Harmonization of color measurements for dental application

Detalhes bibliográficos
Autor(a) principal: Paravina, Rade D.
Data de Publicação: 2020
Outros Autores: Pereira Sanchez, Natalie A., Tango, Rubens N. [UNESP]
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Institucional da UNESP
Texto Completo: http://dx.doi.org/10.1002/col.22553
http://hdl.handle.net/11449/197147
Resumo: This multicenter study evaluated the effect of harmonizing spectrophotometric (d/8 degrees) color measurements by using actual dental materials of clinically relevant thickness and translucency as calibration tiles. Ten tooth-colored, 1-mm thick, resin- and ceramic-based materials were tested at the coordinating center (CC) and six research sites (RS). Correction factors were calculated for each material and each site to obtain the corrected reflection spectra from the non-corrected data set. Non-corrected ((NC)) and corrected ((C)) reflection values were converted into CIEDE2000 and CIELAB color parameters for the following comparisons: CC vs RSNC and CC vs RSC, between RSNC vs RSNC and RSC vs RSC pairs, and among RS separately (alpha = 0.05). Mean color differences for CC vs RSNC and corresponding CC vs RSC values were Delta E-00 = 4.2 (2.6) and Delta E-00 = 0.7 (0.6), respectively. Corresponding values for RSNC vs RSNC and RSC vs RSC specimen pairs were Delta E-00 = 5.9 (2.9) and Delta E-00 = 1.0 (0.5), respectively. The CIEDE2000 NC vs C comparison for each research site separately ranged from 0.5 to 8.5. Color differences of CC vs RSC and RSC vs RSC pairs decreased 83% to 84.5% as compared with respective color differences for non-corrected research site measurements and was RS dependent (P <.001). The obtained results validated the proposed method of harmonizing color measurements for dental application as a reliable quality control tool for dental research laboratories and industry.
id UNSP_d2dd2dde76c05c95c63f521762445c8c
oai_identifier_str oai:repositorio.unesp.br:11449/197147
network_acronym_str UNSP
network_name_str Repositório Institucional da UNESP
repository_id_str 2946
spelling Harmonization of color measurements for dental applicationcolor harmonizationcolor tolerancesdentistryaestheticspsychophysicsspectrophotometryThis multicenter study evaluated the effect of harmonizing spectrophotometric (d/8 degrees) color measurements by using actual dental materials of clinically relevant thickness and translucency as calibration tiles. Ten tooth-colored, 1-mm thick, resin- and ceramic-based materials were tested at the coordinating center (CC) and six research sites (RS). Correction factors were calculated for each material and each site to obtain the corrected reflection spectra from the non-corrected data set. Non-corrected ((NC)) and corrected ((C)) reflection values were converted into CIEDE2000 and CIELAB color parameters for the following comparisons: CC vs RSNC and CC vs RSC, between RSNC vs RSNC and RSC vs RSC pairs, and among RS separately (alpha = 0.05). Mean color differences for CC vs RSNC and corresponding CC vs RSC values were Delta E-00 = 4.2 (2.6) and Delta E-00 = 0.7 (0.6), respectively. Corresponding values for RSNC vs RSNC and RSC vs RSC specimen pairs were Delta E-00 = 5.9 (2.9) and Delta E-00 = 1.0 (0.5), respectively. The CIEDE2000 NC vs C comparison for each research site separately ranged from 0.5 to 8.5. Color differences of CC vs RSC and RSC vs RSC pairs decreased 83% to 84.5% as compared with respective color differences for non-corrected research site measurements and was RS dependent (P <.001). The obtained results validated the proposed method of harmonizing color measurements for dental application as a reliable quality control tool for dental research laboratories and industry.Univ Texas Houston, Sch Dent Houston, Dept Restorat Dent & Prosthodont, Houston, TX USAUniv Texas Houston, Sch Dent Houston, Houston Ctr Biomat & Biomimet HCBB, Houston, TX USAState Univ Sao Paulo, Sch Dent Sao Jose Dos Campos, Dept Dent Mat & Prosthodont, Sao Jose Dos Campos, BrazilState Univ Sao Paulo, Sch Dent Sao Jose Dos Campos, Dept Dent Mat & Prosthodont, Sao Jose Dos Campos, BrazilWiley-BlackwellUniv Texas HoustonUniversidade Estadual Paulista (Unesp)Paravina, Rade D.Pereira Sanchez, Natalie A.Tango, Rubens N. [UNESP]2020-12-10T20:07:40Z2020-12-10T20:07:40Z2020-08-06info:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/article7http://dx.doi.org/10.1002/col.22553Color Research And Application. Hoboken: Wiley, 7 p., 2020.0361-2317http://hdl.handle.net/11449/19714710.1002/col.22553WOS:000556132900001Web of Sciencereponame:Repositório Institucional da UNESPinstname:Universidade Estadual Paulista (UNESP)instacron:UNESPengColor Research And Applicationinfo:eu-repo/semantics/openAccess2021-10-23T12:05:09Zoai:repositorio.unesp.br:11449/197147Repositório InstitucionalPUBhttp://repositorio.unesp.br/oai/requestopendoar:29462024-08-05T23:42:45.613839Repositório Institucional da UNESP - Universidade Estadual Paulista (UNESP)false
dc.title.none.fl_str_mv Harmonization of color measurements for dental application
title Harmonization of color measurements for dental application
spellingShingle Harmonization of color measurements for dental application
Paravina, Rade D.
