INFLUENCE OF DIAMETER AND GEOMETRY IN THE TAPPING OF THE PILOT HOLE IN PEDICLE SCREWS
Autor(a) principal: | |
---|---|
Data de Publicação: | 2019 |
Outros Autores: | , , |
Tipo de documento: | Artigo |
Idioma: | eng |
Título da fonte: | Coluna/Columna |
Texto Completo: | http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1808-18512019000100051 |
Resumo: | ABSTRACT Objective: To evaluate the insertion torque and the pulling force of each screw with different diameters and tap. Methods: Polyurethane blocks with a pilot hole of 2.7 mm were used in the study. An experimental group with 5 blocks was formed, the insertion torque was evaluated with a torque meter, and the pullout strength of each Globus screw of 5.5 mm and 6.5 mm was assessed. Results: The comparison of the insertion torque on the 5.5 mm screws with pilot hole without tapping and with a smaller diameter than that of the screw (4.5 mm) and a different thread, and with the tapping with the same diameter as that of the screw (5.5 mm) and equal or different thread presented a statistical difference with a higher value of the insertion torque in the group in which the tapping was not performed. As for the pulling force of the 5.5 mm screw, the non-tapping of the pilot hole resulted in statistical difference with the same diameter of the screw (5.5 mm) and with a different thread of the screw. The pullout force on the 6.5 mm screw was higher in the group where the pilot hole was not tapped according to the non-parametric Kruskal-Wallis test, with significance level of p <0.05 in the comparison of the groups. Conclusions: Pilot hole tapping reduced insertion torque and pullout resistance of the pedicle screw influencing the fixation with tapping with the same screw diameter and different thread design. |
id |
SBCO-1_0941d3ce4baa6b63f75e0db8aec84bc4 |
---|---|
oai_identifier_str |
oai:scielo:S1808-18512019000100051 |
network_acronym_str |
SBCO-1 |
network_name_str |
Coluna/Columna |
repository_id_str |
|
spelling |
INFLUENCE OF DIAMETER AND GEOMETRY IN THE TAPPING OF THE PILOT HOLE IN PEDICLE SCREWSSpineBone screwsSpinal fusionABSTRACT Objective: To evaluate the insertion torque and the pulling force of each screw with different diameters and tap. Methods: Polyurethane blocks with a pilot hole of 2.7 mm were used in the study. An experimental group with 5 blocks was formed, the insertion torque was evaluated with a torque meter, and the pullout strength of each Globus screw of 5.5 mm and 6.5 mm was assessed. Results: The comparison of the insertion torque on the 5.5 mm screws with pilot hole without tapping and with a smaller diameter than that of the screw (4.5 mm) and a different thread, and with the tapping with the same diameter as that of the screw (5.5 mm) and equal or different thread presented a statistical difference with a higher value of the insertion torque in the group in which the tapping was not performed. As for the pulling force of the 5.5 mm screw, the non-tapping of the pilot hole resulted in statistical difference with the same diameter of the screw (5.5 mm) and with a different thread of the screw. The pullout force on the 6.5 mm screw was higher in the group where the pilot hole was not tapped according to the non-parametric Kruskal-Wallis test, with significance level of p <0.05 in the comparison of the groups. Conclusions: Pilot hole tapping reduced insertion torque and pullout resistance of the pedicle screw influencing the fixation with tapping with the same screw diameter and different thread design.Sociedade Brasileira de Coluna2019-03-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S1808-18512019000100051Coluna/Columna v.18 n.1 2019reponame:Coluna/Columnainstname:Sociedade Brasileira de Coluna (SBCO)instacron:SBCO10.1590/s1808-185120191801195390info:eu-repo/semantics/openAccessDefino,Helton Luiz AparecidoMiranda,Raffaello de FreitasPinheiro,Romulo PedrozaShimano,Antonio Carloseng2019-02-19T00:00:00Zoai:scielo:S1808-18512019000100051Revistahttps://www.revistacoluna.org/ONGhttps://old.scielo.br/oai/scielo-oai.phpcoluna.columna@uol.com.br||revistacoluna@uol.com.br2177-014X1808-1851opendoar:2019-02-19T00:00Coluna/Columna - Sociedade Brasileira de Coluna (SBCO)false |
