Evaluation of the effect of irrigants on the accuracy and repeatability of three electronic apex locators: an ex vivo study
Autor(a) principal: | |
---|---|
Data de Publicação: | 2021 |
Outros Autores: | , , , , , |
Tipo de documento: | Artigo |
Idioma: | eng |
Título da fonte: | Research, Society and Development |
Texto Completo: | https://rsdjournal.org/index.php/rsd/article/view/21636 |
Resumo: | Objective: To evaluate the effect of 2% chlorhexidine (CHX) and 2.5% sodium hypochlorite (NaOCl) on the accuracy and repeatability of three Electronic Apex Locators (EALs). Methodology: Thirty one human teeth were connected to a platform, irrigated with 2.0% CHX, and obtained two readings, again irrigated with 2.5% NaOCl and obtained two more readings. The distance between the tip of the file and the apical constriction was measured. Repeatability was calculated using the intraclass correlation coefficient (ICC) and repeatability coefficient (CR). Results: For CHX 2%, with the Mini Apex Locator, 25 of 31 electronic measurements were accurate (80.6%), 28 of 31 measurements for iPex (90.3%) and for Root ZX II 26 (83, 9%). For 2.5% NaOCl, with the Mini Apex Locator, 25 of the 31 measurements were accurate (80.6%), 28 of the 31 measurements for the iPex (90.3%) and for the Root ZX II, 24 (77, 4%). There were no differences between the two irrigants (p>0.05). The values (ICC) were 0.975 and 0.994 for the Mini Apex Locator, 0.981 and 0.971 for the i Pex and 0.994 and 0.995 for the Root ZX II. When using CHX, the (CR) was 0.196 for the Mini Apex Locator, 0.152 for the iPex and 0.088 for the Root ZX II. When using 2.5% NaOCl, the (CR) was 0.088 for the Mini Apex Locator, 0.196 for the iPex and 0.088 for the Root ZX II. Conclusions: The use of 2% CHX and 2.5% NaOCl did not affect the accuracy and repeatability of the three devices. |
id |
UNIFEI_d7bb8b7aba24a3cfe283b879f7c52814 |
---|---|
oai_identifier_str |
oai:ojs.pkp.sfu.ca:article/21636 |
network_acronym_str |
UNIFEI |
network_name_str |
Research, Society and Development |
repository_id_str |
|
spelling |
Evaluation of the effect of irrigants on the accuracy and repeatability of three electronic apex locators: an ex vivo studyEvaluación del efecto de los irrigantes en la precisión y repetibilidad de tres localizadores apicales electrónicos: un estudio ex vitroAvaliação do efeito de irrigantes na precisão e repetibilidade de três localizadores de ápice eletrônicos: um estudo ex vivoEndodontiaTratamento do canal radicularPrecisão da medição dimensionalOdontometria.EndodonciaTratamiento de conductoPrecisión de la medición dimensionalOdontometría.EndodonticsRoot canal treatmentDimensional measurement accuraciesOdontometry.Objective: To evaluate the effect of 2% chlorhexidine (CHX) and 2.5% sodium hypochlorite (NaOCl) on the accuracy and repeatability of three Electronic Apex Locators (EALs). Methodology: Thirty one human teeth were connected to a platform, irrigated with 2.0% CHX, and obtained two readings, again irrigated with 2.5% NaOCl and obtained two more readings. The distance between the tip of the file and the apical constriction was measured. Repeatability was calculated using the intraclass correlation coefficient (ICC) and repeatability coefficient (CR). Results: For CHX 2%, with the Mini Apex Locator, 25 of 31 electronic measurements were accurate (80.6%), 28 of 31 measurements for iPex (90.3%) and for Root ZX II 26 (83, 9%). For 2.5% NaOCl, with the Mini Apex Locator, 25 of the 31 measurements were accurate (80.6%), 28 of the 31 measurements for the iPex (90.3%) and for the Root ZX II, 24 (77, 4%). There were no differences between the two