Technical Note: Accuracy and repeatability of a new portable ultrasound pachymeter

Detalhes bibliográficos
Autor(a) principal: Queirós, A.
Data de Publicação: 2007
Outros Autores: González-Méijome, José Manuel, Fernandes, Paulo, Jorge, Jorge, Almeida, José B., Parafita, M. A.
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
Texto Completo: http://hdl.handle.net/1822/66250
Resumo: Purpose: To assess the accuracy and repeatability of central corneal thickness (CCT) measurements taken with a new portable ultrasound (US) pachymeter.Methods: Central thickness measurements were taken with a portable pachymeter (SP-100 Handy; Tomey, Nagoya, Japan) and a conventional US pachymeter (Nidek UP-1000; Nidek Technologies, Gamagori, Japan) from 57 right corneas of 57 young adults (19 males, 38 females) aged 18-44 years (mean +/- S.D., 22.95 +/- 3.92). Three repeated measures were obtained and then compared to obtain the repeatability of each instrument and the agreement between the pachymeters. The three readings taken with the portable pachymeter were compared against each other in order to evaluate intra-session repeatability and bias of each individual measurement with respect to the mean of three.Results: Mean values of CCT were 537 +/- 35 mu m for conventional and 534 +/- 35 mu m for the new portable pachymeter. A high agreement was found between the two instruments (mean difference = 2.58 mu m; 95% CI 1.41-3.75 mu m) with only two eyes presenting differences larger than +/- 8.6 mu m which represents 95% CI in the Bland-Altman plots which represents 1.6% of the mean CCT. The first reading taken showed the highest agreement with the mean value for the portable pachymeter.Conclusions: The instrument tested in this study is able to take reliable measurements of corneal thickness even if a single reading is considered. Intra-session repeatability was very high, as was also the agreement between the average of three readings taken with the two US pachymeters.
id RCAP_f29aff4274b0004dea7e39a06035bf8f
oai_identifier_str oai:repositorium.sdum.uminho.pt:1822/66250
network_acronym_str RCAP
network_name_str Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
repository_id_str 7160
spelling Technical Note: Accuracy and repeatability of a new portable ultrasound pachymetercorneal thicknesspachymeter comparisonportable ultrasound pachymetryScience & TechnologyPurpose: To assess the accuracy and repeatability of central corneal thickness (CCT) measurements taken with a new portable ultrasound (US) pachymeter.Methods: Central thickness measurements were taken with a portable pachymeter (SP-100 Handy; Tomey, Nagoya, Japan) and a conventional US pachymeter (Nidek UP-1000; Nidek Technologies, Gamagori, Japan) from 57 right corneas of 57 young adults (19 males, 38 females) aged 18-44 years (mean +/- S.D., 22.95 +/- 3.92). Three repeated measures were obtained and then compared to obtain the repeatability of each instrument and the agreement between the pachymeters. The three readings taken with the portable pachymeter were compared against each other in order to evaluate intra-session repeatability and bias of each individual measurement with respect to the mean of three.Results: Mean values of CCT were 537 +/- 35 mu m for conventional and 534 +/- 35 mu m for the new portable pachymeter. A high agreement was found between the two instruments (mean difference = 2.58 mu m; 95% CI 1.41-3.75 mu m) with only two eyes presenting differences larger than +/- 8.6 mu m which represents 95% CI in the Bland-Altman plots which represents 1.6% of the mean CCT. The first reading taken showed the highest agreement with the mean value for the portable pachymeter.Conclusions: The instrument tested in this study is able to take reliable measurements of corneal thickness even if a single reading is considered. Intra-session repeatability was very high, as was also the agreement between the average of three readings taken with the two US pachymeters.- (undefined)WileyUniversidade do MinhoQueirós, A.González-Méijome, José ManuelFernandes, PauloJorge, JorgeAlmeida, José B.Parafita, M. A.20072007-01-01T00:00:00Zinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articleapplication/pdfhttp://hdl.handle.net/1822/66250eng0275-540810.1111/j.1475-1313.2006.00429.x17324209info:eu-repo/semantics/openAccessreponame:Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)instname:Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informaçãoinstacron:RCAAP2023-07-21T12:30:14Zoai:repositorium.sdum.uminho.pt:1822/66250Portal AgregadorONGhttps://www.rcaap.pt/oai/openaireopendoar:71602024-03-19T19:25:22.138176Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos) - Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informaçãofalse
dc.title.none.fl_str_mv Technical Note: Accuracy and repeatability of a new portable ultrasound pachymeter
title Technical Note: Accuracy and repeatability of a new portable ultrasound pachymeter
spellingShingle Technical Note: Accuracy and repeatability of a new portable ultrasound pachymeter
Queirós, A.
