Effect of axial loads on implant-supported partial fixed prostheses by strain gauge analysis
Autor(a) principal: | |
---|---|
Data de Publicação: | 2011 |
Outros Autores: | , , |
Tipo de documento: | Artigo |
Idioma: | eng |
Título da fonte: | Journal of applied oral science (Online) |
Texto Completo: | https://www.revistas.usp.br/jaos/article/view/3962 |
Resumo: | OBJECTIVES: The present study used strain gauge analysis to perform an in vitro evaluation of the effect of axial loading on 3 elements of implant-supported partial fixed prostheses, varying the type of prosthetic cylinder and the loading points. MATERIAL AND METHODS: Three internal hexagon implants were linearly embedded in a polyurethane block. Microunit abutments were connected to the implants applying a torque of 20 Ncm, and prefabricated Co-Cr cylinders and plastic prosthetic cylinders were screwed onto the abutments, which received standard patterns cast in Co-Cr alloy (n=5). Four strain gauges (SG) were bonded onto the surface of the block tangentially to the implants, SG 01 mesially to implant 1, SG 02 and SG 03 mesially and distally to implant 2, respectively, and SG 04 distally to implant 3. Each metallic structure was screwed onto the abutments with a 10 Ncm torque and an axial load of 30 kg was applied at five predetermined points (A, B, C, D, E). The data obtained from the strain gauge analyses were analyzed statistically by RM ANOVA and Tukey's test, with a level of significance of p |
id |
USP-17_d1230f32147b41df8d1ace59bd672d5e |
---|---|
oai_identifier_str |
oai:revistas.usp.br:article/3962 |
network_acronym_str |
USP-17 |
network_name_str |
Journal of applied oral science (Online) |
repository_id_str |
|
spelling |
Effect of axial loads on implant-supported partial fixed prostheses by strain gauge analysis BiomechanicsDental implantsDental prosthesisImplant-supported dental prosthesis OBJECTIVES: The present study used strain gauge analysis to perform an in vitro evaluation of the effect of axial loading on 3 elements of implant-supported partial fixed prostheses, varying the type of prosthetic cylinder and the loading points. MATERIAL AND METHODS: Three internal hexagon implants were linearly embedded in a polyurethane block. Microunit abutments were connected to the implants applying a torque of 20 Ncm, and prefabricated Co-Cr cylinders and plastic prosthetic cylinders were screwed onto the abutments, which received standard patterns cast in Co-Cr alloy (n=5). Four strain gauges (SG) were bonded onto the surface of the block tangentially to the implants, SG 01 mesially to implant 1, SG 02 and SG 03 mesially and distally to implant 2, respectively, and SG 04 distally to implant 3. Each metallic structure was screwed onto the abutments with a 10 Ncm torque and an axial load of 30 kg was applied at five predetermined points (A, B, C, D, E). The data obtained from the strain gauge analyses were analyzed statistically by RM ANOVA and Tukey's test, with a level of significance of pUniversidade de São Paulo. Faculdade de Odontologia de Bauru2011-12-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionapplication/pdfhttps://www.revistas.usp.br/jaos/article/view/396210.1590/S1678-77572011000600011Journal of Applied Oral Science; Vol. 19 No. 6 (2011); 610-615 Journal of Applied Oral Science; Vol. 19 Núm. 6 (2011); 610-615 Journal of Applied Oral Science; v. 19 n. 6 (2011); 610-615 1678-77651678-7757reponame:Journal of applied oral science (Online)instname:Universidade de São Paulo (USP)instacron:USPenghttps://www.revistas.usp.br/jaos/article/view/3962/4652Copyright (c) 2011 Journal of Applied Oral Scienceinfo:eu-repo/semantics/openAccessVasconcellos, Luis Gustavo Oliveira deNishioka, Renato SussumuVasconcellos, Luana Marotta Reis deNishioka, Lea Nogueira Braulino de Melo2012-04-27T12:16:27Zoai:revistas.usp.br:article/3962Revistahttp://www.scielo.br/jaosPUBhttps://www.revistas.usp.br/jaos/oai||jaos@usp.br1678-77651678-7757opendoar:2012-04-27T12:16:27Journal of applied oral science (Online) - Universidade de São Paulo (USP)false |
dc.title.none.fl_str_mv |
Effect of axial loads on implant-supported partial fixed prostheses by strain gauge analysis |
title |
Effect of axial loads on implant-supported partial fixed prostheses by strain gauge analysis |
spellingShingle |
