Validation of a new fiber electrode prototype for clinical electroretinography

Detalhes bibliográficos
Autor(a) principal: Berezovsky,Adriana
Data de Publicação: 2008
Outros Autores: Pereira,Josenilson Martins, Salomão,Solange Rios, Santos,Vagner Rogério dos, Schor,Paulo
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Arquivos brasileiros de oftalmologia (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492008000300003
Resumo: PURPOSE: To validate a new fiber electrode prototype for clinical electroretinography (ERG). METHODS: A recently developed prototype of a disposable reference-coupled fiber electrode (patent pending Brazilian Institute of Industrial Property # PI0602186-7), including one fiber for corneal signals and a second fiber acting as reference was tested in a group of 20 healthy volunteers (17-31 years; mean 22.7 ± 4.5; 8 males). Standard electroretinography rod and cone responses were recorded from a fully dilated pupil simultaneously in both eyes with a reference-coupled fiber electrode prototype in one randomly assigned eye and a DTL® electrode in the other eye after 30 min of dark-adaptation. After presenting dark- and light-adapted stimuli, each response was analyzed for a- and b-wave amplitude and implicit time. The VERIS 5.1.9 system was used for electroretinography data acquisition and analysis. Electroretinography outcomes were analyzed by Mann-Whitney test. Slit-lamp examination was performed in both eyes right after electroretinography session to evaluate possible adverse effects. RESULTS: Responses recorded with reference-coupled fiber electrode prototypes were comparable to commercially available DTL® fiber electrodes. On a qualitative analysis, reference-coupled fiber electrodes provided recordings with less amount of noise. On average, scotopic electroretinography amplitude and b-wave implicit time recorded using DTL® were, respectively, 287.6 µV and 36.3 ms with similar findings for the reference-coupled fiber electrode prototype (287.9 µV and 36.3 ms). Under photopic conditions DTL® mean amplitude and implicit time were, respectively 108.9 µV and 24.5 ms with similar results for the reference-coupled fiber electrodes prototypes (116.4 µV and 24.5 ms). No corneal abrasions or any other significant adverse effects were found after electroretinography recording with both electrodes. CONCLUSIONS: The reference-coupled fiber electrode prototype provided stable and safe recordings of corneal electroretinograms compared to the commercially availabe DTL® electrode in healthy human subjects. The prototype is a feasible alternative instrument for clinical electroretinography recording to assess retinal function, however further analysis is recommended to validate its clinical usefulness in patients with retinal disorders.
id CBO-2_9f5d9aa86dec13d1efd7c87a50d29cc9
oai_identifier_str oai:scielo:S0004-27492008000300003
network_acronym_str CBO-2
network_name_str Arquivos brasileiros de oftalmologia (Online)
repository_id_str
spelling Validation of a new fiber electrode prototype for clinical electroretinographyElectrodesElectroretinography/instrumentationRetina/physiologyRetinal diseasesValidation studiesPURPOSE: To validate a new fiber electrode prototype for clinical electroretinography (ERG). METHODS: A recently developed prototype of a disposable reference-coupled fiber electrode (patent pending Brazilian Institute of Industrial Property # PI0602186-7), including one fiber for corneal signals and a second fiber acting as reference was tested in a group of 20 healthy volunteers (17-31 years; mean 22.7 ± 4.5; 8 males). Standard electroretinography rod and cone responses were recorded from a fully dilated pupil simultaneously in both eyes with a reference-coupled fiber electrode prototype in one randomly assigned eye and a DTL® electrode in the other eye after 30 min of dark-adaptation. After presenting dark- and light-adapted stimuli, each response was analyzed for a- and b-wave amplitude and implicit time. The VERIS 5.1.9 system was used for electroretinography data acquisition and analysis. Electroretinography outcomes were analyzed by Mann-Whitney test. Slit-lamp examination was performed in both eyes right after electroretinography session to evaluate possible adverse effects. RESULTS: Responses recorded with reference-coupled fiber electrode prototypes were comparable to commercially available DTL® fiber electrodes. On a qualitative analysis, reference-coupled fiber electrodes provided recordings with less amount of noise. On average, scotopic electroretinography amplitude and b-wave implicit time recorded using DTL® were, respectively, 287.6 µV and 36.3 ms with similar findings for the reference-coupled fiber electrode prototype (287.9 µV and 36.3 ms). Under photopic conditions DTL® mean amplitude and implicit time were, respectively 108.9 µV and 24.5 ms with similar results for the reference-coupled fiber electrodes prototypes (116.4 µV and 24.5 ms). No corneal abrasions or any other significant adverse effects were found after electroretinography recording with both electrodes. CONCLUSIONS: The reference-coupled fiber electrode prototype provided stable and safe recordings of corneal electroretinograms compared to the commercially availabe DTL® electrode in healthy human subjects. The prototype is a feasible alternative instrument for clinical electroretinography recording to assess retinal function, however further analysis is recommended to validate its clinical usefulness in patients with retinal disorders.Conselho Brasileiro de Oftalmologia2008-06-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492008000300003Arquivos Brasileiros de Oftalmologia v.71 n.3 2008reponame:Arquivos brasileiros de oftalmologia (Online)instname:Conselho Brasileiro de Oftalmologia (CBO)instacron:CBO10.1590/S0004-27492008000300003info:eu-repo/semantics/openAccessBerezovsky,AdrianaPereira,Josenilson MartinsSalomão,Solange RiosSantos,Vagner Rogério dosSchor,Pauloeng2008-07-15T00:00:00Zoai:scielo:S0004-27492008000300003Revistahttp://aboonline.org.br/https://old.scielo.br/oai/scielo-oai.phpaboonline@cbo.com.br||abo@cbo.com.br1678-29250004-2749opendoar:2008-07-15T00:00Arquivos brasileiros de oftalmologia (Online) - Conselho Brasileiro de Oftalmologia (CBO)false
