Electrical Immunosensor Made with Antigenic Peptide NS5A-1 Immobilized onto Silk Fibroin for Diagnosing Hepatitis C

Detalhes bibliográficos
Autor(a) principal: Lima, Lais R. [UNESP]
Data de Publicação: 2018
Outros Autores: Goncalves, Alem-Mar B., Paulovich, Fernando, Oliveira Jr, Osvaldo N., Ribeiro, Sidney J. L. [UNESP], Moraes, Marli L.
Tipo de documento: Artigo
Idioma: eng
Título da fonte: Repositório Institucional da UNESP
Texto Completo: http://dx.doi.org/10.21577/0103-5053.20180080
http://hdl.handle.net/11449/164825
Resumo: Immunosensors based on impedance spectroscopy for diagnosing hepatitis C are reported where the sensing units were made with the antigenic peptide PPLLESWKDPDYVPPWHG (NS5A-1) derived from the NS5A protein of the hepatitis C virus (HCV) immobilized in layer-by-layer (LbL) films with silk fibroin (SF) and deposited on gold interdigitated electrodes. The electrical response of the sensing units varied upon immersion into solutions containing the antibody anti-HCV owing to the biomolecular recognition of NS5A-1. This was associated with morphological changes on the LbL films caused by adsorption of NS5A-1 and inferred from atomic force microscopy images. Buffer solutions with different anti-HCV concentrations down to 2 ng mL(-1) could be clearly distinguished by analyzing the impedance spectroscopy data with a multidimensional projection technique. The specificity toward anti-HCV antibodies was confirmed in control experiments where no significant changes in the electrical response were measured by exposing the sensing units to solutions containing an anti-human immunodeficiency virus (HIV) antibody. The high sensitivity and selectivity of the units made with LbL films demonstrate the feasibility of measuring electrical impedance as an immunosensing strategy to detect hepatitis C.
id UNSP_fcea04ccd421de8ce404215b3e3f1167
oai_identifier_str oai:repositorio.unesp.br:11449/164825
network_acronym_str UNSP
network_name_str Repositório Institucional da UNESP
repository_id_str 2946
spelling Electrical Immunosensor Made with Antigenic Peptide NS5A-1 Immobilized onto Silk Fibroin for Diagnosing Hepatitis Cimmunosensorlayer-by-layer filmantigenic peptidehepatitis Cimpedance spectroscopyImmunosensors based on impedance spectroscopy for diagnosing hepatitis C are reported where the sensing units were made with the antigenic peptide PPLLESWKDPDYVPPWHG (NS5A-1) derived from the NS5A protein of the hepatitis C virus (HCV) immobilized in layer-by-layer (LbL) films with silk fibroin (SF) and deposited on gold interdigitated electrodes. The electrical response of the sensing units varied upon immersion into solutions containing the antibody anti-HCV owing to the biomolecular recognition of NS5A-1. This was associated with morphological changes on the LbL films caused by adsorption of NS5A-1 and inferred from atomic force microscopy images. Buffer solutions with different anti-HCV concentrations down to 2 ng mL(-1) could be clearly distinguished by analyzing the impedance spectroscopy data with a multidimensional projection technique. The specificity toward anti-HCV antibodies was confirmed in control experiments where no significant changes in the electrical response were measured by exposing the sensing units to solutions containing an anti-human immunodeficiency virus (HIV) antibody. The high sensitivity and selectivity of the units made with LbL films demonstrate the feasibility of measuring electrical impedance as an immunosensing strategy to detect hepatitis C.Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq)Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (CAPES)LMF/LNNano, BrazilUniv Estadual Paulista, Inst Quim, BR-14800060 Araraquara, SP, BrazilUniv Fed Mato Grosso do Sul, Inst Fis, CP-549, BR-79070900 Campo Grande, MS, BrazilUniv Sao Paulo, Inst Ciencias Matemat & Comp, BR-13566590 Sao Carlos, SP, BrazilUniv Sao Paulo, Inst Fis Sao Carlos, BR-13566590 Sao Carlos, SP, BrazilUniv Fed Sao Paulo, Inst Ciencia & Tecnol, BR-12231280 Sao Jose Dos Campos, SP, BrazilUniv Estadual Paulista, Inst Quim, BR-14800060 Araraquara, SP, BrazilCAPES: 26LMF/LNNano, Brazil: LMF 12298Soc Brasileira QuimicaUniversidade Estadual Paulista (Unesp)Universidade Federal de Mato Grosso do Sul (UFMS)Universidade de São Paulo (USP)Universidade Federal de São Paulo (UNIFESP)Lima, Lais R. [UNESP]Goncalves, Alem-Mar B.Paulovich, FernandoOliveira Jr, Osvaldo N.Ribeiro, Sidney J. L. [UNESP]Moraes, Marli L.2018-11-26T22:40:49Z2018-11-26T22:40:49Z2018-10-01info:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/article2054-2059application/pdfhttp://dx.doi.org/10.21577/0103-5053.20180080Journal Of The Brazilian Chemical Society. Sao Paulo: Soc Brasileira Quimica, v. 29, n. 10, p. 2054-2059, 2018.0103-5053http://hdl.handle.net/11449/16482510.21577/0103-5053.20180080S0103-50532018001002054WOS:000442564900007S0103-50532018001002054.pdfWeb of Sciencereponame:Repositório Institucional da UNESPinstname:Universidade Estadual Paulista (UNESP)instacron:UNESPengJournal Of The Brazilian Chemical Societyinfo:eu-repo/semantics/openAccess2023-10-29T06:07:25Zoai:repositorio.unesp.br:11449/164825Repositório InstitucionalPUBhttp://repositorio.unesp.br/oai/requestopendoar:29462024-08-05T16:20:53.912875Repositório Institucional da UNESP - Universidade Estadual Paulista (UNESP)false
dc.title.none.fl_str_mv Electrical Immunosensor Made with Antigenic Peptide NS5A-1 Immobilized onto Silk Fibroin for Diagnosing Hepatitis C
title Electrical Immunosensor Made with Antigenic Peptide NS5A-1 Immobilized onto Silk Fibroin for Diagnosing Hepatitis C
spellingShingle Electrical Immunosensor Made with Antigenic Peptide NS5A-1 Immobilized onto Silk Fibroin for Diagnosing Hepatitis C
Lima, Lais R. [UNESP]
immunosensor
layer-by-layer film
antigenic peptide
hepatitis C
impedance spectroscopy
title_short Electrical Immunosensor Made with Antigenic Peptide NS5A-1 Immobilized onto Silk Fibroin for Diagnosing Hepatitis C
title_full Electrical Immunosensor Made with Antigenic Peptide NS5A-1 Immobilized onto Silk Fibroin for Diagnosing Hepatitis C
title_fullStr Electrical Immunosensor Made with Antigenic Peptide NS5A-1 Immobilized onto Silk Fibroin for Diagnosing Hepatitis C
title_full_unstemmed Electrical Immunosensor Made with Antigenic Peptide NS5A-1 Immobilized onto Silk Fibroin for Diagnosing Hepatitis C
title_sort Electrical Immunosensor Made with Antigenic Peptide NS5A-1 Immobilized onto Silk Fibroin for Diagnosing Hepatitis C
author Lima, Lais R. [UNESP]
author_facet Lima, Lais R. [UNESP]
Goncalves, Alem-Mar B.
