Validation of the mindray BC6000 hematology analyzer for erythroblast counting in peripheral blood

Detalhes bibliográficos
Autor(a) principal: Brito Junior,Lacy
Data de Publicação: 2022
Outros Autores: Caldas,Adna dos Santos, Pereira,Vanessa Ingrid Cardoso, Ferreira,Roberta Isabella Senna, Paixão,Ana Paula Silveira
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Jornal Brasileiro de Patologia e Medicina Laboratorial (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1676-24442022000100105
Resumo: ABSTRACT Introduction Differential counting of erythroblasts in blood samples by hematology analyzers still has limitations. Technological advances in blood count equipment have proposed the fully automated counting of these cells, however, not without its validation. Objective Validate the automated count of erythroblasts in peripheral blood in the Mindray BC6000 hematology analyzer and verify the existence of correlation of the maturation stages of erythroblasts with the equipment's graphics. Material and Methods Prospective study with peripheral blood samples from the Clinical Pathology Laboratory Dr Paulo C. Azevedo, regardless of age and gender, to validate the erythroblast count in the Mindray BC6000 hematology analyzer compared to the manual method (gold standard), in the period of June 2019 to December 2020. Results Seventeen peripheral blood samples were analyzed from newborns (09/17 - 52.3%) and from patients older than 2 years (08/17 - 47.7%) who had more than 18% of erythroblasts after morphological analysis of the sample. Statistical analysis of erythroblast counts by the two methodologies showed that the Mindray BC6000 hematological counter has good reproducibility, precision and linearity. There was no correlation between the maturation stages of erythroblasts and the equipment graphics. Conclusion The proposed validation showed that the Mindray BC6000 hematological counter has good analytical performance for counting erythroblasts in peripheral blood. However, there is no correlation between the maturation stages of erythroblasts with the graphics generated by the equipment.
id SBP-1_9e98f4860f321201b0fd5b62bbc66e51
oai_identifier_str oai:scielo:S1676-24442022000100105
network_acronym_str SBP-1
network_name_str Jornal Brasileiro de Patologia e Medicina Laboratorial (Online)
repository_id_str
spelling Validation of the mindray BC6000 hematology analyzer for erythroblast counting in peripheral bloodhematologyerythroblastslaboratory automationABSTRACT Introduction Differential counting of erythroblasts in blood samples by hematology analyzers still has limitations. Technological advances in blood count equipment have proposed the fully automated counting of these cells, however, not without its validation. Objective Validate the automated count of erythroblasts in peripheral blood in the Mindray BC6000 hematology analyzer and verify the existence of correlation of the maturation stages of erythroblasts with the equipment's graphics. Material and Methods Prospective study with peripheral blood samples from the Clinical Pathology Laboratory Dr Paulo C. Azevedo, regardless of age and gender, to validate the erythroblast count in the Mindray BC6000 hematology analyzer compared to the manual method (gold standard), in the period of June 2019 to December 2020. Results Seventeen peripheral blood samples were analyzed from newborns (09/17 - 52.3%) and from patients older than 2 years (08/17 - 47.7%) who had more than 18% of erythroblasts after morphological analysis of the sample. Statistical analysis of erythroblast counts by the two methodologies showed that the Mindray BC6000 hematological counter has good reproducibility, precision and linearity. There was no correlation between the maturation stages of erythroblasts and the equipment graphics. Conclusion The proposed validation showed that the Mindray BC6000 hematological counter has good analytical performance for counting erythroblasts in peripheral blood. However, there is no correlation between the maturation stages of erythroblasts with the graphics generated by the equipment.Sociedade Brasileira de Patologia Clínica2022-01-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S1676-24442022000100105Jornal Brasileiro de Patologia e Medicina Laboratorial v.58 2022reponame:Jornal Brasileiro de Patologia e Medicina Laboratorial (Online)instname:Sociedade Brasileira de Patologia (SBP)instacron:SBP10.1900/jbpml.2022.58.442info:eu-repo/semantics/openAccessBrito Junior,LacyCaldas,Adna dos SantosPereira,Vanessa Ingrid CardosoFerreira,Roberta Isabella SennaPaixão,Ana Paula Silveiraeng2022-05-26T00:00:00Zoai:scielo:S1676-24442022000100105Revistahttp://www.scielo.br/jbpmlhttps://old.scielo.br/oai/scielo-oai.php||jbpml@sbpc.org.br1678-47741676-2444opendoar:2022-05-26T00:00Jornal Brasileiro de Patologia e Medicina Laboratorial (Online) - Sociedade Brasileira de Patologia (SBP)false
