Evaluation of removal torque values for titanium screws of prosthetic abutments after different torque application techniques: in vitro study

Detalhes bibliográficos
Autor(a) principal: Medina, Carolina Enith Bullen
Data de Publicação: 2016
Outros Autores: Rafael, Caroline Freitas, Volpato, Claudia Angela Maria, Özcan, Mutlu, Vasconcelos, Diego Klee de
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Brazilian Dental Science
Texto Completo: https://ojs.ict.unesp.br/index.php/cob/article/view/1295
Resumo: Objective: To evaluate the removal torque values on abutment screws after different torque application techniques. Materials and methods: Thirty specimens of an external hexagon implant, a prefabricated abutment and a titanium screw were distributed randomly in three groups. In group 1, the screws received a torque of 30 N.cm; group 2 received a torque of 30N.cm and a second torque after 10 min, in accordance with the technique proposed by Brending et al. and Dixon et al.; in group 3, a torque of 30 N.cm was applied and maintained for 20 s, in accordance with the technique recommended by Sella et al. The specimens were attached to a universal testing machine and a digital key was coupled to the load cell in order to control the torque value at a velocity of 1 N.cm/s. Removal torque was performed 10 minutes after torque application. Values were statistically analysed using one-way ANOVA and Tukey’s HSD test (? = 0.05). Results: The mean and standard deviations (±SD) of removal torque values found were 27.95±0.99N.cm for group 1, 28.32±0.68N.cm for group 2 and 26.89±1.03N.cm for group 3. Groups 1 and 2 exhibited statistically higher values of torque than group 3 (p ? 0.05). Conclusion: The technique recommended by Breeding et al. and Dixon et al. seemed to be the best option when considering the removal torque values of external hexagon implants. Keywords: Abutment screw; Dental implants; Removal torque.
id UNESP-20_b7a67450bb13f38d675fe06b3cbf1661
oai_identifier_str oai:ojs.pkp.sfu.ca:article/1295
network_acronym_str UNESP-20
network_name_str Brazilian Dental Science
repository_id_str
spelling Evaluation of removal torque values for titanium screws of prosthetic abutments after different torque application techniques: in vitro studyObjective: To evaluate the removal torque values on abutment screws after different torque application techniques. Materials and methods: Thirty specimens of an external hexagon implant, a prefabricated abutment and a titanium screw were distributed randomly in three groups. In group 1, the screws received a torque of 30 N.cm; group 2 received a torque of 30N.cm and a second torque after 10 min, in accordance with the technique proposed by Brending et al. and Dixon et al.; in group 3, a torque of 30 N.cm was applied and maintained for 20 s, in accordance with the technique recommended by Sella et al. The specimens were attached to a universal testing machine and a digital key was coupled to the load cell in order to control the torque value at a velocity of 1 N.cm/s. Removal torque was performed 10 minutes after torque application. Values were statistically analysed using one-way ANOVA and Tukey’s HSD test (? = 0.05). Results: The mean and standard deviations (±SD) of removal torque values found were 27.95±0.99N.cm for group 1, 28.32±0.68N.cm for group 2 and 26.89±1.03N.cm for group 3. Groups 1 and 2 exhibited statistically higher values of torque than group 3 (p ? 0.05). Conclusion: The technique recommended by Breeding et al. and Dixon et al. seemed to be the best option when considering the removal torque values of external hexagon implants. Keywords: Abutment screw; Dental implants; Removal torque.Institute of Science and Technology of São José dos Campos2016-11-29info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionapplication/pdfimage/jpegapplication/vnd.openxmlformats-officedocument.wordprocessingml.documentapplication/vnd.openxmlformats-officedocument.wordprocessingml.documenthttps://ojs.ict.unesp.br/index.php/cob/article/view/129510.14295/bds.2016.v19i4.1295Brazilian Dental Science; Vol. 19 No. 4 (2016): Oct. - Dec. / 2016 - Published November 2016; 12-18Brazilian Dental Science; v. 19 n. 4 (2016): Oct. - Dec. / 2016 - Published November 2016; 12-182178-6011reponame:Brazilian Dental Scienceinstname:Universidade Estadual Paulista Júlio de Mesquita Filho (UNESP)instacron:UNESPenghttps://ojs.ict.unesp.br/index.php/cob/article/view/1295/1064https://ojs.ict.unesp.br/index.php/cob/article/view/1295/2769https://ojs.ict.unesp.br/index.php/cob/article/view/1295/2770https://ojs.ict.unesp.br/index.php/cob/article/view/1295/2821Copyright (c) 2016 Brazilian Dental Scienceinfo:eu-repo/semantics/openAccessMedina, Carolina Enith BullenRafael, Caroline FreitasVolpato, Claudia Angela MariaÖzcan, MutluVasconcelos, Diego Klee de2020-01-28T12:17:16Zoai:ojs.pkp.sfu.ca:article/1295Revistahttp://bds.ict.unesp.br/PUBhttp://ojs.fosjc.unesp.br/index.php/index/oaisergio@fosjc.unesp.br||sergio@fosjc.unesp.br2178-60112178-6011opendoar:2022-11-08T16:30:08.318672Brazilian Dental Science - Universidade Estadual Paulista Júlio de Mesquita Filho (UNESP)true
