Accuracy evaluation of computer-guided implant surgery associated with prototyped surgical guides

Detalhes bibliográficos
Autor(a) principal: Cunha, Rodrigo Moreira
Data de Publicação: 2020
Outros Autores: Souza, Francisley Ávila [UNESP], Hadad, Henrique [UNESP], Poli, Pier Paolo, Maiorana, Carlo, Carvalho, Paulo Sérgio Perri
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Institucional da UNESP
Texto Completo: http://dx.doi.org/10.1016/j.prosdent.2019.07.010
http://hdl.handle.net/11449/198568
Resumo: Statement of problem: Correct implant placement by means of a pretreatment planning software program is still subject to deviations between the planned and achieved implant positions. Inaccuracy at this level may have drastic consequences, including neurovascular trauma. Further data are therefore needed to evaluate the accuracy of such computer-guided implant planning software programs. Purpose: The purpose of this clinical study was to evaluate the accuracy of computer-guided implant surgery associated with prototyped surgical guides. Material and methods: Cone beam computed tomography (CBCT) scans were made of the participants with a tomographic guide to merge anatomic and prosthetic data. This allowed virtual planning with a prosthetically guided approach respecting the anatomy of the participant. A prototyped surgical guide was then fabricated from the virtual plan, determining the intrasurgical position of the implants. Flapless guided implant surgery was carried out according to the manufacturer's instructions. A second CBCT was made 30 days after the surgery, to enable overlapping of the data from before and after the implant placement. The angular, coronal, central, and apical deviations of the placed implants were measured and compared with those virtually planned. The data were submitted to descriptive statistic and intraclass correlation coefficient (ICC), analysis of variance, and the Student t test (α=.05). Results: A total of 61 implants were analyzed. The mean angular deviation was 2.04 degrees. The mean coronal, central, and apical linear deviations were 0.68 mm, 0.72 mm, and 0.82 mm, respectively. No statistically significant difference was found between the virtual and the real position of the implants inserted. A tendency toward a greater absolute error was observed in the mandible than in the maxilla in terms of angular (P=.047), central (P=.043), and coronal error (P=.031). Conclusions: Flapless computer-guided implant surgery with virtual planning had some angular and linear deviations; nevertheless, this technique should be acceptably safe and accurate.
id UNSP_e3c5ff58c794f7b322779aee03a457c6
oai_identifier_str oai:repositorio.unesp.br:11449/198568
network_acronym_str UNSP
network_name_str Repositório Institucional da UNESP
repository_id_str 2946
spelling Accuracy evaluation of computer-guided implant surgery associated with prototyped surgical guidesStatement of problem: Correct implant placement by means of a pretreatment planning software program is still subject to deviations between the planned and achieved implant positions. Inaccuracy at this level may have drastic consequences, including neurovascular trauma. Further data are therefore needed to evaluate the accuracy of such computer-guided implant planning software programs. Purpose: The purpose of this clinical study was to evaluate the accuracy of computer-guided implant surgery associated with prototyped surgical guides. Material and methods: Cone beam computed tomography (CBCT) scans were made of the participants with a tomographic guide to merge anatomic and prosthetic data. This allowed virtual planning with a prosthetically guided approach respecting the anatomy of the participant. A prototyped surgical guide was then fabricated from the virtual plan, determining the intrasurgical position of the implants. Flapless guided implant surgery was carried out according to the manufacturer's instructions. A second CBCT was made 30 days after the surgery, to enable overlapping of the data from before and after the implant placement. The angular, coronal, central, and apical deviations of the placed implants were measured and compared with those virtually planned. The data were submitted to descriptive statistic and intraclass correlation coefficient (ICC), analysis of variance, and the Student t test (α=.05). Results: A total of 61 implants were analyzed. The mean angular deviation was 2.04 degrees. The mean coronal, central, and apical linear deviations were 0.68 mm, 0.72 mm, and 0.82 mm, respectively. No statistically significant difference was found between the virtual and the real position of the implants inserted. A tendency toward a greater absolute error was observed in the mandible than in the maxilla in terms of angular (P=.047), central (P=.043), and coronal error (P=.031). Conclusions: Flapless computer-guided implant surgery with virtual planning had some angular and linear deviations; nevertheless, this technique should be acceptably safe and accurate.Postgraduate student Implant Dentistry Post-Graduation Program São Leopoldo Mandic School of Dentistry and Research CenterProfessor Department of Surgery and Integrated Clinic São Paulo State University (UNESP) School of DentistryPostgraduate student Department of Surgery and Integrated Clinic São Paulo State University (UNESP) School of DentistryPostgraduate student Implant Center for Edentulism and Jawbone Atrophies Maxillofacial Surgery and Odontostomatology Unit Fondazione IRCCS Cà Granda Maggiore Policlinico Hospital University of MilanProfessor Implant Center for Edentulism and Jawbone Atrophies Maxillofacial Surgery and Odontostomatology Unit Fondazione IRCCS Cà Granda Maggiore Policlinico Hospital University of MilanProfessor Implant Dentistry Post-Graduation Program São Leopoldo Mandic School of Dentistry and Research CenterProfessor Department of Surgery and Integrated Clinic São Paulo State University (UNESP) School