Paulovich, Fernando
Oliveira Jr, Osvaldo N.
Ribeiro, Sidney J. L. [UNESP]
Moraes, Marli L.
author_role author
author2 Goncalves, Alem-Mar B.
Paulovich, Fernando
Oliveira Jr, Osvaldo N.
Ribeiro, Sidney J. L. [UNESP]
Moraes, Marli L.
author2_role author
author
author
author
author
dc.contributor.none.fl_str_mv Universidade Estadual Paulista (Unesp)
Universidade Federal de Mato Grosso do Sul (UFMS)
Universidade de São Paulo (USP)
Universidade Federal de São Paulo (UNIFESP)
dc.contributor.author.fl_str_mv Lima, Lais R. [UNESP]
Goncalves, Alem-Mar B.
Paulovich, Fernando
Oliveira Jr, Osvaldo N.
Ribeiro, Sidney J. L. [UNESP]
Moraes, Marli L.
dc.subject.por.fl_str_mv immunosensor
layer-by-layer film
antigenic peptide
hepatitis C
impedance spectroscopy
topic immunosensor
layer-by-layer film
antigenic peptide
hepatitis C
impedance spectroscopy
description Immunosensors based on impedance spectroscopy for diagnosing hepatitis C are reported where the sensing units were made with the antigenic peptide PPLLESWKDPDYVPPWHG (NS5A-1) derived from the NS5A protein of the hepatitis C virus (HCV) immobilized in layer-by-layer (LbL) films with silk fibroin (SF) and deposited on gold interdigitated electrodes. The electrical response of the sensing units varied upon immersion into solutions containing the antibody anti-HCV owing to the biomolecular recognition of NS5A-1. This was associated with morphological changes on the LbL films caused by adsorption of NS5A-1 and inferred from atomic force microscopy images. Buffer solutions with different anti-HCV concentrations down to 2 ng mL(-1) could be clearly distinguished by analyzing the impedance spectroscopy data with a multidimensional projection technique. The specificity toward anti-HCV antibodies was confirmed in control experiments where no significant changes in the electrical response were measured by exposing the sensing units to solutions containing an anti-human immunodeficiency virus (HIV) antibody. The high sensitivity and selectivity of the units made with LbL films demonstrate the feasibility of measuring electrical impedance as an immunosensing strategy to detect hepatitis C.
publishDate 2018
dc.date.none.fl_str_mv 2018-11-26T22:40:49Z
2018-11-26T22:40:49Z
2018-10-01
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://dx.doi.org/10.21577/0103-5053.20180080
Journal Of The Brazilian Chemical Society. Sao Paulo: Soc Brasileira Quimica, v. 29, n. 10, p. 2054-2059, 2018.
0103-5053
http://hdl.handle.net/11449/164825
10.21577/0103-5053.20180080
S0103-50532018001002054
WOS:000442564900007
S0103-50532018001002054.pdf
url http://dx.doi.org/10.21577/0103-5053.20180080
http://hdl.handle.net/11449/164825
identifier_str_mv Journal Of The Brazilian Chemical Society. Sao Paulo: Soc Brasileira Quimica, v. 29, n. 10, p. 2054-2059, 2018.
0103-5053
10.21577/0103-5053.20180080
S0103-50532018001002054
WOS:000442564900007
S0103-50532018001002054.pdf
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv Journal Of The Brazilian Chemical Society
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv 2054-2059
application/pdf
dc.publisher.none.fl_str_mv Soc Brasileira Quimica
publisher.none.fl_str_mv Soc Brasileira Quimica
dc.source.none.fl_str_mv Web of Science
reponame:Repositório Institucional da UNESP
instname:Universidade Estadual Paulista (UNESP)
instacron:UNESP
instname_str Universidade Estadual Paulista (UNESP)
instacron_str UNESP
institution UNESP
reponame_str Repositório Institucional da UNESP
collection Repositório Institucional da UNESP
repository.name.fl_str_mv Repositório Institucional da UNESP - Universidade Estadual Paulista (UNESP)
repository.mail.fl_str_mv
_version_ 1808128636461514752