Development and validation of a High Performance Liquid Chromatographic method for determination of etoposide in biodegradable polymeric implants

Detalhes bibliográficos
Autor(a) principal: Solano,Ana Gabriela Reis
Data de Publicação: 2012
Outros Autores: Silva,Gisele Rodrigues da, Fialho,Silvia Ligório, Cunha Júnior,Armando da Silva, Pianetti,Gérson Antônio
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Química Nova (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422012000600032
Resumo: A method using HPLC-UV was developed and validated for the determination of etoposide incorporated into polycaprolactone implants. The method was carried out in isocratic mode using a C18 column (250 x 4.6 mm; 5 µm), at 25 ºC, with acetonitrile and acetic acid 4% (70:30) as mobile phase, a flow rate of 2 mL/min, and UV detection at 285 nm. The method was linear (r² &gt; 0.99) over the range of 5 to 65 µg/mL, precise (RSD < 5%), accurate (recovery of 98.7%), robust, selective regarding excipient of the sample, and had a quantitation limit equal to 1.76 µg/mL. The validated method can be successfully employed for routine quality control analyses.
id SBQ-3_3879948892be0fa9f7bdf37f3dd9f933
oai_identifier_str oai:scielo:S0100-40422012000600032
network_acronym_str SBQ-3
network_name_str Química Nova (Online)
repository_id_str
spelling Development and validation of a High Performance Liquid Chromatographic method for determination of etoposide in biodegradable polymeric implantsvalidationetoposidepoly(ε-caprolactone)intraocular implantA method using HPLC-UV was developed and validated for the determination of etoposide incorporated into polycaprolactone implants. The method was carried out in isocratic mode using a C18 column (250 x 4.6 mm; 5 µm), at 25 ºC, with acetonitrile and acetic acid 4% (70:30) as mobile phase, a flow rate of 2 mL/min, and UV detection at 285 nm. The method was linear (r² &gt; 0.99) over the range of 5 to 65 µg/mL, precise (RSD < 5%), accurate (recovery of 98.7%), robust, selective regarding excipient of the sample, and had a quantitation limit equal to 1.76 µg/mL. The validated method can be successfully employed for routine quality control analyses.Sociedade Brasileira de Química2012-01-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422012000600032Química Nova v.35 n.6 2012reponame:Química Nova (Online)instname:Sociedade Brasileira de Química (SBQ)instacron:SBQ10.1590/S0100-40422012000600032info:eu-repo/semantics/openAccessSolano,Ana Gabriela ReisSilva,Gisele Rodrigues daFialho,Silvia LigórioCunha Júnior,Armando da SilvaPianetti,Gérson Antônioeng2012-07-25T00:00:00Zoai:scielo:S0100-40422012000600032Revistahttps://www.scielo.br/j/qn/ONGhttps://old.scielo.br/oai/scielo-oai.phpquimicanova@sbq.org.br1678-70640100-4042opendoar:2012-07-25T00:00Química Nova (Online) - Sociedade Brasileira de Química (SBQ)false
dc.title.none.fl_str_mv Development and validation of a High Performance Liquid Chromatographic method for determination of etoposide in biodegradable polymeric implants
title Development and validation of a High Performance Liquid Chromatographic method for determination of etoposide in biodegradable polymeric implants
spellingShingle Development and validation of a High Performance Liquid Chromatographic method for determination of etoposide in biodegradable polymeric implants
Solano,Ana Gabriela Reis
validation
etoposide
poly(ε-caprolactone)
intraocular implant
title_short Development and validation of a High Performance Liquid Chromatographic method for determination of etoposide in biodegradable polymeric implants
title_full Development and validation of a High Performance Liquid Chromatographic method for determination of etoposide in biodegradable polymeric implants
title_fullStr Development and validation of a High Performance Liquid Chromatographic method for determination of etoposide in biodegradable polymeric implants
title_full_unstemmed Development and validation of a High Performance Liquid Chromatographic method for determination of etoposide in biodegradable polymeric implants
title_sort Development and validation of a High Performance Liquid Chromatographic method for determination of etoposide in biodegradable polymeric implants
author Solano,Ana Gabriela Reis
author_facet Solano,Ana Gabriela Reis
Silva,Gisele Rodrigues da
Fialho,Silvia Ligório
Cunha Júnior,Armando da Silva
Pianetti,Gérson Antônio
author_role author
author2 Silva,Gisele Rodrigues da
Fialho,Silvia Ligório
Cunha Júnior,Armando da Silva
Pianetti,Gérson Antônio
author2_role author
author
author
author
dc.contributor.author.fl_str_mv Solano,Ana Gabriela Reis
Silva,Gisele Rodrigues da
Fialho,Silvia Ligório
Cunha Júnior,Armando da Silva
Pianetti,Gérson Antônio
dc.subject.por.fl_str_mv validation
etoposide
poly(ε-caprolactone)
intraocular implant
topic validation
etoposide
poly(ε-caprolactone)
intraocular implant
description A method using HPLC-UV was developed and validated for the determination of etoposide incorporated into polycaprolactone implants. The method was carried out in isocratic mode using a C18 column (250 x 4.6 mm; 5 µm), at 25 ºC, with acetonitrile and acetic acid 4% (70:30) as mobile phase, a flow rate of 2 mL/min, and UV detection at 285 nm. The method was linear (r² &gt; 0.99) over the range of 5 to 65 µg/mL, precise (RSD < 5%), accurate (recovery of 98.7%), robust, selective regarding excipient of the sample, and had a quantitation limit equal to 1.76 µg/mL. The validated method can be successfully employed for routine quality control analyses.
publishDate 2012
dc.date.none.fl_str_mv 2012-01-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422012000600032
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422012000600032
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/S0100-40422012000600032
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Sociedade Brasileira de Química
publisher.none.fl_str_mv Sociedade Brasileira de Química
dc.source.none.fl_str_mv Química Nova v.35 n.6 2012
reponame:Química Nova (Online)
instname:Sociedade Brasileira de Química (SBQ)
instacron:SBQ
instname_str Sociedade Brasileira de Química (SBQ)
instacron_str SBQ
institution SBQ
reponame_str Química Nova (Online)
collection Química Nova (Online)
repository.name.fl_str_mv Química Nova (Online) - Sociedade Brasileira de Química (SBQ)
repository.mail.fl_str_mv quimicanova@sbq.org.br
_version_ 1750318113570684928