The performance of an in-house loop-mediated isothermal amplification for the rapid detection of Mycobacterium tuberculosis in sputum samples in comparison with Xpert MTB/RIF, microscopy and culture

Detalhes bibliográficos
Autor(a) principal: Phetsuksiri, Benjawan
Data de Publicação: 2020
Outros Autores: Klayut, Wiphat, Rudeeaneksin, Janisara, Srisungngam, Sopa, Bunchoo, Supranee, Toonkomdang, Sarawut, Wongchai, Thanee, Nakajima, Chie, Suzuki, Yasuhiko
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Revista do Instituto de Medicina Tropical de São Paulo
Texto Completo: https://www.revistas.usp.br/rimtsp/article/view/172367
Resumo: Simple, low-cost and effective diagnostic tests for tuberculosis (TB) are needed especially in TB-high burden settings. The present study evaluated the performance of an in-house loop-mediated isothermal amplification (LAMP) for diagnosing TB by comparing it to Xpert MTB/RIF, microscopy and culture. In Thailand, a total of 204 excess sputum samples volume after the processing of cultures were used for Mycobacterium tuberculosis (MTB) detection by Xpert MTB/RIF and LAMP. Based on culture results as the gold standard, the overall sensitivity of LAMP and Xpert MTB/RIF were 82.1% (126/153; 95% confidential interval [CI]: 75.4-88.98%) and 86.9 % (133/153; 95% CI: 80.5-90.8%) respectively, and the specificity of both tests was 100% (51/51; 95% CI: 93.0-100.0%). In comparison with Xpert MTB/RIF, the sensitivity and specificity of LAMP were 94.7% (126/133; 95% CI: 89.5-97.9%), and 100.0% (73/73; 95% CI: 94.9-100.0%), respectively. The average threshold cycle (Ct) of Xpert MTB/RIF detection for positive and negative LAMP results was statistically different, of 18.4 and 27.0, respectively (p < 0.05). In comparison with the acid-fast staining technique, and analyzing LAMP and Xpert MTB/RIF in smear-negative/ culture-positive specimens, there was an increase of the detection rate by 47.7% (21/44) and 54.6% (24/44). The diagnostic sensitivity and specificity of LAMP appeared to be comparable to those of Xpert MTB/RIF. We claim that this LAMP has potential to provide a sensitive diagnostic test for the rapid TB diagnosis. It allowed a fast detection of MTB before the cultures and it could be used in resource-limited laboratory settings.
id IMT-1_010bcf3b37a34fc330465889b00bc28f
oai_identifier_str oai:revistas.usp.br:article/172367
network_acronym_str IMT-1
network_name_str Revista do Instituto de Medicina Tropical de São Paulo
repository_id_str
spelling The performance of an in-house loop-mediated isothermal amplification for the rapid detection of Mycobacterium tuberculosis in sputum samples in comparison with Xpert MTB/RIF, microscopy and cultureLAMPTuberculosisDiagnosisSputumSimple, low-cost and effective diagnostic tests for tuberculosis (TB) are needed especially in TB-high burden settings. The present study evaluated the performance of an in-house loop-mediated isothermal amplification (LAMP) for diagnosing TB by comparing it to Xpert MTB/RIF, microscopy and culture. In Thailand, a total of 204 excess sputum samples volume after the processing of cultures were used for Mycobacterium tuberculosis (MTB) detection by Xpert MTB/RIF and LAMP. Based on culture results as the gold standard, the overall sensitivity of LAMP and Xpert MTB/RIF were 82.1% (126/153; 95% confidential interval [CI]: 75.4-88.98%) and 86.9 % (133/153; 95% CI: 80.5-90.8%) respectively, and the specificity of both tests was 100% (51/51; 95% CI: 93.0-100.0%). In comparison with Xpert MTB/RIF, the sensitivity and specificity of LAMP were 94.7% (126/133; 95% CI: 89.5-97.9%), and 100.0% (73/73; 95% CI: 94.9-100.0%), respectively. The average threshold cycle (Ct) of Xpert MTB/RIF detection for positive and negative LAMP results was statistically different, of 18.4 and 27.0, respectively (p < 0.05). In comparison with the acid-fast