irrigants (p>0.05). The values (ICC) were 0.975 and 0.994 for the Mini Apex Locator, 0.981 and 0.971 for the i Pex and 0.994 and 0.995 for the Root ZX II. When using CHX, the (CR) was 0.196 for the Mini Apex Locator, 0.152 for the iPex and 0.088 for the Root ZX II. When using 2.5% NaOCl, the (CR) was 0.088 for the Mini Apex Locator, 0.196 for the iPex and 0.088 for the Root ZX II. Conclusions: The use of 2% CHX and 2.5% NaOCl did not affect the accuracy and repeatability of the three devices.Objetivo: Evaluar el efecto de la clorhexidina (CHX) al 2% y el hipoclorito de sodio (NaOCl) al 2,5% sobre la precisión y repetibilidad de tres localizadores apicales electrónicos (LAEs). Metodología: Se conectaron 31 dientes humanos a una plataforma, se irrigaron los conductos con CHX al 2%, se obtuvieron dos lecturas, se irrigaron nuevamente con NaOCl al 2.5% y se obtuvieron dos lecturas más. Se midió la distancia entre la punta de la lima y la constricción apical. La repetibilidad se calculó utilizando el coeficiente de correlación intraclase (ICC) y el coeficiente de repetibilidad (CR). Resultados: Para la CHX al 2%, con el Mini Apex Locator, 25 de las 31 mediciones electrónicas fueron precisas (80,6%); con el iPex, 28 de las 31 mediciones fueron precisas (90,3%), y 26 com el Root ZX II (83, 9%). Para el NaOCl al 2.5%, con el Mini Apex Locator, 25 de las 31 mediciones fueron precisas (80,6%), 28 de las 31 mediciones para el iPex (90,3%) y para el Root ZX II, 24 (77, 4%). No hubo diferencias entre los dos irrigantes (p> 0.05). Los valores (ICC) fueron 0.975 y 0.994 para el Mini Apex Locator, 0.981 y 0.971 para iPex, y 0.994 a 0.995 para el Root ZX II. Al usar CHX, el (CR) fué 0.196 para el Mini Apex Locator; 0.152 para el iPex y 0.088 para el Root ZX II. Cuando se usó NaOCl al 2.5%, el (CR) fue 0.088 para el Mini Apex Locator, 0.196 para el iPex y 0.088 para el Root ZX II. Conclusiones: El uso de CHX al 2% y NaOCl al 2,5% no afectó la precisión y repetibilidad de los tres dispositivos.Objetivo: Avaliar o efeito da clorexidina 2% (CHX) e hipoclorito de sódio 2,5% (NaOCl) na precisão e repetibilidade de três Localizadores apicais eletrônicos (LAEs). Metodologia: trinta e um dentes humanos foram conectados a uma plataforma, irrigados com CHX 2,0%, e duas leituras obtidas, irrigou-se novamente com NaOCl 2,5% e obteve-se mais duas leituras. A distância entre a ponta da lima e a Constrição apical foi medida. Calculou-se a repetibilidade usando coeficiente de correlação intraclasse (ICC) e o coeficiente de repetibilidade (CR). Resultados: Para CHX 2%, com o Mini Apex Locator, 25 das 31 medidas eletrônicas foram precisas (80,6%), 28 de 31 medidas para o iPex (90,3%) e para o Root ZX II 26 (83,9%). Para NaOCl 2,5%, com o Mini Apex Locator, 25 das 31 medidas foram precisas (80,6%), 28 das 31 medidas para o iPex (90,3%) e para o Root ZX II, 24 (77,4%). Não houve diferenças entre os dois irrigantes (p> 0,05). Os valores (ICC) foram 0,975 e 0,994 para o Mini Apex Locator, 0,981 e 0,971 para o i Pex e 0,994 a 0,995 para o Root ZX II. Ao utilizar CHX, o (CR) foi 0,196 para o Mini Apex Locator, 0,152 para o iPex e 0,088 para o Root ZX II. Ao utilizar NaOCl 2,5%, o (CR) foi de 0,088 para o Mini Apex Locator, 0,196 para o iPex e 0,088 para o Root ZX II. Conclusões: O uso de CHX a 2% e NaOCl a 2,5% não afetou a precisão e repetibilidade dos três dispositivos.Research, Society and Development2021-10-22info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersionapplication/pdfhttps://rsdjournal.org/index.php/rsd/article/view/2163610.33448/rsd-v10i13.21636Research, Society and Development; Vol. 10 No. 