Microdevice for aqueous humor drainage Maldonado Bas - Pförtner

Detalhes bibliográficos
Autor(a) principal: Maldonado-Bas,Arturo
Data de Publicação: 2011
Outros Autores: Maldonado-Junyent,Ana, Maldonado-Junyent,Arturo
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Arquivos brasileiros de oftalmologia (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492011000300011
Resumo: PURPOSE: A microdevice for the treatment of refractory glaucoma is presented. The underlying concepts, its mechanisms of action and the surgical technique for implanting are explained and the results are analyzed. The microdevice was developed and the surgeries were performed at the Maldonado-Bas Eye Clinic (Cordoba, Argentina), under the rules established in the protocol approved by the provisions of the National Administration of Drugs, Food and Medical Technology 430/7. File No.:1-47-25-649-07-1. METHODS: In a prospective study, following the protocol, 16 eyes with refractory glaucoma were included and operated. Intraocular pressure <21mmHg with or without additional medication was considered successful. The follow-up was one year. Averages, percentages and their 95% confidence bands were calculated. Analysis of variance for repeated measures was used to compare averages. RESULTS: The average preoperative intraocular pressure was 32.81 mmHg, SD ± 10.94 mmHg in a range of 14 to 50 mmHg. The average post-surgical intraocular pressure at one year was 12.43 mmHg, SD ± 2.85 mmHg in a range of 7 to 19 mmHg. The difference between the pre-and post-surgery average intraocular pressure was 20.38 mmHg. The number of successes was 14 eyes (87.5%, confidence interval (CI) 95% 61.6% - 98.6%). The number of failures was two eyes (12.5%, CI 95% 1.43% - 38.4%). CONCLUSIONS: The results show that the microdevice is successful for the treatment of refractory glaucoma.
id CBO-2_d47f54e3be1df81c770d2488229425d6
oai_identifier_str oai:scielo:S0004-27492011000300011
network_acronym_str CBO-2
network_name_str Arquivos brasileiros de oftalmologia (Online)
repository_id_str
spelling Microdevice for aqueous humor drainage Maldonado Bas - PförtnerGlaucoma/surgeryFiltering surgeryIntraocular pressureTrabeculectomy/methodsAqueous humorOptical devicesSclera/surgerySclerostomyPURPOSE: A microdevice for the treatment of refractory glaucoma is presented. The underlying concepts, its mechanisms of action and the surgical technique for implanting are explained and the results are analyzed. The microdevice was developed and the surgeries were performed at the Maldonado-Bas Eye Clinic (Cordoba, Argentina), under the rules established in the protocol approved by the provisions of the National Administration of Drugs, Food and Medical Technology 430/7. File No.:1-47-25-649-07-1. METHODS: In a prospective study, following the protocol, 16 eyes with refractory glaucoma were included and operated. Intraocular pressure <21mmHg with or without additional medication was considered successful. The follow-up was one year. Averages, percentages and their 95% confidence bands were calculated. Analysis of variance for repeated measures was used to compare averages. RESULTS: The average preoperative intraocular pressure was 32.81 mmHg, SD ± 10.94 mmHg in a range of 14 to 50 mmHg. The average post-surgical intraocular pressure at one year was 12.43 mmHg, SD ± 2.85 mmHg in a range of 7 to 19 mmHg. The difference between the pre-and post-surgery average intraocular pressure was 20.38 mmHg. The number of successes was 14 eyes (87.5%, confidence interval (CI) 95% 61.6% - 98.6%). The number of failures was two eyes (12.5%, CI 95% 1.43% - 38.4%). CONCLUSIONS: The results show that the microdevice is successful for the treatment of refractory glaucoma.Conselho Brasileiro de Oftalmologia2011-06-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492011000300011Arquivos Brasileiros de Oftalmologia v.74 n.3 2011reponame:Arquivos brasileiros de oftalmologia (Online)instname:Conselho Brasileiro de Oftalmologia (CBO)instacron:CBO10.1590/S0004-27492011000300011info:eu-repo/semantics/openAccessMaldonado-Bas,ArturoMaldonado-Junyent,AnaMaldonado-Junyent,Arturoeng2011-09-02T00:00:00Zoai:scielo:S0004-27492011000300011Revistahttp://aboonline.org.br/https://old.scielo.br/oai/scielo-oai.phpaboonline@cbo.com.br||abo@cbo.com.br1678-29250004-2749opendoar:2011-09-02T00:00Arquivos brasileiros de oftalmologia (Online) - Conselho Brasileiro de Oftalmologia (CBO)false
