Bilateral fracture of L5 pedicles in a patient with total disc replacement of L5-S1: a case report
Autor(a) principal: | |
---|---|
Data de Publicação: | 2014 |
Outros Autores: | , , , |
Tipo de documento: | Artigo |
Idioma: | eng |
Título da fonte: | Coluna/Columna |
Texto Completo: | http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1808-18512014000200153 |
Resumo: | Report of a rare complication - fracture of the pedicles - in a patient with total disc replacement of L5-S1, a surgical resolution, and a biomechanical explanation. To the authors' knowledge, there is only one previous report of bilateral fracture of the pedicles in the literature, as a complication in total disc replacement of the lumbar spine. In this case, no direct repair was made to the fracture site; instead intersomatic fusion was performed by the anterior approach. A 40-year-old male, a martial arts practitioner who had undergone L5-S1 (ProDisc(r)) disc replacement nine months earlier, with complete resolution of the preoperative symptoms and no complications, sudden pain during physical activity, without neurological symptoms. Computed axial tomography showed a fracture of the L5 pedicles and anterior luxation of the polyethylene insert. Surgery was performed by the posterior approach, with direct repair of the fractures and posterolateral fusion of L5-S1 with transpedicular screws (Schanz) and USS(r) internal fixator. A follow-up axial CT scan at 6 months after surgery showed complete consolidation of the fractures, and the patient was asymptomatic. Due to the alteration in weight transmission through the anterior part of the spine in the total replacement of the lumbar disc, which preserves the movement but not the absorption of forces, the pedicle becomes more susceptible to fracture. It is important to bear this complication in patients submitted to this procedure. |
id |
SBCO-1_704aec449f1c08123840f60e14f0b8f8 |
---|---|
oai_identifier_str |
oai:scielo:S1808-18512014000200153 |
network_acronym_str |
SBCO-1 |
network_name_str |
Coluna/Columna |
repository_id_str |
|
spelling |
Bilateral fracture of L5 pedicles in a patient with total disc replacement of L5-S1: a case reportIntervertebral discSpineSpinal fusionBone screwsReport of a rare complication - fracture of the pedicles - in a patient with total disc replacement of L5-S1, a surgical resolution, and a biomechanical explanation. To the authors' knowledge, there is only one previous report of bilateral fracture of the pedicles in the literature, as a complication in total disc replacement of the lumbar spine. In this case, no direct repair was made to the fracture site; instead intersomatic fusion was performed by the anterior approach. A 40-year-old male, a martial arts practitioner who had undergone L5-S1 (ProDisc(r)) disc replacement nine months earlier, with complete resolution of the preoperative symptoms and no complications, sudden pain during physical activity, without neurological symptoms. Computed axial tomography showed a fracture of the L5 pedicles and anterior luxation of the polyethylene insert. Surgery was performed by the posterior approach, with direct repair of the fractures and posterolateral fusion of L5-S1 with transpedicular screws (Schanz) and USS(r) internal fixator. A follow-up axial CT scan at 6 months after surgery showed complete consolidation of the fractures, and the patient was asymptomatic. Due to the alteration in weight transmission through the anterior part of the spine in the total replacement of the lumbar disc, which preserves the movement but not the absorption of forces, the pedicle becomes more susceptible to fracture. It is important to bear this complication in patients submitted to this procedure.Sociedade Brasileira de Coluna2014-01-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S1808-18512014000200153Coluna/Columna v.13 n.2 2014reponame:Coluna/Columnainstname:Sociedade Brasileira de Coluna (SBCO)instacron:SBCO10.1590/S1808-18512014130200456info:eu-repo/semantics/openAccessEstrada,Jorge Alberto De HaroFernandez,Javier RamosGarnier,Juan Carlos AlvarezJohnson,Michael DittmarLopez,Francisco Cruzeng2015-08-18T00:00:00Zoai:scielo:S1808-18512014000200153Revistahttps://www.revistacoluna.org/ONGhttps://old.scielo.br/oai/scielo-oai.phpcoluna.columna@uol.com.br||revistacoluna@uol.com.br2177-014X1808-1851opendoar:2015-08-18T00:00Coluna/Columna - Sociedade Brasileira de Coluna (SBCO)false |
dc.title.none.fl_str_mv |
