What s new in the Visionaire system® in second-generation for TKA? a comparative study of accuracy, efficiency and functional results

Detalhes bibliográficos
Autor(a) principal: Fontes, Ana Paula
Data de Publicação: 2022
Outros Autores: Cintra, Rui Miguel, Gomes, Luis, Sousa, Joao Paulo
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
Texto Completo: http://hdl.handle.net/10400.1/18055
Resumo: Background: The aim of the study was to analyze and compare the accuracy, efficiency and functional evolution between the first- and second-generation Patient-specific instrumentation (PSI). Methods: We report our experience in TKA using PSI Visionaire System® (Smith and Nephew®) in 456 procedures. The patients were divided into first- (N=272) and second-generation PSI design (N=184). For the accuracy was analyzed the mechanical results, namely Hip-knee-ankle (HKA) alignment post-TKA and outliers’ frequency; for the efficiency was analyzed the length time surgery, the length of stay and satisfaction, for the functional outcomes was compared the evolution at pain, range of motion, gait perimeter and the domains of the Western ontario and mcmaster osteoarthritis index (WOMAC). The t-test for independent samples was applied in the continuous numeric variables and Qui square tests in the dichotomic nominal variables. Statistical significance was p value<0.05. Results: The alignment of the femur worsened significantly in the second-generation (p=0.002), but the alignment of the tibia was improved (p=0.010). However, not enough to improve significantly the HKA. No significant differences were observed in the frequency of outliers. The length of stay decreased in the second-generation (<0.001). The functioning achieved at first and third months after surgery was lower in second-generation PSI; at first month in the pain and stiffness WOMAC (p=0.030; p>0.001) and in the third in WOMAC stiffness (p<0.001). Conclusions: The second-generation PSI of Visionaire system® improved the alignment of the tibia, but not the alignment of the femur. This change did not produce any improvements in functioning.
id RCAP_bbae0c44b73daa84dcdaf86cd2f9b01a
oai_identifier_str oai:sapientia.ualg.pt:10400.1/18055
network_acronym_str RCAP
network_name_str Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
repository_id_str 7160
spelling What s new in the Visionaire system® in second-generation for TKA? a comparative study of accuracy, efficiency and functional resultsPatient-specific instrumentationTotal knee arthroplastyTotal knee replacementFunctional outcomesNavigation systemVisionaire systemBackground: The aim of the study was to analyze and compare the accuracy, efficiency and functional evolution between the first- and second-generation Patient-specific instrumentation (PSI). Methods: We report our experience in TKA using PSI Visionaire System® (Smith and Nephew®) in 456 procedures. The patients were divided into first- (N=272) and second-generation PSI design (N=184). For the accuracy was analyzed the mechanical results, namely Hip-knee-ankle (HKA) alignment post-TKA and outliers’ frequency; for the efficiency was analyzed the length time surgery, the length of stay and satisfaction, for the functional outcomes was compared the evolution at pain, range of motion, gait perimeter and the domains of the Western ontario and mcmaster osteoarthritis index (WOMAC). The t-test for independent samples was applied in the continuous numeric variables and Qui square tests in the dichotomic nominal variables. Statistical significance was p value<0.05. Results: The alignment of the femur worsened significantly in the second-generation (p=0.002), but the alignment of the tibia was improved (p=0.010). However, not enough to improve significantly the HKA. No significant differences were observed in the frequency of outliers. The length of stay decreased in the second-generation (<0.001). The functioning achieved at first and third months after surgery was lower in second-generation PSI; at first month in the pain and stiffness WOMAC (p=0.030; p>0.001) and in the third in WOMAC stiffness (p<0.001). Conclusions: The second-generation PSI of Visionaire system® improved the alignment of the tibia, but not the alignment of the femur. This change did not produce any improvements in functioning.Medip AcademySapientiaFontes, Ana PaulaCintra, Rui MiguelGomes, LuisSousa, Joao Paulo2022-07-21T10:45:39Z2022-04-252022-07-20T16:52:10Z2022-04-25T00:00:00Zinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articleapplication/pdfhttp://hdl.handle.net/10400.1/18055eng2455-4510cv-prod-299443210.18203/issn.2455-4510.intjresorthop20221117info:eu-repo/semantics/openAccessreponame:Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)instname:Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informaçãoinstacron:RCAAP2023-07-24T10:30:17Zoai:sapientia.ualg.pt:10400.1/18055Portal AgregadorONGhttps://www.rcaap.pt/oai/openaireopendoar:71602024-03-19T20:07:52.392601Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos) - Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informaçãofalse