corneal thickness
pachymeter comparison
portable ultrasound pachymetry
Science & Technology
title_short Technical Note: Accuracy and repeatability of a new portable ultrasound pachymeter
title_full Technical Note: Accuracy and repeatability of a new portable ultrasound pachymeter
title_fullStr Technical Note: Accuracy and repeatability of a new portable ultrasound pachymeter
title_full_unstemmed Technical Note: Accuracy and repeatability of a new portable ultrasound pachymeter
title_sort Technical Note: Accuracy and repeatability of a new portable ultrasound pachymeter
author Queirós, A.
author_facet Queirós, A.
González-Méijome, José Manuel
Fernandes, Paulo
Jorge, Jorge
Almeida, José B.
Parafita, M. A.
author_role author
author2 González-Méijome, José Manuel
Fernandes, Paulo
Jorge, Jorge
Almeida, José B.
Parafita, M. A.
author2_role author
author
author
author
author
dc.contributor.none.fl_str_mv Universidade do Minho
dc.contributor.author.fl_str_mv Queirós, A.
González-Méijome, José Manuel
Fernandes, Paulo
Jorge, Jorge
Almeida, José B.
Parafita, M. A.
dc.subject.por.fl_str_mv corneal thickness
pachymeter comparison
portable ultrasound pachymetry
Science & Technology
topic corneal thickness
pachymeter comparison
portable ultrasound pachymetry
Science & Technology
description Purpose: To assess the accuracy and repeatability of central corneal thickness (CCT) measurements taken with a new portable ultrasound (US) pachymeter.Methods: Central thickness measurements were taken with a portable pachymeter (SP-100 Handy; Tomey, Nagoya, Japan) and a conventional US pachymeter (Nidek UP-1000; Nidek Technologies, Gamagori, Japan) from 57 right corneas of 57 young adults (19 males, 38 females) aged 18-44 years (mean +/- S.D., 22.95 +/- 3.92). Three repeated measures were obtained and then compared to obtain the repeatability of each instrument and the agreement between the pachymeters. The three readings taken with the portable pachymeter were compared against each other in order to evaluate intra-session repeatability and bias of each individual measurement with respect to the mean of three.Results: Mean values of CCT were 537 +/- 35 mu m for conventional and 534 +/- 35 mu m for the new portable pachymeter. A high agreement was found between the two instruments (mean difference = 2.58 mu m; 95% CI 1.41-3.75 mu m) with only two eyes presenting differences larger than +/- 8.6 mu m which represents 95% CI in the Bland-Altman plots which represents 1.6% of the mean CCT. The first reading taken showed the highest agreement with the mean value for the portable pachymeter.Conclusions: The instrument tested in this study is able to take reliable measurements of corneal thickness even if a single reading is considered. Intra-session repeatability was very high, as was also the agreement between the average of three readings taken with the two US pachymeters.
publishDate 2007
dc.date.none.fl_str_mv 2007
2007-01-01T00:00:00Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://hdl.handle.net/1822/66250
url http://hdl.handle.net/1822/66250
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 0275-5408
10.1111/j.1475-1313.2006.00429.x
17324209
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Wiley
publisher.none.fl_str_mv Wiley
dc.source.none.fl_str_mv reponame:Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
instname:Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
instacron:RCAAP
instname_str Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
instacron_str RCAAP
institution RCAAP
reponame_str Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
collection Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos)
repository.name.fl_str_mv Repositório Científico de Acesso Aberto de Portugal (Repositórios Cientìficos) - Agência para a Sociedade do Conhecimento (UMIC) - FCT - Sociedade da Informação
repository.mail.fl_str_mv
_version_ 1799132737560379392