Performance analysis of commercial rapid tests for serum troponin detection

Detalhes bibliográficos
Autor(a) principal: Ferreira,Raquel Ida
Data de Publicação: 2020
Outros Autores: Amaral,Marcela R, Brito,Maytê S. R, Pinto,Matheus P, Maranhão,Ana Carolina R, Costa,Lucas R. R, Melo,Rômulo C. Vaz de
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Jornal Brasileiro de Patologia e Medicina Laboratorial (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1676-24442020000100414
Resumo: ABSTRACT Introduction: Rapid tests represent an important diagnostic tool, providing results in a short period of time and eliminating the use of large automated equipment. Objective: To evaluate the performance of five rapid test kits for troponin I detection in serum. Materials and methods: Samples from 100 patients with suspected acute myocardial infarction (AMI) were selected from a hospital in Barbacena, MG, Brazil. They were tested in the five troponin I commercial rapid test kits and the results were compared with the study reference method (quantitative chemiluminescent immunoassay). Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV) and correlation coefficients of the rapid tests against the reference method were defined. Results: The kits from Abon®, Interteck®, Medtest® and Wama® presented the same performance, reaching equal levels of sensitivity (98.08%), specificity (100%), PPV (100%), NPV (97.96%), and correlation coefficient (0.99). The Eco Diagnostica® kit presented lower sensitivity (82.69%), low correlation coefficient (0.91) and NPV (84.21%), not reaching ideal levels even though lying within the confidence interval. Discussion: The results obtained with the Abon®, Interteck®, Medtest® and Wama® test kits corroborate pre-existing literature data on the diagnostic accuracy of rapid tests for troponin detection. These tests are allowed for immediate decision making by physicians and can be used to reduce unnecessary hospitalization time and costs associated with a suspected AMI. Conclusion: The kits from Abon®, Interteck®, Medtest® and Wama® showed excellent diagnostic performance, what makes them an important diagnostic tool in smaller laboratories without automated equipment.
id SBP-1_a91f76e227d408fc78529fb644ccfedc
oai_identifier_str oai:scielo:S1676-24442020000100414
network_acronym_str SBP-1
network_name_str Jornal Brasileiro de Patologia e Medicina Laboratorial (Online)
repository_id_str
spelling Performance analysis of commercial rapid tests for serum troponin detectionmyocardial infarctiontroponin Idiagnostic techniques, cardiovascularABSTRACT Introduction: Rapid tests represent an important diagnostic tool, providing results in a short period of time and eliminating the use of large automated equipment. Objective: To evaluate the performance of five rapid test kits for troponin I detection in serum. Materials and methods: Samples from 100 patients with suspected acute myocardial infarction (AMI) were selected from a hospital in Barbacena, MG, Brazil. They were tested in the five troponin I commercial rapid test kits and the results were compared with the study reference method (quantitative chemiluminescent immunoassay). Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV) and correlation coefficients of the rapid tests against the reference method were defined. Results: The kits from Abon®, Interteck®, Medtest® and Wama® presented the same performance, reaching equal levels of sensitivity (98.08%), specificity (100%), PPV (100%), NPV (97.96%), and correlation coefficient (0.99). The Eco Diagnostica® kit presented lower sensitivity (82.69%), low correlation coefficient (0.91) and NPV (84.21%), not reaching ideal levels even though lying within the confidence interval. Discussion: The results obtained with the Abon®, Interteck®, Medtest® and Wama® test kits corroborate pre-existing literature data on the diagnostic accuracy of rapid tests for troponin detection. These tests are allowed for immediate decision making by physicians and can be used to reduce unnecessary hospitalization time and costs associated with a suspected AMI. Conclusion: The kits from Abon®, Interteck®, Medtest® and Wama® showed excellent diagnostic performance, what makes them an important diagnostic tool in smaller laboratories without automated equipment.Sociedade Brasileira de Patologia Clínica2020-01-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S1676-24442020000100414Jornal Brasileiro de Patologia e Medicina Laboratorial v.56 2020reponame:Jornal Brasileiro de Patologia e Medicina Laboratorial (Online)instname:Sociedade Brasileira de Patologia (SBP)instacron:SBP10.5935/1676-2444.20200020info:eu-repo/semantics/openAccessFerreira,Raquel IdaAmaral,Marcela RBrito,Maytê S. RPinto,Matheus PMaranhão,Ana Carolina RCosta,Lucas R. RMelo,Rômulo C. Vaz deeng2020-05-06T00:00:00Zoai:scielo:S1676-24442020000100414Revistahttp://www.scielo.br/jbpmlhttps://old.scielo.br/oai/scielo-oai.php||jbpml@sbpc.org.br1678-47741676-2444opendoar:2020-05-06T00:00Jornal Brasileiro de Patologia e Medicina Laboratorial (Online) - Sociedade Brasileira de Patologia (SBP)false
