Evaluation of implants with different macrostructures in type I Bone-Pre-Clinical study in rabbits

Detalhes bibliográficos
Autor(a) principal: de Carvalho Silva Leocádio, Amanda [UNESP]
Data de Publicação: 2020
Outros Autores: Júnior, Matusalém Silva, de Oliveira, Guilherme José Pimentel Lopes, da Col Santos Pinto, Gustavo [UNESP], Faeda, Rafael Silveira, Padovan, Luis Eduardo Marques, Júnior, Élcio Marcantonio [UNESP]
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Institucional da UNESP
Texto Completo: http://dx.doi.org/10.3390/ma13071521
http://hdl.handle.net/11449/200251
Resumo: The objective of this study was to assess the primary stability and the osseointegration process in implants with different macrostructures (Cylindrical vs. Hybrid Conical) in rabbit tibiae. Twenty-four (24) rabbits were used, divided into 3 experimental periods (2, 4 and 8 weeks) with 8 animals each. Each animal bilaterally received 2 implants from each group in the tibial metaphysis: Cylindrical Implant (CI) and Hybrid Conical Implant (HCI). All implants were assessed for insertion torque. After the experimental periods, one of the implants in each group was submitted to the removal counter-torque test and descriptive histological analysis while the other implant was used for microtomographic and histometric analysis (%Bone-Implant Contact). HCI implants showed higher insertion torque (32.93 ± 10.61 Ncm vs. 27.99 ± 7.80 Ncm) and higher % of bone-implant contact in the 8-week period (79.08 ± 11.31% vs. 59.72 ± 11.29%) than CI implants. However, CI implants showed higher values of removal counter-torque than HCI implants in the 8-week period (91.05 ± 9.32 Ncm vs. 68.62 ± 13.70 Ncm). There were no differences between groups regarding microtomographic data. It can be concluded that HCI implants showed greater insertion torque and bone-implant contact in relation to CI implants in the period of 8 weeks when installed in cortical bone of rabbits.
id UNSP_6d9472f2d67b57611a815f1a6b659c55
oai_identifier_str oai:repositorio.unesp.br:11449/200251
network_acronym_str UNSP
network_name_str Repositório Institucional da UNESP
repository_id_str 2946
spelling Evaluation of implants with different macrostructures in type I Bone-Pre-Clinical study in rabbitsDental implantsMacrostructureOsseointegrationThe objective of this study was to assess the primary stability and the osseointegration process in implants with different macrostructures (Cylindrical vs. Hybrid Conical) in rabbit tibiae. Twenty-four (24) rabbits were used, divided into 3 experimental periods (2, 4 and 8 weeks) with 8 animals each. Each animal bilaterally received 2 implants from each group in the tibial metaphysis: Cylindrical Implant (CI) and Hybrid Conical Implant (HCI). All implants were assessed for insertion torque. After the experimental periods, one of the implants in each group was submitted to the removal counter-torque test and descriptive histological analysis while the other implant was used for microtomographic and histometric analysis (%Bone-Implant Contact). HCI implants showed higher insertion torque (32.93 ± 10.61 Ncm vs. 27.99 ± 7.80 Ncm) and higher % of bone-implant contact in the 8-week period (79.08 ± 11.31% vs. 59.72 ± 11.29%) than CI implants. However, CI implants showed higher values of removal counter-torque than HCI implants in the 8-week period (91.05 ± 9.32 Ncm vs. 68.62 ± 13.70 Ncm). There were no differences between groups regarding microtomographic data. It can be concluded that HCI implants showed greater insertion torque and bone-implant contact in relation to CI implants in the period of 8 weeks when installed in cortical bone of rabbits.Department of Diagnosis and Surgery School of Dentistry at Araraquara Sao Paulo State University (UNESP)Post Graduation Course in Implantology Latin American Institute of Dental Research (ILAPEO)Department of Periodontology School of Dentistry at Uberlândia Federal University of UberlândiaPost Graduation Course in Odontology University of Araraquara/UNIARADepartment of Diagnosis and Surgery School of Dentistry at Araraquara Sao Paulo State University (UNESP)Universidade Estadual Paulista (Unesp)Latin American Institute of Dental Research (ILAPEO)Universidade Federal de Uberlândia (UFU)University of Araraquara/UNIARAde Carvalho Silva Leocádio, Amanda [UNESP]Júnior, Matusalém Silvade Oliveira, Guilherme José Pimentel Lopesda Col Santos Pinto, Gustavo [UNESP]Faeda, Rafael SilveiraPadovan, Luis Eduardo MarquesJúnior, Élcio Marcantonio [UNESP]2020-12-12T02:01:36Z2020-12-12T02:01:36Z2020-04-01info:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articlehttp://dx.doi.org/10.3390/ma13071521Materials, v. 13, n. 7, 2020.1996-1944http://hdl.handle.net/11449/20025110.3390/ma130715212-s2.0-85082968375Scopusreponame:Repositório Institucional da UNESPinstname:Universidade Estadual Paulista (UNESP)instacron:UNESPengMaterialsinfo:eu-repo/semantics/openAccess2021-10-23T12:39:35Zoai:repositorio.unesp.br:11449/200251Repositório InstitucionalPUBhttp://repositorio.unesp.br/oai/requestopendoar:29462021-10-23T12:39:35Repositório Institucional da UNESP - Universidade Estadual Paulista (UNESP)false