Effect of axial loads on implant-supported partial fixed prostheses by strain gauge analysis Vasconcellos, Luis Gustavo Oliveira de Biomechanics Dental implants Dental prosthesis Implant-supported dental prosthesis |
title_short |
Effect of axial loads on implant-supported partial fixed prostheses by strain gauge analysis |
title_full |
Effect of axial loads on implant-supported partial fixed prostheses by strain gauge analysis |
title_fullStr |
Effect of axial loads on implant-supported partial fixed prostheses by strain gauge analysis |
title_full_unstemmed |
Effect of axial loads on implant-supported partial fixed prostheses by strain gauge analysis |
title_sort |
Effect of axial loads on implant-supported partial fixed prostheses by strain gauge analysis |
author |
Vasconcellos, Luis Gustavo Oliveira de |
author_facet |
Vasconcellos, Luis Gustavo Oliveira de Nishioka, Renato Sussumu Vasconcellos, Luana Marotta Reis de Nishioka, Lea Nogueira Braulino de Melo |
author_role |
author |
author2 |
Nishioka, Renato Sussumu Vasconcellos, Luana Marotta Reis de Nishioka, Lea Nogueira Braulino de Melo |
author2_role |
author author author |
dc.contributor.author.fl_str_mv |
Vasconcellos, Luis Gustavo Oliveira de Nishioka, Renato Sussumu Vasconcellos, Luana Marotta Reis de Nishioka, Lea Nogueira Braulino de Melo |
dc.subject.por.fl_str_mv |
Biomechanics Dental implants Dental prosthesis Implant-supported dental prosthesis |
topic |
Biomechanics Dental implants Dental prosthesis Implant-supported dental prosthesis |
description |
OBJECTIVES: The present study used strain gauge analysis to perform an in vitro evaluation of the effect of axial loading on 3 elements of implant-supported partial fixed prostheses, varying the type of prosthetic cylinder and the loading points. MATERIAL AND METHODS: Three internal hexagon implants were linearly embedded in a polyurethane block. Microunit abutments were connected to the implants applying a torque of 20 Ncm, and prefabricated Co-Cr cylinders and plastic prosthetic cylinders were screwed onto the abutments, which received standard patterns cast in Co-Cr alloy (n=5). Four strain gauges (SG) were bonded onto the surface of the block tangentially to the implants, SG 01 mesially to implant 1, SG 02 and SG 03 mesially and distally to implant 2, respectively, and SG 04 distally to implant 3. Each metallic structure was screwed onto the abutments with a 10 Ncm torque and an axial load of 30 kg was applied at five predetermined points (A, B, C, D, E). The data obtained from the strain gauge analyses were analyzed statistically by RM ANOVA and Tukey's test, with a level of significance of p |
publishDate |
2011 |
dc.date.none.fl_str_mv |
2011-12-01 |
dc.type.driver.fl_str_mv |
info:eu-repo/semantics/article info:eu-repo/semantics/publishedVersion |
format |
article |
status_str |
publishedVersion |
dc.identifier.uri.fl_str_mv |
https://www.revistas.usp.br/jaos/article/view/3962 10.1590/S1678-77572011000600011 |
url |
https://www.revistas.usp.br/jaos/article/view/3962 |
identifier_str_mv |
10.1590/S1678-77572011000600011 |
dc.language.iso.fl_str_mv |
eng |
language |
eng |
dc.relation.none.fl_str_mv |
https://www.revistas.usp.br/jaos/article/view/3962/4652 |
dc.rights.driver.fl_str_mv |
Copyright (c) 2011 Journal of Applied Oral Science info:eu-repo/semantics/openAccess |
rights_invalid_str_mv |
Copyright (c) 2011 Journal of Applied Oral Science |
eu_rights_str_mv |
openAccess |
dc.format.none.fl_str_mv |
application/pdf |
dc.publisher.none.fl_str_mv |
Universidade de São Paulo. Faculdade de Odontologia de Bauru |
publisher.none.fl_str_mv |
Universidade de São Paulo. Faculdade de Odontologia de Bauru |
dc.source.none.fl_str_mv |
Journal of Applied Oral Science; Vol. 19 No. 6 (2011); 610-615 Journal of Applied Oral Science; Vol. 19 Núm. 6 (2011); 610-615 Journal of Applied Oral Science; v. 19 n. 6 (2011); 610-615 1678-7765 1678-7757 reponame:Journal of applied oral science (Online) instname:Universidade de São Paulo (USP) instacron:USP |
instname_str |
Universidade de São Paulo (USP) |
instacron_str |
USP |
institution |
USP |
reponame_str |
Journal of applied oral science (Online) |
collection |
Journal of applied oral science (Online) |
repository.name.fl_str_mv |
Journal of applied oral science (Online) - Universidade de São Paulo (USP) |
repository.mail.fl_str_mv |
||jaos@usp.br |
_version_ |
1800221676888129536 |