of DentistryPostgraduate student Department of Surgery and Integrated Clinic São Paulo State University (UNESP) School of DentistrySão Leopoldo Mandic School of Dentistry and Research CenterUniversidade Estadual Paulista (Unesp)University of MilanCunha, Rodrigo MoreiraSouza, Francisley Ávila [UNESP]Hadad, Henrique [UNESP]Poli, Pier PaoloMaiorana, CarloCarvalho, Paulo Sérgio Perri2020-12-12T01:16:21Z2020-12-12T01:16:21Z2020-01-01info:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articlehttp://dx.doi.org/10.1016/j.prosdent.2019.07.010Journal of Prosthetic Dentistry.0022-3913http://hdl.handle.net/11449/19856810.1016/j.prosdent.2019.07.0102-s2.0-85080058513Scopusreponame:Repositório Institucional da UNESPinstname:Universidade Estadual Paulista (UNESP)instacron:UNESPengJournal of Prosthetic Dentistryinfo:eu-repo/semantics/openAccess2021-10-22T16:54:19Zoai:repositorio.unesp.br:11449/198568Repositório InstitucionalPUBhttp://repositorio.unesp.br/oai/requestopendoar:29462024-08-05T20:06:39.123364Repositório Institucional da UNESP - Universidade Estadual Paulista (UNESP)false
dc.title.none.fl_str_mv Accuracy evaluation of computer-guided implant surgery associated with prototyped surgical guides
title Accuracy evaluation of computer-guided implant surgery associated with prototyped surgical guides
spellingShingle Accuracy evaluation of computer-guided implant surgery associated with prototyped surgical guides
Cunha, Rodrigo Moreira
title_short Accuracy evaluation of computer-guided implant surgery associated with prototyped surgical guides
title_full Accuracy evaluation of computer-guided implant surgery associated with prototyped surgical guides
title_fullStr Accuracy evaluation of computer-guided implant surgery associated with prototyped surgical guides
title_full_unstemmed Accuracy evaluation of computer-guided implant surgery associated with prototyped surgical guides
title_sort Accuracy evaluation of computer-guided implant surgery associated with prototyped surgical guides
author Cunha, Rodrigo Moreira
author_facet Cunha, Rodrigo Moreira
Souza, Francisley Ávila [UNESP]
Hadad, Henrique [UNESP]
Poli, Pier Paolo
Maiorana, Carlo
Carvalho, Paulo Sérgio Perri
author_role author
author2 Souza, Francisley Ávila [UNESP]
Hadad, Henrique [UNESP]
Poli, Pier Paolo
Maiorana, Carlo
Carvalho, Paulo Sérgio Perri
author2_role author
author
author
author
author
dc.contributor.none.fl_str_mv São Leopoldo Mandic School of Dentistry and Research Center
Universidade Estadual Paulista (Unesp)
University of Milan
dc.contributor.author.fl_str_mv Cunha, Rodrigo Moreira
Souza, Francisley Ávila [UNESP]
Hadad, Henrique [UNESP]
Poli, Pier Paolo
Maiorana, Carlo
Carvalho, Paulo Sérgio Perri
description Statement of problem: Correct implant placement by means of a pretreatment planning software program is still subject to deviations between the planned and achieved implant positions. Inaccuracy at this level may have drastic consequences, including neurovascular trauma. Further data are therefore needed to evaluate the accuracy of such computer-guided implant planning software programs. Purpose: The purpose of this clinical study was to evaluate the accuracy of computer-guided implant surgery associated with prototyped surgical guides. Material and methods: Cone beam computed tomography (CBCT) scans were made of the participants with a tomographic guide to merge anatomic and prosthetic data. This allowed virtual planning with a prosthetically guided approach respecting the anatomy of the participant. A prototyped surgical guide was then fabricated from the virtual plan, determining the intrasurgical position of the implants. Flapless guided implant surgery was carried out according to the manufacturer's instructions. A second CBCT was made 30 days after the surgery, to enable overlapping of the data from before and after the implant placement. The angular, coronal, central, and apical deviations of the placed implants were measured and compared with those virtually planned. The data were submitted to descriptive statistic and intraclass correlation coefficient (ICC), analysis of variance, and the Student t test (α=.05). Results: A total of 61 implants were analyzed. The mean angular deviation was 2.04 degrees. The mean coronal, central, and apical linear deviations were 0.68 mm, 0.72 mm, and 0.82 mm, respectively. No statistically significant difference was found between the virtual and the real position of the implants inserted. A tendency toward a greater absolute error was observed in the mandible than in the maxilla in terms of angular (P=.047), central (P=.043), and coronal error (P=.031). Conclusions: Flapless computer-guided implant surgery with virtual planning had some angular and linear deviations; nevertheless, this technique should be acceptably safe and accurate.
publishDate 2020
dc.date.none.fl_str_mv 2020-12-12T01:16:21Z
2020-12-12T01:16:21Z
2020-01-01
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://dx.doi.org/10.1016/j.prosdent.2019.07.010
Journal of Prosthetic Dentistry.
0022-3913
http://hdl.handle.net/11449/198568
10.1016/j.prosdent.2019.07.010
2-s2.0-85080058513
url http://dx.doi.org/10.1016/j.prosdent.2019.07.010
http://hdl.handle.net/11449/198568
identifier_str_mv Journal of Prosthetic Dentistry.
0022-3913
10.1016/j.prosdent.2019.07.010
2-s2.0-85080058513
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv Journal of Prosthetic Dentistry
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.source.none.fl_str_mv Scopus
reponame:Repositório Institucional da UNESP
instname:Universidade Estadual Paulista (UNESP)
instacron:UNESP
instname_str Universidade Estadual Paulista (UNESP)
instacron_str UNESP
institution UNESP
reponame_str Repositório Institucional da UNESP
collection Repositório Institucional da UNESP
repository.name.fl_str_mv Repositório Institucional da UNESP - Universidade Estadual Paulista (UNESP)
repository.mail.fl_str_mv
_version_ 1808129160649900032