dc.title.none.fl_str_mv |
INFLUENCE OF DIAMETER AND GEOMETRY IN THE TAPPING OF THE PILOT HOLE IN PEDICLE SCREWS |
title |
INFLUENCE OF DIAMETER AND GEOMETRY IN THE TAPPING OF THE PILOT HOLE IN PEDICLE SCREWS |
spellingShingle |
INFLUENCE OF DIAMETER AND GEOMETRY IN THE TAPPING OF THE PILOT HOLE IN PEDICLE SCREWS Defino,Helton Luiz Aparecido Spine Bone screws Spinal fusion |
title_short |
INFLUENCE OF DIAMETER AND GEOMETRY IN THE TAPPING OF THE PILOT HOLE IN PEDICLE SCREWS |
title_full |
INFLUENCE OF DIAMETER AND GEOMETRY IN THE TAPPING OF THE PILOT HOLE IN PEDICLE SCREWS |
title_fullStr |
INFLUENCE OF DIAMETER AND GEOMETRY IN THE TAPPING OF THE PILOT HOLE IN PEDICLE SCREWS |
title_full_unstemmed |
INFLUENCE OF DIAMETER AND GEOMETRY IN THE TAPPING OF THE PILOT HOLE IN PEDICLE SCREWS |
title_sort |
INFLUENCE OF DIAMETER AND GEOMETRY IN THE TAPPING OF THE PILOT HOLE IN PEDICLE SCREWS |
author |
Defino,Helton Luiz Aparecido |
author_facet |
Defino,Helton Luiz Aparecido Miranda,Raffaello de Freitas Pinheiro,Romulo Pedroza Shimano,Antonio Carlos |
author_role |
author |
author2 |
Miranda,Raffaello de Freitas Pinheiro,Romulo Pedroza Shimano,Antonio Carlos |
author2_role |
author author author |
dc.contributor.author.fl_str_mv |
Defino,Helton Luiz Aparecido Miranda,Raffaello de Freitas Pinheiro,Romulo Pedroza Shimano,Antonio Carlos |
dc.subject.por.fl_str_mv |
Spine Bone screws Spinal fusion |
topic |
Spine Bone screws Spinal fusion |
description |
ABSTRACT Objective: To evaluate the insertion torque and the pulling force of each screw with different diameters and tap. Methods: Polyurethane blocks with a pilot hole of 2.7 mm were used in the study. An experimental group with 5 blocks was formed, the insertion torque was evaluated with a torque meter, and the pullout strength of each Globus screw of 5.5 mm and 6.5 mm was assessed. Results: The comparison of the insertion torque on the 5.5 mm screws with pilot hole without tapping and with a smaller diameter than that of the screw (4.5 mm) and a different thread, and with the tapping with the same diameter as that of the screw (5.5 mm) and equal or different thread presented a statistical difference with a higher value of the insertion torque in the group in which the tapping was not performed. As for the pulling force of the 5.5 mm screw, the non-tapping of the pilot hole resulted in statistical difference with the same diameter of the screw (5.5 mm) and with a different thread of the screw. The pullout force on the 6.5 mm screw was higher in the group where the pilot hole was not tapped according to the non-parametric Kruskal-Wallis test, with significance level of p <0.05 in the comparison of the groups. Conclusions: Pilot hole tapping reduced insertion torque and pullout resistance of the pedicle screw influencing the fixation with tapping with the same screw diameter and different thread design. |
publishDate |
2019 |
dc.date.none.fl_str_mv |
2019-03-01 |
dc.type.driver.fl_str_mv |
info:eu-repo/semantics/article |
dc.type.status.fl_str_mv |
info:eu-repo/semantics/publishedVersion |
format |
article |
status_str |
publishedVersion |
dc.identifier.uri.fl_str_mv |
http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1808-18512019000100051 |
url |
http://old.scielo.br/scielo.php?script=sci_arttext&pid=S1808-18512019000100051 |
dc.language.iso.fl_str_mv |
eng |
language |
eng |
dc.relation.none.fl_str_mv |
10.1590/s1808-185120191801195390 |
dc.rights.driver.fl_str_mv |
info:eu-repo/semantics/openAccess |
eu_rights_str_mv |
openAccess |
dc.format.none.fl_str_mv |
text/html |
dc.publisher.none.fl_str_mv |
Sociedade Brasileira de Coluna |
publisher.none.fl_str_mv |
Sociedade Brasileira de Coluna |
dc.source.none.fl_str_mv |
Coluna/Columna v.18 n.1 2019 reponame:Coluna/Columna instname:Sociedade Brasileira de Coluna (SBCO) instacron:SBCO |
instname_str |
Sociedade Brasileira de Coluna (SBCO) |
instacron_str |
SBCO |
institution |
SBCO |
reponame_str |
Coluna/Columna |
collection |
Coluna/Columna |
repository.name.fl_str_mv |
Coluna/Columna - Sociedade Brasileira de Coluna (SBCO) |
repository.mail.fl_str_mv |
coluna.columna@uol.com.br||revistacoluna@uol.com.br |
_version_ |
1752126615981129728 |