Resumo: | Introduction: Electronic Apex Locators (EAL) may interfere with Cardiac Pacemakers (PM) by electromangnetic interference (EMI). Objective: To evaluate, in vitro, the EMI caused by EAL on a PM. Methods: Four different EAL (Apex DSP, NovApex N-21, RomiApex A-15, C-ROOT I) had their electrodes immersed in a saline solution at distances of 10, 15, 20, 25 and 30 cm from the electrode of the PM (Medtronic Sensia SEDR01) monitored by telemetry. Results: With 0.2 mV sensitivity, the PM had interference from all 4 EAL, with inhibition, stimulation recycle, and deflagration. With 2mV sensitivity, no interference was observed. Regardless the PM sensitivity, no EAL was able to lead, even at short distances, reversion to asynchronous or deprogramming. Conclusion: When set to maximum sensitivity, the PM suffered interference, although that wouldn’t be a real in vivo situation. |
id |
ABO-1_0ed4b837f689aa631526ca43cb0a532e |
---|---|
oai_identifier_str |
oai:revista_aborj_org_br.www.revista.aborj.org.br:article/372 |
network_acronym_str |
ABO-1 |
network_name_str |
Revista Brasileira de Odontologia |
repository_id_str |
|
spelling |
In vitro interference of Electronic Apex Locator on Implantable Cardiac PacemakerInterferência in vitro do localizador apical eletrônico em marcapasso cardíaco implantávelMEDICINA DENTÁRIA (ODONTOLOGIA)Marcapasso cardíaco; Localizador apical eletrônico; Interferência eletromagnéticaEndodontia, Cardiologia, Marcapasso Cardíaco Artificial,Introduction: Electronic Apex Locators (EAL) may interfere with Cardiac Pacemakers (PM) by electromangnetic interference (EMI). Objective: To evaluate, in vitro, the EMI caused by EAL on a PM. Methods: Four different EAL (Apex DSP, NovApex N-21, RomiApex A-15, C-ROOT I) had their electrodes immersed in a saline solution at distances of 10, 15, 20, 25 and 30 cm from the electrode of the PM (Medtronic Sensia SEDR01) monitored by telemetry. Results: With 0.2 mV sensitivity, the PM had interference from all 4 EAL, with inhibition, stimulation recycle, and deflagration. With 2mV sensitivity, no interference was observed. Regardless the PM sensitivity, no EAL was able to lead, even at short distances, reversion to asynchronous or deprogramming. Conclusion: When set to maximum sensitivity, the PM suffered interference, although that wouldn’t be a real in vivo situation.Introdução: Localizadores Apicais Eletrônicos (LAE) podem interferir com Marcapasso Cardíaco (MP) por gerarem interferência eletromagnética (IEM). Objetivo: Avaliar, in vitro, a IEM causada por LAE em um MP. Métodos: Quatro LAE (Apex DSP, NovApex N-21, RomiApex A-15, C-ROOT I) tiveram seus eletrodos imersos em uma solução salina a distâncias de 10, 15, 20, 25 e 30 cm do eletrodo de um MP (Medtronic Sensia SEDR01) monitorado por telemetria. Resultados: Com 0,2mV de sensibilidade, o MP sofreu interferência dos 4 LAE, apresentando inibição, reinício do ciclo ou deflagração. Com 2mV de sensibilidade, não houve interferências. Independente da sensibilidade do MP, nenhum LAE provocou reversão para assíncrono ou desprogramação. Conclusão: Quando em máxima sensibilidade, o MP sofreu interferências, porém esta não é uma situação observada in vivo.Associação Brasileira de Odontologia - Seção Rio de Janeiro2013-06-24info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionPesquisa Laboratorialapplication/pdfhttps://revista.aborj.org.br/index.php/rbo/article/view/37210.18363/rbo.v69n2.p.260Revista Brasileira de Odontologia; v. 69, n. 2 (2012); 260Revista Brasileira de Odontologia; v. 69, n. 2 (2012); 2601984-3747reponame:Revista Brasileira de Odontologiainstname:Associação Brasileira de Odontologia (ABO)instacron:ABOporhttps://revista.aborj.org.br/index.php/rbo/article/view/372/324Juiz de Fora; Brasil; AméricaContemporâneaDireitos autorais 2016 Revistasinfo:eu-repo/semantics/openAccessBrito, Daniel IbrahimDaibert, Frederico KleinsorgeMedeiros, Álvaro Augusto Machado deEgidio, Jhonny FerrazSantos, Patrícia de CarvalhoGouvêa, Polyana Vitoi Barbosa2013-07-22T15:16:41Zoai:revista_aborj_org_br.www.revista.aborj.org.br:article/372Revistahttps://revista.aborj.org.br/index.php/rboONGhttp://www.revista.aborj.org.br/index.php/rbo/oaihttps://revista.aborj.org.br/index.php/rbo/oai0034-72721984-3747opendoar:2013-07-22T15:16:41Revista Brasileira de Odontologia - Associação Brasileira de Odontologia (ABO)false |
dc.title.none.fl_str_mv |
In vitro interference of Electronic Apex Locator on Implantable Cardiac Pacemaker Interferência in vitro do localizador apical eletrônico em marcapasso cardíaco implantável |
title |
