Avaliação biomecânica da resistência ao torque de inserção de diferentes sistemas de implantes

Detalhes bibliográficos
Autor(a) principal: Sedmak, José Felício
Data de Publicação: 2015
Tipo de documento: Dissertação
Idioma: por
Título da fonte: Biblioteca Digital de Teses e Dissertações da USC
Texto Completo: http://tede2.usc.br:8080/jspui/handle/tede/295
Resumo: Objective: to, in vitro resistance to insertion torque of different implant systems, for the purpose of identifying possible risks to clinical use. Methods: A total of three groups (five units each), as follows: A) Implant Osseotite, 4 x 13 mm, external hexagonal (Biomet 3i); B) Deploy Bonelike 4 x 13 mm, external hexagonal (Biotechnology); C) Implant Mk III RP, 3.75 x 13 mm, external hexagonal (Nobel Biocare). Disposable insert keys were used, and each set was subjected to the torque test (ASTM F543: 2007) for determining the rupture torque, break angle flow in two degrees torque and maximum torque. Results: we obtained the average values of the parameters analyzed for the respective groups: Break torque (N / cm2): 218.00; 132.00; 132.00; angular deformation of rupture (°): 29.04; 39.94; 31.92; torque flow at 2 ° (N / cm2): 177.00; 153.00; 140.00; Maximum torque (N / cm2): 222.00; 172.00; 142.00. Conclusions: the resistance limits the torque of the analyzed systems were superior to clinical needs, offering security to their applications, since their mechanical limits respected, and, Group A (Biomet 3i) showed the best results among the groups.
id USC_2149f1137a912ba2559bf8b3b02d2a9a
oai_identifier_str oai:localhost:tede/295
network_acronym_str USC
network_name_str Biblioteca Digital de Teses e Dissertações da USC
repository_id_str
spelling Nary Filho, Hugo08701441892http://lattes.cnpq.br/756407112710780613593539845http://lattes.cnpq.br/5531767232831283Sedmak, José Felício2016-01-29T21:55:18Z2015-09-08SEDMAK, José Felício. Avaliação biomecânica da resistência ao torque de inserção de diferentes sistemas de implantes. 2015. 41 f. Dissertação (Mestrado em Biologia Oral - Implantologia) - Universidade do Sagrado Coração, Bauru, 2015.http://tede2.usc.br:8080/jspui/handle/tede/295Objective: to, in vitro resistance to insertion torque of different implant systems, for the purpose of identifying possible risks to clinical use. Methods: A total of three groups (five units each), as follows: A) Implant Osseotite, 4 x 13 mm, external hexagonal (Biomet 3i); B) Deploy Bonelike 4 x 13 mm, external hexagonal (Biotechnology); C) Implant Mk III RP, 3.75 x 13 mm, external hexagonal (Nobel Biocare). Disposable insert keys were used, and each set was subjected to the torque test (ASTM F543: 2007) for determining the rupture torque, break angle flow in two degrees torque and maximum torque. Results: we obtained the average values of the parameters analyzed for the respective groups: Break torque (N / cm2): 218.00; 132.00; 132.00; angular deformation of rupture (°): 29.04; 39.94; 31.92; torque flow at 2 ° (N / cm2): 177.00; 153.00; 140.00; Maximum torque (N / cm2): 222.00; 172.00; 142.00. Conclusions: the resistance limits the torque of the analyzed systems were superior to clinical needs, offering security to their applications, since their mechanical limits respected, and, Group A (Biomet 3i) showed the best results among the groups.Objetivo: conhecer, in vitro, a resistência ao torque de inserção, de diferentes sistemas de implantes, com a finalidade de se identificar possíveis riscos ao uso clínico. Materiais e métodos: foram utilizados três grupos (cinco unidades cada), da seguinte forma: A) Implante Osseotite, 4 x 13 mm, hexagonal externo (Biomet 3i); B) Implante Bonelike 4 x 13 mm, hexagonal externo (Biotechnology); C) Implante Mk III RP, 3,75 x 13 mm, hexagonal externo (Nobel Biocare). Foram usadas chaves de inserção descartáveis, sendo que cada conjunto foi submetido ao teste de torção (ASTM F543:2007) para a determinação do