Gel and porous polyethylene implants for anophthalmic cavity reconstruction-evaluation using the B scan ultrasound

Detalhes bibliográficos
Autor(a) principal: Barbarini Ferraz, Lucieni Cristina [UNESP]
Data de Publicação: 2009
Outros Autores: Schellini, Silvana Artioli [UNESP], Hoyama, Erika [UNESP], Wludarski, Sheila Lordelo [UNESP], Padovani, Carlos Roberto [UNESP]
Tipo de documento: Artigo de conferência
Idioma: eng
Título da fonte: Repositório Institucional da UNESP
Texto Completo: http://www.ijo.cn/en_publish/2009/4/200904003.pdf
http://hdl.handle.net/11449/12630
Resumo: AIM: To evaluate the host response of the gel and porous polyethylene implants in anophthalmic cavities using the B scan ultrasound.METHODS: Thirty-six white rabbits underwent unilateral enucleation with placement of gel or porous polyethylene spheres implants. The animals were submitted to clinical examination weekly and to ultrasound evaluation on 30, 60 and 90 days after surgery.RESULTS: All rabbits with gel polyethylene spheres, except one, showed implant extrusion probably because the gel spheres have hydrated and increased in volume. The B ultrasound of the gel polyethylene implant did not show vessels inside during the following period. Five animals (27.8%) with porous polyethylene spheres presented implant extrusion after 30 days of surgery. According to B ultrasound, the porous polyethylene implant showed irregular and heterogeneous architecture and reflective peaks similar to vascularized tissues.CONCLUSION: More studies are required to determine the ideal volume of gel polyethylene implant necessary to correct the diminished orbital content in the anophthalmic cavity. The B ultrasound effectiveness showed in this study for anophthalmic socket implants evaluation provides useful information for further in vivo studies and might substitute expensive methods of implants vascularization evaluation,
id UNSP_3379679b13eefc7fb46d45c890359638
oai_identifier_str oai:repositorio.unesp.br:11449/12630
network_acronym_str UNSP
network_name_str Repositório Institucional da UNESP
repository_id_str 2946
spelling Gel and porous polyethylene implants for anophthalmic cavity reconstruction-evaluation using the B scan ultrasoundpolyethylene implantB ultrasoundrabbitanophthalmic cavityAIM: To evaluate the host response of the gel and porous polyethylene implants in anophthalmic cavities using the B scan ultrasound.METHODS: Thirty-six white rabbits underwent unilateral enucleation with placement of gel or porous polyethylene spheres implants. The animals were submitted to clinical examination weekly and to ultrasound evaluation on 30, 60 and 90 days after surgery.RESULTS: All rabbits with gel polyethylene spheres, except one, showed implant extrusion probably because the gel spheres have hydrated and increased in volume. The B ultrasound of the gel polyethylene implant did not show vessels inside during the following period. Five animals (27.8%) with porous polyethylene spheres presented implant extrusion after 30 days of surgery. According to B ultrasound, the porous polyethylene implant showed irregular and heterogeneous architecture and reflective peaks similar to vascularized tissues.CONCLUSION: More studies are required to determine the ideal volume of gel polyethylene implant necessary to correct the diminished orbital content in the anophthalmic cavity. The B ultrasound effectiveness showed in this study for anophthalmic socket implants evaluation provides useful information for further in vivo studies and might substitute expensive methods of implants vascularization evaluation,Univ Estadual Paulista, Fac Med Botucatu, Dept Ophthalmol, São Paulo, SP, BrazilUniv Estadual Paulista, Fac Med Botucatu, Dept Pathol, São Paulo, SP, BrazilUniv Estadual Paulista, Fac Med Botucatu, Dept Biostat, São Paulo, SP, BrazilUniv Estadual Paulista, Fac Med Botucatu, Dept Ophthalmol, São Paulo, SP, BrazilUniv Estadual Paulista, Fac Med Botucatu, Dept Pathol, São Paulo, SP, BrazilUniv Estadual Paulista, Fac Med Botucatu, Dept Biostat, São Paulo, SP, BrazilIjo PressUniversidade Estadual Paulista (Unesp)Barbarini Ferraz, Lucieni Cristina [UNESP]Schellini, Silvana Artioli [UNESP]Hoyama, Erika [UNESP]Wludarski, Sheila Lordelo [UNESP]Padovani, Carlos Roberto [UNESP]2014-05-20T13:36:41Z2014-05-20T13:36:41Z2009-12-18info:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/conferenceObject291-293http://www.ijo.cn/en_publish/2009/4/200904003.pdfInternational Journal of Ophthalmology. Xi An: Ijo Press, v. 2, n. 4, p. 291-293, 2009.1672-5123http://hdl.handle.net/11449/12630WOS:00028293400000394202491008354928727897080522289Web of Sciencereponame:Repositório Institucional da UNESPinstname:Universidade Estadual Paulista (UNESP)instacron:UNESPengInternational Journal of Ophthalmology0,109info:eu-repo/semantics/openAccess2021-10-22T18:27:07Zoai:repositorio.unesp.br:11449/12630Repositório InstitucionalPUBhttp://repositorio.unesp.br/oai/requestopendoar:29462021-10-22T18:27:07Repositório Institucional da UNESP - Universidade Estadual Paulista (UNESP)false
