EFEITOS DA LUZ DE UM FOTOPOLIMERIZADOR LED NA TEMPERATURA PULPAR BASAL IN VIVO EM PRÉ-MOLARES HUMANOS E SUA CORRELAÇÃO COM AS CARACTERÍSTICAS ANATÔMICAS DENTAIS EX VIVO

Detalhes bibliográficos
Autor(a) principal: Runnacles, Patricio
Data de Publicação: 2014
Tipo de documento: Tese
Idioma: por
Título da fonte: Biblioteca Digital de Teses e Dissertações da UEPG
Texto Completo: http://tede2.uepg.br/jspui/handle/prefix/1696
Resumo: Objective: This in vivo study evaluated the baseline pulp temperature in human intact upper premolars and the pulp temperature rise induced by light emitted by a high power polywave Light Emitting Diode (LED) light curing unit (LCU); also, the possible correlation between the change in temperature (ΔT) and the volume of the crown and the pulp chamber ex vivo. Material and methods: after approval of the local Ethics Committee was obtained (protocol #255,945), upper right and left first premolars requiring extraction for orthodontic reasons from 8 volunteers (n=15) received infiltrative and intraligamental anesthesia and were isolated using rubber dam. A small, occlusal preparation was made using high-speed hand piece, under constant air-water spray, until a minute pulp exposure was attained. The sterile probe from a wireless, NIST-traceable, temperature acquisition system (Temperature Data Acquisition - Thermes WFI, Physitemp) was inserted directly into the coronal pulp. Once the baseline pulp temperature was stable, the buccal surface was exposed to polywave light from a LED LCU (Bluephase 20i, Ivoclar Vivadent) using selected EMs: 10-s either in low (10-s/L) or High (10-s/H); 5-s-Turbo (5-s/T); and 60-s-High (60-s/H) intensities, allowing a 7-min span between each exposure. Peak PT values and PT increases from baseline (ΔT) after exposure were subjected to 1-Way, repeated measures ANOVAs, and Bonferroni's post-hoc tests (α=0.05). Linear regression analysis was performed to establish the relationship between applied radiant exposure and ΔT. Linear regression analysis was performed to establish the relationship between applied radiant exposure and ΔT. the teeth were extracted according to treatment plan. The crown and pulp volumes were measured using micro-tomography scanner, and pulp chamber / crown volume ratio was determined. Afterwards, premolar buccal-lingual, cervico-occlusal, mesio-distal dimensions as well as between cusp distances were measured using a digital caliper. Results: consistent temperature below body temperature, resulting in an average core temperature of 35 ° C (+ 0.7 ° C) was found. All modes of activation of the LED promoted significant temperature rise in the pulp chamber. A positive correlation between energy density and thermal variation was observed. No correlation was observed between coronal volume, the pulp chamber with temperature changes. A significant negative correlation was noted between pulp temperature changes when 10-s/L was used and VL.
