Assessment of a glycated hemoglobin point-of-care analyzer (A1CNow+) in comparison with an immunoturbidimetric method: a diagnostic accuracy study

Detalhes bibliográficos
Autor(a) principal: Affret,Aurélie
Data de Publicação: 2015
Outros Autores: Griz,Luiz Henrique Maciel, Cesse,Eduarda Ângela Pessoa, Specht,Yuri da Silva, Carvalho,Eduardo Maia Freese de, Fontbonne,Annick
Tipo de documento: Artigo
Idioma: eng
Título da fonte: São Paulo medical journal (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1516-31802015000600460
Resumo: CONTEXT AND OBJECTIVE: To monitor glycemic control in diabetic patients, regular measurement of glycated hemoglobin (HbA1c) is recommended, but this can be difficult in remote places without access to laboratories. Portable point-of-care testing devices can prove a useful alternative. Our study aimed to assess the performance of one of them: A1CNow+, from Bayer. DESIGN AND SETTING: Cross-sectional accuracy study conducted at a university hospital in Brazil. METHODS: We made three successive measurements of capillary HbA1c using the A1CNow+ in 55 diabetic volunteers, while the same measurement was made on venous blood using the hospital reference method (Vitros 5,1 FS). We used the Bland-Altman graphical method to assess the A1CNow+ in relation to the Vitros 5,1 FS method. We also evaluated clinical usefulness by calculating the sensitivity and specificity of A1CNow+ for detecting patients with HbA1c lower than 7%, which is the usual limit for good glycemic control. RESULTS: The coefficient of variation between repeat testing for the A1CNow+ was 3.6%. The mean difference between A1CNow+ and Vitros 5,1 FS was +0.67% (95% confidence interval, CI: +0.52 to +0.81). The agreement limits of our Bland-Altman graph were -0.45 (95% CI: -0.71 to -0.19) and +1.82 (95% CI: +1.52 to +2.05). The sensitivity and specificity in relation to the 7% limit were respectively 100% and 67.7%. CONCLUSIONS: Although the A1CNow+ had good sensitivity, its accuracy was insufficient for use as a replacement for laboratory measurements of HbA1c, for glycemic control monitoring in diabetic patients.
id APM-1_a5f4dd179534ea4f4fca9bfafa269a73
oai_identifier_str oai:scielo:S1516-31802015000600460
network_acronym_str APM-1
network_name_str São Paulo medical journal (Online)
repository_id_str
spelling Assessment of a glycated hemoglobin point-of-care analyzer (A1CNow+) in comparison with an immunoturbidimetric method: a diagnostic accuracy studyHemoglobin A, glycosylatedPoint-of-care systemsTechnology assessment, biomedicalDiabetes mellitusPrimary health care CONTEXT AND OBJECTIVE: To monitor glycemic control in diabetic patients, regular measurement of glycated hemoglobin (HbA1c) is recommended, but this can be difficult in remote places without access to laboratories. Portable point-of-care testing devices can prove a useful alternative. Our study aimed to assess the performance of one of them: A1CNow+, from Bayer. DESIGN AND SETTING: Cross-sectional accuracy study conducted at a university hospital in Brazil. METHODS: We made three successive measurements of capillary HbA1c using the A1CNow+ in 55 diabetic volunteers, while the same measurement was made on venous blood using the hospital reference method (Vitros 5,1 FS). We used the Bland-Altman graphical method to assess the A1CNow+ in relation to the Vitros 5,1 FS method. We also evaluated clinical usefulness by calculating the sensitivity and specificity of A1CNow+ for detecting patients with HbA1c lower than 7%, which is the usual limit for good glycemic control. RESULTS: The coefficient of variation between repeat testing for the A1CNow+ was 3.6%. The mean difference between A1CNow+ and Vitros 5,1 FS was +0.67% (95% confidence interval, CI: +0.52 to +0.81). The agreement limits of our Bland-Altman graph were -0.45 (95% CI: -0.71 to -0.19) and +1.82 (95% CI: +1.52 to +2.05). The sensitivity and specificity in relation to the 7% limit were respectively 100% and 67.7%. CONCLUSIONS: Although the A1CNow+ had good sensitivity, its accuracy was insufficient for use as a replacement for laboratory measurements of HbA1c, for glycemic control monitoring in diabetic patients.Associação Paulista de Medicina - APM2015-12-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S1516-31802015000600460Sao Paulo Medical Journal v.133 n.6 2015reponame:São Paulo medical journal (Online)instname:Associação Paulista de Medicinainstacron:APM10.1590/1516-3180.2013.9110911info:eu-repo/semantics/openAccessAffret,AurélieGriz,Luiz Henrique MacielCesse,Eduarda Ângela PessoaSpecht,Yuri da SilvaCarvalho,Eduardo Maia Freese deFontbonne,Annickeng2016-01-06T00:00:00Zoai:scielo:S1516-31802015000600460Revistahttp://www.scielo.br/spmjhttps://old.scielo.br/oai/scielo-oai.phprevistas@apm.org.br1806-94601516-3180opendoar:2016-01-06T00:00São Paulo medical journal (Online) - Associação Paulista de Medicinafalse
