Evaluation of removal torque values for titanium screws of prosthetic abutments after different torque application techniques: in vitro study
Autor(a) principal: | |
---|---|
Data de Publicação: | 2016 |
Outros Autores: | , , , |
Tipo de documento: | Artigo |
Idioma: | eng |
Título da fonte: | Brazilian Dental Science |
Texto Completo: | https://ojs.ict.unesp.br/index.php/cob/article/view/1295 |
Resumo: | Objective: To evaluate the removal torque values on abutment screws after different torque application techniques. Materials and methods: Thirty specimens of an external hexagon implant, a prefabricated abutment and a titanium screw were distributed randomly in three groups. In group 1, the screws received a torque of 30 N.cm; group 2 received a torque of 30N.cm and a second torque after 10 min, in accordance with the technique proposed by Brending et al. and Dixon et al.; in group 3, a torque of 30 N.cm was applied and maintained for 20 s, in accordance with the technique recommended by Sella et al. The specimens were attached to a universal testing machine and a digital key was coupled to the load cell in order to control the torque value at a velocity of 1 N.cm/s. Removal torque was performed 10 minutes after torque application. Values were statistically analysed using one-way ANOVA and Tukey’s HSD test (? = 0.05). Results: The mean and standard deviations (±SD) of removal torque values found were 27.95±0.99N.cm for group 1, 28.32±0.68N.cm for group 2 and 26.89±1.03N.cm for group 3. Groups 1 and 2 exhibited statistically higher values of torque than group 3 (p ? 0.05). Conclusion: The technique recommended by Breeding et al. and Dixon et al. seemed to be the best option when considering the removal torque values of external hexagon implants. Keywords: Abutment screw; Dental implants; Removal torque. |
id |
UNESP-20_b7a67450bb13f38d675fe06b3cbf1661 |
---|---|
oai_identifier_str |
oai:ojs.pkp.sfu.ca:article/1295 |
network_acronym_str |
UNESP-20 |
network_name_str |
Brazilian Dental Science |
repository_id_str |
|
spelling |
Evaluation of removal torque values for titanium screws of prosthetic abutments after different torque application techniques: in vitro studyObjective: To evaluate the removal torque values on abutment screws after different torque application techniques. Materials and methods: Thirty specimens of an external hexagon implant, a prefabricated abutment and a titanium screw were distributed randomly in three groups. In group 1, the screws received a torque of 30 N.cm; group 2 received a torque of 30N.cm and a second torque after 10 min, in accordance with the technique proposed by Brending et al. and Dixon et al.; in group 3, a torque of 30 N.cm was applied and maintained for 20 s, in accordance with the technique recommended by Sella et al. The specimens were attached to a universal testing machine and a digital key was coupled to the load cell in order to control the torque value at a velocity of 1 N.cm/s. Removal torque was performed 10 minutes after torque application. Values were statistically analysed using one-way ANOVA and Tukey’s HSD test (? = 0.05). Results: The mean and standard deviations (±SD) of removal torque values found were 27.95±0.99N.cm for group 1, 28.32±0.68N.cm for group 2 and 26.89±1.03N.cm for group 3. Groups 1 and 2 exhibited statistically higher values of torque than group 3 (p ? 0.05). Conclusion: The technique recommended by Breeding et al. and Dixon et al. seemed to be the best option when considering the removal torque values of external hexagon implants. Keywords: Abutment screw; Dental implants; Removal torque.Institute of Science and Technology of São José dos Campos2016-11-29info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionapplication/pdfimage/jpegapplication/vnd.openxmlformats-officedocument.wordprocessingml.documentapplication/vnd.openxmlformats-officedocument.wordprocessingml.documenthttps://ojs.ict.unesp.br/index.php/cob/article/view/129510.14295/bds.2016.v19i4.1295Brazilian Dental Science; Vol. 19 No. 4 (2016): Oct. - Dec. / 2016 - Published November 2016; 12-18Brazilian Dental Science; v. 19 n. 4 (2016): Oct. - Dec. / 2016 - Published November 2016; 12-182178-6011reponame:Brazilian Dental Scienceinstname:Universidade Estadual Paulista Júlio de Mesquita Filho (UNESP)instacron:UNESPenghttps://ojs.ict.unesp.br/index.php/cob/article/view/1295/1064https://ojs.ict.unesp.br/index.php/cob/article/view/1295/2769https://ojs.ict.unesp.br/index.php/cob/article/view/1295/2770https://ojs.ict.unesp.br/index.php/cob/article/view/1295/2821Copyright (c) 2016 Brazilian Dental Scienceinfo:eu-repo/semantics/openAccessMedina, Carolina Enith BullenRafael, Caroline FreitasVolpato, Claudia Angela MariaÖzcan, MutluVasconcelos, Diego Klee de2020-01-28T12:17:16Zoai:ojs.pkp.sfu.ca:article/1295Revistahttp://bds.ict.unesp.br/PUBhttp://ojs.fosjc.unesp.br/index.php/index/oaisergio@fosjc.unesp.br||sergio@fosjc.unesp.br2178-60112178-6011opendoar:2022-11-08T16:30:08.318672Brazilian Dental Science - Universidade Estadual Paulista Júlio de Mesquita Filho (UNESP)true |
dc.title.none.fl_str_mv |
Evaluation of removal torque values for titanium screws of prosthetic abutments after different torque application techniques: in vitro study |