id UEPG_2cf1ed6be9f0bc145f225aab1925fb12
oai_identifier_str oai:tede2.uepg.br:prefix/1696
network_acronym_str UEPG
network_name_str Biblioteca Digital de Teses e Dissertações da UEPG
repository_id_str
spelling Gomes, Osnara Maria MongruelCPF:47308257991http://buscatextual.cnpq.br/buscatextual/visualizacv.do?id=K4773647J1Arrais, Cesar Augusto GalvãoCPF:20064607810http://buscatextual.cnpq.br/buscatextual/visualizacv.do?id=K4770689J6Reis, AlessandraCPF:25713648800REIS, A.Campanha, Nara HellenCPF:11993069836http://buscatextual.cnpq.br/buscatextual/visualizacv.do?id=K4705349Z2Pupo, Yasmine MendesCPF:04694480910http://buscatextual.cnpq.br/buscatextual/visualizacv.do?id=K4734244Z4Gonzaga, Carla CastigliaCPF:27051979825http://buscatextual.cnpq.br/buscatextual/visualizacv.do?id=K4735565H6CPF:00411125982http://buscatextual.cnpq.br/buscatextual/visualizacv.do?id=K4359313A4Runnacles, Patricio2017-07-24T19:21:57Z2015-07-152017-07-24T19:21:57Z2014-12-17RUNNACLES, Patricio. Effects of ligth emitted by a LED light curing unit on the in vivo baseline pulp temperature of intact human premolars and their correlation within ex vivo dental anatomic caracteristics. 2014. 102 f. Tese (Doutorado em Clinica Integrada, Dentística Restauradora e Periodontia) - UNIVERSIDADE ESTADUAL DE PONTA GROSSA, Ponta Grossa, 2014.http://tede2.uepg.br/jspui/handle/prefix/1696Objective: This in vivo study evaluated the baseline pulp temperature in human intact upper premolars and the pulp temperature rise induced by light emitted by a high power polywave Light Emitting Diode (LED) light curing unit (LCU); also, the possible correlation between the change in temperature (ΔT) and the volume of the crown and the pulp chamber ex vivo. Material and methods: after approval of the local Ethics Committee was obtained (protocol #255,945), upper right and left first premolars requiring extraction for orthodontic reasons from 8 volunteers (n=15) received infiltrative and intraligamental anesthesia and were isolated using rubber dam. A small, occlusal preparation was made using high-speed hand piece, under constant air-water spray, until a minute pulp exposure was attained. The sterile probe from a wireless, NIST-traceable, temperature acquisition system (Temperature Data Acquisition - Thermes WFI, Physitemp) was inserted directly into the coronal pulp. Once the baseline pulp temperature was stable, the buccal surface was exposed to polywave light from a LED LCU (Bluephase 20i, Ivoclar Vivadent) using selected EMs: 10-s either in low (10-s/L) or High (10-s/H); 5-s-Turbo (5-s/T); and 60-s-High (60-s/H) intensities, allowing a 7-min span between each exposure. Peak PT values and PT increases from baseline (ΔT) after exposure were subjected to 1-Way, repeated measures ANOVAs, and Bonferroni's post-hoc tests (α=0.05). Linear regression analysis was performed to establish the relationship between applied radiant exposure and ΔT. Linear regression analysis was performed to establish the relationship between applied radiant exposure and ΔT. the teeth were extracted according to treatment plan. The crown and pulp volumes were measured using micro-tomography scanner, and pulp chamber / crown volume ratio was determined. Afterwards, premolar buccal-lingual, cervico-occlusal, mesio-distal dimensions as well as between cusp distances were measured using a digital caliper. Results: consistent temperature below body temperature, resulting in an average core temperature of 35 ° C (+ 0.7 ° C) was found. All modes of activation of the LED promoted significant temperature rise in the pulp chamber. A positive