Contributions on the Use of Bismuth as Internal Standard for Lead Determinations Using ICP-Based Techniques

Detalhes bibliográficos
Autor(a) principal: Bechlin,Marcos A.
Data de Publicação: 2015
Outros Autores: Ferreira,Edilene C., Gomes Neto,José A., Ramos,Juliano C., Borges,Daniel L. G.
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Journal of the Brazilian Chemical Society (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-50532015000901879
Resumo: Bismuth was evaluated as internal standard for Pb determinations and showed usefulness to minimize variations due to changes in the experimental conditions of the inductively coupled plasma optical emission spectrometry (ICP OES), inductively coupled plasma mass spectrometry (ICP-MS) and electrothermal vaporization inductively coupled plasma mass spectrometry (ETV-ICP-MS). Thirty four samples were analyzed and the Pb contents obtained by ICP-MS and ETV-ICP-MS were in agreement at 95% confidence level. Plant, biological tissue and soil reference materials were also analyzed and the results were close to certified values upon the use of the internal standard. Recoveries for Pb spikes (ICP OES: 95-110%; ICP-MS: 95-106%; ETV-ICP-MS: 92-119%) were better than those obtained with external calibration (ICP OES: 56-91%; ICP-MS: 36-136%; ETV‑ICP-MS: 5-123%). The relative standard deviations (n = 12) was improved from 1-9% to 0.2-6% (ICP OES), from 0.5-11% to 0.1-7% (ICP-MS) and from 0.4-7% to 0.1-4% (ETV‑ICP‑MS) by using external and internal standard calibrations, respectively.
id SBQ-2_6612193d9da2008aa0d4fdb82cb5f2c8
oai_identifier_str oai:scielo:S0103-50532015000901879
network_acronym_str SBQ-2
network_name_str Journal of the Brazilian Chemical Society (Online)
repository_id_str
spelling Contributions on the Use of Bismuth as Internal Standard for Lead Determinations Using ICP-Based Techniqueslead determinationBi internal standardICP OESICP-MSETV-ICP-MSBismuth was evaluated as internal standard for Pb determinations and showed usefulness to minimize variations due to changes in the experimental conditions of the inductively coupled plasma optical emission spectrometry (ICP OES), inductively coupled plasma mass spectrometry (ICP-MS) and electrothermal vaporization inductively coupled plasma mass spectrometry (ETV-ICP-MS). Thirty four samples were analyzed and the Pb contents obtained by ICP-MS and ETV-ICP-MS were in agreement at 95% confidence level. Plant, biological tissue and soil reference materials were also analyzed and the results were close to certified values upon the use of the internal standard. Recoveries for Pb spikes (ICP OES: 95-110%; ICP-MS: 95-106%; ETV-ICP-MS: 92-119%) were better than those obtained with external calibration (ICP OES: 56-91%; ICP-MS: 36-136%; ETV‑ICP-MS: 5-123%). The relative standard deviations (n = 12) was improved from 1-9% to 0.2-6% (ICP OES), from 0.5-11% to 0.1-7% (ICP-MS) and from 0.4-7% to 0.1-4% (ETV‑ICP‑MS) by using external and internal standard calibrations, respectively.Sociedade Brasileira de Química2015-09-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-50532015000901879Journal of the Brazilian Chemical Society v.26 n.9 2015reponame:Journal of the Brazilian Chemical Society (Online)instname:Sociedade Brasileira de Química (SBQ)instacron:SBQ10.5935/0103-5053.20150165info:eu-repo/semantics/openAccessBechlin,Marcos A.Ferreira,Edilene C.Gomes Neto,José A.Ramos,Juliano C.Borges,Daniel L. G.eng2015-09-11T00:00:00Zoai:scielo:S0103-50532015000901879Revistahttp://jbcs.sbq.org.brONGhttps://old.scielo.br/oai/scielo-oai.php||office@jbcs.sbq.org.br1678-47900103-5053opendoar:2015-09-11T00:00Journal of the Brazilian Chemical Society (Online) - Sociedade Brasileira de Química (SBQ)false
dc.title.none.fl_str_mv Contributions on the Use of Bismuth as Internal Standard for Lead Determinations Using ICP-Based Techniques
title Contributions on the Use of Bismuth as Internal Standard for Lead Determinations Using ICP-Based Techniques
spellingShingle Contributions on the Use of Bismuth as Internal Standard for Lead Determinations Using ICP-Based Techniques
Bechlin,Marcos A.
