmHg), but not human standards (≤5 ± 8 mmHg), with cuffs 1 and 2 placed on the thoracic limb (7.4 ± 13.9 and –5.8 ± 9.5 mmHg, respectively), and with cuff 2 placed proximal to the tarsus (7.2 ± 12.4 mmHg). Cuff width-to-limb circumference ratios resulting in acceptable agreement between ABPDoppler and SAPinvasive were 0.31 ± 0.04 (cuff 1) and 0.42 ± 0.05 (cuff 2) on the thoracic limb, and 0.43 ± 0.05 (cuff 2) above the tarsus. ABPDoppler showed no acceptable agreement with MAPinvasive by any reference standard. Conclusions and clinical relevance: The agreement between ABPDoppler and SAPinvasive can be optimized by placing the occlusive cuff on the distal third of the antebrachium and above the tarsus. In these locations, cuff width should approach 40% of limb circumference to provide clinically acceptable estimations of SAPinvasive. Doppler ultrasound cannot be used to estimate MAPinvasive in cats.Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)Faculdade de Medicina Veterinária e Zootecnia Universidade Estadual Paulista (UNESP)Escola de Veterinária Universidade Federal de Minas Gerais (UFMG)Faculdade de Medicina Veterinária e Zootecnia Universidade Estadual Paulista (UNESP)FAPESP: 2012/09837-8Universidade Estadual Paulista (Unesp)Universidade Federal de Minas Gerais (UFMG)Cerejo, Sofia A. [UNESP]Teixeira-Neto, Francisco J. [UNESP]Garofalo, Natache A. [UNESP]Pimenta, Eutálio LM.Zanuzzo, Felipe S. [UNESP]Klein, Adriana V. [UNESP]2020-12-12T01:13:40Z2020-12-12T01:13:40Z2020-03-01info:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/article191-199http://dx.doi.org/10.1016/j.vaa.2019.11.001Veterinary Anaesthesia and Analgesia, v. 47, n. 2, p. 191-199, 2020.1467-29951467-2987http://hdl.handle.net/11449/19846810.1016/j.vaa.2019.11.0012-s2.0-85078757132Scopusreponame:Repositório Institucional da UNESPinstname:Universidade Estadual Paulista (UNESP)instacron:UNESPengVeterinary Anaesthesia and Analgesiainfo:eu-repo/semantics/openAccess2021-10-22T12:25:16Zoai:repositorio.unesp.br:11449/198468Repositório InstitucionalPUBhttp://repositorio.unesp.br/oai/requestopendoar:29462024-08-05T22:55:33.786099Repositório Institucional da UNESP - Universidade Estadual Paulista (UNESP)false
dc.title.none.fl_str_mv Effects of cuff size and position on the agreement between arterial blood pressure measured by Doppler ultrasound and through a dorsal pedal artery catheter in anesthetized cats
title Effects of cuff size and position on the agreement between arterial blood pressure measured by Doppler ultrasound and through a dorsal pedal artery catheter in anesthetized cats
spellingShingle Effects of cuff size and position on the agreement between arterial blood pressure measured by Doppler ultrasound and through a dorsal pedal artery catheter in anesthetized cats
Cerejo, Sofia A. [UNESP]
cats
Doppler
noninvasive blood pressure
title_short Effects of cuff size and position on the agreement between arterial blood pressure measured by Doppler ultrasound and through a dorsal pedal artery catheter in anesthetized cats
title_full Effects of cuff size and position on the agreement between arterial blood pressure measured by Doppler ultrasound and through a dorsal pedal artery catheter in anesthetized cats
title_fullStr Effects of cuff size and position on the agreement between arterial blood pressure measured by Doppler ultrasound and through a dorsal pedal artery catheter in anesthetized cats
title_full_unstemmed Effects of cuff size and position on the agreement between arterial blood pressure measured by Doppler ultrasound and through a dorsal pedal artery catheter in anesthetized cats
title_sort Effects of cuff size and position on the agreement between arterial blood pressure measured by Doppler ultrasound and through a dorsal pedal artery catheter in anesthetized cats
author Cerejo, Sofia A. [UNESP]
author_facet Cerejo, Sofia A. [UNESP]
Teixeira-Neto, Francisco J. [UNESP]
Garofalo, Natache A. [UNESP]
Pimenta, Eutálio LM.