correlation between energy density and thermal variation was observed. No correlation was observed between coronal volume, the pulp chamber with temperature changes. A significant negative correlation was noted between pulp temperature changes when 10-s/L was used and VL.Objetivo: Este estudo in vivo avaliou a temperatura basal da polpa em pré-molares hígidos superiores em humanos e o aumento da temperatura induzida pela luz emitida por um fotopolimerizador à base de Diodo Emissor de Luz (LED) polywave de alta potência; além disso, a possível correlação entre a variação de temperatura (ΔT) e o volume da coroa e da câmara pulpar ex vivo. Material e métodos: após aprovação da Comissão de Ética da Universidade Estadual de Ponta Grossa (UEPG) (protocolo #255,945), pré-molares direito e esquerdo de 8 pacientes voluntários (n=15), que necessitavam de extrações por motivos ortodônticos, receberam anestesia infiltrativa e intraligamentar e isolamento absoluto. Um pequeno acesso oclusal com exposição pulpar foi realizado usando alta-rotação, sob spray de ar e água constante, até atingir uma mínima exposição da polpa. Inseriu-se uma sonda estéril de um termopar wireless (Temperature Data Acquisition - Thermes WFI, Physitemp) diretamente no interior da câmara pulpar. Após a estabilização da temperatura basal da polpa, a face vestibular foi exposta à luz de um fotopolimerizador LED polywave (Bluephase 20i, Ivoclar Vivadent), usando modos de exposição selecionados: 10s em baixa potência (10s/L), 10s em alta potência (10s/H), 5s em potência turbo (5s/T) e 60s em alta potência (60s/H), respeitando um intervalo de 7 min entre cada exposição. A temperatura pulpar foi registrada a cada 0,2 s. Os valores absolutos e a variação da temperatura pulpar após exposição foram submetidos a análise estatística fator único (ANOVA) com medidas repetidas seguida pelo teste de Bonferroni (α=0,05). A correlação entre as densidades de energia aplicadas e os valores das variações de aumento de temperatura foi estabelecida por uma análise de regressão linear. Após as exodontias realizadas de acordo com os planos de tratamento, os elementos dentários (n=11) foram armazenados em Timol 0,1% e as dimensões mésio-distal da face oclusal (MD), cérvico-oclusal (CO), e vestíbulo-lingual (VL) da coroa dos dentes foram medidas com um paquímetro digital. Os volumes coronário e da câmara pulpar foram determinados por meio de microtomografia computadorizada (μCT) dos espécimes. Resultados: foi encontrada uma temperatura basal média consistente abaixo da temperatura corporal, de 35°C (+ 0,7°C). Todos os modos de ativação do LED promoveram aumento significativo de temperatura na câmara pulpar. Foi observada uma correlação positiva entre densidade de energia e variação térmica. Não houve correlação dos volumes coronal e da câmara pulpar com as variações térmicas, mas foi encontrada uma correlação negativa entre a medida VL das coroas e o ΔT no modo de exposição 10s/L.Made available in DSpace on 2017-07-24T19:21:57Z (GMT). No. of bitstreams: 1 Patricio Runnacles.pdf: 1899405 bytes, checksum: f12074a5a86f43aa25ef7c17d1ba8e87 (MD5) Previous issue date: 2014-12-17application/pdfporUNIVERSIDADE ESTADUAL DE PONTA GROSSAPrograma de Pós-Graduação em OdontologiaUEPGBRClinica Integrada, Dentística Restauradora e Periodontiadente pré-molarpolpa dentáriatemperatura corporalcura luminosa de adesivos dentáriodental pulppulp temperaturepremolarLED dental curing lightsPremolar volumeMicro ctCNPQ::CIENCIAS DA SAUDE::ODONTOLOGIAEFEITOS DA LUZ DE UM FOTOPOLIMERIZADOR LED NA TEMPERATURA PULPAR BASAL IN VIVO EM