Topographic paracentral corneal thickness with pentacam and orbscan: effect of acoustic factor

Detalhes bibliográficos
Autor(a) principal: González-Pérez, Javier
Data de Publicação: 2011
Outros Autores: González-Méijome, José Manuel, Rodríguez Ares, María T., Parafita, Manuel A.
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
Texto Completo: http://hdl.handle.net/1822/17154
Resumo: Purpose: To evaluate the effect of an acoustic factor (AF) on the comparison of central corneal thickness (CCT) and peripheral corneal thickness (PCT) measurements with Orbscan II and Pentacam. Methods: The CCT and PCT at 1, 2, and 3 radial distances from the corneal apex were measured using Orbscan II and Pentacam in 22 right eyes of 22 healthy adults (7 men, 15 women). Three measures were obtained from each 1 of the 25 points measured and then compared to gauge the agreement between both devices at the corneal center and anular areas located at 1-, 2-, and 3-mm distances from the central measurement. Orbscan II readings were considered with and without an AF correction. Results: Pentacam provides statistically significant higher values than Orbscan II does at all the 25 locations analyzed (P,0.001). With a few exceptions, the average difference was fairly constant between 20 and 40 mm for all the corneal locations. A high correlation existed between central readings (r2=0.927; P,0.001) and average thickness at each one of the annular areas being analyzed (r2=0.897 at 1 mm, r2=0.876 at 2 mm, and r2=0.870 at 3 mm); Pentacam minus Orbscan II value averaged for all the points changed from 228610 to 22269 mm after the removal of the AF in Orbscan II. Conclusions: Central and peripheral measurements of the corneal thickness obtained with Orbscan II and Pentacam are significantly different. The removal of the AF in Orbscan II renders lower mean differences but decreases the agreement between both systems and potentially induces an overestimation of CCT and PCT by Orbsan II compared with Pentacam.
id RCAP_85a7f1e82cdcc5d8c6e301df7ffd43f7
oai_identifier_str oai:repositorium.sdum.uminho.pt:1822/17154
network_acronym_str RCAP
network_name_str Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
repository_id_str 7160
spelling Topographic paracentral corneal thickness with pentacam and orbscan: effect of acoustic factorPachometryPeripheral corneal thicknessPentacamOrbscan IIBlandAltmanBland-AltmanScience & TechnologyPurpose: To evaluate the effect of an acoustic factor (AF) on the comparison of central corneal thickness (CCT) and peripheral corneal thickness (PCT) measurements with Orbscan II and Pentacam. Methods: The CCT and PCT at 1, 2, and 3 radial distances from the corneal apex were measured using Orbscan II and Pentacam in 22 right eyes of 22 healthy adults (7 men, 15 women). Three measures were obtained from each 1 of the 25 points measured and then compared to gauge the agreement between both devices at the corneal center and anular areas located at 1-, 2-, and 3-mm distances from the central measurement. Orbscan II readings were considered with and without an AF correction. Results: Pentacam provides statistically significant higher values than Orbscan II does at all the 25 locations analyzed (P,0.001). With a few exceptions, the average difference was fairly constant between 20 and 40 mm for all the corneal locations. A high correlation existed between central readings (r2=0.927; P,0.001) and average thickness at each one of the annular areas being analyzed (r2=0.897 at 1 mm, r2=0.876 at 2 mm, and r2=0.870 at 3 mm); Pentacam minus Orbscan II value averaged for all the points changed from 228610 to 22269 mm after the removal of the AF in Orbscan II. Conclusions: Central and peripheral measurements of the corneal thickness obtained with Orbscan II and Pentacam are significantly different. The removal of the AF in Orbscan II renders lower mean differences but decreases the agreement between both systems and potentially induces an overestimation of CCT and PCT by Orbsan II compared with Pentacam.Supported by national grant PI081380 from Instituto de Salud Carlos III, Spain granted to Ocular Surface and Contact Lens Research Group GI-1750.Lippincott, Williams & WilkinsUniversidade do MinhoGonzález-Pérez, JavierGonzález-Méijome, José ManuelRodríguez Ares, María T.Parafita, Manuel A.20112011-01-01T00:00:00Zinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articleapplication/pdfhttp://hdl.handle.net/1822/17154eng1542-232110.1097/ICL.0b013e3182323dcf21983548http://journals.lww.com/pages/default.aspxinfo:eu-repo/semantics/openAccessreponame:Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)instname:Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informaçãoinstacron:RCAAP2023-07-21T12:00:38Zoai:repositorium.sdum.uminho.pt:1822/17154Portal AgregadorONGhttps://www.rcaap.pt/oai/openaireopendoar:71602024-03-19T18:50:29.833878Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos) - Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informaçãofalse