dc.title.none.fl_str_mv Performance analysis of commercial rapid tests for serum troponin detection
title Performance analysis of commercial rapid tests for serum troponin detection
spellingShingle Performance analysis of commercial rapid tests for serum troponin detection
Ferreira,Raquel Ida
myocardial infarction
troponin I
diagnostic techniques, cardiovascular
title_short Performance analysis of commercial rapid tests for serum troponin detection
title_full Performance analysis of commercial rapid tests for serum troponin detection
title_fullStr Performance analysis of commercial rapid tests for serum troponin detection
title_full_unstemmed Performance analysis of commercial rapid tests for serum troponin detection
title_sort Performance analysis of commercial rapid tests for serum troponin detection
author Ferreira,Raquel Ida
author_facet Ferreira,Raquel Ida
Amaral,Marcela R
Brito,Maytê S. R
Pinto,Matheus P
Maranhão,Ana Carolina R
Costa,Lucas R. R
Melo,Rômulo C. Vaz de
author_role author
author2 Amaral,Marcela R
Brito,Maytê S. R
Pinto,Matheus P
Maranhão,Ana Carolina R
Costa,Lucas R. R
Melo,Rômulo C. Vaz de
author2_role author
author
author
author
author
author
dc.contributor.author.fl_str_mv Ferreira,Raquel Ida
Amaral,Marcela R
Brito,Maytê S. R
Pinto,Matheus P
Maranhão,Ana Carolina R
Costa,Lucas R. R
Melo,Rômulo C. Vaz de
dc.subject.por.fl_str_mv myocardial infarction
troponin I
diagnostic techniques, cardiovascular
topic myocardial infarction
troponin I
diagnostic techniques, cardiovascular
description ABSTRACT Introduction: Rapid tests represent an important diagnostic tool, providing results in a short period of time and eliminating the use of large automated equipment. Objective: To evaluate the performance of five rapid test kits for troponin I detection in serum. Materials and methods: Samples from 100 patients with suspected acute myocardial infarction (AMI) were selected from a hospital in Barbacena, MG, Brazil. They were tested in the five troponin I commercial rapid test kits and the results were compared with the study reference method (quantitative chemiluminescent immunoassay). Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV) and correlation coefficients of the rapid tests against the reference method were defined. Results: The kits from Abon®, Interteck®, Medtest® and Wama® presented the same performance, reaching equal levels of sensitivity (98.08%), specificity (100%), PPV (100%), NPV (97.96%), and correlation coefficient (0.99). The Eco Diagnostica® kit presented lower sensitivity (82.69%), low correlation coefficient (0.91) and NPV (84.21%), not reaching ideal levels even though lying within the confidence interval. Discussion: The results obtained with the Abon®, Interteck®, Medtest® and Wama® test kits corroborate pre-existing literature data on the diagnostic accuracy of rapid tests for troponin detection. These tests are allowed for immediate decision making by physicians and can be used to reduce unnecessary hospitalization time and costs associated with a suspected AMI. Conclusion: The kits from Abon®, Interteck®, Medtest® and Wama® showed excellent diagnostic performance, what makes them an important diagnostic tool in smaller laboratories without automated equipment.
publishDate 2020
dc.date.none.fl_str_mv 2020-01-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1676-24442020000100414
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1676-24442020000100414
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.5935/1676-2444.20200020
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv
Sociedade Brasileira de Patologia Clínica
publisher.none.fl_str_mv
Sociedade Brasileira de Patologia Clínica
dc.source.none.fl_str_mv Jornal Brasileiro de Patologia e Medicina Laboratorial v.56 2020
reponame:Jornal Brasileiro de Patologia e Medicina Laboratorial (Online)
instname:Sociedade Brasileira de Patologia (SBP)
instacron:SBP
instname_str Sociedade Brasileira de Patologia (SBP)
instacron_str SBP
institution SBP
reponame_str Jornal Brasileiro de Patologia e Medicina Laboratorial (Online)
collection Jornal Brasileiro de Patologia e Medicina Laboratorial (Online)
repository.name.fl_str_mv Jornal Brasileiro de Patologia e Medicina Laboratorial (Online) - Sociedade Brasileira de Patologia (SBP)
repository.mail.fl_str_mv ||jbpml@sbpc.org.br
_version_ 1752122297611714560