dc.title.none.fl_str_mv What s new in the Visionaire system® in second-generation for TKA? a comparative study of accuracy, efficiency and functional results
title What s new in the Visionaire system® in second-generation for TKA? a comparative study of accuracy, efficiency and functional results
spellingShingle What s new in the Visionaire system® in second-generation for TKA? a comparative study of accuracy, efficiency and functional results
Fontes, Ana Paula
Patient-specific instrumentation
Total knee arthroplasty
Total knee replacement
Functional outcomes
Navigation system
Visionaire system
title_short What s new in the Visionaire system® in second-generation for TKA? a comparative study of accuracy, efficiency and functional results
title_full What s new in the Visionaire system® in second-generation for TKA? a comparative study of accuracy, efficiency and functional results
title_fullStr What s new in the Visionaire system® in second-generation for TKA? a comparative study of accuracy, efficiency and functional results
title_full_unstemmed What s new in the Visionaire system® in second-generation for TKA? a comparative study of accuracy, efficiency and functional results
title_sort What s new in the Visionaire system® in second-generation for TKA? a comparative study of accuracy, efficiency and functional results
author Fontes, Ana Paula
author_facet Fontes, Ana Paula
Cintra, Rui Miguel
Gomes, Luis
Sousa, Joao Paulo
author_role author
author2 Cintra, Rui Miguel
Gomes, Luis
Sousa, Joao Paulo
author2_role author
author
author
dc.contributor.none.fl_str_mv Sapientia
dc.contributor.author.fl_str_mv Fontes, Ana Paula
Cintra, Rui Miguel
Gomes, Luis
Sousa, Joao Paulo
dc.subject.por.fl_str_mv Patient-specific instrumentation
Total knee arthroplasty
Total knee replacement
Functional outcomes
Navigation system
Visionaire system
topic Patient-specific instrumentation
Total knee arthroplasty
Total knee replacement
Functional outcomes
Navigation system
Visionaire system
description Background: The aim of the study was to analyze and compare the accuracy, efficiency and functional evolution between the first- and second-generation Patient-specific instrumentation (PSI). Methods: We report our experience in TKA using PSI Visionaire System® (Smith and Nephew®) in 456 procedures. The patients were divided into first- (N=272) and second-generation PSI design (N=184). For the accuracy was analyzed the mechanical results, namely Hip-knee-ankle (HKA) alignment post-TKA and outliers’ frequency; for the efficiency was analyzed the length time surgery, the length of stay and satisfaction, for the functional outcomes was compared the evolution at pain, range of motion, gait perimeter and the domains of the Western ontario and mcmaster osteoarthritis index (WOMAC). The t-test for independent samples was applied in the continuous numeric variables and Qui square tests in the dichotomic nominal variables. Statistical significance was p value<0.05. Results: The alignment of the femur worsened significantly in the second-generation (p=0.002), but the alignment of the tibia was improved (p=0.010). However, not enough to improve significantly the HKA. No significant differences were observed in the frequency of outliers. The length of stay decreased in the second-generation (<0.001). The functioning achieved at first and third months after surgery was lower in second-generation PSI; at first month in the pain and stiffness WOMAC (p=0.030; p>0.001) and in the third in WOMAC stiffness (p<0.001). Conclusions: The second-generation PSI of Visionaire system® improved the alignment of the tibia, but not the alignment of the femur. This change did not produce any improvements in functioning.
publishDate 2022
dc.date.none.fl_str_mv 2022-07-21T10:45:39Z
2022-04-25
2022-07-20T16:52:10Z
2022-04-25T00:00:00Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://hdl.handle.net/10400.1/18055
url http://hdl.handle.net/10400.1/18055
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 2455-4510
cv-prod-2994432
10.18203/issn.2455-4510.intjresorthop20221117
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Medip Academy
publisher.none.fl_str_mv Medip Academy
dc.source.none.fl_str_mv reponame:Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
instname:Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
instacron:RCAAP
instname_str Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
instacron_str RCAAP
institution RCAAP
reponame_str Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
collection Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
repository.name.fl_str_mv Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos) - Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
repository.mail.fl_str_mv
_version_ 1799133324390694912