dc.title.none.fl_str_mv Assessment of a glycated hemoglobin point-of-care analyzer (A1CNow+) in comparison with an immunoturbidimetric method: a diagnostic accuracy study
title Assessment of a glycated hemoglobin point-of-care analyzer (A1CNow+) in comparison with an immunoturbidimetric method: a diagnostic accuracy study
spellingShingle Assessment of a glycated hemoglobin point-of-care analyzer (A1CNow+) in comparison with an immunoturbidimetric method: a diagnostic accuracy study
Affret,Aurélie
Hemoglobin A, glycosylated
Point-of-care systems
Technology assessment, biomedical
Diabetes mellitus
Primary health care
title_short Assessment of a glycated hemoglobin point-of-care analyzer (A1CNow+) in comparison with an immunoturbidimetric method: a diagnostic accuracy study
title_full Assessment of a glycated hemoglobin point-of-care analyzer (A1CNow+) in comparison with an immunoturbidimetric method: a diagnostic accuracy study
title_fullStr Assessment of a glycated hemoglobin point-of-care analyzer (A1CNow+) in comparison with an immunoturbidimetric method: a diagnostic accuracy study
title_full_unstemmed Assessment of a glycated hemoglobin point-of-care analyzer (A1CNow+) in comparison with an immunoturbidimetric method: a diagnostic accuracy study
title_sort Assessment of a glycated hemoglobin point-of-care analyzer (A1CNow+) in comparison with an immunoturbidimetric method: a diagnostic accuracy study
author Affret,Aurélie
author_facet Affret,Aurélie
Griz,Luiz Henrique Maciel
Cesse,Eduarda Ângela Pessoa
Specht,Yuri da Silva
Carvalho,Eduardo Maia Freese de
Fontbonne,Annick
author_role author
author2 Griz,Luiz Henrique Maciel
Cesse,Eduarda Ângela Pessoa
Specht,Yuri da Silva
Carvalho,Eduardo Maia Freese de
Fontbonne,Annick
author2_role author
author
author
author
author
dc.contributor.author.fl_str_mv Affret,Aurélie
Griz,Luiz Henrique Maciel
Cesse,Eduarda Ângela Pessoa
Specht,Yuri da Silva
Carvalho,Eduardo Maia Freese de
Fontbonne,Annick
dc.subject.por.fl_str_mv Hemoglobin A, glycosylated
Point-of-care systems
Technology assessment, biomedical
Diabetes mellitus
Primary health care
topic Hemoglobin A, glycosylated
Point-of-care systems
Technology assessment, biomedical
Diabetes mellitus
Primary health care
description CONTEXT AND OBJECTIVE: To monitor glycemic control in diabetic patients, regular measurement of glycated hemoglobin (HbA1c) is recommended, but this can be difficult in remote places without access to laboratories. Portable point-of-care testing devices can prove a useful alternative. Our study aimed to assess the performance of one of them: A1CNow+, from Bayer. DESIGN AND SETTING: Cross-sectional accuracy study conducted at a university hospital in Brazil. METHODS: We made three successive measurements of capillary HbA1c using the A1CNow+ in 55 diabetic volunteers, while the same measurement was made on venous blood using the hospital reference method (Vitros 5,1 FS). We used the Bland-Altman graphical method to assess the A1CNow+ in relation to the Vitros 5,1 FS method. We also evaluated clinical usefulness by calculating the sensitivity and specificity of A1CNow+ for detecting patients with HbA1c lower than 7%, which is the usual limit for good glycemic control. RESULTS: The coefficient of variation between repeat testing for the A1CNow+ was 3.6%. The mean difference between A1CNow+ and Vitros 5,1 FS was +0.67% (95% confidence interval, CI: +0.52 to +0.81). The agreement limits of our Bland-Altman graph were -0.45 (95% CI: -0.71 to -0.19) and +1.82 (95% CI: +1.52 to +2.05). The sensitivity and specificity in relation to the 7% limit were respectively 100% and 67.7%. CONCLUSIONS: Although the A1CNow+ had good sensitivity, its accuracy was insufficient for use as a replacement for laboratory measurements of HbA1c, for glycemic control monitoring in diabetic patients.
publishDate 2015
dc.date.none.fl_str_mv 2015-12-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1516-31802015000600460
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1516-31802015000600460
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/1516-3180.2013.9110911
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Associação Paulista de Medicina - APM
publisher.none.fl_str_mv Associação Paulista de Medicina - APM
dc.source.none.fl_str_mv Sao Paulo Medical Journal v.133 n.6 2015
reponame:São Paulo medical journal (Online)
instname:Associação Paulista de Medicina
instacron:APM
instname_str Associação Paulista de Medicina
instacron_str APM
institution APM
reponame_str São Paulo medical journal (Online)
collection São Paulo medical journal (Online)
repository.name.fl_str_mv São Paulo medical journal (Online) - Associação Paulista de Medicina
repository.mail.fl_str_mv revistas@apm.org.br
_version_ 1754209264484745216