Multiple-unit implant frames: one-piece casting vs. laser welding and brazing

Detalhes bibliográficos
Autor(a) principal: Costa, Elza Maria Valadares Da [UNESP]
Data de Publicação: 2004
Outros Autores: Neisser, Maximiliano Piero [UNESP], Bottino, Marco Antônio [UNESP]
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Institucional da UNESP
Texto Completo: http://dx.doi.org/10.1590/S1678-77572004000300013
http://hdl.handle.net/11449/213057
Resumo: The linear distortion of prostheses over implants, one-piece casting and cast in sections followed by laser welding by laser and brazing was evaluated in an edentulous mandibular model with five parallel abutments, with a distance of 10mm from center to center. Seventy five gold cylinders were tightened with screws on the abutments with 10Ncm torque. The cylinder/analogue assemblies were measured by microscopic examination (0.001mm accuracy) and the obtained results were compared with the GC (control group). Fifteen metal frames were waxed and cast in a gold alloy (Stabilor, Degussa Hulls, Brazil) and divided into three groups with five elements each, as followed: GM (one-piece casting), GB (section and brazing) and GL (section and laser welding). In all groups, measurements were taken at the right, left, buccal and lingual sides of the cylinder/analogue interface and the results were submitted to analysis of variance (ANOVA) and to the Tukey test (5%). The smallest amount of distortion was seen in the laser group (GL with a mean value of 13.58), followed by the brazing group (GB with a mean value of 24.33) and one-piece (GM with a mean value of 40.00). The greatest distortion was found in the one-piece group (GM).
id UNSP_e038276603316840257ea71178e34b9b
oai_identifier_str oai:repositorio.unesp.br:11449/213057
network_acronym_str UNSP
network_name_str Repositório Institucional da UNESP
repository_id_str 2946
spelling Multiple-unit implant frames: one-piece casting vs. laser welding and brazingSobre estruturas de implantes múltiplos: fundição em monobloco versus soldagem a laser e brasagemImplant prosthesisEndosseous dental implantsWeldingImplante sobre próteseImplante dentário endoósseoBrasagemThe linear distortion of prostheses over implants, one-piece casting and cast in sections followed by laser welding by laser and brazing was evaluated in an edentulous mandibular model with five parallel abutments, with a distance of 10mm from center to center. Seventy five gold cylinders were tightened with screws on the abutments with 10Ncm torque. The cylinder/analogue assemblies were measured by microscopic examination (0.001mm accuracy) and the obtained results were compared with the GC (control group). Fifteen metal frames were waxed and cast in a gold alloy (Stabilor, Degussa Hulls, Brazil) and divided into three groups with five elements each, as followed: GM (one-piece casting), GB (section and brazing) and GL (section and laser welding). In all groups, measurements were taken at the right, left, buccal and lingual sides of the cylinder/analogue interface and the results were submitted to analysis of variance (ANOVA) and to the Tukey test (5%). The smallest amount of distortion was seen in the laser group (GL with a mean value of 13.58), followed by the brazing group (GB with a mean value of 24.33) and one-piece (GM with a mean value of 40.00). The greatest distortion was found in the one-piece group (GM).A distorção linear das próteses sobre implantes, fundidas em monobloco e fundidas em secções e soldadas a laser e por brasagem foi avaliada em um modelo de uma mandíbula edêntula com cinco análogos de pilares de implantes, paralelos entre si com distância de 10mm medidos de centro a centro. Sobre estes foram parafusados novos cilindros de ouro, com torque de 10N/cm. Então foram executadas 15 sobre-estruturas metálicas que foram divididas: GC - Adaptação passiva dos cilindros de ouro; GM - monobloco, GB - segmentos soldados por brasagem e, GL - segmentos soldados a laser. Observaram-se as estruturas fixadas com parafusos novos sob microscópio de mensuração. A medição foi realizada na vestibular direita e esquerda e da mesma forma na lingual da interface cilindro/análogo e dados anotados em tabela apropriada. Para o eixo y, a técnica de soldagem a laser (Média=12,41 e Dp=11,12) produziu menos distorção significativamente do que a técnica de brasagem (Média=23,06 e Dp=21,88) seguida pela técnica de monobloco (Média=39,83 e Dp=21,88) p=0.05. Para o eixo x, somente o grupo Monobloco apresentou diferença significativa para as demais técnicas (Média=45,66 e Dp=21,56).Universidade Estadual Paulista, Dental SchoolUniversidade Estadual Paulista, Dental SchoolFaculdade De Odontologia De Bauru - USPUniversidade Estadual Paulista (Unesp)Costa, Elza Maria Valadares Da [UNESP]Neisser, Maximiliano Piero [UNESP]Bottino, Marco Antônio [UNESP]2021-07-14T10:49:18Z2021-07-14T10:49:18Z2004-09info:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/article227-231application/pdfhttp://dx.doi.org/10.1590/S1678-77572004000300013Journal of Applied Oral Science. Bauru, SP, Brazil: Faculdade De Odontologia De Bauru - USP, v. 12, n. 3, p. 227-231, 2004.1678-77571678-7765http://hdl.handle.net/11449/21305710.1590/S1678-77572004000300013S1678-77572004000300013S1678-77572004000300013.pdfSciELOreponame:Repositório Institucional da UNESPinstname:Universidade Estadual Paulista (UNESP)instacron:UNESPengJournal of Applied Oral Scienceinfo:eu-repo/semantics/openAccess2023-10-31T06:08:10Zoai:repositorio.unesp.br:11449/213057Repositório InstitucionalPUBhttp://repositorio.unesp.br/oai/requestopendoar:29462024-08-05T16:31:09.056904Repositório Institucional da UNESP - Universidade Estadual Paulista (UNESP)false