13; e599101321636Research, Society and Development; Vol. 10 Núm. 13; e599101321636Research, Society and Development; v. 10 n. 13; e5991013216362525-3409reponame:Research, Society and Developmentinstname:Universidade Federal de Itajubá (UNIFEI)instacron:UNIFEIenghttps://rsdjournal.org/index.php/rsd/article/view/21636/19268Copyright (c) 2021 Fábio Luiz Cunha D'Assunção; Juan Ramon Salazar Silva; Karen Christine Lacerda Arnaud; Thiago Bezerra Leite; Jozemar Pereira dos Santos; Verônica Cabral dos Santos Cunha D'Assunção; Andressa Cartaxo de Almeidahttps://creativecommons.org/licenses/by/4.0info:eu-repo/semantics/openAccessD'Assunção, Fábio Luiz CunhaSalazar Silva, Juan RamonArnaud, Karen Christine Lacerda Leite, Thiago Bezerra Santos, Jozemar Pereira dos D'Assunção, Verônica Cabral dos Santos CunhaAlmeida, Andressa Cartaxo de2021-11-21T18:26:28Zoai:ojs.pkp.sfu.ca:article/21636Revistahttps://rsdjournal.org/index.php/rsd/indexPUBhttps://rsdjournal.org/index.php/rsd/oairsd.articles@gmail.com2525-34092525-3409opendoar:2024-01-17T09:41:00.848178Research, Society and Development - Universidade Federal de Itajubá (UNIFEI)false |
dc.title.none.fl_str_mv |
Evaluation of the effect of irrigants on the accuracy and repeatability of three electronic apex locators: an ex vivo study Evaluación del efecto de los irrigantes en la precisión y repetibilidad de tres localizadores apicales electrónicos: un estudio ex vitro Avaliação do efeito de irrigantes na precisão e repetibilidade de três localizadores de ápice eletrônicos: um estudo ex vivo |
title |
Evaluation of the effect of irrigants on the accuracy and repeatability of three electronic apex locators: an ex vivo study |
spellingShingle |
Evaluation of the effect of irrigants on the accuracy and repeatability of three electronic apex locators: an ex vivo study D'Assunção, Fábio Luiz Cunha Endodontia Tratamento do canal radicular Precisão da medição dimensional Odontometria. Endodoncia Tratamiento de conducto Precisión de la medición dimensional Odontometría. Endodontics Root canal treatment Dimensional measurement accuracies Odontometry. |
title_short |
Evaluation of the effect of irrigants on the accuracy and repeatability of three electronic apex locators: an ex vivo study |
title_full |
Evaluation of the effect of irrigants on the accuracy and repeatability of three electronic apex locators: an ex vivo study |
title_fullStr |
Evaluation of the effect of irrigants on the accuracy and repeatability of three electronic apex locators: an ex vivo study |
title_full_unstemmed |
Evaluation of the effect of irrigants on the accuracy and repeatability of three electronic apex locators: an ex vivo study |
title_sort |
Evaluation of the effect of irrigants on the accuracy and repeatability of three electronic apex locators: an ex vivo study |
author |
D'Assunção, Fábio Luiz Cunha |
author_facet |
D'Assunção, Fábio Luiz Cunha Salazar Silva, Juan Ramon Arnaud, Karen Christine Lacerda Leite, Thiago Bezerra Santos, Jozemar Pereira dos D'Assunção, Verônica Cabral dos Santos Cunha Almeida, Andressa Cartaxo de |
author_role |
author |
author2 |
Salazar Silva, Juan Ramon Arnaud, Karen Christine Lacerda Leite, Thiago Bezerra Santos, Jozemar Pereira dos D'Assunção, Verônica Cabral dos Santos Cunha Almeida, Andressa Cartaxo de |
author2_role |
author author author author author author |
dc.contributor.author.fl_str_mv |
D'Assunção, Fábio Luiz Cunha Salazar Silva, Juan Ramon Arnaud, Karen Christine Lacerda Leite, Thiago Bezerra Santos, Jozemar Pereira dos D'Assunção, Verônica Cabral dos Santos Cunha Almeida, Andressa Cartaxo de |
dc.subject.por.fl_str_mv |