dc.title.none.fl_str_mv Validation of a new fiber electrode prototype for clinical electroretinography
title Validation of a new fiber electrode prototype for clinical electroretinography
spellingShingle Validation of a new fiber electrode prototype for clinical electroretinography
Berezovsky,Adriana
Electrodes
Electroretinography/instrumentation
Retina/physiology
Retinal diseases
Validation studies
title_short Validation of a new fiber electrode prototype for clinical electroretinography
title_full Validation of a new fiber electrode prototype for clinical electroretinography
title_fullStr Validation of a new fiber electrode prototype for clinical electroretinography
title_full_unstemmed Validation of a new fiber electrode prototype for clinical electroretinography
title_sort Validation of a new fiber electrode prototype for clinical electroretinography
author Berezovsky,Adriana
author_facet Berezovsky,Adriana
Pereira,Josenilson Martins
Salomão,Solange Rios
Santos,Vagner Rogério dos
Schor,Paulo
author_role author
author2 Pereira,Josenilson Martins
Salomão,Solange Rios
Santos,Vagner Rogério dos
Schor,Paulo
author2_role author
author
author
author
dc.contributor.author.fl_str_mv Berezovsky,Adriana
Pereira,Josenilson Martins
Salomão,Solange Rios
Santos,Vagner Rogério dos
Schor,Paulo
dc.subject.por.fl_str_mv Electrodes
Electroretinography/instrumentation
Retina/physiology
Retinal diseases
Validation studies
topic Electrodes
Electroretinography/instrumentation
Retina/physiology
Retinal diseases
Validation studies
description PURPOSE: To validate a new fiber electrode prototype for clinical electroretinography (ERG). METHODS: A recently developed prototype of a disposable reference-coupled fiber electrode (patent pending Brazilian Institute of Industrial Property # PI0602186-7), including one fiber for corneal signals and a second fiber acting as reference was tested in a group of 20 healthy volunteers (17-31 years; mean 22.7 ± 4.5; 8 males). Standard electroretinography rod and cone responses were recorded from a fully dilated pupil simultaneously in both eyes with a reference-coupled fiber electrode prototype in one randomly assigned eye and a DTL® electrode in the other eye after 30 min of dark-adaptation. After presenting dark- and light-adapted stimuli, each response was analyzed for a- and b-wave amplitude and implicit time. The VERIS 5.1.9 system was used for electroretinography data acquisition and analysis. Electroretinography outcomes were analyzed by Mann-Whitney test. Slit-lamp examination was performed in both eyes right after electroretinography session to evaluate possible adverse effects. RESULTS: Responses recorded with reference-coupled fiber electrode prototypes were comparable to commercially available DTL® fiber electrodes. On a qualitative analysis, reference-coupled fiber electrodes provided recordings with less amount of noise. On average, scotopic electroretinography amplitude and b-wave implicit time recorded using DTL® were, respectively, 287.6 µV and 36.3 ms with similar findings for the reference-coupled fiber electrode prototype (287.9 µV and 36.3 ms). Under photopic conditions DTL® mean amplitude and implicit time were, respectively 108.9 µV and 24.5 ms with similar results for the reference-coupled fiber electrodes prototypes (116.4 µV and 24.5 ms). No corneal abrasions or any other significant adverse effects were found after electroretinography recording with both electrodes. CONCLUSIONS: The reference-coupled fiber electrode prototype provided stable and safe recordings of corneal electroretinograms compared to the commercially availabe DTL® electrode in healthy human subjects. The prototype is a feasible alternative instrument for clinical electroretinography recording to assess retinal function, however further analysis is recommended to validate its clinical usefulness in patients with retinal disorders.
publishDate 2008
dc.date.none.fl_str_mv 2008-06-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492008000300003
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492008000300003
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/S0004-27492008000300003
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Conselho Brasileiro de Oftalmologia
publisher.none.fl_str_mv Conselho Brasileiro de Oftalmologia
dc.source.none.fl_str_mv Arquivos Brasileiros de Oftalmologia v.71 n.3 2008
reponame:Arquivos brasileiros de oftalmologia (Online)
instname:Conselho Brasileiro de Oftalmologia (CBO)
instacron:CBO
instname_str Conselho Brasileiro de Oftalmologia (CBO)
instacron_str CBO
institution CBO
reponame_str Arquivos brasileiros de oftalmologia (Online)
collection Arquivos brasileiros de oftalmologia (Online)
repository.name.fl_str_mv Arquivos brasileiros de oftalmologia (Online) - Conselho Brasileiro de Oftalmologia (CBO)
repository.mail.fl_str_mv aboonline@cbo.com.br||abo@cbo.com.br
_version_ 1754209025058144256