title |
Evaluation of removal torque values for titanium screws of prosthetic abutments after different torque application techniques: in vitro study |
spellingShingle |
Evaluation of removal torque values for titanium screws of prosthetic abutments after different torque application techniques: in vitro study Medina, Carolina Enith Bullen |
title_short |
Evaluation of removal torque values for titanium screws of prosthetic abutments after different torque application techniques: in vitro study |
title_full |
Evaluation of removal torque values for titanium screws of prosthetic abutments after different torque application techniques: in vitro study |
title_fullStr |
Evaluation of removal torque values for titanium screws of prosthetic abutments after different torque application techniques: in vitro study |
title_full_unstemmed |
Evaluation of removal torque values for titanium screws of prosthetic abutments after different torque application techniques: in vitro study |
title_sort |
Evaluation of removal torque values for titanium screws of prosthetic abutments after different torque application techniques: in vitro study |
author |
Medina, Carolina Enith Bullen |
author_facet |
Medina, Carolina Enith Bullen Rafael, Caroline Freitas Volpato, Claudia Angela Maria Özcan, Mutlu Vasconcelos, Diego Klee de |
author_role |
author |
author2 |
Rafael, Caroline Freitas Volpato, Claudia Angela Maria Özcan, Mutlu Vasconcelos, Diego Klee de |
author2_role |
author author author author |
dc.contributor.author.fl_str_mv |
Medina, Carolina Enith Bullen Rafael, Caroline Freitas Volpato, Claudia Angela Maria Özcan, Mutlu Vasconcelos, Diego Klee de |
description |
Objective: To evaluate the removal torque values on abutment screws after different torque application techniques. Materials and methods: Thirty specimens of an external hexagon implant, a prefabricated abutment and a titanium screw were distributed randomly in three groups. In group 1, the screws received a torque of 30 N.cm; group 2 received a torque of 30N.cm and a second torque after 10 min, in accordance with the technique proposed by Brending et al. and Dixon et al.; in group 3, a torque of 30 N.cm was applied and maintained for 20 s, in accordance with the technique recommended by Sella et al. The specimens were attached to a universal testing machine and a digital key was coupled to the load cell in order to control the torque value at a velocity of 1 N.cm/s. Removal torque was performed 10 minutes after torque application. Values were statistically analysed using one-way ANOVA and Tukey’s HSD test (? = 0.05). Results: The mean and standard deviations (±SD) of removal torque values found were 27.95±0.99N.cm for group 1, 28.32±0.68N.cm for group 2 and 26.89±1.03N.cm for group 3. Groups 1 and 2 exhibited statistically higher values of torque than group 3 (p ? 0.05). Conclusion: The technique recommended by Breeding et al. and Dixon et al. seemed to be the best option when considering the removal torque values of external hexagon implants. Keywords: Abutment screw; Dental implants; Removal torque. |
publishDate |
2016 |
dc.date.none.fl_str_mv |
2016-11-29 |
dc.type.driver.fl_str_mv |
info:eu-repo/semantics/article info:eu-repo/semantics/publishedVersion |
format |
article |
status_str |
publishedVersion |
dc.identifier.uri.fl_str_mv |
https://ojs.ict.unesp.br/index.php/cob/article/view/1295 10.14295/bds.2016.v19i4.1295 |
url |
https://ojs.ict.unesp.br/index.php/cob/article/view/1295 |
identifier_str_mv |
10.14295/bds.2016.v19i4.1295 |
dc.language.iso.fl_str_mv |
eng |
language |
eng |
dc.relation.none.fl_str_mv |
https://ojs.ict.unesp.br/index.php/cob/article/view/1295/1064 https://ojs.ict.unesp.br/index.php/cob/article/view/1295/2769 https://ojs.ict.unesp.br/index.php/cob/article/view/1295/2770 https://ojs.ict.unesp.br/index.php/cob/article/view/1295/2821 |
dc.rights.driver.fl_str_mv |
Copyright (c) 2016 Brazilian Dental Science info:eu-repo/semantics/openAccess |
rights_invalid_str_mv |
Copyright (c) 2016 Brazilian Dental Science |
eu_rights_str_mv |
openAccess |
dc.format.none.fl_str_mv |
application/pdf image/jpeg application/vnd.openxmlformats-officedocument.wordprocessingml.document application/vnd.openxmlformats-officedocument.wordprocessingml.document |
dc.publisher.none.fl_str_mv |
Institute of Science and Technology of São José dos Campos |
publisher.none.fl_str_mv |
Institute of Science and Technology of São José dos Campos |
dc.source.none.fl_str_mv |
Brazilian Dental Science; Vol. 19 No. 4 (2016): Oct. - Dec. / 2016 - Published November 2016; 12-18 Brazilian Dental Science; v. 19 n. 4 (2016): Oct. - Dec. / 2016 - Published November 2016; 12-18 2178-6011 reponame:Brazilian Dental Science instname:Universidade Estadual Paulista Júlio de Mesquita Filho (UNESP) instacron:UNESP |
instname_str |
Universidade Estadual Paulista Júlio de Mesquita Filho (UNESP) |
instacron_str |
UNESP |
institution |
UNESP |
reponame_str |
Brazilian Dental Science |
collection |
Brazilian Dental Science |
repository.name.fl_str_mv |
Brazilian Dental Science - Universidade Estadual Paulista Júlio de Mesquita Filho (UNESP) |
repository.mail.fl_str_mv |
sergio@fosjc.unesp.br||sergio@fosjc.unesp.br |
_version_ |
1788346899758055424 |