staining technique, and analyzing LAMP and Xpert MTB/RIF in smear-negative/ culture-positive specimens, there was an increase of the detection rate by 47.7% (21/44) and 54.6% (24/44). The diagnostic sensitivity and specificity of LAMP appeared to be comparable to those of Xpert MTB/RIF. We claim that this LAMP has potential to provide a sensitive diagnostic test for the rapid TB diagnosis. It allowed a fast detection of MTB before the cultures and it could be used in resource-limited laboratory settings.Universidade de São Paulo. Instituto de Medicina Tropical de São Paulo2020-07-13info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionapplication/pdfapplication/xmlhttps://www.revistas.usp.br/rimtsp/article/view/17236710.1590/s1678-9946202062036Revista do Instituto de Medicina Tropical de São Paulo; Vol. 62 (2020); e36Revista do Instituto de Medicina Tropical de São Paulo; Vol. 62 (2020); e36Revista do Instituto de Medicina Tropical de São Paulo; v. 62 (2020); e361678-99460036-4665reponame:Revista do Instituto de Medicina Tropical de São Pauloinstname:Instituto de Medicina Tropical (IMT)instacron:IMTenghttps://www.revistas.usp.br/rimtsp/article/view/172367/161985https://www.revistas.usp.br/rimtsp/article/view/172367/161986Copyright (c) 2020 Revista do Instituto de Medicina Tropical de São Pauloinfo:eu-repo/semantics/openAccessPhetsuksiri, BenjawanKlayut, WiphatRudeeaneksin, JanisaraSrisungngam, SopaBunchoo, SupraneeToonkomdang, SarawutWongchai, ThaneeNakajima, ChieSuzuki, Yasuhiko2020-10-26T17:39:25Zoai:revistas.usp.br:article/172367Revistahttp://www.revistas.usp.br/rimtsp/indexPUBhttps://www.revistas.usp.br/rimtsp/oai||revimtsp@usp.br1678-99460036-4665opendoar:2022-12-13T16:52:52.606616Revista do Instituto de Medicina Tropical de São Paulo - Instituto de Medicina Tropical (IMT)true
dc.title.none.fl_str_mv The performance of an in-house loop-mediated isothermal amplification for the rapid detection of Mycobacterium tuberculosis in sputum samples in comparison with Xpert MTB/RIF, microscopy and culture
title The performance of an in-house loop-mediated isothermal amplification for the rapid detection of Mycobacterium tuberculosis in sputum samples in comparison with Xpert MTB/RIF, microscopy and culture
spellingShingle The performance of an in-house loop-mediated isothermal amplification for the rapid detection of Mycobacterium tuberculosis in sputum samples in comparison with Xpert MTB/RIF, microscopy and culture
Phetsuksiri, Benjawan
LAMP
Tuberculosis
Diagnosis
Sputum
title_short The performance of an in-house loop-mediated isothermal amplification for the rapid detection of Mycobacterium tuberculosis in sputum samples in comparison with Xpert MTB/RIF, microscopy and culture
title_full The performance of an in-house loop-mediated isothermal amplification for the rapid detection of Mycobacterium tuberculosis in sputum samples in comparison with Xpert MTB/RIF, microscopy and culture
title_fullStr The performance of an in-house loop-mediated isothermal amplification for the rapid detection of Mycobacterium tuberculosis in sputum samples in comparison with Xpert MTB/RIF, microscopy and culture
title_full_unstemmed The performance of an in-house loop-mediated isothermal amplification for the rapid detection of Mycobacterium tuberculosis in sputum samples in comparison with Xpert MTB/RIF, microscopy and culture
title_sort The performance of an in-house loop-mediated isothermal amplification for the rapid detection of Mycobacterium tuberculosis in sputum samples in comparison with Xpert MTB/RIF, microscopy and culture
author Phetsuksiri, Benjawan
author_facet Phetsuksiri, Benjawan
Klayut, Wiphat
Rudeeaneksin, Janisara
Srisungngam, Sopa
Bunchoo, Supranee
Toonkomdang, Sarawut
Wongchai, Thanee
Nakajima, Chie
Suzuki, Yasuhiko
author_role author
author2 Klayut, Wiphat
Rudeeaneksin, Janisara
Srisungngam, Sopa