PRÉ-MOLARES HUMANOS E SUA CORRELAÇÃO COM AS CARACTERÍSTICAS ANATÔMICAS DENTAIS EX VIVOEffects of ligth emitted by a LED light curing unit on the in vivo baseline pulp temperature of intact human premolars and their correlation within ex vivo dental anatomic caracteristicsinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/doctoralThesisinfo:eu-repo/semantics/openAccessreponame:Biblioteca Digital de Teses e Dissertações da UEPGinstname:Universidade Estadual de Ponta Grossa (UEPG)instacron:UEPGORIGINALPatricio Runnacles.pdfapplication/pdf1899405http://tede2.uepg.br/jspui/bitstream/prefix/1696/1/Patricio%20Runnacles.pdff12074a5a86f43aa25ef7c17d1ba8e87MD51prefix/16962017-07-24 16:21:57.107oai:tede2.uepg.br:prefix/1696Biblioteca Digital de Teses e Dissertaçõeshttps://tede2.uepg.br/jspui/PUBhttp://tede2.uepg.br/oai/requestbicen@uepg.br||mv_fidelis@yahoo.com.bropendoar:2017-07-24T19:21:57Biblioteca Digital de Teses e Dissertações da UEPG - Universidade Estadual de Ponta Grossa (UEPG)false
dc.title.por.fl_str_mv EFEITOS DA LUZ DE UM FOTOPOLIMERIZADOR LED NA TEMPERATURA PULPAR BASAL IN VIVO EM PRÉ-MOLARES HUMANOS E SUA CORRELAÇÃO COM AS CARACTERÍSTICAS ANATÔMICAS DENTAIS EX VIVO
dc.title.alternative.eng.fl_str_mv Effects of ligth emitted by a LED light curing unit on the in vivo baseline pulp temperature of intact human premolars and their correlation within ex vivo dental anatomic caracteristics
title EFEITOS DA LUZ DE UM FOTOPOLIMERIZADOR LED NA TEMPERATURA PULPAR BASAL IN VIVO EM PRÉ-MOLARES HUMANOS E SUA CORRELAÇÃO COM AS CARACTERÍSTICAS ANATÔMICAS DENTAIS EX VIVO
spellingShingle EFEITOS DA LUZ DE UM FOTOPOLIMERIZADOR LED NA TEMPERATURA PULPAR BASAL IN VIVO EM PRÉ-MOLARES HUMANOS E SUA CORRELAÇÃO COM AS CARACTERÍSTICAS ANATÔMICAS DENTAIS EX VIVO
Runnacles, Patricio
dente pré-molar
polpa dentária
temperatura corporal
cura luminosa de adesivos dentário
dental pulp
pulp temperature
premolar
LED dental curing lights
Premolar volume
Micro ct
CNPQ::CIENCIAS DA SAUDE::ODONTOLOGIA
title_short EFEITOS DA LUZ DE UM FOTOPOLIMERIZADOR LED NA TEMPERATURA PULPAR BASAL IN VIVO EM PRÉ-MOLARES HUMANOS E SUA CORRELAÇÃO COM AS CARACTERÍSTICAS ANATÔMICAS DENTAIS EX VIVO
title_full EFEITOS DA LUZ DE UM FOTOPOLIMERIZADOR LED NA TEMPERATURA PULPAR BASAL IN VIVO EM PRÉ-MOLARES HUMANOS E SUA CORRELAÇÃO COM AS CARACTERÍSTICAS ANATÔMICAS DENTAIS EX VIVO
title_fullStr EFEITOS DA LUZ DE UM FOTOPOLIMERIZADOR LED NA TEMPERATURA PULPAR BASAL IN VIVO EM PRÉ-MOLARES HUMANOS E SUA CORRELAÇÃO COM AS CARACTERÍSTICAS ANATÔMICAS DENTAIS EX VIVO
title_full_unstemmed EFEITOS DA LUZ DE UM FOTOPOLIMERIZADOR LED NA TEMPERATURA PULPAR BASAL IN VIVO EM PRÉ-MOLARES HUMANOS E SUA CORRELAÇÃO COM AS CARACTERÍSTICAS ANATÔMICAS DENTAIS EX VIVO
title_sort EFEITOS DA LUZ DE UM FOTOPOLIMERIZADOR LED NA TEMPERATURA PULPAR BASAL IN VIVO EM PRÉ-MOLARES HUMANOS E SUA CORRELAÇÃO COM AS CARACTERÍSTICAS ANATÔMICAS DENTAIS EX VIVO
author Runnacles, Patricio
author_facet Runnacles, Patricio
author_role author
dc.contributor.advisor1.fl_str_mv Gomes, Osnara Maria Mongruel
dc.contributor.advisor1ID.fl_str_mv CPF:47308257991
dc.contributor.advisor1Lattes.fl_str_mv http://buscatextual.cnpq.br/buscatextual/visualizacv.do?id=K4773647J1
dc.contributor.advisor-co1.fl_str_mv Arrais, Cesar Augusto Galvão
dc.contributor.advisor-co1ID.fl_str_mv CPF:20064607810
dc.contributor.advisor-co1Lattes.fl_str_mv http://buscatextual.cnpq.br/buscatextual/visualizacv.do?id=K4770689J6
dc.contributor.referee1.fl_str_mv Reis, Alessandra
dc.contributor.referee1ID.fl_str_mv CPF:25713648800
dc.contributor.referee1Lattes.fl_str_mv REIS, A.