Zanuzzo, Felipe S. [UNESP]
Klein, Adriana V. [UNESP]
author_role author
author2 Teixeira-Neto, Francisco J. [UNESP]
Garofalo, Natache A. [UNESP]
Pimenta, Eutálio LM.
Zanuzzo, Felipe S. [UNESP]
Klein, Adriana V. [UNESP]
author2_role author
author
author
author
author
dc.contributor.none.fl_str_mv Universidade Estadual Paulista (Unesp)
Universidade Federal de Minas Gerais (UFMG)
dc.contributor.author.fl_str_mv Cerejo, Sofia A. [UNESP]
Teixeira-Neto, Francisco J. [UNESP]
Garofalo, Natache A. [UNESP]
Pimenta, Eutálio LM.
Zanuzzo, Felipe S. [UNESP]
Klein, Adriana V. [UNESP]
dc.subject.por.fl_str_mv cats
Doppler
noninvasive blood pressure
topic cats
Doppler
noninvasive blood pressure
description Objective: To compare the effects of cuff size/position on the agreement between arterial blood pressure measured by Doppler ultrasound (ABPDoppler) and dorsal pedal artery catheter measurements of systolic (SAPinvasive) and mean arterial pressure (MAPinvasive) in anesthetized cats. Study design: Prospective study. Animals: A total of eight cats (3.0–3.8 kg) for neutering. Methods: During isoflurane anesthesia, before surgery, changes in end-tidal isoflurane concentrations and/or administration of dopamine were performed to achieve SAPinvasive within 60–150 mmHg. Cuff sizes 1, 2 and 3 (bladder width: 20, 25 and 35 mm, respectively) were placed on distal third of the antebrachium, above the tarsus and below the tarsus for ABPDoppler measurements. Agreement between ABPDoppler and SAPinvasive or between ABPDoppler and MAPinvasive was compared with reference standards for noninvasive blood pressure devices used in humans and small animals. Results: Mean bias and precision (±standard deviation) between ABPDoppler and SAPinvasive met veterinary standards (≤10 ± 15 mmHg), but not human standards (≤5 ± 8 mmHg), with cuffs 1 and 2 placed on the thoracic limb (7.4 ± 13.9 and –5.8 ± 9.5 mmHg, respectively), and with cuff 2 placed proximal to the tarsus (7.2 ± 12.4 mmHg). Cuff width-to-limb circumference ratios resulting in acceptable agreement between ABPDoppler and SAPinvasive were 0.31 ± 0.04 (cuff 1) and 0.42 ± 0.05 (cuff 2) on the thoracic limb, and 0.43 ± 0.05 (cuff 2) above the tarsus. ABPDoppler showed no acceptable agreement with MAPinvasive by any reference standard. Conclusions and clinical relevance: The agreement between ABPDoppler and SAPinvasive can be optimized by placing the occlusive cuff on the distal third of the antebrachium and above the tarsus. In these locations, cuff width should approach 40% of limb circumference to provide clinically acceptable estimations of SAPinvasive. Doppler ultrasound cannot be used to estimate MAPinvasive in cats.
publishDate 2020
dc.date.none.fl_str_mv 2020-12-12T01:13:40Z
2020-12-12T01:13:40Z
2020-03-01
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://dx.doi.org/10.1016/j.vaa.2019.11.001
Veterinary Anaesthesia and Analgesia, v. 47, n. 2, p. 191-199, 2020.
1467-2995
1467-2987
http://hdl.handle.net/11449/198468
10.1016/j.vaa.2019.11.001
2-s2.0-85078757132
url http://dx.doi.org/10.1016/j.vaa.2019.11.001
http://hdl.handle.net/11449/198468
identifier_str_mv Veterinary Anaesthesia and Analgesia, v. 47, n. 2, p. 191-199, 2020.
1467-2995
1467-2987
10.1016/j.vaa.2019.11.001
2-s2.0-85078757132
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv Veterinary Anaesthesia and Analgesia
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv 191-199
dc.source.none.fl_str_mv Scopus
reponame:Repositório Institucional da UNESP
instname:Universidade Estadual Paulista (UNESP)
instacron:UNESP
instname_str Universidade Estadual Paulista (UNESP)
instacron_str UNESP
institution UNESP
reponame_str Repositório Institucional da UNESP
collection Repositório Institucional da UNESP
repository.name.fl_str_mv Repositório Institucional da UNESP - Universidade Estadual Paulista (UNESP)
repository.mail.fl_str_mv
_version_ 1808129474008449024