dc.title.none.fl_str_mv Evaluation of removal torque values for titanium screws of prosthetic abutments after different torque application techniques: in vitro study
title Evaluation of removal torque values for titanium screws of prosthetic abutments after different torque application techniques: in vitro study
spellingShingle Evaluation of removal torque values for titanium screws of prosthetic abutments after different torque application techniques: in vitro study
Medina, Carolina Enith Bullen
title_short Evaluation of removal torque values for titanium screws of prosthetic abutments after different torque application techniques: in vitro study
title_full Evaluation of removal torque values for titanium screws of prosthetic abutments after different torque application techniques: in vitro study
title_fullStr Evaluation of removal torque values for titanium screws of prosthetic abutments after different torque application techniques: in vitro study
title_full_unstemmed Evaluation of removal torque values for titanium screws of prosthetic abutments after different torque application techniques: in vitro study
title_sort Evaluation of removal torque values for titanium screws of prosthetic abutments after different torque application techniques: in vitro study
author Medina, Carolina Enith Bullen
author_facet Medina, Carolina Enith Bullen
Rafael, Caroline Freitas
Volpato, Claudia Angela Maria
Özcan, Mutlu
Vasconcelos, Diego Klee de
author_role author
author2 Rafael, Caroline Freitas
Volpato, Claudia Angela Maria
Özcan, Mutlu
Vasconcelos, Diego Klee de
author2_role author
author
author
author
dc.contributor.author.fl_str_mv Medina, Carolina Enith Bullen
Rafael, Caroline Freitas
Volpato, Claudia Angela Maria
Özcan, Mutlu
Vasconcelos, Diego Klee de
description Objective: To evaluate the removal torque values on abutment screws after different torque application techniques. Materials and methods: Thirty specimens of an external hexagon implant, a prefabricated abutment and a titanium screw were distributed randomly in three groups. In group 1, the screws received a torque of 30 N.cm; group 2 received a torque of 30N.cm and a second torque after 10 min, in accordance with the technique proposed by Brending et al. and Dixon et al.; in group 3, a torque of 30 N.cm was applied and maintained for 20 s, in accordance with the technique recommended by Sella et al. The specimens were attached to a universal testing machine and a digital key was coupled to the load cell in order to control the torque value at a velocity of 1 N.cm/s. Removal torque was performed 10 minutes after torque application. Values were statistically analysed using one-way ANOVA and Tukey’s HSD test (? = 0.05). Results: The mean and standard deviations (±SD) of removal torque values found were 27.95±0.99N.cm for group 1, 28.32±0.68N.cm for group 2 and 26.89±1.03N.cm for group 3. Groups 1 and 2 exhibited statistically higher values of torque than group 3 (p ? 0.05). Conclusion: The technique recommended by Breeding et al. and Dixon et al. seemed to be the best option when considering the removal torque values of external hexagon implants. Keywords: Abutment screw; Dental implants; Removal torque.
publishDate 2016
dc.date.none.fl_str_mv 2016-11-29
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv https://ojs.ict.unesp.br/index.php/cob/article/view/1295
10.14295/bds.2016.v19i4.1295
url https://ojs.ict.unesp.br/index.php/cob/article/view/1295
identifier_str_mv 10.14295/bds.2016.v19i4.1295
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv https://ojs.ict.unesp.br/index.php/cob/article/view/1295/1064
https://ojs.ict.unesp.br/index.php/cob/article/view/1295/2769
https://ojs.ict.unesp.br/index.php/cob/article/view/1295/2770
https://ojs.ict.unesp.br/index.php/cob/article/view/1295/2821
dc.rights.driver.fl_str_mv Copyright (c) 2016 Brazilian Dental Science
info:eu-repo/semantics/openAccess
rights_invalid_str_mv Copyright (c) 2016 Brazilian Dental Science
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
image/jpeg
application/vnd.openxmlformats-officedocument.wordprocessingml.document
application/vnd.openxmlformats-officedocument.wordprocessingml.document
dc.publisher.none.fl_str_mv Institute of Science and Technology of São José dos Campos
publisher.none.fl_str_mv Institute of Science and Technology of São José dos Campos
dc.source.none.fl_str_mv Brazilian Dental Science; Vol. 19 No. 4 (2016): Oct. - Dec. / 2016 - Published November 2016; 12-18
Brazilian Dental Science; v. 19 n. 4 (2016): Oct. - Dec. / 2016 - Published November 2016; 12-18
2178-6011
reponame:Brazilian Dental Science
instname:Universidade Estadual Paulista Júlio de Mesquita Filho (UNESP)
instacron:UNESP
instname_str Universidade Estadual Paulista Júlio de Mesquita Filho (UNESP)
instacron_str UNESP
institution UNESP
reponame_str Brazilian Dental Science
collection Brazilian Dental Science
repository.name.fl_str_mv Brazilian Dental Science - Universidade Estadual Paulista Júlio de Mesquita Filho (UNESP)
repository.mail.fl_str_mv sergio@fosjc.unesp.br||sergio@fosjc.unesp.br
_version_ 1788346899758055424