In vitro interference of Electronic Apex Locator on Implantable Cardiac Pacemaker |
spellingShingle |
In vitro interference of Electronic Apex Locator on Implantable Cardiac Pacemaker Brito, Daniel Ibrahim MEDICINA DENTÁRIA (ODONTOLOGIA) Marcapasso cardíaco; Localizador apical eletrônico; Interferência eletromagnética Endodontia, Cardiologia, Marcapasso Cardíaco Artificial, |
title_short |
In vitro interference of Electronic Apex Locator on Implantable Cardiac Pacemaker |
title_full |
In vitro interference of Electronic Apex Locator on Implantable Cardiac Pacemaker |
title_fullStr |
In vitro interference of Electronic Apex Locator on Implantable Cardiac Pacemaker |
title_full_unstemmed |
In vitro interference of Electronic Apex Locator on Implantable Cardiac Pacemaker |
title_sort |
In vitro interference of Electronic Apex Locator on Implantable Cardiac Pacemaker |
author |
Brito, Daniel Ibrahim |
author_facet |
Brito, Daniel Ibrahim Daibert, Frederico Kleinsorge Medeiros, Álvaro Augusto Machado de Egidio, Jhonny Ferraz Santos, Patrícia de Carvalho Gouvêa, Polyana Vitoi Barbosa |
author_role |
author |
author2 |
Daibert, Frederico Kleinsorge Medeiros, Álvaro Augusto Machado de Egidio, Jhonny Ferraz Santos, Patrícia de Carvalho Gouvêa, Polyana Vitoi Barbosa |
author2_role |
author author author author author |
dc.contributor.author.fl_str_mv |
Brito, Daniel Ibrahim Daibert, Frederico Kleinsorge Medeiros, Álvaro Augusto Machado de Egidio, Jhonny Ferraz Santos, Patrícia de Carvalho Gouvêa, Polyana Vitoi Barbosa |
dc.subject.por.fl_str_mv |
MEDICINA DENTÁRIA (ODONTOLOGIA) Marcapasso cardíaco; Localizador apical eletrônico; Interferência eletromagnética Endodontia, Cardiologia, Marcapasso Cardíaco Artificial, |
topic |
MEDICINA DENTÁRIA (ODONTOLOGIA) Marcapasso cardíaco; Localizador apical eletrônico; Interferência eletromagnética Endodontia, Cardiologia, Marcapasso Cardíaco Artificial, |
description |
Introduction: Electronic Apex Locators (EAL) may interfere with Cardiac Pacemakers (PM) by electromangnetic interference (EMI). Objective: To evaluate, in vitro, the EMI caused by EAL on a PM. Methods: Four different EAL (Apex DSP, NovApex N-21, RomiApex A-15, C-ROOT I) had their electrodes immersed in a saline solution at distances of 10, 15, 20, 25 and 30 cm from the electrode of the PM (Medtronic Sensia SEDR01) monitored by telemetry. Results: With 0.2 mV sensitivity, the PM had interference from all 4 EAL, with inhibition, stimulation recycle, and deflagration. With 2mV sensitivity, no interference was observed. Regardless the PM sensitivity, no EAL was able to lead, even at short distances, reversion to asynchronous or deprogramming. Conclusion: When set to maximum sensitivity, the PM suffered interference, although that wouldn’t be a real in vivo situation. |
publishDate |
2013 |
dc.date.none.fl_str_mv |
2013-06-24 |
dc.type.none.fl_str_mv |
|
dc.type.driver.fl_str_mv |
info:eu-repo/semantics/article info:eu-repo/semantics/publishedVersion Pesquisa Laboratorial |
format |
article |
status_str |
publishedVersion |
dc.identifier.uri.fl_str_mv |
https://revista.aborj.org.br/index.php/rbo/article/view/372 10.18363/rbo.v69n2.p.260 |
url |
https://revista.aborj.org.br/index.php/rbo/article/view/372 |
identifier_str_mv |
10.18363/rbo.v69n2.p.260 |
dc.language.iso.fl_str_mv |
por |
language |
por |
dc.relation.none.fl_str_mv |
https://revista.aborj.org.br/index.php/rbo/article/view/372/324 |
dc.rights.driver.fl_str_mv |
Direitos autorais 2016 Revistas info:eu-repo/semantics/openAccess |
rights_invalid_str_mv |
Direitos autorais 2016 Revistas |
eu_rights_str_mv |
openAccess |
dc.format.none.fl_str_mv |
application/pdf |
dc.coverage.none.fl_str_mv |
Juiz de Fora; Brasil; América Contemporânea |
dc.publisher.none.fl_str_mv |
Associação Brasileira de Odontologia - Seção Rio de Janeiro |
publisher.none.fl_str_mv |
Associação Brasileira de Odontologia - Seção Rio de Janeiro |
dc.source.none.fl_str_mv |
Revista Brasileira de Odontologia; v. 69, n. 2 (2012); 260 Revista Brasileira de Odontologia; v. 69, n. 2 (2012); 260 1984-3747 reponame:Revista Brasileira de Odontologia instname:Associação Brasileira de Odontologia (ABO) instacron:ABO |
instname_str |
Associação Brasileira de Odontologia (ABO) |
instacron_str |
ABO |
institution |
ABO |
reponame_str |
Revista Brasileira de Odontologia |
collection |
Revista Brasileira de Odontologia |
repository.name.fl_str_mv |
Revista Brasileira de Odontologia - Associação Brasileira de Odontologia (ABO) |
repository.mail.fl_str_mv |
https://revista.aborj.org.br/index.php/rbo/oai |
_version_ |
1798313205614772224 |