dc.title.none.fl_str_mv Microdevice for aqueous humor drainage Maldonado Bas - Pförtner
title Microdevice for aqueous humor drainage Maldonado Bas - Pförtner
spellingShingle Microdevice for aqueous humor drainage Maldonado Bas - Pförtner
Maldonado-Bas,Arturo
Glaucoma/surgery
Filtering surgery
Intraocular pressure
Trabeculectomy/methods
Aqueous humor
Optical devices
Sclera/surgery
Sclerostomy
title_short Microdevice for aqueous humor drainage Maldonado Bas - Pförtner
title_full Microdevice for aqueous humor drainage Maldonado Bas - Pförtner
title_fullStr Microdevice for aqueous humor drainage Maldonado Bas - Pförtner
title_full_unstemmed Microdevice for aqueous humor drainage Maldonado Bas - Pförtner
title_sort Microdevice for aqueous humor drainage Maldonado Bas - Pförtner
author Maldonado-Bas,Arturo
author_facet Maldonado-Bas,Arturo
Maldonado-Junyent,Ana
Maldonado-Junyent,Arturo
author_role author
author2 Maldonado-Junyent,Ana
Maldonado-Junyent,Arturo
author2_role author
author
dc.contributor.author.fl_str_mv Maldonado-Bas,Arturo
Maldonado-Junyent,Ana
Maldonado-Junyent,Arturo
dc.subject.por.fl_str_mv Glaucoma/surgery
Filtering surgery
Intraocular pressure
Trabeculectomy/methods
Aqueous humor
Optical devices
Sclera/surgery
Sclerostomy
topic Glaucoma/surgery
Filtering surgery
Intraocular pressure
Trabeculectomy/methods
Aqueous humor
Optical devices
Sclera/surgery
Sclerostomy
description PURPOSE: A microdevice for the treatment of refractory glaucoma is presented. The underlying concepts, its mechanisms of action and the surgical technique for implanting are explained and the results are analyzed. The microdevice was developed and the surgeries were performed at the Maldonado-Bas Eye Clinic (Cordoba, Argentina), under the rules established in the protocol approved by the provisions of the National Administration of Drugs, Food and Medical Technology 430/7. File No.:1-47-25-649-07-1. METHODS: In a prospective study, following the protocol, 16 eyes with refractory glaucoma were included and operated. Intraocular pressure <21mmHg with or without additional medication was considered successful. The follow-up was one year. Averages, percentages and their 95% confidence bands were calculated. Analysis of variance for repeated measures was used to compare averages. RESULTS: The average preoperative intraocular pressure was 32.81 mmHg, SD ± 10.94 mmHg in a range of 14 to 50 mmHg. The average post-surgical intraocular pressure at one year was 12.43 mmHg, SD ± 2.85 mmHg in a range of 7 to 19 mmHg. The difference between the pre-and post-surgery average intraocular pressure was 20.38 mmHg. The number of successes was 14 eyes (87.5%, confidence interval (CI) 95% 61.6% - 98.6%). The number of failures was two eyes (12.5%, CI 95% 1.43% - 38.4%). CONCLUSIONS: The results show that the microdevice is successful for the treatment of refractory glaucoma.
publishDate 2011
dc.date.none.fl_str_mv 2011-06-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492011000300011
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492011000300011
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/S0004-27492011000300011
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Conselho Brasileiro de Oftalmologia
publisher.none.fl_str_mv Conselho Brasileiro de Oftalmologia
dc.source.none.fl_str_mv Arquivos Brasileiros de Oftalmologia v.74 n.3 2011
reponame:Arquivos brasileiros de oftalmologia (Online)
instname:Conselho Brasileiro de Oftalmologia (CBO)
instacron:CBO
instname_str Conselho Brasileiro de Oftalmologia (CBO)
instacron_str CBO
institution CBO
reponame_str Arquivos brasileiros de oftalmologia (Online)
collection Arquivos brasileiros de oftalmologia (Online)
repository.name.fl_str_mv Arquivos brasileiros de oftalmologia (Online) - Conselho Brasileiro de Oftalmologia (CBO)
repository.mail.fl_str_mv aboonline@cbo.com.br||abo@cbo.com.br
_version_ 1754209026849112064