dc.contributor.referee2.fl_str_mv Campanha, Nara Hellen
dc.contributor.referee2ID.fl_str_mv CPF:11993069836
dc.contributor.referee2Lattes.fl_str_mv http://buscatextual.cnpq.br/buscatextual/visualizacv.do?id=K4705349Z2
dc.contributor.referee3.fl_str_mv Pupo, Yasmine Mendes
dc.contributor.referee3ID.fl_str_mv CPF:04694480910
dc.contributor.referee3Lattes.fl_str_mv http://buscatextual.cnpq.br/buscatextual/visualizacv.do?id=K4734244Z4
dc.contributor.referee4.fl_str_mv Gonzaga, Carla Castiglia
dc.contributor.referee4ID.fl_str_mv CPF:27051979825
dc.contributor.referee4Lattes.fl_str_mv http://buscatextual.cnpq.br/buscatextual/visualizacv.do?id=K4735565H6
dc.contributor.authorID.fl_str_mv CPF:00411125982
dc.contributor.authorLattes.fl_str_mv http://buscatextual.cnpq.br/buscatextual/visualizacv.do?id=K4359313A4
dc.contributor.author.fl_str_mv Runnacles, Patricio
contributor_str_mv Gomes, Osnara Maria Mongruel
Arrais, Cesar Augusto Galvão
Reis, Alessandra
Campanha, Nara Hellen
Pupo, Yasmine Mendes
Gonzaga, Carla Castiglia
dc.subject.por.fl_str_mv dente pré-molar
polpa dentária
temperatura corporal
cura luminosa de adesivos dentário
topic dente pré-molar
polpa dentária
temperatura corporal
cura luminosa de adesivos dentário
dental pulp
pulp temperature
premolar
LED dental curing lights
Premolar volume
Micro ct
CNPQ::CIENCIAS DA SAUDE::ODONTOLOGIA
dc.subject.eng.fl_str_mv dental pulp
pulp temperature
premolar
LED dental curing lights
Premolar volume
Micro ct
dc.subject.cnpq.fl_str_mv CNPQ::CIENCIAS DA SAUDE::ODONTOLOGIA
description Objective: This in vivo study evaluated the baseline pulp temperature in human intact upper premolars and the pulp temperature rise induced by light emitted by a high power polywave Light Emitting Diode (LED) light curing unit (LCU); also, the possible correlation between the change in temperature (ΔT) and the volume of the crown and the pulp chamber ex vivo. Material and methods: after approval of the local Ethics Committee was obtained (protocol #255,945), upper right and left first premolars requiring extraction for orthodontic reasons from 8 volunteers (n=15) received infiltrative and intraligamental anesthesia and were isolated using rubber dam. A small, occlusal preparation was made using high-speed hand piece, under constant air-water spray, until a minute pulp exposure was attained. The sterile probe from a wireless, NIST-traceable, temperature acquisition system (Temperature Data Acquisition - Thermes WFI, Physitemp) was inserted directly into the coronal pulp. Once the baseline pulp temperature was stable, the buccal surface was exposed to polywave light from a LED LCU (Bluephase 20i, Ivoclar Vivadent) using selected EMs: 10-s either in low (10-s/L) or High (10-s/H); 5-s-Turbo (5-s/T); and 60-s-High (60-s/H) intensities, allowing a 7-min span between each exposure. Peak PT values and PT increases from baseline (ΔT) after exposure were subjected to 1-Way, repeated measures ANOVAs, and Bonferroni's post-hoc tests (α=0.05). Linear regression analysis was performed to establish the relationship between applied radiant exposure and ΔT. Linear regression analysis was performed to establish the relationship between applied radiant exposure and ΔT. the teeth were extracted according to treatment plan. The crown and pulp volumes were measured using micro-tomography scanner, and pulp chamber / crown volume ratio was determined. Afterwards, premolar buccal-lingual, cervico-occlusal, mesio-distal dimensions as well as between cusp distances were measured using a digital caliper. Results: consistent temperature below body temperature, resulting in an average core temperature of 35 ° C (+ 0.7 ° C) was found. All modes of activation of the LED promoted significant temperature rise in the pulp chamber. A positive correlation between energy density and thermal variation was observed. No correlation was observed between coronal volume, the pulp chamber with temperature changes. A significant negative correlation was noted between pulp temperature changes when 10-s/L was used and VL.