Bilateral fracture of L5 pedicles in a patient with total disc replacement of L5-S1: a case report |
title |
Bilateral fracture of L5 pedicles in a patient with total disc replacement of L5-S1: a case report |
spellingShingle |
Bilateral fracture of L5 pedicles in a patient with total disc replacement of L5-S1: a case report Estrada,Jorge Alberto De Haro Intervertebral disc Spine Spinal fusion Bone screws |
title_short |
Bilateral fracture of L5 pedicles in a patient with total disc replacement of L5-S1: a case report |
title_full |
Bilateral fracture of L5 pedicles in a patient with total disc replacement of L5-S1: a case report |
title_fullStr |
Bilateral fracture of L5 pedicles in a patient with total disc replacement of L5-S1: a case report |
title_full_unstemmed |
Bilateral fracture of L5 pedicles in a patient with total disc replacement of L5-S1: a case report |
title_sort |
Bilateral fracture of L5 pedicles in a patient with total disc replacement of L5-S1: a case report |
author |
Estrada,Jorge Alberto De Haro |
author_facet |
Estrada,Jorge Alberto De Haro Fernandez,Javier Ramos Garnier,Juan Carlos Alvarez Johnson,Michael Dittmar Lopez,Francisco Cruz |
author_role |
author |
author2 |
Fernandez,Javier Ramos Garnier,Juan Carlos Alvarez Johnson,Michael Dittmar Lopez,Francisco Cruz |
author2_role |
author author author author |
dc.contributor.author.fl_str_mv |
Estrada,Jorge Alberto De Haro Fernandez,Javier Ramos Garnier,Juan Carlos Alvarez Johnson,Michael Dittmar Lopez,Francisco Cruz |
dc.subject.por.fl_str_mv |
Intervertebral disc Spine Spinal fusion Bone screws |
topic |
Intervertebral disc Spine Spinal fusion Bone screws |
description |
Report of a rare complication - fracture of the pedicles - in a patient with total disc replacement of L5-S1, a surgical resolution, and a biomechanical explanation. To the authors' knowledge, there is only one previous report of bilateral fracture of the pedicles in the literature, as a complication in total disc replacement of the lumbar spine. In this case, no direct repair was made to the fracture site; instead intersomatic fusion was performed by the anterior approach. A 40-year-old male, a martial arts practitioner who had undergone L5-S1 (ProDisc(r)) disc replacement nine months earlier, with complete resolution of the preoperative symptoms and no complications, sudden pain during physical activity, without neurological symptoms. Computed axial tomography showed a fracture of the L5 pedicles and anterior luxation of the polyethylene insert. Surgery was performed by the posterior approach, with direct repair of the fractures and posterolateral fusion of L5-S1 with transpedicular screws (Schanz) and USS(r) internal fixator. A follow-up axial CT scan at 6 months after surgery showed complete consolidation of the fractures, and the patient was asymptomatic. Due to the alteration in weight transmission through the anterior part of the spine in the total replacement of the lumbar disc, which preserves the movement but not the absorption of forces, the pedicle becomes more susceptible to fracture. It is important to bear this complication in patients submitted to this procedure. |
publishDate |
2014 |
dc.date.none.fl_str_mv |
2014-01-01 |
dc.type.driver.fl_str_mv |
info:eu-repo/semantics/article |
dc.type.status.fl_str_mv |
info:eu-repo/semantics/publishedVersion |
format |
article |
status_str |
publishedVersion |
dc.identifier.uri.fl_str_mv |
http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1808-18512014000200153 |
url |
http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1808-18512014000200153 |
dc.language.iso.fl_str_mv |
eng |
language |
eng |
dc.relation.none.fl_str_mv |
10.1590/S1808-18512014130200456 |
dc.rights.driver.fl_str_mv |
info:eu-repo/semantics/openAccess |
eu_rights_str_mv |
openAccess |
dc.format.none.fl_str_mv |
text/html |
dc.publisher.none.fl_str_mv |
Sociedade Brasileira de Coluna |
publisher.none.fl_str_mv |
Sociedade Brasileira de Coluna |
dc.source.none.fl_str_mv |
Coluna/Columna v.13 n.2 2014 reponame:Coluna/Columna instname:Sociedade Brasileira de Coluna (SBCO) instacron:SBCO |
instname_str |
Sociedade Brasileira de Coluna (SBCO) |
instacron_str |
SBCO |
institution |
SBCO |
reponame_str |
Coluna/Columna |
collection |
Coluna/Columna |
repository.name.fl_str_mv |
Coluna/Columna - Sociedade Brasileira de Coluna (SBCO) |
repository.mail.fl_str_mv |
coluna.columna@uol.com.br||revistacoluna@uol.com.br |
_version_ |
1752126614705012736 |