dc.title.none.fl_str_mv Gel and porous polyethylene implants for anophthalmic cavity reconstruction-evaluation using the B scan ultrasound
title Gel and porous polyethylene implants for anophthalmic cavity reconstruction-evaluation using the B scan ultrasound
spellingShingle Gel and porous polyethylene implants for anophthalmic cavity reconstruction-evaluation using the B scan ultrasound
Barbarini Ferraz, Lucieni Cristina [UNESP]
polyethylene implant
B ultrasound
rabbit
anophthalmic cavity
title_short Gel and porous polyethylene implants for anophthalmic cavity reconstruction-evaluation using the B scan ultrasound
title_full Gel and porous polyethylene implants for anophthalmic cavity reconstruction-evaluation using the B scan ultrasound
title_fullStr Gel and porous polyethylene implants for anophthalmic cavity reconstruction-evaluation using the B scan ultrasound
title_full_unstemmed Gel and porous polyethylene implants for anophthalmic cavity reconstruction-evaluation using the B scan ultrasound
title_sort Gel and porous polyethylene implants for anophthalmic cavity reconstruction-evaluation using the B scan ultrasound
author Barbarini Ferraz, Lucieni Cristina [UNESP]
author_facet Barbarini Ferraz, Lucieni Cristina [UNESP]
Schellini, Silvana Artioli [UNESP]
Hoyama, Erika [UNESP]
Wludarski, Sheila Lordelo [UNESP]
Padovani, Carlos Roberto [UNESP]
author_role author
author2 Schellini, Silvana Artioli [UNESP]
Hoyama, Erika [UNESP]
Wludarski, Sheila Lordelo [UNESP]
Padovani, Carlos Roberto [UNESP]
author2_role author
author
author
author
dc.contributor.none.fl_str_mv Universidade Estadual Paulista (Unesp)
dc.contributor.author.fl_str_mv Barbarini Ferraz, Lucieni Cristina [UNESP]
Schellini, Silvana Artioli [UNESP]
Hoyama, Erika [UNESP]
Wludarski, Sheila Lordelo [UNESP]
Padovani, Carlos Roberto [UNESP]
dc.subject.por.fl_str_mv polyethylene implant
B ultrasound
rabbit
anophthalmic cavity
topic polyethylene implant
B ultrasound
rabbit
anophthalmic cavity
description AIM: To evaluate the host response of the gel and porous polyethylene implants in anophthalmic cavities using the B scan ultrasound.METHODS: Thirty-six white rabbits underwent unilateral enucleation with placement of gel or porous polyethylene spheres implants. The animals were submitted to clinical examination weekly and to ultrasound evaluation on 30, 60 and 90 days after surgery.RESULTS: All rabbits with gel polyethylene spheres, except one, showed implant extrusion probably because the gel spheres have hydrated and increased in volume. The B ultrasound of the gel polyethylene implant did not show vessels inside during the following period. Five animals (27.8%) with porous polyethylene spheres presented implant extrusion after 30 days of surgery. According to B ultrasound, the porous polyethylene implant showed irregular and heterogeneous architecture and reflective peaks similar to vascularized tissues.CONCLUSION: More studies are required to determine the ideal volume of gel polyethylene implant necessary to correct the diminished orbital content in the anophthalmic cavity. The B ultrasound effectiveness showed in this study for anophthalmic socket implants evaluation provides useful information for further in vivo studies and might substitute expensive methods of implants vascularization evaluation,
publishDate 2009
dc.date.none.fl_str_mv 2009-12-18
2014-05-20T13:36:41Z
2014-05-20T13:36:41Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/conferenceObject
format conferenceObject
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://www.ijo.cn/en_publish/2009/4/200904003.pdf
International Journal of Ophthalmology. Xi An: Ijo Press, v. 2, n. 4, p. 291-293, 2009.
1672-5123
http://hdl.handle.net/11449/12630
WOS:000282934000003
9420249100835492
8727897080522289
url http://www.ijo.cn/en_publish/2009/4/200904003.pdf
http://hdl.handle.net/11449/12630
identifier_str_mv International Journal of Ophthalmology. Xi An: Ijo Press, v. 2, n. 4, p. 291-293, 2009.
1672-5123
WOS:000282934000003
9420249100835492
8727897080522289
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv International Journal of Ophthalmology
0,109
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv 291-293
dc.publisher.none.fl_str_mv Ijo Press
publisher.none.fl_str_mv Ijo Press
dc.source.none.fl_str_mv Web of Science
reponame:Repositório Institucional da UNESP
instname:Universidade Estadual Paulista (UNESP)
instacron:UNESP
instname_str Universidade Estadual Paulista (UNESP)
instacron_str UNESP
institution UNESP
reponame_str Repositório Institucional da UNESP
collection Repositório Institucional da UNESP
repository.name.fl_str_mv Repositório Institucional da UNESP - Universidade Estadual Paulista (UNESP)
repository.mail.fl_str_mv
_version_ 1799964845411401728