torque de ruptura, ângulo de ruptura, torque de escoamento em dois graus e torque máximo. Resultados: foram obtidos os valores médios dos parâmetros analisados para os respectivos grupos: torque de ruptura (N/cm2): 218,00; 132,00; 132,00; deformação angular de ruptura (°): 29,04; 39,94; 31,92; torque de escoamento em 2° (N/cm2): 177,00; 153,00; 140,00; torque máximo (N/cm2): 222,00; 172,00; 142,00. Conclusões: os limites de resistência ao torque dos sistemas analisados foram superiores às necessidades clínicas, oferecendo segurança às suas aplicações, desde que respeitados os seus limites mecânicos, sendo que, o Grupo A (Biomet 3i) apresentou os melhores resultados entre os grupos.Submitted by Valquiria Tavares (valquiria.tavares@usc.br) on 2016-01-29T21:55:18Z No. of bitstreams: 1 dissertacao_jose_sedmak.pdf: 565553 bytes, checksum: b90326ed1dea87bbb8d19c69f48afd6a (MD5)Made available in DSpace on 2016-01-29T21:55:18Z (GMT). No. of bitstreams: 1 dissertacao_jose_sedmak.pdf: 565553 bytes, checksum: b90326ed1dea87bbb8d19c69f48afd6a (MD5) Previous issue date: 2015-09-08application/pdfporUniversidade do Sagrado CoraçãoImplantologiaUSCBrasilCiências da Saúde e BiológicasEstresse mecânicoImplantes dentáriosTorqueMechanical stressDental implantsTorqueCIENCIAS DA SAUDE::ODONTOLOGIAAvaliação biomecânica da resistência ao torque de inserção de diferentes sistemas de implantesBiomechanical evaluation of the resistance to insertion torque of different implant systemsinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/masterThesis1715202039863940245600600600-753428887677182963-2070498469879244349info:eu-repo/semantics/openAccessreponame:Biblioteca Digital de Teses e Dissertações da USCinstname:Universidade do Sagrado Coração (USC)instacron:USCLICENSElicense.txtlicense.txttext/plain; charset=utf-82165http://localhost:8080/tede/bitstream/tede/295/1/license.txtbd3efa91386c1718a7f26a329fdcb468MD51ORIGINALdissertacao_jose_sedmak.pdfdissertacao_jose_sedmak.pdfapplication/pdf565553http://localhost:8080/tede/bitstream/tede/295/2/dissertacao_jose_sedmak.pdfb90326ed1dea87bbb8d19c69f48afd6aMD52tede/2952016-01-30 11:07:59.405oai:localhost: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Biblioteca Digital de Teses e Dissertaçõeshttps://tede2.usc.br:8443/http://tede2.usc.br:8080/oai/requestbiblicorjesu@unisagrado.edu.br||normalizacao@usc.bropendoar:2016-01-30T13:07:59Biblioteca Digital de Teses e Dissertações da USC - Universidade do Sagrado Coração (USC)false
dc.title.eng.fl_str_mv Avaliação biomecânica da resistência ao torque de inserção de diferentes sistemas de implantes
dc.title.alternative.eng.fl_str_mv Biomechanical evaluation of the resistance to insertion torque of different implant systems
title Avaliação biomecânica da resistência ao torque de inserção de diferentes sistemas de implantes
spellingShingle Avaliação biomecânica da resistência ao torque de inserção de diferentes sistemas de implantes
Sedmak, José Felício
Estresse mecânico
Implantes dentários
Torque
Mechanical stress
Dental implants
Torque
CIENCIAS DA SAUDE::ODONTOLOGIA
title_short Avaliação biomecânica da resistência ao torque de inserção de diferentes sistemas de implantes
title_full Avaliação biomecânica da resistência ao torque de inserção de diferentes sistemas de implantes
title_fullStr Avaliação biomecânica da resistência ao torque de inserção de diferentes sistemas de implantes
title_full_unstemmed Avaliação biomecânica da resistência ao torque de inserção de diferentes sistemas de implantes
title_sort Avaliação biomecânica da resistência ao torque de inserção de diferentes sistemas de implantes
author Sedmak, José Felício
author_facet Sedmak, José Felício
author_role author
dc.contributor.advisor1.fl_str_mv Nary Filho, Hugo
dc.contributor.advisor1ID.fl_str_mv 08701441892
dc.contributor.advisor1Lattes.fl_str_mv http://lattes.cnpq.br/7564071127107806
dc.contributor.authorID.fl_str_mv 13593539845