dc.title.none.fl_str_mv Validation of the mindray BC6000 hematology analyzer for erythroblast counting in peripheral blood
title Validation of the mindray BC6000 hematology analyzer for erythroblast counting in peripheral blood
spellingShingle Validation of the mindray BC6000 hematology analyzer for erythroblast counting in peripheral blood
Brito Junior,Lacy
hematology
erythroblasts
laboratory automation
title_short Validation of the mindray BC6000 hematology analyzer for erythroblast counting in peripheral blood
title_full Validation of the mindray BC6000 hematology analyzer for erythroblast counting in peripheral blood
title_fullStr Validation of the mindray BC6000 hematology analyzer for erythroblast counting in peripheral blood
title_full_unstemmed Validation of the mindray BC6000 hematology analyzer for erythroblast counting in peripheral blood
title_sort Validation of the mindray BC6000 hematology analyzer for erythroblast counting in peripheral blood
author Brito Junior,Lacy
author_facet Brito Junior,Lacy
Caldas,Adna dos Santos
Pereira,Vanessa Ingrid Cardoso
Ferreira,Roberta Isabella Senna
Paixão,Ana Paula Silveira
author_role author
author2 Caldas,Adna dos Santos
Pereira,Vanessa Ingrid Cardoso
Ferreira,Roberta Isabella Senna
Paixão,Ana Paula Silveira
author2_role author
author
author
author
dc.contributor.author.fl_str_mv Brito Junior,Lacy
Caldas,Adna dos Santos
Pereira,Vanessa Ingrid Cardoso
Ferreira,Roberta Isabella Senna
Paixão,Ana Paula Silveira
dc.subject.por.fl_str_mv hematology
erythroblasts
laboratory automation
topic hematology
erythroblasts
laboratory automation
description ABSTRACT Introduction Differential counting of erythroblasts in blood samples by hematology analyzers still has limitations. Technological advances in blood count equipment have proposed the fully automated counting of these cells, however, not without its validation. Objective Validate the automated count of erythroblasts in peripheral blood in the Mindray BC6000 hematology analyzer and verify the existence of correlation of the maturation stages of erythroblasts with the equipment's graphics. Material and Methods Prospective study with peripheral blood samples from the Clinical Pathology Laboratory Dr Paulo C. Azevedo, regardless of age and gender, to validate the erythroblast count in the Mindray BC6000 hematology analyzer compared to the manual method (gold standard), in the period of June 2019 to December 2020. Results Seventeen peripheral blood samples were analyzed from newborns (09/17 - 52.3%) and from patients older than 2 years (08/17 - 47.7%) who had more than 18% of erythroblasts after morphological analysis of the sample. Statistical analysis of erythroblast counts by the two methodologies showed that the Mindray BC6000 hematological counter has good reproducibility, precision and linearity. There was no correlation between the maturation stages of erythroblasts and the equipment graphics. Conclusion The proposed validation showed that the Mindray BC6000 hematological counter has good analytical performance for counting erythroblasts in peripheral blood. However, there is no correlation between the maturation stages of erythroblasts with the graphics generated by the equipment.
publishDate 2022
dc.date.none.fl_str_mv 2022-01-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1676-24442022000100105
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1676-24442022000100105
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1900/jbpml.2022.58.442
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv
Sociedade Brasileira de Patologia Clínica
publisher.none.fl_str_mv
Sociedade Brasileira de Patologia Clínica
dc.source.none.fl_str_mv Jornal Brasileiro de Patologia e Medicina Laboratorial v.58 2022
reponame:Jornal Brasileiro de Patologia e Medicina Laboratorial (Online)
instname:Sociedade Brasileira de Patologia (SBP)
instacron:SBP
instname_str Sociedade Brasileira de Patologia (SBP)
instacron_str SBP
institution SBP
reponame_str Jornal Brasileiro de Patologia e Medicina Laboratorial (Online)
collection Jornal Brasileiro de Patologia e Medicina Laboratorial (Online)
repository.name.fl_str_mv Jornal Brasileiro de Patologia e Medicina Laboratorial (Online) - Sociedade Brasileira de Patologia (SBP)
repository.mail.fl_str_mv ||jbpml@sbpc.org.br
_version_ 1752122297910558720