dc.title.none.fl_str_mv Evaluation of implants with different macrostructures in type I Bone-Pre-Clinical study in rabbits
title Evaluation of implants with different macrostructures in type I Bone-Pre-Clinical study in rabbits
spellingShingle Evaluation of implants with different macrostructures in type I Bone-Pre-Clinical study in rabbits
de Carvalho Silva Leocádio, Amanda [UNESP]
Dental implants
Macrostructure
Osseointegration
title_short Evaluation of implants with different macrostructures in type I Bone-Pre-Clinical study in rabbits
title_full Evaluation of implants with different macrostructures in type I Bone-Pre-Clinical study in rabbits
title_fullStr Evaluation of implants with different macrostructures in type I Bone-Pre-Clinical study in rabbits
title_full_unstemmed Evaluation of implants with different macrostructures in type I Bone-Pre-Clinical study in rabbits
title_sort Evaluation of implants with different macrostructures in type I Bone-Pre-Clinical study in rabbits
author de Carvalho Silva Leocádio, Amanda [UNESP]
author_facet de Carvalho Silva Leocádio, Amanda [UNESP]
Júnior, Matusalém Silva
de Oliveira, Guilherme José Pimentel Lopes
da Col Santos Pinto, Gustavo [UNESP]
Faeda, Rafael Silveira
Padovan, Luis Eduardo Marques
Júnior, Élcio Marcantonio [UNESP]
author_role author
author2 Júnior, Matusalém Silva
de Oliveira, Guilherme José Pimentel Lopes
da Col Santos Pinto, Gustavo [UNESP]
Faeda, Rafael Silveira
Padovan, Luis Eduardo Marques
Júnior, Élcio Marcantonio [UNESP]
author2_role author
author
author
author
author
author
dc.contributor.none.fl_str_mv Universidade Estadual Paulista (Unesp)
Latin American Institute of Dental Research (ILAPEO)
Universidade Federal de Uberlândia (UFU)
University of Araraquara/UNIARA
dc.contributor.author.fl_str_mv de Carvalho Silva Leocádio, Amanda [UNESP]
Júnior, Matusalém Silva
de Oliveira, Guilherme José Pimentel Lopes
da Col Santos Pinto, Gustavo [UNESP]
Faeda, Rafael Silveira
Padovan, Luis Eduardo Marques
Júnior, Élcio Marcantonio [UNESP]
dc.subject.por.fl_str_mv Dental implants
Macrostructure
Osseointegration
topic Dental implants
Macrostructure
Osseointegration
description The objective of this study was to assess the primary stability and the osseointegration process in implants with different macrostructures (Cylindrical vs. Hybrid Conical) in rabbit tibiae. Twenty-four (24) rabbits were used, divided into 3 experimental periods (2, 4 and 8 weeks) with 8 animals each. Each animal bilaterally received 2 implants from each group in the tibial metaphysis: Cylindrical Implant (CI) and Hybrid Conical Implant (HCI). All implants were assessed for insertion torque. After the experimental periods, one of the implants in each group was submitted to the removal counter-torque test and descriptive histological analysis while the other implant was used for microtomographic and histometric analysis (%Bone-Implant Contact). HCI implants showed higher insertion torque (32.93 ± 10.61 Ncm vs. 27.99 ± 7.80 Ncm) and higher % of bone-implant contact in the 8-week period (79.08 ± 11.31% vs. 59.72 ± 11.29%) than CI implants. However, CI implants showed higher values of removal counter-torque than HCI implants in the 8-week period (91.05 ± 9.32 Ncm vs. 68.62 ± 13.70 Ncm). There were no differences between groups regarding microtomographic data. It can be concluded that HCI implants showed greater insertion torque and bone-implant contact in relation to CI implants in the period of 8 weeks when installed in cortical bone of rabbits.
publishDate 2020
dc.date.none.fl_str_mv 2020-12-12T02:01:36Z
2020-12-12T02:01:36Z
2020-04-01
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://dx.doi.org/10.3390/ma13071521
Materials, v. 13, n. 7, 2020.
1996-1944
http://hdl.handle.net/11449/200251
10.3390/ma13071521
2-s2.0-85082968375
url http://dx.doi.org/10.3390/ma13071521
http://hdl.handle.net/11449/200251
identifier_str_mv Materials, v. 13, n. 7, 2020.
1996-1944
10.3390/ma13071521
2-s2.0-85082968375
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv Materials
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.source.none.fl_str_mv Scopus
reponame:Repositório Institucional da UNESP
instname:Universidade Estadual Paulista (UNESP)
instacron:UNESP
instname_str Universidade Estadual Paulista (UNESP)
instacron_str UNESP
institution UNESP
reponame_str Repositório Institucional da UNESP
collection Repositório Institucional da UNESP
repository.name.fl_str_mv Repositório Institucional da UNESP - Universidade Estadual Paulista (UNESP)
repository.mail.fl_str_mv
_version_ 1799964882605441024