Validation, repeatability and reproducibility of a noninvasive instrument for measuring thoracic and lumbar curvature of the spine in the sagittal plane

Detalhes bibliográficos
Autor(a) principal: Chaise,Fabiana O.
Data de Publicação: 2011
Outros Autores: Candotti,Cláudia T., Torre,Marcelo L., Furlanetto,Tássia S., Pelinson,Patricia P. T., Loss,Jefferson F.
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Brazilian Journal of Physical Therapy
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1413-35552011000600013
Resumo: BACKGROUND: The need for early identification of postural abnormalities without exposing patients to constant radiation has stimulated the development of instruments aiming to measure the spinal curvatures. OBJECTIVE: To verify the validity, repeatability and reproducibility of angular measures of sagittal curvatures of the spine obtained using an adapted arcometer, by comparing them with Cobb angles of the respective curvatures obtained by using X-rays. METHODS: 52 participants were submitted to two procedures designed to evaluate the thoracic and lumbar curvatures: (1) X-ray examination from which the Cobb angles (CA) of both curvatures were obtained, and (2) measuring the angles with the arcometer (AA). Two evaluators collected the data using the arcometer, with the rods placed at T1, T12, L1 and L5 spinous processes levels in a way as to permit linear measurements which, with aid of trigonometry, supplied the AA. RESULTS: There was a very strong and significant correlation between AA and CA (r=0.94; p<0.01), with no-significant difference (p=0.32), for the thoracic curvature. There was a strong and significant correlation for the lumbar curvature (r=0.71; p<0.01) between AA and CA, with no-significant difference (p=0.30). There is a very strong correlation between intra-evaluator and inter-evaluator AA. CONCLUSION: It was possible to quantify reliably the thoracic and lumbar curvatures with the arcometer and it can thus be considered valid and reliable and for use in evaluating spinal curvatures in the sagittal plane.
id ABRA-FT-1_06e7ead6d459c1795b86a7f508d720b6
oai_identifier_str oai:scielo:S1413-35552011000600013
network_acronym_str ABRA-FT-1
network_name_str Brazilian Journal of Physical Therapy
repository_id_str
spelling Validation, repeatability and reproducibility of a noninvasive instrument for measuring thoracic and lumbar curvature of the spine in the sagittal planeevaluationspineX-raysvalidity of testsphysical therapyBACKGROUND: The need for early identification of postural abnormalities without exposing patients to constant radiation has stimulated the development of instruments aiming to measure the spinal curvatures. OBJECTIVE: To verify the validity, repeatability and reproducibility of angular measures of sagittal curvatures of the spine obtained using an adapted arcometer, by comparing them with Cobb angles of the respective curvatures obtained by using X-rays. METHODS: 52 participants were submitted to two procedures designed to evaluate the thoracic and lumbar curvatures: (1) X-ray examination from which the Cobb angles (CA) of both curvatures were obtained, and (2) measuring the angles with the arcometer (AA). Two evaluators collected the data using the arcometer, with the rods placed at T1, T12, L1 and L5 spinous processes levels in a way as to permit linear measurements which, with aid of trigonometry, supplied the AA. RESULTS: There was a very strong and significant correlation between AA and CA (r=0.94; p<0.01), with no-significant difference (p=0.32), for the thoracic curvature. There was a strong and significant correlation for the lumbar curvature (r=0.71; p<0.01) between AA and CA, with no-significant difference (p=0.30). There is a very strong correlation between intra-evaluator and inter-evaluator AA. CONCLUSION: It was possible to quantify reliably the thoracic and lumbar curvatures with the arcometer and it can thus be considered valid and reliable and for use in evaluating spinal curvatures in the sagittal plane.Associação Brasileira de Pesquisa e Pós-Graduação em Fisioterapia 2011-12-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S1413-35552011000600013Brazilian Journal of Physical Therapy v.15 n.6 2011reponame:Brazilian Journal of Physical Therapyinstname:Associação Brasileira de Pesquisa e Pós-Graduação em Fisioterapia (ABRAPG-FT)instacron:ABRAPG-FT10.1590/S1413-35552011005000031info:eu-repo/semantics/openAccessChaise,Fabiana O.Candotti,Cláudia T.Torre,Marcelo L.Furlanetto,Tássia S.Pelinson,Patricia P. T.Loss,Jefferson F.eng2012-01-02T00:00:00Zoai:scielo:S1413-35552011000600013Revistahttps://www.scielo.br/j/rbfis/https://old.scielo.br/oai/scielo-oai.phpcontato@rbf-bjpt.org.br||contato@rbf-bjpt.org.br1809-92461413-3555opendoar:2012-01-02T00:00Brazilian Journal of Physical Therapy - Associação Brasileira de Pesquisa e Pós-Graduação em Fisioterapia (ABRAPG-FT)false
dc.title.none.fl_str_mv Validation, repeatability and reproducibility of a noninvasive instrument for measuring thoracic and lumbar curvature of the spine in the sagittal plane
title Validation, repeatability and reproducibility of a noninvasive instrument for measuring thoracic and lumbar curvature of the spine in the sagittal plane
spellingShingle Validation, repeatability and reproducibility of a noninvasive instrument for measuring thoracic and lumbar curvature of the spine in the sagittal plane
Chaise,Fabiana O.