dc.contributor.authorLattes.fl_str_mv http://lattes.cnpq.br/5531767232831283
dc.contributor.author.fl_str_mv Sedmak, José Felício
contributor_str_mv Nary Filho, Hugo
dc.subject.por.fl_str_mv Estresse mecânico
Implantes dentários
Torque
topic Estresse mecânico
Implantes dentários
Torque
Mechanical stress
Dental implants
Torque
CIENCIAS DA SAUDE::ODONTOLOGIA
dc.subject.eng.fl_str_mv Mechanical stress
Dental implants
Torque
dc.subject.cnpq.fl_str_mv CIENCIAS DA SAUDE::ODONTOLOGIA
description Objective: to, in vitro resistance to insertion torque of different implant systems, for the purpose of identifying possible risks to clinical use. Methods: A total of three groups (five units each), as follows: A) Implant Osseotite, 4 x 13 mm, external hexagonal (Biomet 3i); B) Deploy Bonelike 4 x 13 mm, external hexagonal (Biotechnology); C) Implant Mk III RP, 3.75 x 13 mm, external hexagonal (Nobel Biocare). Disposable insert keys were used, and each set was subjected to the torque test (ASTM F543: 2007) for determining the rupture torque, break angle flow in two degrees torque and maximum torque. Results: we obtained the average values of the parameters analyzed for the respective groups: Break torque (N / cm2): 218.00; 132.00; 132.00; angular deformation of rupture (°): 29.04; 39.94; 31.92; torque flow at 2 ° (N / cm2): 177.00; 153.00; 140.00; Maximum torque (N / cm2): 222.00; 172.00; 142.00. Conclusions: the resistance limits the torque of the analyzed systems were superior to clinical needs, offering security to their applications, since their mechanical limits respected, and, Group A (Biomet 3i) showed the best results among the groups.
publishDate 2015
dc.date.issued.fl_str_mv 2015-09-08
dc.date.accessioned.fl_str_mv 2016-01-29T21:55:18Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/masterThesis
format masterThesis
status_str publishedVersion
dc.identifier.citation.fl_str_mv SEDMAK, José Felício. Avaliação biomecânica da resistência ao torque de inserção de diferentes sistemas de implantes. 2015. 41 f. Dissertação (Mestrado em Biologia Oral - Implantologia) - Universidade do Sagrado Coração, Bauru, 2015.
dc.identifier.uri.fl_str_mv http://tede2.usc.br:8080/jspui/handle/tede/295
identifier_str_mv SEDMAK, José Felício. Avaliação biomecânica da resistência ao torque de inserção de diferentes sistemas de implantes. 2015. 41 f. Dissertação (Mestrado em Biologia Oral - Implantologia) - Universidade do Sagrado Coração, Bauru, 2015.
url http://tede2.usc.br:8080/jspui/handle/tede/295
dc.language.iso.fl_str_mv por
language por
dc.relation.program.fl_str_mv 1715202039863940245
dc.relation.confidence.fl_str_mv 600
600
600
dc.relation.department.fl_str_mv -753428887677182963
dc.relation.cnpq.fl_str_mv -2070498469879244349
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Universidade do Sagrado Coração
dc.publisher.program.fl_str_mv Implantologia
dc.publisher.initials.fl_str_mv USC
dc.publisher.country.fl_str_mv Brasil
dc.publisher.department.fl_str_mv Ciências da Saúde e Biológicas
publisher.none.fl_str_mv Universidade do Sagrado Coração
dc.source.none.fl_str_mv reponame:Biblioteca Digital de Teses e Dissertações da USC
instname:Universidade do Sagrado Coração (USC)
instacron:USC
instname_str Universidade do Sagrado Coração (USC)
instacron_str USC
institution USC
reponame_str Biblioteca Digital de Teses e Dissertações da USC
collection Biblioteca Digital de Teses e Dissertações da USC
bitstream.url.fl_str_mv http://localhost:8080/tede/bitstream/tede/295/1/license.txt
http://localhost:8080/tede/bitstream/tede/295/2/dissertacao_jose_sedmak.pdf
bitstream.checksum.fl_str_mv bd3efa91386c1718a7f26a329fdcb468
b90326ed1dea87bbb8d19c69f48afd6a
bitstream.checksumAlgorithm.fl_str_mv MD5
MD5
repository.name.fl_str_mv Biblioteca Digital de Teses e Dissertações da USC - Universidade do Sagrado Coração (USC)
repository.mail.fl_str_mv biblicorjesu@unisagrado.edu.br||normalizacao@usc.br
_version_ 1815438144957317120