Assessment of the interface between implant and abutments of five systems by Scanning Electronic Microscopy

Detalhes bibliográficos
Autor(a) principal: Fujiwara, Carlos Alberto [UNESP]
Data de Publicação: 2009
Outros Autores: Filho, Osvaldo Magro, Oliveira, Nilson T. C. [UNESP], Queiroz, Thallita Pereira, Abla, Marcelo Sabagg [UNESP], Pardini, Luiz Carlos
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Institucional da UNESP
Texto Completo: http://hdl.handle.net/11449/219889
Resumo: Aim: The objective of this study was to assess, by Scanning Electronic Microscopy (SEM), the interface between the platforms of five implant systems and the conic abutments (I) gold ucla type (II) and multi-unit type (III). Materials and methods: The implant systems used and their respective abutments studied were: AS Technology Titanium Fix (I, II, III), Conexão (I, II, III), Neodent (I, II, III), Sterngold Implamed (I, II) and 3i Implant Innovation (I, II). The abutments were screwed to the implants with a 20 Ncm torque. For each sample, 6 points were selected for measurement, 3 on the right and 3 three on the left side: the most external point, the middle point and the most internal point at the interface formed between the implant platform and the abutment. Results: On the outer point there was a variation in measurement from 0 to 11.173 μm, on the middle point there was variation from 0 to 8.314 μm and in the internal point there was a variation from 0 to 15.267 μm. The smallest gaps for abutments I, II and III were obtained with Neodent (0.733 μm), Sterngold Implamed (0.513 μm) and Conexão (0.503 μm), respectively. The Neodent system showed statistically significant differences with the other systems analyzed in the right internal point (p<0.05), while on the left ones differences were not significant only with the the Conexão system (p=0.168); the latter showed statistically significant differences with the other systems except for the 3i system (p=0.311). Conclusion: It is concluded that the Neodent system differed from the other ones in the right and left internal points (p<0.05) except for the Conexão system (p=0.168), which also differed from the other ones, and the 3i system (p=0.311). Maladjustment values obtained are similar to those reported in the literature.
id UNSP_919f445fe1227827f5d861127bfed07c
oai_identifier_str oai:repositorio.unesp.br:11449/219889
network_acronym_str UNSP
network_name_str Repositório Institucional da UNESP
repository_id_str 2946
spelling Assessment of the interface between implant and abutments of five systems by Scanning Electronic MicroscopyAbutmentGapImplantPeri-implantAim: The objective of this study was to assess, by Scanning Electronic Microscopy (SEM), the interface between the platforms of five implant systems and the conic abutments (I) gold ucla type (II) and multi-unit type (III). Materials and methods: The implant systems used and their respective abutments studied were: AS Technology Titanium Fix (I, II, III), Conexão (I, II, III), Neodent (I, II, III), Sterngold Implamed (I, II) and 3i Implant Innovation (I, II). The abutments were screwed to the implants with a 20 Ncm torque. For each sample, 6 points were selected for measurement, 3 on the right and 3 three on the left side: the most external point, the middle point and the most internal point at the interface formed between the implant platform and the abutment. Results: On the outer point there was a variation in measurement from 0 to 11.173 μm, on the middle point there was variation from 0 to 8.314 μm and in the internal point there was a variation from 0 to 15.267 μm. The smallest gaps for abutments I, II and III were obtained with Neodent (0.733 μm), Sterngold Implamed (0.513 μm) and Conexão (0.503 μm), respectively. The Neodent system showed statistically significant differences with the other systems analyzed in the right internal point (p<0.05), while on the left ones differences were not significant only with the the Conexão system (p=0.168); the latter showed statistically significant differences with the other systems except for the 3i system (p=0.311). Conclusion: It is concluded that the Neodent system differed from the other ones in the right and left internal points (p<0.05) except for the Conexão system (p=0.168), which also differed from the other ones, and the 3i system (p=0.311). Maladjustment values obtained are similar to those reported in the literature.São Paulo University State Dental School of Araçatuba UNESP, São PauloOral and Maxillofacial SurgeryDepartment of Morphology, Dentistry and Physiology Dental School of Ribeirão Preto University of São Paulo, USPBiomaterials Group Unesp-AraraquaraSão Paulo University State Dental School of Araçatuba UNESP, São PauloBiomaterials Group Unesp-AraraquaraUniversidade Estadual Paulista (UNESP)Oral and Maxillofacial SurgeryUniversidade de São Paulo (USP)Fujiwara, Carlos Alberto [UNESP]Filho, Osvaldo MagroOliveira, Nilson T. C. [UNESP]Queiroz, Thallita PereiraAbla, Marcelo Sabagg [UNESP]Pardini, Luiz Carlos2022-04-28T18:58:12Z2022-04-28T18:58:12Z2009-08-01info:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/article60-66Journal of Osseointegration, v. 1, n. 2, p. 60-66, 2009.2036-413X2036-4121http://hdl.handle.net/11449/2198892-s2.0-84873932577Scopusreponame:Repositório Institucional da UNESPinstname:Universidade Estadual Paulista (UNESP)instacron:UNESPengJournal of Osseointegrationinfo:eu-repo/semantics/openAccess2022-04-28T18:58:12Zoai:repositorio.unesp.br:11449/219889Repositório InstitucionalPUBhttp://repositorio.unesp.br/oai/requestopendoar:29462024-08-05T21:50:07.049512Repositório Institucional da UNESP - Universidade Estadual Paulista (UNESP)false
