Evaluation of the rupture of silicone breast implants by mammography, ultrasonography and magnetic resonance imaging in asymptomatic patients: correlation with surgical findings

Detalhes bibliográficos
Autor(a) principal: Scaranelo,Anabel Medeiros
Data de Publicação: 2004
Outros Autores: Marques,Américo Ferreira, Smialowski,Elizabeth Brenda, Lederman,Henriquel Manoel
Tipo de documento: Artigo
Idioma: eng
Título da fonte: São Paulo medical journal (Online)
Texto Completo: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1516-31802004000200002
Resumo: CONTEXT: Different imaging methods can identify the integrity of breast implants and also the extent of possible silicone leakage. Mammography, ultrasonography and magnetic resonance imaging are often used to evaluate the integrity of breast implants, usually in patients that are symptomatic for rupture. A group of clinically asymptomatic patients was taken as a sample. These patients wanted to remove or change their breast implants for psychological or cosmetic reasons. OBJECTIVE: The purpose of this study was to compare the efficacy of mammography, sonography and magnetic resonance imaging in the detection of breast implant rupture in an asymptomatic population. TYPE OF STUDY: Prospective study. SETTING: Department of Diagnostic Imaging, Universidade de São Paulo, São Paulo, Brazil. METHODS: The participants were 44 asymptomatic patients who subsequently had implants surgically removed. Eighty-three implants were evaluated by both film-screen mammography and high-resolution sonography and 77 implants were evaluated by magnetic resonance imaging. The sensitivity and specificity of mammography, ultrasonography and magnetic resonance imaging were assessed using predetermined diagnostic criteria for implant rupture. All radiological signs were discussed and false positives and false negatives were retrospectively evaluated to identify the pitfalls in the investigations. RESULTS: The respective sensitivity and specificity of mammography were 20% and 89%; sonography, 30% and 81%; and magnetic resonance imaging, 64% and 77%. The differences between patients with breast implants for cosmetic and oncological reasons were discussed. CONCLUSIONS: Our experience suggests that magnetic resonance imaging seems to be the best imaging method on its own for the evaluation of rupturing among asymptomatic patients.
id APM-1_e730e1f0077034c2a2fc2efa03235e87
oai_identifier_str oai:scielo:S1516-31802004000200002
network_acronym_str APM-1
network_name_str São Paulo medical journal (Online)
repository_id_str
spelling Evaluation of the rupture of silicone breast implants by mammography, ultrasonography and magnetic resonance imaging in asymptomatic patients: correlation with surgical findingsSiliconeBreast implantsRuptureMammographyUltrasonographyMagnetic resonance imagingCONTEXT: Different imaging methods can identify the integrity of breast implants and also the extent of possible silicone leakage. Mammography, ultrasonography and magnetic resonance imaging are often used to evaluate the integrity of breast implants, usually in patients that are symptomatic for rupture. A group of clinically asymptomatic patients was taken as a sample. These patients wanted to remove or change their breast implants for psychological or cosmetic reasons. OBJECTIVE: The purpose of this study was to compare the efficacy of mammography, sonography and magnetic resonance imaging in the detection of breast implant rupture in an asymptomatic population. TYPE OF STUDY: Prospective study. SETTING: Department of Diagnostic Imaging, Universidade de São Paulo, São Paulo, Brazil. METHODS: The participants were 44 asymptomatic patients who subsequently had implants surgically removed. Eighty-three implants were evaluated by both film-screen mammography and high-resolution sonography and 77 implants were evaluated by magnetic resonance imaging. The sensitivity and specificity of mammography, ultrasonography and magnetic resonance imaging were assessed using predetermined diagnostic criteria for implant rupture. All radiological signs were discussed and false positives and false negatives were retrospectively evaluated to identify the pitfalls in the investigations. RESULTS: The respective sensitivity and specificity of mammography were 20% and 89%; sonography, 30% and 81%; and magnetic resonance imaging, 64% and 77%. The differences between patients with breast implants for cosmetic and oncological reasons were discussed. CONCLUSIONS: Our experience suggests that magnetic resonance imaging seems to be the best imaging method on its own for the evaluation of rupturing among asymptomatic patients.Associação Paulista de Medicina - APM2004-04-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S1516-31802004000200002Sao Paulo Medical Journal v.122 n.2 2004reponame:São Paulo medical journal (Online)instname:Associação Paulista de Medicinainstacron:APM10.1590/S1516-31802004000200002info:eu-repo/semantics/openAccessScaranelo,Anabel MedeirosMarques,Américo FerreiraSmialowski,Elizabeth BrendaLederman,Henriquel Manoeleng2004-07-05T00:00:00Zoai:scielo:S1516-31802004000200002Revistahttp://www.scielo.br/spmjhttps://old.scielo.br/oai/scielo-oai.phprevistas@apm.org.br1806-94601516-3180opendoar:2004-07-05T00:00São Paulo medical journal (Online) - Associação Paulista de Medicinafalse