lead determination
Bi internal standard
ICP OES
ICP-MS
ETV-ICP-MS
title_short Contributions on the Use of Bismuth as Internal Standard for Lead Determinations Using ICP-Based Techniques
title_full Contributions on the Use of Bismuth as Internal Standard for Lead Determinations Using ICP-Based Techniques
title_fullStr Contributions on the Use of Bismuth as Internal Standard for Lead Determinations Using ICP-Based Techniques
title_full_unstemmed Contributions on the Use of Bismuth as Internal Standard for Lead Determinations Using ICP-Based Techniques
title_sort Contributions on the Use of Bismuth as Internal Standard for Lead Determinations Using ICP-Based Techniques
author Bechlin,Marcos A.
author_facet Bechlin,Marcos A.
Ferreira,Edilene C.
Gomes Neto,José A.
Ramos,Juliano C.
Borges,Daniel L. G.
author_role author
author2 Ferreira,Edilene C.
Gomes Neto,José A.
Ramos,Juliano C.
Borges,Daniel L. G.
author2_role author
author
author
author
dc.contributor.author.fl_str_mv Bechlin,Marcos A.
Ferreira,Edilene C.
Gomes Neto,José A.
Ramos,Juliano C.
Borges,Daniel L. G.
dc.subject.por.fl_str_mv lead determination
Bi internal standard
ICP OES
ICP-MS
ETV-ICP-MS
topic lead determination
Bi internal standard
ICP OES
ICP-MS
ETV-ICP-MS
description Bismuth was evaluated as internal standard for Pb determinations and showed usefulness to minimize variations due to changes in the experimental conditions of the inductively coupled plasma optical emission spectrometry (ICP OES), inductively coupled plasma mass spectrometry (ICP-MS) and electrothermal vaporization inductively coupled plasma mass spectrometry (ETV-ICP-MS). Thirty four samples were analyzed and the Pb contents obtained by ICP-MS and ETV-ICP-MS were in agreement at 95% confidence level. Plant, biological tissue and soil reference materials were also analyzed and the results were close to certified values upon the use of the internal standard. Recoveries for Pb spikes (ICP OES: 95-110%; ICP-MS: 95-106%; ETV-ICP-MS: 92-119%) were better than those obtained with external calibration (ICP OES: 56-91%; ICP-MS: 36-136%; ETV‑ICP-MS: 5-123%). The relative standard deviations (n = 12) was improved from 1-9% to 0.2-6% (ICP OES), from 0.5-11% to 0.1-7% (ICP-MS) and from 0.4-7% to 0.1-4% (ETV‑ICP‑MS) by using external and internal standard calibrations, respectively.
publishDate 2015
dc.date.none.fl_str_mv 2015-09-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-50532015000901879
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0103-50532015000901879
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.5935/0103-5053.20150165
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Sociedade Brasileira de Química
publisher.none.fl_str_mv Sociedade Brasileira de Química
dc.source.none.fl_str_mv Journal of the Brazilian Chemical Society v.26 n.9 2015
reponame:Journal of the Brazilian Chemical Society (Online)
instname:Sociedade Brasileira de Química (SBQ)
instacron:SBQ
instname_str Sociedade Brasileira de Química (SBQ)
instacron_str SBQ
institution SBQ
reponame_str Journal of the Brazilian Chemical Society (Online)
collection Journal of the Brazilian Chemical Society (Online)
repository.name.fl_str_mv Journal of the Brazilian Chemical Society (Online) - Sociedade Brasileira de Química (SBQ)
repository.mail.fl_str_mv ||office@jbcs.sbq.org.br
_version_ 1750318177486635008