Repeatability of central corneal thickness measurement with the Pentacam HR system

Detalhes bibliográficos
Autor(a) principal: Alonso,Ruiz Simonato
Data de Publicação: 2012
Outros Autores: Fontes,Bruno Machado, Ventura,Marcelo Palis, Ambrósio Jr,Renato
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Revista Brasileira de Oftalmologia (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0034-72802012000100003
Resumo: PURPOSE: To assess the repeatability of central corneal thickness measurement at the geometrical center (Central Corneal Thickness - CCT) given by the Pentacam High Resolution (HR) Comprehensive Eye Scanner (Oculus, Wetzlar, Germany) over time. METHODS: Prospective, single center, observational study. Two separate CCT measurements were taken by the Pentacam corneal tomography exam (CTm) 3 to 12 months apart, and compared. RESULTS: One hundred and sixteen eyes (n=116) of 62 health patients were included in this study. Average CCT in first and last visits was 541.6±37 µm and 543.6±36.9 µm respectively. Mean difference between both measurements was 9.2±6.4 µm, and there was no statistically significant difference in CCT measurement between visits, with good correlation between them (P = 0.057, r² = 0,9209). CONCLUSION: Pentacam (HR) CTm gives repeatable CCT measurements over time.
id SBO-1_eced60b8fd32b200e17c86cc60609fd7
oai_identifier_str oai:scielo:S0034-72802012000100003
network_acronym_str SBO-1
network_name_str Revista Brasileira de Oftalmologia (Online)
repository_id_str
spelling Repeatability of central corneal thickness measurement with the Pentacam HR systemCornea/anatomy & histologyCorneal topographyBiometry/methodsReproducibility of resultsDiagnostic techniquesophthalmological/instrumentationPURPOSE: To assess the repeatability of central corneal thickness measurement at the geometrical center (Central Corneal Thickness - CCT) given by the Pentacam High Resolution (HR) Comprehensive Eye Scanner (Oculus, Wetzlar, Germany) over time. METHODS: Prospective, single center, observational study. Two separate CCT measurements were taken by the Pentacam corneal tomography exam (CTm) 3 to 12 months apart, and compared. RESULTS: One hundred and sixteen eyes (n=116) of 62 health patients were included in this study. Average CCT in first and last visits was 541.6±37 µm and 543.6±36.9 µm respectively. Mean difference between both measurements was 9.2±6.4 µm, and there was no statistically significant difference in CCT measurement between visits, with good correlation between them (P = 0.057, r² = 0,9209). CONCLUSION: Pentacam (HR) CTm gives repeatable CCT measurements over time.Sociedade Brasileira de Oftalmologia2012-02-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S0034-72802012000100003Revista Brasileira de Oftalmologia v.71 n.1 2012reponame:Revista Brasileira de Oftalmologia (Online)instname:Sociedade Brasileira de Oftalmologia (SBO)instacron:SBO10.1590/S0034-72802012000100003info:eu-repo/semantics/openAccessAlonso,Ruiz SimonatoFontes,Bruno MachadoVentura,Marcelo PalisAmbrósio Jr,Renatoeng2012-03-12T00:00:00Zoai:scielo:S0034-72802012000100003Revistahttps://rbo.emnuvens.com.br/rbo/indexhttps://old.scielo.br/oai/scielo-oai.phpsob@sboportal.org.br||rbo@sboportal.org.br1982-85510034-7280opendoar:2012-03-12T00:00Revista Brasileira de Oftalmologia (Online) - Sociedade Brasileira de Oftalmologia (SBO)false
dc.title.none.fl_str_mv Repeatability of central corneal thickness measurement with the Pentacam HR system
title Repeatability of central corneal thickness measurement with the Pentacam HR system
spellingShingle Repeatability of central corneal thickness measurement with the Pentacam HR system
Alonso,Ruiz Simonato
Cornea/anatomy & histology
Corneal topography
Biometry/methods
Reproducibility of results
Diagnostic techniques
ophthalmological/instrumentation
title_short Repeatability of central corneal thickness measurement with the Pentacam HR system
title_full Repeatability of central corneal thickness measurement with the Pentacam HR system
title_fullStr Repeatability of central corneal thickness measurement with the Pentacam HR system
title_full_unstemmed Repeatability of central corneal thickness measurement with the Pentacam HR system
title_sort Repeatability of central corneal thickness measurement with the Pentacam HR system
author Alonso,Ruiz Simonato
author_facet Alonso,Ruiz Simonato
Fontes,Bruno Machado
Ventura,Marcelo Palis
Ambrósio Jr,Renato
author_role author
author2 Fontes,Bruno Machado
Ventura,Marcelo Palis
Ambrósio Jr,Renato
author2_role author
author
author
dc.contributor.author.fl_str_mv Alonso,Ruiz Simonato
Fontes,Bruno Machado
Ventura,Marcelo Palis
Ambrósio Jr,Renato
dc.subject.por.fl_str_mv Cornea/anatomy & histology
Corneal topography
Biometry/methods
Reproducibility of results
Diagnostic techniques
ophthalmological/instrumentation
topic Cornea/anatomy & histology
Corneal topography
Biometry/methods
Reproducibility of results
Diagnostic techniques
ophthalmological/instrumentation
description PURPOSE: To assess the repeatability of central corneal thickness measurement at the geometrical center (Central Corneal Thickness - CCT) given by the Pentacam High Resolution (HR) Comprehensive Eye Scanner (Oculus, Wetzlar, Germany) over time. METHODS: Prospective, single center, observational study. Two separate CCT measurements were taken by the Pentacam corneal tomography exam (CTm) 3 to 12 months apart, and compared. RESULTS: One hundred and sixteen eyes (n=116) of 62 health patients were included in this study. Average CCT in first and last visits was 541.6±37 µm and 543.6±36.9 µm respectively. Mean difference between both measurements was 9.2±6.4 µm, and there was no statistically significant difference in CCT measurement between visits, with good correlation between them (P = 0.057, r² = 0,9209). CONCLUSION: Pentacam (HR) CTm gives repeatable CCT measurements over time.
publishDate 2012
dc.date.none.fl_str_mv 2012-02-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0034-72802012000100003
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0034-72802012000100003
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/S0034-72802012000100003
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Sociedade Brasileira de Oftalmologia
publisher.none.fl_str_mv Sociedade Brasileira de Oftalmologia
dc.source.none.fl_str_mv Revista Brasileira de Oftalmologia v.71 n.1 2012
reponame:Revista Brasileira de Oftalmologia (Online)
instname:Sociedade Brasileira de Oftalmologia (SBO)
instacron:SBO
instname_str Sociedade Brasileira de Oftalmologia (SBO)
instacron_str SBO
institution SBO
reponame_str Revista Brasileira de Oftalmologia (Online)
collection Revista Brasileira de Oftalmologia (Online)
repository.name.fl_str_mv Revista Brasileira de Oftalmologia (Online) - Sociedade Brasileira de Oftalmologia (SBO)
repository.mail.fl_str_mv sob@sboportal.org.br||rbo@sboportal.org.br
_version_ 1752122336026296320