dc.title.none.fl_str_mv Assessment of the interface between implant and abutments of five systems by Scanning Electronic Microscopy
title Assessment of the interface between implant and abutments of five systems by Scanning Electronic Microscopy
spellingShingle Assessment of the interface between implant and abutments of five systems by Scanning Electronic Microscopy
Fujiwara, Carlos Alberto [UNESP]
Abutment
Gap
Implant
Peri-implant
title_short Assessment of the interface between implant and abutments of five systems by Scanning Electronic Microscopy
title_full Assessment of the interface between implant and abutments of five systems by Scanning Electronic Microscopy
title_fullStr Assessment of the interface between implant and abutments of five systems by Scanning Electronic Microscopy
title_full_unstemmed Assessment of the interface between implant and abutments of five systems by Scanning Electronic Microscopy
title_sort Assessment of the interface between implant and abutments of five systems by Scanning Electronic Microscopy
author Fujiwara, Carlos Alberto [UNESP]
author_facet Fujiwara, Carlos Alberto [UNESP]
Filho, Osvaldo Magro
Oliveira, Nilson T. C. [UNESP]
Queiroz, Thallita Pereira
Abla, Marcelo Sabagg [UNESP]
Pardini, Luiz Carlos
author_role author
author2 Filho, Osvaldo Magro
Oliveira, Nilson T. C. [UNESP]
Queiroz, Thallita Pereira
Abla, Marcelo Sabagg [UNESP]
Pardini, Luiz Carlos
author2_role author
author
author
author
author
dc.contributor.none.fl_str_mv Universidade Estadual Paulista (UNESP)
Oral and Maxillofacial Surgery
Universidade de São Paulo (USP)
dc.contributor.author.fl_str_mv Fujiwara, Carlos Alberto [UNESP]
Filho, Osvaldo Magro
Oliveira, Nilson T. C. [UNESP]
Queiroz, Thallita Pereira
Abla, Marcelo Sabagg [UNESP]
Pardini, Luiz Carlos
dc.subject.por.fl_str_mv Abutment
Gap
Implant
Peri-implant
topic Abutment
Gap
Implant
Peri-implant
description Aim: The objective of this study was to assess, by Scanning Electronic Microscopy (SEM), the interface between the platforms of five implant systems and the conic abutments (I) gold ucla type (II) and multi-unit type (III). Materials and methods: The implant systems used and their respective abutments studied were: AS Technology Titanium Fix (I, II, III), Conexão (I, II, III), Neodent (I, II, III), Sterngold Implamed (I, II) and 3i Implant Innovation (I, II). The abutments were screwed to the implants with a 20 Ncm torque. For each sample, 6 points were selected for measurement, 3 on the right and 3 three on the left side: the most external point, the middle point and the most internal point at the interface formed between the implant platform and the abutment. Results: On the outer point there was a variation in measurement from 0 to 11.173 μm, on the middle point there was variation from 0 to 8.314 μm and in the internal point there was a variation from 0 to 15.267 μm. The smallest gaps for abutments I, II and III were obtained with Neodent (0.733 μm), Sterngold Implamed (0.513 μm) and Conexão (0.503 μm), respectively. The Neodent system showed statistically significant differences with the other systems analyzed in the right internal point (p<0.05), while on the left ones differences were not significant only with the the Conexão system (p=0.168); the latter showed statistically significant differences with the other systems except for the 3i system (p=0.311). Conclusion: It is concluded that the Neodent system differed from the other ones in the right and left internal points (p<0.05) except for the Conexão system (p=0.168), which also differed from the other ones, and the 3i system (p=0.311). Maladjustment values obtained are similar to those reported in the literature.
publishDate 2009
dc.date.none.fl_str_mv 2009-08-01
2022-04-28T18:58:12Z
2022-04-28T18:58:12Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv Journal of Osseointegration, v. 1, n. 2, p. 60-66, 2009.
2036-413X
2036-4121
http://hdl.handle.net/11449/219889
2-s2.0-84873932577
identifier_str_mv Journal of Osseointegration, v. 1, n. 2, p. 60-66, 2009.
2036-413X
2036-4121
2-s2.0-84873932577
url http://hdl.handle.net/11449/219889
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv Journal of Osseointegration
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv 60-66
dc.source.none.fl_str_mv Scopus
reponame:Repositório Institucional da UNESP
instname:Universidade Estadual Paulista (UNESP)
instacron:UNESP
instname_str Universidade Estadual Paulista (UNESP)
instacron_str UNESP
institution UNESP
reponame_str Repositório Institucional da UNESP
collection Repositório Institucional da UNESP
repository.name.fl_str_mv Repositório Institucional da UNESP - Universidade Estadual Paulista (UNESP)
repository.mail.fl_str_mv
_version_ 1808129364389265408