color harmonization
color tolerances
dentistry
aesthetics
psychophysics
spectrophotometry
title_short Harmonization of color measurements for dental application
title_full Harmonization of color measurements for dental application
title_fullStr Harmonization of color measurements for dental application
title_full_unstemmed Harmonization of color measurements for dental application
title_sort Harmonization of color measurements for dental application
author Paravina, Rade D.
author_facet Paravina, Rade D.
Pereira Sanchez, Natalie A.
Tango, Rubens N. [UNESP]
author_role author
author2 Pereira Sanchez, Natalie A.
Tango, Rubens N. [UNESP]
author2_role author
author
dc.contributor.none.fl_str_mv Univ Texas Houston
Universidade Estadual Paulista (Unesp)
dc.contributor.author.fl_str_mv Paravina, Rade D.
Pereira Sanchez, Natalie A.
Tango, Rubens N. [UNESP]
dc.subject.por.fl_str_mv color harmonization
color tolerances
dentistry
aesthetics
psychophysics
spectrophotometry
topic color harmonization
color tolerances
dentistry
aesthetics
psychophysics
spectrophotometry
description This multicenter study evaluated the effect of harmonizing spectrophotometric (d/8 degrees) color measurements by using actual dental materials of clinically relevant thickness and translucency as calibration tiles. Ten tooth-colored, 1-mm thick, resin- and ceramic-based materials were tested at the coordinating center (CC) and six research sites (RS). Correction factors were calculated for each material and each site to obtain the corrected reflection spectra from the non-corrected data set. Non-corrected ((NC)) and corrected ((C)) reflection values were converted into CIEDE2000 and CIELAB color parameters for the following comparisons: CC vs RSNC and CC vs RSC, between RSNC vs RSNC and RSC vs RSC pairs, and among RS separately (alpha = 0.05). Mean color differences for CC vs RSNC and corresponding CC vs RSC values were Delta E-00 = 4.2 (2.6) and Delta E-00 = 0.7 (0.6), respectively. Corresponding values for RSNC vs RSNC and RSC vs RSC specimen pairs were Delta E-00 = 5.9 (2.9) and Delta E-00 = 1.0 (0.5), respectively. The CIEDE2000 NC vs C comparison for each research site separately ranged from 0.5 to 8.5. Color differences of CC vs RSC and RSC vs RSC pairs decreased 83% to 84.5% as compared with respective color differences for non-corrected research site measurements and was RS dependent (P <.001). The obtained results validated the proposed method of harmonizing color measurements for dental application as a reliable quality control tool for dental research laboratories and industry.
publishDate 2020
dc.date.none.fl_str_mv 2020-12-10T20:07:40Z
2020-12-10T20:07:40Z
2020-08-06
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://dx.doi.org/10.1002/col.22553
Color Research And Application. Hoboken: Wiley, 7 p., 2020.
0361-2317
http://hdl.handle.net/11449/197147
10.1002/col.22553
WOS:000556132900001
url http://dx.doi.org/10.1002/col.22553
http://hdl.handle.net/11449/197147
identifier_str_mv Color Research And Application. Hoboken: Wiley, 7 p., 2020.
0361-2317
10.1002/col.22553
WOS:000556132900001
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv Color Research And Application
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv 7
dc.publisher.none.fl_str_mv Wiley-Blackwell
publisher.none.fl_str_mv Wiley-Blackwell
dc.source.none.fl_str_mv Web of Science
reponame:Repositório Institucional da UNESP
instname:Universidade Estadual Paulista (UNESP)
instacron:UNESP
instname_str Universidade Estadual Paulista (UNESP)
instacron_str UNESP
institution UNESP
reponame_str Repositório Institucional da UNESP
collection Repositório Institucional da UNESP
repository.name.fl_str_mv Repositório Institucional da UNESP - Universidade Estadual Paulista (UNESP)
repository.mail.fl_str_mv
_version_ 1808129544644722688