publishDate 2014
dc.date.issued.fl_str_mv 2014-12-17
dc.date.available.fl_str_mv 2015-07-15
2017-07-24T19:21:57Z
dc.date.accessioned.fl_str_mv 2017-07-24T19:21:57Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/doctoralThesis
format doctoralThesis
status_str publishedVersion
dc.identifier.citation.fl_str_mv RUNNACLES, Patricio. Effects of ligth emitted by a LED light curing unit on the in vivo baseline pulp temperature of intact human premolars and their correlation within ex vivo dental anatomic caracteristics. 2014. 102 f. Tese (Doutorado em Clinica Integrada, Dentística Restauradora e Periodontia) - UNIVERSIDADE ESTADUAL DE PONTA GROSSA, Ponta Grossa, 2014.
dc.identifier.uri.fl_str_mv http://tede2.uepg.br/jspui/handle/prefix/1696
identifier_str_mv RUNNACLES, Patricio. Effects of ligth emitted by a LED light curing unit on the in vivo baseline pulp temperature of intact human premolars and their correlation within ex vivo dental anatomic caracteristics. 2014. 102 f. Tese (Doutorado em Clinica Integrada, Dentística Restauradora e Periodontia) - UNIVERSIDADE ESTADUAL DE PONTA GROSSA, Ponta Grossa, 2014.
url http://tede2.uepg.br/jspui/handle/prefix/1696
dc.language.iso.fl_str_mv por
language por
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv UNIVERSIDADE ESTADUAL DE PONTA GROSSA
dc.publisher.program.fl_str_mv Programa de Pós-Graduação em Odontologia
dc.publisher.initials.fl_str_mv UEPG
dc.publisher.country.fl_str_mv BR
dc.publisher.department.fl_str_mv Clinica Integrada, Dentística Restauradora e Periodontia
publisher.none.fl_str_mv UNIVERSIDADE ESTADUAL DE PONTA GROSSA
dc.source.none.fl_str_mv reponame:Biblioteca Digital de Teses e Dissertações da UEPG
instname:Universidade Estadual de Ponta Grossa (UEPG)
instacron:UEPG
instname_str Universidade Estadual de Ponta Grossa (UEPG)
instacron_str UEPG
institution UEPG
reponame_str Biblioteca Digital de Teses e Dissertações da UEPG
collection Biblioteca Digital de Teses e Dissertações da UEPG
bitstream.url.fl_str_mv http://tede2.uepg.br/jspui/bitstream/prefix/1696/1/Patricio%20Runnacles.pdf
bitstream.checksum.fl_str_mv f12074a5a86f43aa25ef7c17d1ba8e87
bitstream.checksumAlgorithm.fl_str_mv MD5
repository.name.fl_str_mv Biblioteca Digital de Teses e Dissertações da UEPG - Universidade Estadual de Ponta Grossa (UEPG)
repository.mail.fl_str_mv bicen@uepg.br||mv_fidelis@yahoo.com.br
_version_ 1800232802133737472