dc.title.none.fl_str_mv Multiple-unit implant frames: one-piece casting vs. laser welding and brazing
Sobre estruturas de implantes múltiplos: fundição em monobloco versus soldagem a laser e brasagem
title Multiple-unit implant frames: one-piece casting vs. laser welding and brazing
spellingShingle Multiple-unit implant frames: one-piece casting vs. laser welding and brazing
Costa, Elza Maria Valadares Da [UNESP]
Implant prosthesis
Endosseous dental implants
Welding
Implante sobre prótese
Implante dentário endoósseo
Brasagem
title_short Multiple-unit implant frames: one-piece casting vs. laser welding and brazing
title_full Multiple-unit implant frames: one-piece casting vs. laser welding and brazing
title_fullStr Multiple-unit implant frames: one-piece casting vs. laser welding and brazing
title_full_unstemmed Multiple-unit implant frames: one-piece casting vs. laser welding and brazing
title_sort Multiple-unit implant frames: one-piece casting vs. laser welding and brazing
author Costa, Elza Maria Valadares Da [UNESP]
author_facet Costa, Elza Maria Valadares Da [UNESP]
Neisser, Maximiliano Piero [UNESP]
Bottino, Marco Antônio [UNESP]
author_role author
author2 Neisser, Maximiliano Piero [UNESP]
Bottino, Marco Antônio [UNESP]
author2_role author
author
dc.contributor.none.fl_str_mv Universidade Estadual Paulista (Unesp)
dc.contributor.author.fl_str_mv Costa, Elza Maria Valadares Da [UNESP]
Neisser, Maximiliano Piero [UNESP]
Bottino, Marco Antônio [UNESP]
dc.subject.por.fl_str_mv Implant prosthesis
Endosseous dental implants
Welding
Implante sobre prótese
Implante dentário endoósseo
Brasagem
topic Implant prosthesis
Endosseous dental implants
Welding
Implante sobre prótese
Implante dentário endoósseo
Brasagem
description The linear distortion of prostheses over implants, one-piece casting and cast in sections followed by laser welding by laser and brazing was evaluated in an edentulous mandibular model with five parallel abutments, with a distance of 10mm from center to center. Seventy five gold cylinders were tightened with screws on the abutments with 10Ncm torque. The cylinder/analogue assemblies were measured by microscopic examination (0.001mm accuracy) and the obtained results were compared with the GC (control group). Fifteen metal frames were waxed and cast in a gold alloy (Stabilor, Degussa Hulls, Brazil) and divided into three groups with five elements each, as followed: GM (one-piece casting), GB (section and brazing) and GL (section and laser welding). In all groups, measurements were taken at the right, left, buccal and lingual sides of the cylinder/analogue interface and the results were submitted to analysis of variance (ANOVA) and to the Tukey test (5%). The smallest amount of distortion was seen in the laser group (GL with a mean value of 13.58), followed by the brazing group (GB with a mean value of 24.33) and one-piece (GM with a mean value of 40.00). The greatest distortion was found in the one-piece group (GM).
publishDate 2004
dc.date.none.fl_str_mv 2004-09
2021-07-14T10:49:18Z
2021-07-14T10:49:18Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://dx.doi.org/10.1590/S1678-77572004000300013
Journal of Applied Oral Science. Bauru, SP, Brazil: Faculdade De Odontologia De Bauru - USP, v. 12, n. 3, p. 227-231, 2004.
1678-7757
1678-7765
http://hdl.handle.net/11449/213057
10.1590/S1678-77572004000300013
S1678-77572004000300013
S1678-77572004000300013.pdf
url http://dx.doi.org/10.1590/S1678-77572004000300013
http://hdl.handle.net/11449/213057
identifier_str_mv Journal of Applied Oral Science. Bauru, SP, Brazil: Faculdade De Odontologia De Bauru - USP, v. 12, n. 3, p. 227-231, 2004.
1678-7757
1678-7765
10.1590/S1678-77572004000300013
S1678-77572004000300013
S1678-77572004000300013.pdf
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv Journal of Applied Oral Science
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv 227-231
application/pdf
dc.publisher.none.fl_str_mv Faculdade De Odontologia De Bauru - USP
publisher.none.fl_str_mv Faculdade De Odontologia De Bauru - USP
dc.source.none.fl_str_mv SciELO
reponame:Repositório Institucional da UNESP
instname:Universidade Estadual Paulista (UNESP)
instacron:UNESP
instname_str Universidade Estadual Paulista (UNESP)
instacron_str UNESP
institution UNESP
reponame_str Repositório Institucional da UNESP
collection Repositório Institucional da UNESP
repository.name.fl_str_mv Repositório Institucional da UNESP - Universidade Estadual Paulista (UNESP)
repository.mail.fl_str_mv
_version_ 1808128664888410112