evaluation
spine
X-rays
validity of tests
physical therapy
title_short Validation, repeatability and reproducibility of a noninvasive instrument for measuring thoracic and lumbar curvature of the spine in the sagittal plane
title_full Validation, repeatability and reproducibility of a noninvasive instrument for measuring thoracic and lumbar curvature of the spine in the sagittal plane
title_fullStr Validation, repeatability and reproducibility of a noninvasive instrument for measuring thoracic and lumbar curvature of the spine in the sagittal plane
title_full_unstemmed Validation, repeatability and reproducibility of a noninvasive instrument for measuring thoracic and lumbar curvature of the spine in the sagittal plane
title_sort Validation, repeatability and reproducibility of a noninvasive instrument for measuring thoracic and lumbar curvature of the spine in the sagittal plane
author Chaise,Fabiana O.
author_facet Chaise,Fabiana O.
Candotti,Cláudia T.
Torre,Marcelo L.
Furlanetto,Tássia S.
Pelinson,Patricia P. T.
Loss,Jefferson F.
author_role author
author2 Candotti,Cláudia T.
Torre,Marcelo L.
Furlanetto,Tássia S.
Pelinson,Patricia P. T.
Loss,Jefferson F.
author2_role author
author
author
author
author
dc.contributor.author.fl_str_mv Chaise,Fabiana O.
Candotti,Cláudia T.
Torre,Marcelo L.
Furlanetto,Tássia S.
Pelinson,Patricia P. T.
Loss,Jefferson F.
dc.subject.por.fl_str_mv evaluation
spine
X-rays
validity of tests
physical therapy
topic evaluation
spine
X-rays
validity of tests
physical therapy
description BACKGROUND: The need for early identification of postural abnormalities without exposing patients to constant radiation has stimulated the development of instruments aiming to measure the spinal curvatures. OBJECTIVE: To verify the validity, repeatability and reproducibility of angular measures of sagittal curvatures of the spine obtained using an adapted arcometer, by comparing them with Cobb angles of the respective curvatures obtained by using X-rays. METHODS: 52 participants were submitted to two procedures designed to evaluate the thoracic and lumbar curvatures: (1) X-ray examination from which the Cobb angles (CA) of both curvatures were obtained, and (2) measuring the angles with the arcometer (AA). Two evaluators collected the data using the arcometer, with the rods placed at T1, T12, L1 and L5 spinous processes levels in a way as to permit linear measurements which, with aid of trigonometry, supplied the AA. RESULTS: There was a very strong and significant correlation between AA and CA (r=0.94; p<0.01), with no-significant difference (p=0.32), for the thoracic curvature. There was a strong and significant correlation for the lumbar curvature (r=0.71; p<0.01) between AA and CA, with no-significant difference (p=0.30). There is a very strong correlation between intra-evaluator and inter-evaluator AA. CONCLUSION: It was possible to quantify reliably the thoracic and lumbar curvatures with the arcometer and it can thus be considered valid and reliable and for use in evaluating spinal curvatures in the sagittal plane.
publishDate 2011
dc.date.none.fl_str_mv 2011-12-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1413-35552011000600013
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1413-35552011000600013
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/S1413-35552011005000031
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Associação Brasileira de Pesquisa e Pós-Graduação em Fisioterapia
publisher.none.fl_str_mv Associação Brasileira de Pesquisa e Pós-Graduação em Fisioterapia
dc.source.none.fl_str_mv Brazilian Journal of Physical Therapy v.15 n.6 2011
reponame:Brazilian Journal of Physical Therapy
instname:Associação Brasileira de Pesquisa e Pós-Graduação em Fisioterapia (ABRAPG-FT)
instacron:ABRAPG-FT
instname_str Associação Brasileira de Pesquisa e Pós-Graduação em Fisioterapia (ABRAPG-FT)
instacron_str ABRAPG-FT
institution ABRAPG-FT
reponame_str Brazilian Journal of Physical Therapy
collection Brazilian Journal of Physical Therapy
repository.name.fl_str_mv Brazilian Journal of Physical Therapy - Associação Brasileira de Pesquisa e Pós-Graduação em Fisioterapia (ABRAPG-FT)
repository.mail.fl_str_mv contato@rbf-bjpt.org.br||contato@rbf-bjpt.org.br
_version_ 1754575948755238912