Endodontia Tratamento do canal radicular Precisão da medição dimensional Odontometria. Endodoncia Tratamiento de conducto Precisión de la medición dimensional Odontometría. Endodontics Root canal treatment Dimensional measurement accuracies Odontometry. |
topic |
Endodontia Tratamento do canal radicular Precisão da medição dimensional Odontometria. Endodoncia Tratamiento de conducto Precisión de la medición dimensional Odontometría. Endodontics Root canal treatment Dimensional measurement accuracies Odontometry. |
description |
Objective: To evaluate the effect of 2% chlorhexidine (CHX) and 2.5% sodium hypochlorite (NaOCl) on the accuracy and repeatability of three Electronic Apex Locators (EALs). Methodology: Thirty one human teeth were connected to a platform, irrigated with 2.0% CHX, and obtained two readings, again irrigated with 2.5% NaOCl and obtained two more readings. The distance between the tip of the file and the apical constriction was measured. Repeatability was calculated using the intraclass correlation coefficient (ICC) and repeatability coefficient (CR). Results: For CHX 2%, with the Mini Apex Locator, 25 of 31 electronic measurements were accurate (80.6%), 28 of 31 measurements for iPex (90.3%) and for Root ZX II 26 (83, 9%). For 2.5% NaOCl, with the Mini Apex Locator, 25 of the 31 measurements were accurate (80.6%), 28 of the 31 measurements for the iPex (90.3%) and for the Root ZX II, 24 (77, 4%). There were no differences between the two irrigants (p>0.05). The values (ICC) were 0.975 and 0.994 for the Mini Apex Locator, 0.981 and 0.971 for the i Pex and 0.994 and 0.995 for the Root ZX II. When using CHX, the (CR) was 0.196 for the Mini Apex Locator, 0.152 for the iPex and 0.088 for the Root ZX II. When using 2.5% NaOCl, the (CR) was 0.088 for the Mini Apex Locator, 0.196 for the iPex and 0.088 for the Root ZX II. Conclusions: The use of 2% CHX and 2.5% NaOCl did not affect the accuracy and repeatability of the three devices. |
publishDate |
2021 |
dc.date.none.fl_str_mv |
2021-10-22 |
dc.type.driver.fl_str_mv |
info:eu-repo/semantics/article info:eu-repo/semantics/publishedVersion |
format |
article |
status_str |
publishedVersion |
dc.identifier.uri.fl_str_mv |
https://rsdjournal.org/index.php/rsd/article/view/21636 10.33448/rsd-v10i13.21636 |
url |
https://rsdjournal.org/index.php/rsd/article/view/21636 |
identifier_str_mv |
10.33448/rsd-v10i13.21636 |
dc.language.iso.fl_str_mv |
eng |
language |
eng |
dc.relation.none.fl_str_mv |
https://rsdjournal.org/index.php/rsd/article/view/21636/19268 |
dc.rights.driver.fl_str_mv |
https://creativecommons.org/licenses/by/4.0 info:eu-repo/semantics/openAccess |
rights_invalid_str_mv |
https://creativecommons.org/licenses/by/4.0 |
eu_rights_str_mv |
openAccess |
dc.format.none.fl_str_mv |
application/pdf |
dc.publisher.none.fl_str_mv |
Research, Society and Development |
publisher.none.fl_str_mv |
Research, Society and Development |
dc.source.none.fl_str_mv |
Research, Society and Development; Vol. 10 No. 13; e599101321636 Research, Society and Development; Vol. 10 Núm. 13; e599101321636 Research, Society and Development; v. 10 n. 13; e599101321636 2525-3409 reponame:Research, Society and Development instname:Universidade Federal de Itajubá (UNIFEI) instacron:UNIFEI |
instname_str |
Universidade Federal de Itajubá (UNIFEI) |
instacron_str |
UNIFEI |
institution |
UNIFEI |
reponame_str |
Research, Society and Development |
collection |
Research, Society and Development |
repository.name.fl_str_mv |
Research, Society and Development - Universidade Federal de Itajubá (UNIFEI) |
repository.mail.fl_str_mv |
rsd.articles@gmail.com |
_version_ |
1797052693285437440 |