dc.title.none.fl_str_mv Topographic paracentral corneal thickness with pentacam and orbscan: effect of acoustic factor
title Topographic paracentral corneal thickness with pentacam and orbscan: effect of acoustic factor
spellingShingle Topographic paracentral corneal thickness with pentacam and orbscan: effect of acoustic factor
González-Pérez, Javier
Pachometry
Peripheral corneal thickness
Pentacam
Orbscan II
Bland
Altman
Bland-Altman
Science & Technology
title_short Topographic paracentral corneal thickness with pentacam and orbscan: effect of acoustic factor
title_full Topographic paracentral corneal thickness with pentacam and orbscan: effect of acoustic factor
title_fullStr Topographic paracentral corneal thickness with pentacam and orbscan: effect of acoustic factor
title_full_unstemmed Topographic paracentral corneal thickness with pentacam and orbscan: effect of acoustic factor
title_sort Topographic paracentral corneal thickness with pentacam and orbscan: effect of acoustic factor
author González-Pérez, Javier
author_facet González-Pérez, Javier
González-Méijome, José Manuel
Rodríguez Ares, María T.
Parafita, Manuel A.
author_role author
author2 González-Méijome, José Manuel
Rodríguez Ares, María T.
Parafita, Manuel A.
author2_role author
author
author
dc.contributor.none.fl_str_mv Universidade do Minho
dc.contributor.author.fl_str_mv González-Pérez, Javier
González-Méijome, José Manuel
Rodríguez Ares, María T.
Parafita, Manuel A.
dc.subject.por.fl_str_mv Pachometry
Peripheral corneal thickness
Pentacam
Orbscan II
Bland
Altman
Bland-Altman
Science & Technology
topic Pachometry
Peripheral corneal thickness
Pentacam
Orbscan II
Bland
Altman
Bland-Altman
Science & Technology
description Purpose: To evaluate the effect of an acoustic factor (AF) on the comparison of central corneal thickness (CCT) and peripheral corneal thickness (PCT) measurements with Orbscan II and Pentacam. Methods: The CCT and PCT at 1, 2, and 3 radial distances from the corneal apex were measured using Orbscan II and Pentacam in 22 right eyes of 22 healthy adults (7 men, 15 women). Three measures were obtained from each 1 of the 25 points measured and then compared to gauge the agreement between both devices at the corneal center and anular areas located at 1-, 2-, and 3-mm distances from the central measurement. Orbscan II readings were considered with and without an AF correction. Results: Pentacam provides statistically significant higher values than Orbscan II does at all the 25 locations analyzed (P,0.001). With a few exceptions, the average difference was fairly constant between 20 and 40 mm for all the corneal locations. A high correlation existed between central readings (r2=0.927; P,0.001) and average thickness at each one of the annular areas being analyzed (r2=0.897 at 1 mm, r2=0.876 at 2 mm, and r2=0.870 at 3 mm); Pentacam minus Orbscan II value averaged for all the points changed from 228610 to 22269 mm after the removal of the AF in Orbscan II. Conclusions: Central and peripheral measurements of the corneal thickness obtained with Orbscan II and Pentacam are significantly different. The removal of the AF in Orbscan II renders lower mean differences but decreases the agreement between both systems and potentially induces an overestimation of CCT and PCT by Orbsan II compared with Pentacam.
publishDate 2011
dc.date.none.fl_str_mv 2011
2011-01-01T00:00:00Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://hdl.handle.net/1822/17154
url http://hdl.handle.net/1822/17154
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 1542-2321
10.1097/ICL.0b013e3182323dcf
21983548
http://journals.lww.com/pages/default.aspx
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Lippincott, Williams & Wilkins
publisher.none.fl_str_mv Lippincott, Williams & Wilkins
dc.source.none.fl_str_mv reponame:Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
instname:Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
instacron:RCAAP
instname_str Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
instacron_str RCAAP
institution RCAAP
reponame_str Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
collection Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
repository.name.fl_str_mv Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos) - Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
repository.mail.fl_str_mv
_version_ 1799132274020581376