dc.title.none.fl_str_mv Evaluation of the rupture of silicone breast implants by mammography, ultrasonography and magnetic resonance imaging in asymptomatic patients: correlation with surgical findings
title Evaluation of the rupture of silicone breast implants by mammography, ultrasonography and magnetic resonance imaging in asymptomatic patients: correlation with surgical findings
spellingShingle Evaluation of the rupture of silicone breast implants by mammography, ultrasonography and magnetic resonance imaging in asymptomatic patients: correlation with surgical findings
Scaranelo,Anabel Medeiros
Silicone
Breast implants
Rupture
Mammography
Ultrasonography
Magnetic resonance imaging
title_short Evaluation of the rupture of silicone breast implants by mammography, ultrasonography and magnetic resonance imaging in asymptomatic patients: correlation with surgical findings
title_full Evaluation of the rupture of silicone breast implants by mammography, ultrasonography and magnetic resonance imaging in asymptomatic patients: correlation with surgical findings
title_fullStr Evaluation of the rupture of silicone breast implants by mammography, ultrasonography and magnetic resonance imaging in asymptomatic patients: correlation with surgical findings
title_full_unstemmed Evaluation of the rupture of silicone breast implants by mammography, ultrasonography and magnetic resonance imaging in asymptomatic patients: correlation with surgical findings
title_sort Evaluation of the rupture of silicone breast implants by mammography, ultrasonography and magnetic resonance imaging in asymptomatic patients: correlation with surgical findings
author Scaranelo,Anabel Medeiros
author_facet Scaranelo,Anabel Medeiros
Marques,Américo Ferreira
Smialowski,Elizabeth Brenda
Lederman,Henriquel Manoel
author_role author
author2 Marques,Américo Ferreira
Smialowski,Elizabeth Brenda
Lederman,Henriquel Manoel
author2_role author
author
author
dc.contributor.author.fl_str_mv Scaranelo,Anabel Medeiros
Marques,Américo Ferreira
Smialowski,Elizabeth Brenda
Lederman,Henriquel Manoel
dc.subject.por.fl_str_mv Silicone
Breast implants
Rupture
Mammography
Ultrasonography
Magnetic resonance imaging
topic Silicone
Breast implants
Rupture
Mammography
Ultrasonography
Magnetic resonance imaging
description CONTEXT: Different imaging methods can identify the integrity of breast implants and also the extent of possible silicone leakage. Mammography, ultrasonography and magnetic resonance imaging are often used to evaluate the integrity of breast implants, usually in patients that are symptomatic for rupture. A group of clinically asymptomatic patients was taken as a sample. These patients wanted to remove or change their breast implants for psychological or cosmetic reasons. OBJECTIVE: The purpose of this study was to compare the efficacy of mammography, sonography and magnetic resonance imaging in the detection of breast implant rupture in an asymptomatic population. TYPE OF STUDY: Prospective study. SETTING: Department of Diagnostic Imaging, Universidade de São Paulo, São Paulo, Brazil. METHODS: The participants were 44 asymptomatic patients who subsequently had implants surgically removed. Eighty-three implants were evaluated by both film-screen mammography and high-resolution sonography and 77 implants were evaluated by magnetic resonance imaging. The sensitivity and specificity of mammography, ultrasonography and magnetic resonance imaging were assessed using predetermined diagnostic criteria for implant rupture. All radiological signs were discussed and false positives and false negatives were retrospectively evaluated to identify the pitfalls in the investigations. RESULTS: The respective sensitivity and specificity of mammography were 20% and 89%; sonography, 30% and 81%; and magnetic resonance imaging, 64% and 77%. The differences between patients with breast implants for cosmetic and oncological reasons were discussed. CONCLUSIONS: Our experience suggests that magnetic resonance imaging seems to be the best imaging method on its own for the evaluation of rupturing among asymptomatic patients.
publishDate 2004
dc.date.none.fl_str_mv 2004-04-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1516-31802004000200002
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1516-31802004000200002
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1590/S1516-31802004000200002
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Associação Paulista de Medicina - APM
publisher.none.fl_str_mv Associação Paulista de Medicina - APM
dc.source.none.fl_str_mv Sao Paulo Medical Journal v.122 n.2 2004
reponame:São Paulo medical journal (Online)
instname:Associação Paulista de Medicina
instacron:APM
instname_str Associação Paulista de Medicina
instacron_str APM
institution APM
reponame_str São Paulo medical journal (Online)
collection São Paulo medical journal (Online)
repository.name.fl_str_mv São Paulo medical journal (Online) - Associação Paulista de Medicina
repository.mail.fl_str_mv revistas@apm.org.br
_version_ 1754209260839895040