Bunchoo, Supranee
Toonkomdang, Sarawut
Wongchai, Thanee
Nakajima, Chie
Suzuki, Yasuhiko
author2_role author
author
author
author
author
author
author
author
dc.contributor.author.fl_str_mv Phetsuksiri, Benjawan
Klayut, Wiphat
Rudeeaneksin, Janisara
Srisungngam, Sopa
Bunchoo, Supranee
Toonkomdang, Sarawut
Wongchai, Thanee
Nakajima, Chie
Suzuki, Yasuhiko
dc.subject.por.fl_str_mv LAMP
Tuberculosis
Diagnosis
Sputum
topic LAMP
Tuberculosis
Diagnosis
Sputum
description Simple, low-cost and effective diagnostic tests for tuberculosis (TB) are needed especially in TB-high burden settings. The present study evaluated the performance of an in-house loop-mediated isothermal amplification (LAMP) for diagnosing TB by comparing it to Xpert MTB/RIF, microscopy and culture. In Thailand, a total of 204 excess sputum samples volume after the processing of cultures were used for Mycobacterium tuberculosis (MTB) detection by Xpert MTB/RIF and LAMP. Based on culture results as the gold standard, the overall sensitivity of LAMP and Xpert MTB/RIF were 82.1% (126/153; 95% confidential interval [CI]: 75.4-88.98%) and 86.9 % (133/153; 95% CI: 80.5-90.8%) respectively, and the specificity of both tests was 100% (51/51; 95% CI: 93.0-100.0%). In comparison with Xpert MTB/RIF, the sensitivity and specificity of LAMP were 94.7% (126/133; 95% CI: 89.5-97.9%), and 100.0% (73/73; 95% CI: 94.9-100.0%), respectively. The average threshold cycle (Ct) of Xpert MTB/RIF detection for positive and negative LAMP results was statistically different, of 18.4 and 27.0, respectively (p < 0.05). In comparison with the acid-fast staining technique, and analyzing LAMP and Xpert MTB/RIF in smear-negative/ culture-positive specimens, there was an increase of the detection rate by 47.7% (21/44) and 54.6% (24/44). The diagnostic sensitivity and specificity of LAMP appeared to be comparable to those of Xpert MTB/RIF. We claim that this LAMP has potential to provide a sensitive diagnostic test for the rapid TB diagnosis. It allowed a fast detection of MTB before the cultures and it could be used in resource-limited laboratory settings.
publishDate 2020
dc.date.none.fl_str_mv 2020-07-13
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv https://www.revistas.usp.br/rimtsp/article/view/172367
10.1590/s1678-9946202062036
url https://www.revistas.usp.br/rimtsp/article/view/172367
identifier_str_mv 10.1590/s1678-9946202062036
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv https://www.revistas.usp.br/rimtsp/article/view/172367/161985
https://www.revistas.usp.br/rimtsp/article/view/172367/161986
dc.rights.driver.fl_str_mv Copyright (c) 2020 Revista do Instituto de Medicina Tropical de São Paulo
info:eu-repo/semantics/openAccess
rights_invalid_str_mv Copyright (c) 2020 Revista do Instituto de Medicina Tropical de São Paulo
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
application/xml
dc.publisher.none.fl_str_mv Universidade de São Paulo. Instituto de Medicina Tropical de São Paulo
publisher.none.fl_str_mv Universidade de São Paulo. Instituto de Medicina Tropical de São Paulo
dc.source.none.fl_str_mv Revista do Instituto de Medicina Tropical de São Paulo; Vol. 62 (2020); e36
Revista do Instituto de Medicina Tropical de São Paulo; Vol. 62 (2020); e36
Revista do Instituto de Medicina Tropical de São Paulo; v. 62 (2020); e36
1678-9946
0036-4665
reponame:Revista do Instituto de Medicina Tropical de São Paulo
instname:Instituto de Medicina Tropical (IMT)
instacron:IMT
instname_str Instituto de Medicina Tropical (IMT)
instacron_str IMT
institution IMT
reponame_str Revista do Instituto de Medicina Tropical de São Paulo
collection Revista do Instituto de Medicina Tropical de São Paulo
repository.name.fl_str_mv Revista do Instituto de Medicina Tropical de São Paulo - Instituto de Medicina Tropical (IMT)
repository.mail.fl_str_mv ||revimtsp@usp.br
_version_ 1798951652698357760